Detection of small amounts of biological compounds is of ever-increasing importance but also remains an experimental challenge. In this context, plasmonic nanoparticles have emerged as strong contenders enabling label-free optical sensing with single-molecule resolution. However, the performance of a plasmonic single-molecule biosensor is not only dependent on its ability to detect a molecule but equally importantly on its efficiency to transport it to the binding site. Here, we present a theoretical study of the impact of downscaling fluidic structures decorated with plasmonic nanoparticles from conventional microfluidics to nanofluidics. We find that for ultrasmall picolitre sample volumes, nanofluidics enables unprecedented binding characteristics inaccessible with conventional microfluidic devices, and that both detection times and number of detected binding events can be improved by several orders of magnitude. Therefore, we propose nanoplasmonic-nanofluidic biosensing platforms as an efficient tool that paves the way for label-free single-molecule detection from ultrasmall volumes, such as single cells.
Detection of small amounts of biological compounds is of ever-increasing importance but also remains an experimental challenge. In this context, plasmonic nanoparticles have emerged as strong contenders enabling label-free optical sensing with single-molecule resolution. However, the performance of a plasmonic single-molecule biosensor is not only dependent on its ability to detect a molecule but equally importantly on its efficiency to transport it to the binding site. Here, we present a theoretical study of the impact of downscaling fluidic structures decorated with plasmonic nanoparticles from conventional microfluidics to nanofluidics. We find that for ultrasmall picolitre sample volumes, nanofluidics enables unprecedented binding characteristics inaccessible with conventional microfluidic devices, and that both detection times and number of detected binding events can be improved by several orders of magnitude. Therefore, we propose nanoplasmonic-nanofluidic biosensing platforms as an efficient tool that paves the way for label-free single-molecule detection from ultrasmall volumes, such as single cells.
Detection
of tiny amounts of
chemical and biological entities is one of the most important tasks
in chemistry, biology, medicine, environmental monitoring, and homeland
security. Optical sensors based on plasmonic nanoantennas supporting
localized surface plasmon resonance (LSPR) are promising to provide
a solution for this challenge since they combine high sensitivity
with label-free detection and vast miniaturization potential.[1] Furthermore, by utilizing individual plasmonic
nanoparticles (NPs) as highly sensitive probes, the detection of single
biomolecule binding events has been demonstrated in multiple studies,[2] as well as in a highly multiplexed fashion.[3] From an application perspective, single-molecule
detection is particularly important when only a limited sample volume
is available and/or when only a very small number of biomolecules
are present in the sample.[4] This is, for
instance, relevant for the study of single cells that have a volume
of a few picolitres only.[5] Such single-cell
analysis enables studies of intracellular variability in many biological
processes and it has been hallmarked as a necessity for progress in
cell biology and tissue engineering.[6] However,
studies at the single-cell level require the capacity to manipulate
small sample volumes down to a picolitre and the capacity to detect
extremely low concentrations of target molecules.[7,8] To
this end, plasmonic biosensors combined with a microfluidic device
have recently been used for the detection of single-cell cytokine
expression.[9]The practical relevance
of nanoplasmonic biosensor devices, however,
does not depend only on the ability of a plasmonic NP to transduce
the single binding event, but also—and equally critically—on
the ability of the device to effectively transport a single biomolecule
to the binding site. Here, the performance of conventional surface-based
biosensors often suffers from inefficient mass transport that influences
the binding kinetics and leads to very long detection times for low
concentrations of target molecules. Therefore, strategies to improve
mass transport by enforcing sensor operation in the diffusion-limited
regime have been reported for plasmonic nanoarchitectures like sparse
arrays of plasmonic nanoparticles,[10] nanoholes,
and nanopores.[11] However, as it was pointed
out in several theoretical mass-transport studies,[12,13] plasmonic nanostructures can be of benefit only if a large number
of sensing entities are involved.Here, we present a theoretical
analysis of the mass-transport characteristics
of nanoplasmonic sensor architectures with particular focus on the
impact of the size of the sample volume on binding kinetics. In particular,
using both an analytical approach and stochastic simulations of molecular
mass transport, we show that for applications where ultrasmall sample
volumes are critical, such as in single-cell analysis, efficient mass
transport to the plasmonic sensor still remains a considerable challenge.
Moreover, we propose a solution to this problem by combining nanoplasmonic
sensing with nanofluidics into a single device. Nanofluidics provides
an efficient tool for the handling of fluids in the subpicoliter range
by downsizing the fluid control to the nanoscale.[14] However, apart from proof-of-principle studies demonstrating
the feasibility of nanoplasmonic–nanofluidic biosensing,[15,16] it has so far mostly been used in the context of fluorescence microscopy.[17,18] As the key result of our study, we identify a regime where the downscaling
of the fluidic system from widely used conventional microfluidics
to the dimensions characteristic of nanofluidic structures enables
the reliable plasmonic label-free detection of single molecules within
a realistic detection time range that has not been accessible previously.
