Astigmatic Results of a Diffractive Trifocal Toric IOL Following Intraoperative Aberrometry Guidance.

John F Blaylock1, Brad Hall2.   

Abstract

PURPOSE: To determine if intraoperative aberrometry (IA) improves astigmatic outcomes for trifocal toric IOL (TTI) cases. PATIENTS AND METHODS: This was a retrospective study examining 137 eyes that underwent cataract extraction and TTI implantation using femtosecond laser, digital registration, and IA. Final cylinder power and axis of placement were determined by IA. Monocular uncorrected distance visual acuity (UDVA), uncorrected intermediate visual acuity (UIVA), uncorrected near visual acuity (UNVA), and refractive data were collected at 3 months. Postoperative residual astigmatism (PRA) determined by manifest refraction was compared to back-calculated residual astigmatism (BRA) using the cylinder power calculated preoperatively.
RESULTS: Postoperatively, 97.8% of eyes had IA PRA ≤ 0.50D and 80.3% had BRA ≤ 0.50 D, a difference of 17.5%. Mean PRA for IA was 0.07 D ± 0.19 (range 0.00-1.00 D) compared to BRA 0.31 D ± 0.33 (range 0.00-1.34 D) (P < 0.001). Cylinder power was changed in 50.4% of cases based upon IA. Postoperative mean UDVA (LogMAR) was 0.04 ± 0.09 (range -0.12-0.30 logMAR), and 65% of eyes were ≤ 0.0, 85% ≤ 0.1, and 99% ≤ 0.18.
CONCLUSION: The proportion of eyes with PRA ≤ 0.50 D and mean PRA was significantly lower using IA versus the preoperative planned cylinder power.
© 2020 Blaylock and Hall.

Entities:  

Keywords:  PanOptix; cataract surgery; toric IOL; trifocal IOL

Year:  2020        PMID: 33363357      PMCID: PMC7754259          DOI: 10.2147/OPTH.S285711

Source DB:  PubMed          Journal:  Clin Ophthalmol        ISSN: 1177-5467


Plain Language Summary

Cataract surgery involves removing the opaque natural lens of the eye and replacing it with an artificial intraocular lens. Many cataract surgery patients have astigmatism, which occurs when the front of the eye (cornea) is curved differently in one direction compared to the other (eg the cornea is shaped like a football instead of a soccer ball). Toric intraocular lenses (IOL) are used to correct astigmatism. By measuring several features of the eye, including eye length and the curvature of the cornea, a surgeon can select the best power for the toric IOL. Successful power selection often depends on the accuracy of these measurements. Appropriate power choice is crucial for reducing patient need for glasses to see clearly after surgery. An intraoperative aberrometer (IA) is a device that allows surgeons to perform measurements of the eye during cataract surgery and confirm or refine their power selection. This study was a chart review and was designed to compare results using an IA to theoretical results had the IA not been used. Results show significant benefits to using an IA during surgery to confirm or refine IOL power selection. There was less astigmatism for patients compared to theoretical results had the IA not been used, which provides clearer vision after surgery.

