Feasibility of non-contact cardiorespiratory monitoring using impulse-radio ultra-wideband radar in the neonatal intensive care unit.

Won Hyuk Lee1, Yonggu Lee2, Jae Yoon Na3, Seung Hyun Kim3, Hyun Ju Lee3, Young-Hyo Lim2, Seok Hyun Cho4, Sung Ho Cho1, Hyun-Kyung Park3.   

Abstract

BACKGROUND: Current cardiorespiratory monitoring equipment can cause injuries and infections in neonates with fragile skin. Impulse-radio ultra-wideband (IR-UWB) radar was recently demonstrated to be an effective contactless vital sign monitor in adults. The purpose of this study was to assess heart rates (HRs) and respiratory rates (RRs) in the neonatal intensive care unit (NICU) using IR-UWB radar and to evaluate its accuracy and reliability compared to conventional electrocardiography (ECG)/impedance pneumography (IPG).
METHODS: The HR and RR were recorded in 34 neonates between 3 and 72 days of age during minimal movement (51 measurements in total) using IR-UWB radar (HRRd, RRRd) and ECG/IPG (HRECG, RRIPG) simultaneously. The radar signals were processed in real time using algorithms for neonates. Radar and ECG/IPG measurements were compared using concordance correlation coefficients (CCCs) and Bland-Altman plots.
RESULTS: From the 34 neonates, 12,530 HR samples and 3,504 RR samples were measured. Both the HR and RR measured using the two methods were highly concordant when the neonates had minimal movements (CCC = 0.95 between the RRRd and RRIPG, CCC = 0.97 between the HRRd and HRECG). In the Bland-Altman plot, the mean biases were 0.17 breaths/min (95% limit of agreement [LOA] -7.0-7.3) between the RRRd and RRIPG and -0.23 bpm (95% LOA -5.3-4.8) between the HRRd and HRECG. Moreover, the agreement for the HR and RR measurements between the two modalities was consistently high regardless of neonate weight.
CONCLUSIONS: A cardiorespiratory monitor using IR-UWB radar may provide accurate non-contact HR and RR estimates without wires and electrodes for neonates in the NICU.

Entities:  

Year:  2020        PMID: 33370375      PMCID: PMC7769476          DOI: 10.1371/journal.pone.0243939

Source DB:  PubMed          Journal:  PLoS One        ISSN: 1932-6203            Impact factor:   3.240


  29 in total

1.  Cardiorespiratory events recorded on home monitors: Comparison of healthy infants with those at increased risk for SIDS.

Authors:  R Ramanathan; M J Corwin; C E Hunt; G Lister; L R Tinsley; T Baird; J M Silvestri; D H Crowell; D Hufford; R J Martin; M R Neuman; D E Weese-Mayer; L A Cupples; M Peucker; M Willinger; T G Keens
Journal:  JAMA       Date:  2001-05-02       Impact factor: 56.272

Review 2.  A systematic review of novel technology for monitoring infant and newborn heart rate.

Authors:  Ajay C Kevat; Denise V R Bullen; Peter G Davis; C Omar F Kamlin
Journal:  Acta Paediatr       Date:  2017-03-06       Impact factor: 2.299

3.  A framework for UWB-based communication and location tracking systems for wireless sensor networks.

Authors:  Juan Chóliz; Angela Hernández; Antonio Valdovinos
Journal:  Sensors (Basel)       Date:  2011-09-21       Impact factor: 3.576

4.  Optimal waveforms design for ultra-wideband impulse radio sensors.

Authors:  Bin Li; Zheng Zhou; Weixia Zou; Dejian Li; Chong Zhao
Journal:  Sensors (Basel)       Date:  2010-12-06       Impact factor: 3.576

5.  A Detailed Algorithm for Vital Sign Monitoring of a Stationary/Non-Stationary Human through IR-UWB Radar.

Authors:  Faheem Khan; Sung Ho Cho
Journal:  Sensors (Basel)       Date:  2017-02-04       Impact factor: 3.576

6.  'There were more wires than him': the potential for wireless patient monitoring in neonatal intensive care.