Mass Transport in Nanoplasmonic Biosensors
In this
study, we assume a nanoplasmonic biosensor comprising a
fluidic structure containing a solution of analyte molecules that
is transported via diffusion and convection when
pressure-driven flow is induced through the structure. One of the
walls of the fluidic structure is decorated with an array of plasmonic
NPs functionalized with immobilized receptors with a surface density,
Γ0 (Figure ). We assume that the analyte molecules exclusively bind to
the receptors (i.e., there is no unspecific binding)
in a 1:1 ratio and the interaction is characterized by the association
and dissociation constants, kon and koff. It has to be noted, that this represents
an idealized situation, where unspecific binding is not taken into
account. In practice, suppression of unspecific binding represents
a considerable challenge, especially for biosensing in complex media.
However, we propose that it can be efficiently mitigated using established
biochemical surface functionalization methods.[19] Moreover, we assume that the array of plasmonic NPs, as
well as the optical readout, is designed in a way that single-binding
and unbinding events can be resolved. In this setting, the binding
and unbinding of analyte molecules is controlled by the complex interplay
between convection, diffusion, and reaction of the analyte with the
immobilized receptors and can be described by the convection–diffusion
equation coupled with the first-order Langmuir kinetic equationwhere D is the diffusivity
of the analyte molecule, v is the flow velocity field
defined by the architecture of the fluidic structure, c is the (bulk) concentration of the analyte molecules inside the
fluidic structure, cS is the (bulk) concentration
of unbound analyte molecules at the NP surface, and Γ is the
surface concentration of the bound analyte molecules.
Figure 1
(a) Schematic depiction
of a nanoplasmonic biosensor comprising
an array of plasmonic NPs functionalized with analyte-specific receptors.
The transport of the analyte molecules present in the sample volume
to the surface of the particle is defined by the flow velocity, v, and the diffusivity, D. The binding and
unbinding of the analyte molecules are defined by the kinetic rate
constants kon and koff, respectively. (b–e) Typical scenarios related to
single-molecule detection: (b) a large sample volume is flown through
a fluidic channel, (c) an ultrasmall sample volume is pipetted into
a fluidic well, and (d) a single cell is inserted into a fluidic well.
The cell can then either be lysed to enable subsequent analysis of
the intracellular content or secrete metabolites that are analyzed
and (e) a single cell is immobilized inside a fluidic structure adjacent
to a nanofluidic channel and the intracellular content or its metabolites
are transported into the sensing region, where they are detected.
(a) Schematic depiction
of a nanoplasmonic biosensor comprising
an array of plasmonic NPs functionalized with analyte-specific receptors.
The transport of the analyte molecules present in the sample volume
to the surface of the particle is defined by the flow velocity, v, and the diffusivity, D. The binding and
unbinding of the analyte molecules are defined by the kinetic rate
constants kon and koff, respectively. (b–e) Typical scenarios related to
single-molecule detection: (b) a large sample volume is flown through
a fluidic channel, (c) an ultrasmall sample volume is pipetted into
a fluidic well, and (d) a single cell is inserted into a fluidic well.
The cell can then either be lysed to enable subsequent analysis of
the intracellular content or secrete metabolites that are analyzed
and (e) a single cell is immobilized inside a fluidic structure adjacent
to a nanofluidic channel and the intracellular content or its metabolites
are transported into the sensing region, where they are detected.In the following sections, we have explored the
binding characteristics
for a series of typical scenarios related to single-molecule detection
by focusing on the impact of characteristic dimensions of the used
fluidic structures to enable the direct comparison of micro- and nanofluidics
in this respect. As we will demonstrate further, the key characteristic
that leads to qualitatively different conclusions is the sample volume
(volume of the analyte solution, Vs) relative
to the volume of the fluidic structure. Therefore, each section of
our work reported below describes a different scenario in this respect. Section analyzes the situation
that considers an infinite sample volume that flows through a fluidic
channel. In practice, this situation can be encountered when a sample
is taken from a patient and flushed through the fluidic structure
(Figure b). The volume
is large enough (i.e., μL or mL) that the detection
time is not limited by the time needed to flow the whole volume through
the fluidic system.[20]Sections 3 and 4 analyze a different
scenario that considers a finite sample volume, containing only a
limited number of molecules, where the means of loading the sample
into the device is critical. In practice, this situation corresponds
to applications dealing with ultrasmall sample volumes (pL or nL).
Specifically, Section considers a confined finite sample volume in an enclosed fluidic
well without flow. This situation is encountered when a tiny amount
of a sample is pipetted inside a well, followed by enclosing the well
(Figure c); alternatively,
a single cell is inserted into the fluidic well, and then the cell
can be either lysed and intracellular content is subsequently analyzed[9] or its metabolite secretion is studied[21,22] (Figure d). Section considers a finite
sample volume that flows through a fluidic channel. This scenario
corresponds to a situation where, for e.g., a single
cell is immobilized close to a nanofluidic channel and the intracellular
content or its metabolites are transported into the nanofluidic sensing
region, where they are detected[7] (Figure e).