Introduction

Cataract surgery remains one of the most common surgical procedures performed worldwide. It is a refractive procedure, and patients increasingly expect to reduce their spectacle dependence following intraocular lens (IOL) implantation. Monofocal IOLs allow patients to reduce their spectacle dependence at distance, however patients may still require spectacles to see clearly at near and intermediate. Multifocal IOLs were developed to overcome the limitations of using only one focus. Initially, multifocal IOLs were primarily bifocal and provided adequate vision at near and distance, however patients were not always able to see clearly at intermediate.1 Trifocal IOLs are designed to split light into 3 foci (distance, intermediate, and near) and to provide good visual acuity at all ranges of vision. The AcrySof® PanOptix® IOL (Alcon Laboratories, Inc., Fort Worth, TX, USA) is currently the only trifocal IOL approved by the US FDA. Studies of this trifocal IOL have reported good visual acuity at all ranges of vision (distance, intermediate, and near),2,3 although, as with other multifocal IOLs, mild visual disturbances have also been reported.4,5 Postoperative astigmatism can cause suboptimal visual outcomes in patients with multifocal IOLs. Each diopter of postoperative astigmatism may decrease uncorrected distance visual acuity (UDVA) by 1.5 lines.6 Surgical approaches to correct astigmatism include corneal incisions, corneal refractive surgery, and toric IOL implantation.7–9 However, better visual outcomes with the use of toric IOLs have been reported.10 With toric IOL implantation, precise cylinder power selection and axis placement are necessary for optimal refractive outcomes. Intraoperative aberrometry (IA) can be used to measure aphakic and pseudophakic eyes during cataract surgery. Recommendations for spherical and cylindrical IOL power can be obtained by measuring the aphakic state, while examining the pseudophakic state assists the surgeon to determine if adjusting the orientation of the implanted IOL could improve refractive outcomes.11 The ORA System (Alcon Laboratories, Inc., Fort Worth, TX, USA), is commonly used and is reported to provide improved refractive outcomes compared to conventional preoperative planning.12 Studies have reported on refractive outcomes of the AcrySof® PanOptix® toric IOL.13,14 However, to our knowledge, there are few studies that have compared outcomes after trifocal toric implantation with and without IA. The purpose of this study is to determine if IA improves astigmatic outcomes for trifocal toric IOL cases and to compare to conventional preoperative planning.

Patients and Methods

This was a retrospective chart review of astigmatic outcomes at a single site (Valley Laser Eye Centre, Abbotsford, BC, Canada) of a diffractive trifocal toric IOL with FLACS, DT, and IA utilized during implantation. Patient chart data were de-identified. This study was conducted in compliance with Good Clinical Practice (GCP), International Harmonization (ICH) guidelines, and the tenets of the Declaration of Helsinki. An institutional review board (Advarra IRB, Columbia, MD) approved the study (Pro00035969). Patient informed consent was obtained for the use of chart data. Subjects were included if they presented with visually significant cataracts or as a candidate for clear lens extraction (clear lens extraction is off-label), showed interest in and were eligible for implantation of a diffractive toric IOL. Exclusion criteria were ocular comorbidity that may affect post-operative visual acuity, or irregular corneal astigmatism and keratoconus. Data records were used to identify subjects that fit the inclusion and exclusion criteria above. The chart review identified 137 eyes that fit the inclusion and exclusion criteria above, and all 137 eyes were included in the study. De-identified data included data from the preoperative and postoperative examinations. Preoperative and postoperative data included sex, age, refractive error, and visual acuity (UDVA, UIVA, UNVA, BCVA). Visual acuities were recorded in Snellen and converted to the equivalent log of the minimum angle of resolution (logMAR) notation for statistical analysis. A LenSx femtosecond laser (Alcon Laboratories, Inc., Fort Worth, TX, USA) was used for corneal incisions, capsulotomy, and lens fragmentation. Patients received preoperative biometry measurements using the IOL Master 500 (Carl Zeiss Meditec, Jena, Germany). Topography was performed with the Atlas 9000 (Carl Zeiss Meditec, Jena, Germany) and tomography was performed using the Pentacam (Oculus, Wetzlar, Germany). To determine IOL power preoperatively, biometry data was entered into the Barrett Universal II online calculator15 and the Alcon-Barrett Toric online calculator.16 The VERION System was used for image-guided digital tracking. The ORA System with Verifeye+ (Alcon Laboratories, Inc., Fort Worth, TX, USA) was used intraoperatively to determine spherical power, cylinder power, and axis of placement. The surgeon’s usual standard of care for patients implanted with advanced technology IOLs was the postoperative regimen. Acrysof IQ Panoptix toric IOL models were implanted (TFNT20, TFNT30, TFNT40, TFNT50, TFNT60; Alcon Laboratories, Inc., Fort Worth, TX, USA). Acrysof Panoptix IQ IOLs are designed as diffractive aspheric opticals.17 They are based on quadrifocal design, with focal points at near (40 cm), intermediate (60 cm), and distance (∞), but functionally is a trifocal IOL.18 The primary outcome measure of interest was the manifest refraction and specifically the percentage of eyes that were ≤0.5 D in their postoperative residual astigmatism (PRA) at the 3 months postoperative visit. Secondary endpoints included manifest refraction spherical equivalent (MRSE), and monocular uncorrected visual acuities at distance (UDVA; 6m), intermediate (UIVA; 60cm), and near (UNVA; 40cm) and best-corrected visual acuity (BCVA). The preoperative planned cylinder power was not always the same power as that suggested by IA. Where there were differences, the cylinder power suggested by IA was chosen. Back-calculations were performed to simulate the refractive error had the preoperative planned cylinder power and axis of placement been used. First described by Hill et al,19 back-calculations can simulate removal of the implanted IOL and replacement with the preoperative planned cylinder power and axis of placement. Residual astigmatism after IOL implantation is a combination of 4 vectors: preoperative corneal astigmatism, surgically induced astigmatism, random surgical effects, and the IOL. By mathematically replacing the IOL vector with another, back-calculations can determine the refractive error had the preoperative planned cylinder power and axis of placement been used. Back-calculations were performed by crossed vector analysis of two vectors. The primary vector was the absolute difference in power at the corneal plane between the implanted and suggested IOLs, and the preoperative steep axis, while the secondary vector was the postoperative cylinder power and axis. The primary and secondary vectors were calculated for each eye, input into the Eye Vectors iPhone Application Version 2.0 (Cambui Labs), and the resulting crossed vector result was recorded. Vector analysis was conducted after descriptive and other analytical statistics. Statistical analyses were performed using the software program R (version 3.6.3; The R Foundation for Statistical Computing, Vienna, Austria). Differences in continuous variables were compared using a paired t-test for two continuous variables or an analysis of variance (ANOVA) for 3 or more groups. In all cases, p ≤ 0.05 was considered significant.