Authors:  Oliver Bonner; Kathryn Beardsall; Nathan Crilly; Joan Lasenby
Journal:  BMJ Innov       Date:  2017-01-04

7.  Detection of Heart Rate through a Wall Using UWB Impulse Radar.

Authors:  Hui-Sup Cho; Young-Jin Park
Journal:  J Healthc Eng       Date:  2018-04-01       Impact factor: 2.682

8.  A Novel Non-contact Heart Rate Monitor Using Impulse-Radio Ultra-Wideband (IR-UWB) Radar Technology.

Authors:  Yonggu Lee; Jun-Young Park; Yeon-Woo Choi; Hyun-Kyung Park; Seok-Hyun Cho; Sung Ho Cho; Young-Hyo Lim
Journal:  Sci Rep       Date:  2018-08-29       Impact factor: 4.379

9.  Non-contact respiration monitoring using impulse radio ultrawideband radar in neonates.

Authors:  Jong Deok Kim; Won Hyuk Lee; Yonggu Lee; Hyun Ju Lee; Teahyen Cha; Seung Hyun Kim; Ki-Min Song; Young-Hyo Lim; Seok Hyun Cho; Sung Ho Cho; Hyun-Kyung Park
Journal:  R Soc Open Sci       Date:  2019-06-05       Impact factor: 2.963

10.  Binodal, wireless epidermal electronic systems with in-sensor analytics for neonatal intensive care.

Authors:  Ha Uk Chung; Bong Hoon Kim; Jong Yoon Lee; Jungyup Lee; Zhaoqian Xie; Erin M Ibler; KunHyuck Lee; Anthony Banks; Ji Yoon Jeong; Jongwon Kim; Christopher Ogle; Dominic Grande; Yongjoon Yu; Hokyung Jang; Pourya Assem; Dennis Ryu; Jean Won Kwak; Myeong Namkoong; Jun Bin Park; Yechan Lee; Do Hoon Kim; Arin Ryu; Jaeseok Jeong; Kevin You; Bowen Ji; Zhuangjian Liu; Qingze Huo; Xue Feng; Yujun Deng; Yeshou Xu; Kyung-In Jang; Jeonghyun Kim; Yihui Zhang; Roozbeh Ghaffari; Casey M Rand; Molly Schau; Aaron Hamvas; Debra E Weese-Mayer; Yonggang Huang; Seung Min Lee; Chi Hwan Lee; Naresh R Shanbhag; Amy S Paller; Shuai Xu; John A Rogers
Journal:  Science       Date:  2019-03-01       Impact factor: 47.728

View more
  4 in total

Review 1.  The newborn delivery room of tomorrow: emerging and future technologies.

Authors:  Natalie Batey; Caroline Henry; Shalabh Garg; Michael Wagner; Atul Malhotra; Michel Valstar; Thomas Smith; Don Sharkey
Journal:  Pediatr Res       Date:  2022-03-03       Impact factor: 3.756

2.  Preclinical evaluation of noncontact vital signs monitoring using real-time IR-UWB radar and factors affecting its accuracy.

Authors:  Jun-Young Park; Yonggu Lee; Ran Heo; Hyun-Kyung Park; Seok-Hyun Cho; Sung Ho Cho; Young-Hyo Lim
Journal:  Sci Rep       Date:  2021-12-08       Impact factor: 4.379

3.  Preclinical trial of noncontact anthropometric measurement using IR-UWB radar.

Authors:  Jinsup Kim; Won Hyuk Lee; Seung Hyun Kim; Jae Yoon Na; Young-Hyo Lim; Seok Hyun Cho; Sung Ho Cho; Hyun-Kyung Park
Journal:  Sci Rep       Date:  2022-05-17       Impact factor: 4.996

4.  Non-contact Sleep/Wake Monitoring Using Impulse-Radio Ultrawideband Radar in Neonates.

Authors:  Won Hyuk Lee; Seung Hyun Kim; Jae Yoon Na; Young-Hyo Lim; Seok Hyun Cho; Sung Ho Cho; Hyun-Kyung Park
Journal:  Front Pediatr       Date:  2021-12-21       Impact factor: 3.418

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.