Infinite Sample Volume
In this section, we consider the
first case of an infinite sample
volume that flows through a fluidic channel with a volumetric flow
rate, Q (Figure a). For this scenario, the problem of mass transport
described by eqs and 2 can be treated analytically in terms of the two
compartment model,[23,24] which states that when binding
occurs quasi-steadily, the analyte concentration at the surface can
be described aswhere Γf = Γ0 – Γ
is the surface concentration of the free receptors, c0 is the initial (bulk) concentration of the
analyte molecules, and km is the mass
transfer coefficient, which can be calculated analytically (see SI Section 1) via parameters related
to both the fluidic conditions and the architecture of the NP array.[25] The term in the denominator, the Damköhler
number konΓf/km, describes the relative rate of analyte reactive
capture with respect to the rate of analyte transport to the sensor
surface. In a regime when km ≪ konΓf, the collection rate is
dictated mainly by the transport of the analyte molecules to the binding
sites, i.e., the system is diffusion-limited. In
the opposite regime, when km ≫ konΓf, mass transport supplies
analyte molecules more quickly than the receptors can bind them, the
system is therefore reaction-limited. In essence and at the general
level, this means that for each specific affinity format (defined
by kon, koff, Γ0 of the used molecules) the maximum collection
rate is achieved when the system is reaction-limited.
Figure 2
(a) Schematic depiction
of the simulated system consisting of an
infinite sample volume, which flows at a volumetric flow rate Q through a sensing region of a fluidic channel of height H, width W, and length L, comprising a functionalized plasmonic NP array of area A. (b) Summary of the parameters used in the calculations.
(c, d) Top view of the two investigated configurations of the fluidic
channel: (c) meandering channel with L = 1000 μm
and W = 10 μm and (d) straight channel with L = W = 100 μm. Both types of fluidic
systems contain the same NP array defined by the area A = 100 × 100 μm2, i.e., the
field of the view of a 100× microscope objective. (e) Calculated
comparison of km and konΓ0 describing the relative rate of
reactive capture with respect to the rate of analyte transport for
both the meandering channel (dashed lines) and the straight channel
(solid lines). The results shown for km correspond to a range of technically relevant flow rates Q, a specific size of analyte molecules defined by D, and they were calculated for the two channel types assuming
the channel height H ranging from 10 μm characteristic
of microfluidics down to 100 nm characteristic of nanofluidics. The
results shown for konΓ0 correspond to the values characteristic for protein interactions,
that is, kon = 104–106 M–1·s–1.[13] Note that for the entire range of assumed parameters km > konΓ0, which means that the system is predominantly reaction-limited.
(a) Schematic depiction
of the simulated system consisting of an
infinite sample volume, which flows at a volumetric flow rate Q through a sensing region of a fluidic channel of height H, width W, and length L, comprising a functionalized plasmonic NP array of area A. (b) Summary of the parameters used in the calculations.
(c, d) Top view of the two investigated configurations of the fluidic
channel: (c) meandering channel with L = 1000 μm
and W = 10 μm and (d) straight channel with L = W = 100 μm. Both types of fluidic
systems contain the same NP array defined by the area A = 100 × 100 μm2, i.e., the
field of the view of a 100× microscope objective. (e) Calculated
comparison of km and konΓ0 describing the relative rate of
reactive capture with respect to the rate of analyte transport for
both the meandering channel (dashed lines) and the straight channel
(solid lines). The results shown for km correspond to a range of technically relevant flow rates Q, a specific size of analyte molecules defined by D, and they were calculated for the two channel types assuming
the channel height H ranging from 10 μm characteristic
of microfluidics down to 100 nm characteristic of nanofluidics. The
results shown for konΓ0 correspond to the values characteristic for protein interactions,
that is, kon = 104–106 M–1·s–1.[13] Note that for the entire range of assumed parameters km > konΓ0, which means that the system is predominantly reaction-limited.For our analysis, we define our system such that
the dimensions
of the sensing region of the fluidic channel are given by its height H, width W, and length L. The sensing region comprises the functionalized NPs that are defined
by their surface area Ap and their shape.
The NPs are arranged into an array with a surface density Γp, covering the area A = 100 × 100 μm2 that fits the field of view of a 100× microscope objective.