Results

The chart review identified 137 eyes of 86 patients for inclusion in this study. Our study included approximately equal numbers of males (44) and females (42), who had an average age of 64.1 ± 7.8 (range 47–86). The demographic and preoperative data are summarized in Table 1. All patients were scheduled for FLACS, however two eyes were not able to receive FLACS due to insufficient pupil dilation, and were converted to conventional phacoemulsification.
Table 1

Preoperative Patient Demographics

ParametersMean ± SD (Range)
Number of eyes (patients)137 (86)
Post-Operative Visit (days)70.1 ± 29.3 (43–222)
Age (yr)64.1 ± 7.8 (47–86)
Sex
 Male (n)44
 Female (n)42
Axial Length (mm)24.23 ± 1.23 (21.74–28.29)
Cylinder (D)1.01 ± 0.50 (0.18–2.77)
Preoperative Patient Demographics A comparison of the distributions of postoperative absolute prediction error for IA and back-calculated absolute prediction error is shown in Figure 1. For IA, 83.9% (115/137) and 95.6% (131/137) of eyes had an absolute prediction error of 0.25 D and 0.5 D, respectively. Compared to IA, the percentage of patients with an absolute prediction error of 0.25 D and 0.5 D was lower for back-calculations using preoperative planned power: 72.3% (99/137) and 90.5% (124/137) respectively. Additionally, using the preoperative planned power could have resulted in 1 eye having MRSE greater than 1 D.
Figure 1

Postoperative refractive prediction errors in eyes using intraoperative aberrometry and preoperative calculations.