To implement this scenario, we consider two different arrangements
of the fluidic channel: (i) a meandering channel with W = 10 μm and L = 1000 μm (Figure c) and (ii) a straight channel
with W = L = 100 μm (Figure d). We note that
the thickness of the walls between the meandering channels is omitted
in the theoretical analysis. In practice, using state-of-the-art nanofabrication,
wall thicknesses down to 300 nm are easily feasible due to well-established
vertical etch processes,[26] thereby not
significantly impacting the conclusions from our calculations. To
investigate the analyte transport to the binding sites, we have calculated
the mass transfer coefficient, km, and
compared it to values of konΓ0 typical for protein interactions (Figure e).[13] Specifically,
we carried out calculations for a NP array with parameters previously
efficiently used for single-molecule biosensing[3] (Figure b) and for the two types of fluidic systems depicted in Figure c,d, by systematically
varying the channel height H from 10 μm, as
characteristic of microfluidics, down to 100 nm, as characteristic
of nanofluidics, and by screening a range of technically relevant
flow rates Q.[27] Details
of the calculations, as well as the results of the extended analysis
for a range of scenarios characteristic for a broad set of parameters
pertaining to the fluidic structure (H, W, L, Q),
NP array, (A, Ap, Γp), and analytes (D, kon), can be found in SI Section 1 and Figures S-1, S-2. We note that for all of the assumed parameters that
are relevant for current state-of-the-art single-molecule nanoplasmonic
biosensors, km ≫ konΓ0 can be achieved for high enough,
but still technically relevant, Q. This also means
that the collection rate can reach its maximum, i.e., the reaction-limited regime, irrespective of the fluidic structure
dimensions. Thus, cs ≈ c0 (eq ), and the collection rate (number of bound analyte molecules
per unit time) can be written asby using eq where As = ΓpApA represents
the active sensing area, that is, the area covered by receptors. From eq it can be seen that in
the reaction-limited regime, for any specific affinity format (defined
by kon, koff, and Γ0) and any specific analyte concentration,
the collection rate can only be improved by enlarging the active sensing
area. However, the size of the active sensing area is limited by a
series of constraints related to the optical readout for single-molecule
detection: (i) The maximal surface area of the NPs in the array (Ap) is limited by the condition of single-molecule
resolution since the optical response to a single-molecule binding
event worsens with increasing dimensions of the plasmonic NP.[28] (ii) The maximal area of the NP array (A) is limited by the field of view of the used microscope.
(iii) The maximal density of the NP array (Γp) is
limited by the optical resolution of the microscope, i.e., the response from each individual NP has to be resolved separately,
avoiding incidental cross-talk between neighboring NPs.In other
words, and as the main conclusion for a fluidic channel
connected to an infinite reservoir of the analyte, changing the channel
dimension from the micro- to the nanofluidic regime has little to
no impact on the collection rate under predominantly reaction-limited
conditions and, therefore, neither offers a handle to significantly
improve the overall binding characteristics nor the sensor response
time.
Finite Sample Volume Inside a Fluidic Well
The situation described above becomes very different when we consider
a scenario where the sample volume, Vs, is very small, that is, in the pL range and thus below the volumes
that effectively can be handled by traditional microfluidics (μL
or mL). In this regime, as we will illustrate in detail, the characteristic
dimensions of the fluidic system have a significant impact on the
binding characteristics and thus on the critical timescales for single-molecule
detection. In the following example, we assume the finite sample volume
(Vs) containing a finite number of molecules, M0 = c0Vs, that is placed in an enclosed fluidic well under stagnant
conditions without applied flow (Figure a). The fluidic well is defined by its volume, V ≥ Vs, and height, H, and comprises an array of functionalized plasmonic NPs
defined in the same manner as in Section . It has to be noted that in the situation
when Vs < V, the analyte
molecules are transported into the entire volume by diffusion and
the sample is therefore diluted.
Figure 3
(a) Schematic depiction of the simulated
system considering a finite
sample volume, Vs, placed inside a fluidic
well with a volume V and height H, decorated with an array of functionalized plasmonic NPs covering
an area A. (b) Dependency of the normalized initial
collection rate (eq ) on the height and volume of the fluidic well in the reaction-limited
regime. The maximal A is limited by Afw = 100 × 100 μm2, i.e., the field of view of a 100× microscope objective. (c–f)
Schematic depictions of the investigated fluidic wells defined by
their heights (c, d) H = 100 nm (nanowell) and (e,
f) H = 10 μm (microwell) and by their volumes
(c, e) V = 1 pL and (d, f) V = 100
pL.
(a) Schematic depiction of the simulated
system considering a finite
sample volume, Vs, placed inside a fluidic
well with a volume V and height H, decorated with an array of functionalized plasmonic NPs covering
an area A. (b) Dependency of the normalized initial
collection rate (eq ) on the height and volume of the fluidic well in the reaction-limited
regime. The maximal A is limited by Afw = 100 × 100 μm2, i.e., the field of view of a 100× microscope objective. (c–f)
Schematic depictions of the investigated fluidic wells defined by
their heights (c, d) H = 100 nm (nanowell) and (e,
f) H = 10 μm (microwell) and by their volumes
(c, e) V = 1 pL and (d, f) V = 100
pL.The analytical theory of mass
transport[25] used in Section is essentially not applicable for such an
enclosed system, as it
is valid only for an open system in a quasi-steady state. Therefore,
we have instead utilized a stochastic diffusion-reaction model that
rigorously describes the behavior of molecules inside fluidic wells
with functionalized surfaces (details in Methods ). We have carried out simulations for a broad range of parameters
pertaining to the dimensions of a fluidic well (H, V), the architecture of an NP array, (A, Ap, Γp),
and properties of analytes (D, kon). As one important general conclusion, the results
show (SI Section 2, Figures S-3, S-4) that
for all scenarios relevant for current state-of-the-art single-molecule
nanoplasmonic biosensors, the mass transport is mostly reaction-limited.
Moreover, the binding characteristics are independent from the vertical
or horizontal position of the molecules inside the well at the beginning
of the detection process. In other words, the binding characteristics
are independent of the sample placement or, in the case Vs < V, the concentration profile during
sample dilution (diffusivity of analyte molecules). Therefore, the
concentration of the unbound analyte molecules at the surface of the
NPs corresponds to the mean concentration of the unbound analyte molecules
inside the fluidic well, i.e., cs(t) = M0/V – N(t)/V. Thus, using eq , the collection rate can be written asEquation suggests
that for each specific affinity format (defined by kon, koff, Γ0), the sample volume and concentration (defined by M0), the collection rate can be improved by maximizing
the As/V ratio. In other
words, enlarging of the active sensing area (As) is beneficial only if it does not require an increase of
the volume of the well (V > Vs), as this would lead to further dilution of the sample.
Thus,
as the key point, in comparison to the case for the infinite sample
volume analyzed first, not only the architecture of the NPs, but also
the size of the fluidic structure is critical in the present scenario.To now showcase this scenario explicitly, we consider different
volumes, V, of a fluidic well that range from subpicolitre
to hundreds of picolitre and two fixed heights, a microwell with H = 10 μm and a nanowell with H =
100 nm (Figure c–f).
All assumed fluidic wells comprise plasmonic NP arrays with the same
Γp and Ap and the same
number of molecules, M0. In both cases,
we assume that the area of the plasmonic NP array is either limited
by Afw = 100 × 100 μm2, the field of view of a 100× microscope objective, or by the
dimensions of the horizontal wall of the fluidic well. To provide
a general comparison of binding characteristics that are valid for
any number of molecules present in the sample volume, affinity formats
and NP arrays, the initial collection rates, J (eq , N(t → 0)
= 0), we thus show as the normalized values (Figure b)For both the micro- and
the nanowell it applies
that for V > HAfw the
area of the NP array is limited by the field of view of the microscope
and thus the normalized initial collection rates increase with decreasing
volume since the A/V ratio increases
(eq ). For V ≤ HAfw, the area of the NP array
matches the area of the wells’ horizontal wall and thus the
normalized initial collection rates are constant since the A/V ratio also is constant. As a consequence,
the normalized initial collection rates for volumes V > 100 pL are equal for the nanowell (Figure d) and the microwell (Figure f), since the areas of the NP array are identical
and match the field of view of the microscope. However, for volumes V < 100 pL, the initial collection rates are remarkably
higher for the nanowell (Figure c) than for the microwell (Figure e), since the area of the NP array inside
the nanowell is larger. Therefore, for ultrasmall sample volumes (Vs < 100 pL), the collection rates can be
improved by up to 2 orders of magnitude by decreasing the dimensions
of the fluidic system from the microfluidic to the nanofluidic range
and by matching the volume of the fluidic well to the sample volume.
In this way, analyte dilution is avoided, while at the same time the
sensing area is maximized. We also note that, analogously to the case
of Section , binding
characteristics are not dependent on the width of the well, as long
as the well is able to host the entire NP array (Figure c,d).To further investigate
the impact of the size of a fluidic well
on the binding characteristics of an analyte from a finite volume,
we have calculated two important characteristics: (i) the mean waiting
time until the first detected binding event by the sensor, t1, and (ii) the total number of binding events, N+, detected as a function time. Using eq , the mean waiting time
until the first binding event can be defined asPlease note that
this parameter is sometimes
referred to as the accumulation time.[11] Assuming that the number of analyte molecules is much smaller than
the number of receptors (N0 = Γ0As), M0 ≪ N0, eq can be solved analytically, and the number
of binding events can be written aswhere K–1 = 1 + koffV/( konN0).Figure shows relevant
examples for t1 and N+ for a sample volume Vs =
1 pL placed inside fluidic wells with different heights—a microwell
with H = 10 μm (Figure e,f) and a nanowell with H = 100 nm (Figure c). The parameters defining the NP array and the binding properties
of the molecules are the same as in the previous example (cf. Figure b). Using these boundary conditions, we executed calculations that
cover the kon–koff space of biomolecular interactions typical of proteins
with kon ranging from 104 to
106 M–1·s–1 and koff from 10–5 to 10–2 s–1. Specifically, we compare the results of the
analytical model pertaining to the reaction-limited regime (eqs and 8, Figure a–f
solid lines) with the results of the corresponding stochastic simulations
(Figure a–f
dashed lines). We note that the results from the analytical model
and the simulations generally agree very well, except for fast binding
kinetics, where the values of t1 are slightly
higher (Figure a)
and the values of N+ are slightly lower
at the beginning of the binding curve (Figure c,e). This suggests a moderate diffusion
limitation in this regime, which, however, does not affect the general
conclusions discussed below. As the key results, from the data it
can be seen that t1 decreases linearly
with increasing concentration of the analyte molecules (i.e., the number of analyte molecules confined inside the volume) and
it is 2 orders of magnitude smaller for the nanowell than for the
microwell (Figure a,b). Specifically, when only a few molecules are confined inside
the microwell, t1 ranges from minutes
(Figure a) to tens
of hours (Figure b),
depending on kon. As the key point, for
most kon, this is far beyond practical
timescales. On the other hand, if the same sample containing only
a few molecules is confined inside the nanowell instead, t1 is substantially reduced to the practically relevant
timescales ranging from seconds (Figure a) to minutes, depending on kon (Figure b).