Postoperative refractive prediction errors in eyes using intraoperative aberrometry and preoperative calculations. Figure 2 shows the distribution of the PRA for IA compared to back-calculated residual astigmatism (BRA). It demonstrates that a higher percentage of IA eyes, 90.5% (124/137) and 97.8% (134/137), had PRA of 0.25 D and 0.50 D, respectively, than back-calculations using preoperative planned power. For the back-calculations, 45.0% (61/137) and 80.0% (109/137) had BCA 0.25 D and 0.50 D, respectively. Additionally, no IA eyes had postoperative astigmatism greater than 1.0 D, compared to 6.0% (8/137) in back-calculations using preoperative planned cylinder power.
Figure 2

Postoperative residual astigmatism in eyes using intraoperative aberrometry and preoperative calculations.

Postoperative residual astigmatism in eyes using intraoperative aberrometry and preoperative calculations. The postoperative monocular visual acuities are summarized in Table 2. Patients had good visual acuity at all ranges of vision—distance, intermediate, and near. Monocular UDVA, BCVA, UIVA, and UNVA were 0.10 logMAR or better in 84.7% (116/137), 98.5% (135/137), 99.3% (136/137), and 73.0% (100/137) of eyes, respectively. Almost all eyes had UDVA, BCVA, UIVA, and UNVA of 0.18 logMAR or better. Mean UDVA was 0.04 ± 0.09 logMAR (range −0.12–0.30 logMAR).
Table 2

Postoperative Monocular Visual Acuity

AcuityMean ± SD (Range) logMAR% 0.40 logMAR or Better% 0.30 logMAR or Better% 0.18 logMAR or Better% 0.10 logMAR or Better% 0.00 logMAR or Better% −0.12 logMAR or Better
UDVA0.04 ± 0.09 (−0.12–0.30)10010098.584.764.710.2
BCVA0.00 ± 0.06 (−0.12–0.18)10010010098.587.610.9
UIVA−0.07 ± 0.08 (−0.12–0.40)10099.399.399.394.964.2
UNVA0.04 ± 0.07 (0.00–0.30)10010098.590.573.00
Postoperative Monocular Visual Acuity A comparison of refractive outcomes between implanted eyes with the power suggested by IA to back-calculated refractive outcomes is shown in Table 3. The preoperative planned power and IA suggested different cylinder power in 50.4% (69/137) of cases. The PRA was significantly lower with IA (0.07 ± 0.19 D; range 0.00–1.00 D) compared to BRA (0.31 ± 0.33 D; range 0.00–1.34 D; paired t-test, P < 0.001). Postoperative MRSE was also significantly lower with IA (0.01 ± 0.27 D; range −0.88–1.00 D) compared to back-calculated MRSE (0.13 ± 0.27 D; range −0.71–1.17 D; paired t-test, P < 0.001). Mean absolute prediction error was significantly lower with IA, at 0.19 ± 0.19 D (range 0.00–0.96 D), compared to preoperative planned power (0.23 ± 0.23 D; range 0–1.09 D); paired t-test, (p=0.038).
Table 3

Postoperative Refractive Outcomes

ParametersIA Mean ± SD (Range)PreOP Planned Power Mean ± SD (Range)P value*
Residual Astigmatism (D)0.07 ± 0.19 (0.00–1.00)0.31 ± 0.33 (0.00–1.34)< 0.001
MRSE (D)0.01 ± 0.27 (−0.88–1.00)0.13 ± 0.27 (−0.71–1.17)< 0.001
Prediction Error0.19 ± 0.19 (0.00–0.96)0.23 ± 0.23 (0.00–1.09)0.038

Note: *Paired t-test.

Abbreviations: IA, intraoperative aberrometry; PreOP, preoperative; MRSE, manifest refraction spherical equivalent.

Postoperative Refractive Outcomes Note: *Paired t-test. Abbreviations: IA, intraoperative aberrometry; PreOP, preoperative; MRSE, manifest refraction spherical equivalent. Figure 3 shows double-angle vector plots for preoperative astigmatism, PRA, and BRA. The standard deviation for PRA and BRA, 0.2 D and 0.45 D, respectively, were lower compared to the preoperative astigmatism (1.08 D). The standard deviation for PRA was also lower than that of the BRA.
Figure 3

Double-angle vector plots of astigmatism vectors for (A) Preoperative, (B) Intraoperative Aberrometry, and (C) Preoperative Planned Power. Each ring represents 1 D. The diamond represents the centroid. N=137 eyes.

Double-angle vector plots of astigmatism vectors for (A) Preoperative, (B) Intraoperative Aberrometry, and (C) Preoperative Planned Power. Each ring represents 1 D. The diamond represents the centroid. N=137 eyes.

Discussion

We found that astigmatic outcomes were improved with IA for trifocal toric IOL cases and were better than the results of back-calculations using preoperative planned cylinder power. The primary outcome measure for this study was the percentage of eyes that were ≤0.5 D in their PRA. Our results appear as good or lower compared to previous reports of TTI with the Acrysof Panoptix IQ IOL.13,14,20,21 Carreño et al13 reported that approximately 66% and 94% of patients were ≤0.25 D and ≤0.5 D in their PRA, compared to 90.5% and 97.8% in our study. In both studies, the VERION and ORA systems were used. A possible explanation for the difference in percentage of eyes that were ≤0.25 D is that the authors reported using FLACS in 53% of eyes, while our study used FLACS in 98.5% percent of eyes. However, Carreño et al13 reported no significant differences between FLACS and standard surgery in their results. Other causes for the differences observed in our study could be the length of follow-up and preoperative refraction. It is interesting to note that the results reported by Carreño et al13 for percentage of eyes ≤0.25 D (66%) appear in between our results of BRA and PRA (45% and 90.5%, respectively). Mean postoperative cylinder after TTI with the Acrysof Panoptix IQ IOL has been reported previously between −0.21 and 0.28.13,14,20,21 Our results with IA 0.07 ± 0.19 D fall close to the middle of this range, while the results of our back-calculations using preoperative planned power, 0.31 ± 0.33 D fall outside it. Postoperative MRSE in our study (0.01 ± 0.27 D) fell close to the middle of the range for previously published results (−0.07–0.12 D)14,20,21 of TTI with the Acrysof Panoptix IQ IOL. Visual acuities in this study were similar to previous reports of implantation of the Acrysof Panoptix IQ toric IOL.14,20,21 Kohnen et al20 reported similar results of monocular UDVA, UIVA, and UNVA of 0.02, 0.12, and 0.10 logMAR respectively with the Acrysof Panoptix IQ toric IOL, compared UDVA, UIVA, and UNVA of 0.04, −0.07, and 0.04 logMAR reported in this study, though the sample size was smaller in their study. We found that our method of back-calculation provided a useful technique to compare refractive outcomes of implantations using the preoperative planned power and the power suggested by IA. A significant benefit of back-calculations is that they eliminate the need for multiple groups of patients (such as those that receive the preoperative planned power and those that receive the IA suggested power) because each patient eye provides data for each group. One drawback of our method of back-calculation is the need to manually input each vector for each crossed vector analysis into the application, which can be time consuming. Our postoperative results and back-calculations show that using IA improved refractive outcomes compared to preoperative planning. The prediction error was significantly higher with preoperative planning, as were the residual astigmatism and postoperative MRSE. A limitation of this study is that it was a retrospective study and not a prospective study. Additionally, the measured outcomes included refractive error and visual acuity, but other measures such as patient satisfaction could have been useful to report.