Figure 4
Binding characteristics of 1 pL sample volume confined inside a
microwell and a nanowell with defined heights, H =
10 μm and H = 100 nm, respectively. The architecture
of the wells corresponds to those described in Figure e for the microwell and Figure c for the nanowell. The parameters
corresponding to the NP array and molecules are summarized in Figure b. (a, b) Dependency
of the mean waiting time until the first binding event on the number
of the analyte molecules present in a 1 pL sample volume, M0, at the concentration c0. (c–f) Mean number of binding events for each molecule
present in a 1 pL volume, N+/M0. The results are shown for the kon–koff space of biomolecular interactions typical
of proteins.
Binding characteristics of 1 pL sample volume confined inside a
microwell and a nanowell with defined heights, H =
10 μm and H = 100 nm, respectively. The architecture
of the wells corresponds to those described in Figure e for the microwell and Figure c for the nanowell. The parameters
corresponding to the NP array and molecules are summarized in Figure b. (a, b) Dependency
of the mean waiting time until the first binding event on the number
of the analyte molecules present in a 1 pL sample volume, M0, at the concentration c0. (c–f) Mean number of binding events for each molecule
present in a 1 pL volume, N+/M0. The results are shown for the kon–koff space of biomolecular interactions typical
of proteins.Furthermore, we see that the binding
rate (number of binding events
per time, dN+/dt) is
substantially increased for the nanowell compared to the microwell
for the whole kon–koff space (Figure c−f). Equilibrium is reached 2 orders of magnitude
faster, and also the binding rate at equilibrium is 2 orders of magnitude
higher for the nanowell. Specifically, for kinetics with low kon and high koff, it can be seen in Figure f that, while in the microwell each molecule in the sample
volume is detected within 1 h with a probability of 3.5% (corresponds
to N+/M0 =
0.0035), in the nanowell, each molecule is detected about 10 times
per hour (N+/M0 = 10). In other words, not only enables the nanowell faster detection
but it actually enables multiple detection events of a single molecule
within reasonably short time. The detection limit of plasmonic biosensors
with single-molecule resolution, defined in terms of the minimal amount
of analyte molecules detected in a reasonable time, is mainly limited
by the binding rate.[12,13,29] This unprecedented performance of the nanowell implies that it,
similar to a photon multiplier, can “multiply” the signal
from a very small number of analyte molecules and thus, in principle,
improve the detection limit significantly. To put this finding into
perspective from an application viewpoint, we note that binding kinetics
with low kon and high koff are a regime characteristic for many biological processes,
such as interactions of cell surface proteins.[30] At the same time, low kon and
high koff result in a very low surface
concentration of bound molecules in equilibrium, which makes their
detection very challenging or even impossible for traditional biodetection
techniques, such as biosensors based on surface plasmon resonance
(SPR), that rely on measuring the mean surface concentration of the
bound molecules.[31] In this context the
predicted multiple detection of the same molecule enabled by the nanowell,
as the molecule binds and unbinds, constitutes a significant advantage.
Finite Sample Volume Inside a Flow-Through Fluidic
Channel
In Section 3, we have demonstrated
the
potential of nanofluidics in the field of single-molecule detection
from ultrasmall sample volumes. However, despite the recent advances
in manipulation of such ultrasmall volumes,[32] enclosing a sample into a nanowell as described in Section is a considerable challenge
to realize in practice. In addition, as the size of a cell typically
exceeds the height of a nanowell by more than 2 orders of magnitude, in vivo single-cell analysis would be very difficult. Therefore,
in this section, we analyze the potential benefits of nanofluidics
compared to microfluidics in a flow-through configuration. In particular,
we assume that the content of an ultrasmall sample volume Vs is transported by diffusion and/or convection
defined by D and Q, respectively,
to the sensing region of a micro- or nanofluidic channel, which contains
an array of functionalized plasmonic NPs defined in the same manner
as in Section (Figure a–c). For
this specific configuration, the pertinent equations describing diffusion,
convection, and reaction (eqs and 2) cannot be solved analytically.
Therefore, we have employed the stochastic diffusion–convection–reaction
model, described in Section .
Figure 5
(a–c) Schematic depiction of the considered system comprising
a finite sample volume Vs = 1 pL, which
flows with a volumetric flow rate Q through (a) a
microchannel with H = 10 μm or (b) a nanochannel
with H = 100 nm. The dimensions of the channels in
both cases are W = 10 μm and L = 1000 μm and they are arranged into meanders to fit an array
of functionalized plasmonic NPs with an area A =
100 × 100 μm2 to match the field of view of
a 100× microscope. At the start of the detection process (t = 0), the sample volume is located outside of the sensing
region toward which it is transported over time by means of diffusive
and convective flow. (c) Top view of the simulated system. (d) Time
evolution of the normalized analyte concentration distribution, c/c0, for the flow rate Q = 1 μm3·s–1 and
diffusivity of the analyte D = 10 μm2·s–1. (e) Time-dependent normalized averaged
concentration over the sensing area, c̅/c0. We note that binding and unbinding is not
considered here.
(a–c) Schematic depiction of the considered system comprising
a finite sample volume Vs = 1 pL, which
flows with a volumetric flow rate Q through (a) a
microchannel with H = 10 μm or (b) a nanochannel
with H = 100 nm. The dimensions of the channels in
both cases are W = 10 μm and L = 1000 μm and they are arranged into meanders to fit an array
of functionalized plasmonic NPs with an area A =
100 × 100 μm2 to match the field of view of
a 100× microscope. At the start of the detection process (t = 0), the sample volume is located outside of the sensing
region toward which it is transported over time by means of diffusive
and convective flow. (c) Top view of the simulated system. (d) Time
evolution of the normalized analyte concentration distribution, c/c0, for the flow rate Q = 1 μm3·s–1 and
diffusivity of the analyte D = 10 μm2·s–1. (e) Time-dependent normalized averaged
concentration over the sensing area, c̅/c0. We note that binding and unbinding is not
considered here.To showcase this scenario,
we present the results of an analysis
of a specific example. However, as discussed below, more general conclusions
can be derived based on this example. To relate directly to the results
presented in Sections 2 and 3, also here we assume a sample volume Vs = 1 pL, which, for example, corresponds to the volume of
a β cell,[33] an NP array with A = 100 × 100 μm2 with parameters
defined in Figure b and a fluidic channel with the width W = 10 μm
and length L = 1000 μm (Figure c). To compare the performance of a nanofluidic
with a microfluidic channel, we again compare the two cases of channel
heights H = 10 μm (microfluidics) and H = 100 nm (nanofluidics) (Figure a,b). Here, we also note that the fluidic
channel anywhere outside the sensing region is assumed to have a height
of 10 μm.In the first step of our analysis, to decouple
the interplay of
diffusion, convection, and reaction, we have studied the transport
of analyte molecules through the fluidic channels without considering
binding in the sensing region (Figure d,e). Under a low flow rate (Q = 1
μm3·s–1), the analyte molecules
are transported mainly by diffusion, which results in a slow increase
of the mean concentration of the analyte molecules in the sensing
region over time, , where V is the volume
of the sensing region of the channel (Figure d). Increasing the flow rate results in faster
transport, but shortens the time the analyte is present in the sensing
region (Figure e).
Furthermore, it can be seen that the analyte transport differs considerably
for the microchannel and the nanochannel. For the microchannel, since
the sensing volume is 100 × higher than the sample volume, the
maximum c̅(t) inside the sensing
volume is 0.01 × c0, which corresponds
to a situation where 100% of the analyte molecules are present inside
the sensing volume. On the other hand, for the nanochannel, since
the volume of the channel is substantially smaller and matches the
sample volume, the maximum c̅(t) is much higher (0.3 × c0), and
corresponds to a situation where maximal 30% of all of the analyte
molecules are present inside the sensing volume. The rest of the molecules
are “lost” due to diffusion into the channel outside
the sensing region.In the second step of our analysis, we now
include the binding
and unbinding of analyte molecules inside the sensing region. The
corresponding results for a kon–koff space characteristic of proteins are presented
as the mean number of binding events for each molecule in the sample
volume in Figure a–d.
It has to be noted that the mean number of binding events for a nano-
and microchannel with any applied flow cannot exceed the limits set
by the previous case of the enclosed nano- and microwell, respectively,
as described in Section (added as dashed lines in Figure ). At the beginning of the detection, the sample volume
is localized mainly outside the sensing region (i.e., c̅ is low, see Figure e), therefore the mean number of binding
events for each molecule is considerably lower compared to the case
of the enclosed fluidic wells. Subsequently, as time passes, more
and more molecules are transported toward the sensing area in the
channel, (i.e., c̅ increases,
see Figure e). Therefore
the mean number of binding events increases rapidly. However, it is
overall lower compared to the case of the enclosed fluidic well due
to the “loss” of molecules because of the applied flow.
Increasing the flow rate results in an increase of the mean number
of binding events at the beginning of the detection but at the cost
of losing more molecules that never bind. Hence, the choice of the
optimal flow rate is dependent on the binding kinetics of the specific
molecules at hand and the required detection time. Specifically, for
high kon and low koff (Figure a), most binding events would be detected at a high flow rate, independent
of the detection time. On the other hand, for the remaining cases
(Figure c,d), for
short detection times more binding events will be detected at high
flow rates and for long detection times more binding events will be
detected at low flow rates.
Figure 6
Mean number of binding events for each molecule
present in a 1
pL volume, N+/M0, detected inside the fluidic micro- or nanochannels depicted in Figure a,b, respectively,
for a range of applied flow rates Q = 0–100
μm3·s–1. For comparison, the
data from the enclosed fluidic micro- and nanowells depicted in Figure c,e, respectively,
are also shown. The results were obtained from stochastic simulations
for a kon–koff space characteristic of proteins.