Conclusions

In summary, IA demonstrated improved accuracy of astigmatic outcomes in cases of trifocal toric implantation. Postoperatively, patients had excellent refractive outcomes and visual performance at all ranges.
  19 in total

1.  Comparison of visual and refractive outcomes of 2 trifocal intraocular lenses.

Authors:  Filomena J Ribeiro; Tiago B Ferreira
Journal:  J Cataract Refract Surg       Date:  2020-05       Impact factor: 3.351

2.  Comparison of bifocal and trifocal diffractive and refractive intraocular lenses using an optical bench.

Authors:  Damien Gatinel; Yvette Houbrechts
Journal:  J Cataract Refract Surg       Date:  2013-05-18       Impact factor: 3.351

3.  Limbal relaxing incisions with cataract surgery.

Authors:  K Budak; N J Friedman; D D Koch
Journal:  J Cataract Refract Surg       Date:  1998-04       Impact factor: 3.351

4.  A comparative study of the visual outcomes between a new trifocal and an extended depth of focus intraocular lens.

Authors:  Ramón Ruiz-Mesa; Antonio Abengózar-Vela; María Ruiz-Santos
Journal:  Eur J Ophthalmol       Date:  2017-08-09       Impact factor: 2.597

5.  Clinical outcomes with a diffractive trifocal intraocular lens.

Authors:  Jorge L Alió; Ana B Plaza-Puche; Jorge L Alió Del Barrio; Pedro Amat-Peral; Vicente Ortuño; Pilar Yébana; Olena Al-Shymali; Alfredo Vega-Estrada
Journal:  Eur J Ophthalmol       Date:  2018-04-05       Impact factor: 2.597

6.  Simulation of toric intraocular lens results: manual keratometry versus dual-zone automated keratometry from an integrated biometer.

Authors:  Warren Hill; Robert Osher; David Cooke; Kerry Solomon; Helga Sandoval; Rosario Salas-Cervantes; Richard Potvin
Journal:  J Cataract Refract Surg       Date:  2011-12       Impact factor: 3.351

Review 7.  Multifocal versus monofocal intraocular lenses after cataract extraction.

Authors:  Samantha R de Silva; Jennifer R Evans; Varo Kirthi; Mohammed Ziaei; Martin Leyland
Journal:  Cochrane Database Syst Rev       Date:  2016-12-12

Review 8.  Toric Intraocular Lenses in the Correction of Astigmatism During Cataract Surgery: A Systematic Review and Meta-analysis.

Authors:  Line Kessel; Jens Andresen; Britta Tendal; Ditte Erngaard; Per Flesner; Jesper Hjortdal
Journal:  Ophthalmology       Date:  2015-11-18       Impact factor: 12.079

9.  Laser in situ keratomileusis for correction of induced astigmatism from cataract surgery.

Authors:  Hamid Norouzi; Mohsen Rahmati-Kamel
Journal:  J Refract Surg       Date:  2003 Jul-Aug       Impact factor: 3.573

10.  Comparison of outcomes with multifocal intraocular lenses: a meta-analysis.

Authors:  Béatrice Cochener; Antoine Lafuma; Babak Khoshnood; Laurène Courouve; Gilles Berdeaux
Journal:  Clin Ophthalmol       Date:  2011-01-07
View more
  4 in total

1.  Refractive Outcomes Following Trifocal Intraocular Lens Implantation in Post-Myopic LASIK and PRK Eyes.

Authors:  John F Blaylock; Brad J Hall
Journal:  Clin Ophthalmol       Date:  2022-07-01

2.  Clinical Outcomes of Monofocal Toric IOLs Using Digital Tracking and Intraoperative Aberrometry.

Authors:  John F Blaylock; Brad J Hall
Journal:  Clin Ophthalmol       Date:  2021-08-26

3.  Patient Expectation, Satisfaction and Clinical Outcomes with a New Multifocal Intraocular Lens.

Authors:  Matthew Ison; Jane Scott; John Apel; Andrew Apel
Journal:  Clin Ophthalmol       Date:  2021-10-13

4.  Evaluation of Refractive Accuracy of ORA and the Factors Impacting Residual Astigmatism in Patients Implanted with Trifocal IOLs During Cataract Surgery: A Retrospective Observational Study.

Authors:  Keizo Watanabe
Journal:  Clin Ophthalmol       Date:  2022-08-10
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.