Mean number of binding events for each molecule
present in a 1
pL volume, N+/M0, detected inside the fluidic micro- or nanochannels depicted in Figure a,b, respectively,
for a range of applied flow rates Q = 0–100
μm3·s–1. For comparison, the
data from the enclosed fluidic micro- and nanowells depicted in Figure c,e, respectively,
are also shown. The results were obtained from stochastic simulations
for a kon–koff space characteristic of proteins.As the key result, we thus find that the overall
binding characteristics in the nanofluidic channel (Figure b) can be improved by at least
an order of magnitude compared to the microfluidic channel (Figure a) or the enclosed
microwell (Figure e) for any specific type of binding kinetics or detection time. Therefore,
we postulate that such a configuration presents an alternative to
single-cell microfluidic devices that are used for the analysis of
the extracellular environment and metabolite secretion[21,22] or for lysis and the subsequent study of the intracellular content.[34] To this end, two experimental studies suggesting
single-cell analysis based on nanofluidic devices using fluorescent
labeling[8] or nanosampling[7] exist.
Conclusions
In conclusion,
our theoretical analysis has shown that decreasing
the dimensions of a fluidic channel from the regime characteristic
of microfluidics to the regime of nanofluidics can reduce the detection
time from ultrasmall volumes (below hundreds of picoliters) by up
to 2 orders of magnitude and thus make single-molecule detection practically
viable. Specifically, the showcased representative example inspired
by a single living cell with 1 pL volume has shown that for the whole
range of association and dissociation constants characteristic of
affinity-based protein detection, most of the molecules in the sensing
volume can be detected within minutes. Furthermore, for kinetics involving
the creation of unstable complexes due to fast unbinding kinetics,
we have demonstrated that every molecule can be detected multiple
times within tens of minutes in a nanofluidic channel and thus improve
the detection limit of the sensor considerably, thanks to this “molecular
multiplier” effect. We therefore predict that nanoplasmonic
sensors integrated with nanofluidics present a promising new paradigm
for single-molecule detection from ultrasmall sample volumes, such
as single cells.
Methods
Stochastic Diffusion–Convection–Reaction
Simulations of Biomolecular Interactions
To describe the
molecular binding rigorously, we have implemented stochastic diffusion–convection–reaction
simulations. Our model is designed to mimic the behavior of a single
molecule inside a fluidic structure with functionalized arrays of
NPs. It is inspired by the previous work from the field of stochastic
simulations of diffusion-controlled reactions[35] and random walk simulations of convective–diffusive transport
inside a fluidic structure.[25]The
Reynolds number is a dimensionless parameter that helps predict flow
patterns in different fluid flow situations and it is defined aswhere DH is the
hydraulic diameter that for channels with a high aspect ratio is equal
to twice the shorter axis of the channel, A is the
channel’s cross-sectional area, and ν is the kinematic
viscosity (ν = 10–3 m2·s–1 for water). Since the Reynolds number pertinent to
assumed geometries and the flow rates (Q = 0–100
μm3·s–1) varies between 0
and 2·10–7, which is far from the turbulent
regime (Re > 2000), only laminar flow can be considered
here. Therefore, for the quasi-two-dimensional (2D) configuration
(W ≫ H), we assume the axial
fluid velocity to bewhere z is the vertical coordinate,
which follows the 2D solution to the Navier–Stokes equations.
For the configuration where W ≥ H, a 2D velocity-field profile was calculated according to the analytical
solution of the 3D Navier–Stokes equations for channels with
a rectangular profile.[36]At time
zero, M0 molecules were randomly
placed inside the defined volume and their movement inside the fluidic
structure was recorded in time. At each time step, the position of
each molecule was updated through the addition of a convective and
diffusive component aswhere R, R, and R were calculated
separately at each time step as a normally distributed random number
with a mean of zero and a standard deviation of . The time
step for all simulations was
set to Δt = 10–5 s, determined
by convergence tests.The surface of NPs was discretized in
a way representing different
molecular receptor locations, i.e., into patches
with area Γ0–1. When the center of a molecule (a molecule of zero
volume is assumed) hits the surface of a plasmonic NP at a specific
point of time, it is decided based on the random number generation,
whether the molecule binds or further diffuses into the volume of
the fluidic structure. Specifically, if a random number R < Pon, where is the probability of the binding,
a molecule
binds to the receptor and the time of the binding event is recorded.
Further, in each time step it is decided if the molecule unbinds.
Specifically, if a random number R < Poff, where Poff = 1 –
exp(koffΔt) is
the probability of unbinding, a molecule unbinds and the time of the
unbinding event is also recorded. When the center of a molecule hits
the wall of a fluidic structure, it is reflected back in a specular
fashion. Thus, by repeating the process Msim times (Msim = 106 for a single
molecule present in 1 pL volume was used), we obtain characteristic
histograms of subsequent binding and unbinding events, n(t). The mean waiting time until the first binding
event, t1, is then calculated as the mean
value of the first binding events detected for all molecules and the
mean number of binding events collected over time is then calculated
as
Authors: Joachim Fritzsche; David Albinsson; Michael Fritzsche; Tomasz J Antosiewicz; Fredrik Westerlund; Christoph Langhammer Journal: Nano Lett Date: 2016-11-21 Impact factor: 11.189