Abstract

BACKGROUND: There are concerns regarding glenoid erosion with metal shoulder hemiarthroplasty. PyroCarbon may offer an alternative because of favorable wear characteristics and preservation of the glenoid. The purpose of this study was to assess in vitro bone wear characteristics of PyroCarbon relative to cobalt chromium alloy hemiarthroplasty in a shoulder wear simulator.
METHODS: Wear of PyroCarbon and cobalt chromium prostheses articulating with bone were characterized by means of bone wear penetration rate, changes to surface roughness, and wear particle analysis.
RESULTS: PyroCarbon prostheses produced significantly less damage to bone and were less damaged by the bone than cobalt chromium prostheses. Cobalt chromium testing was halted at approximately 320,000 cycles because the bone was consumed. Wear testing of PyroCarbon specimens continued through five million cycles. Linearized bone penetration rate, bone volume loss rate, and surface roughness for cobalt chromium test specimens were 30 times greater than for PyroCarbon.
CONCLUSIONS: Results demonstrate significantly less damage to bone in simulated shoulder function testing for PyroCarbon hemiarthroplasty implants relative to conventional cobalt chromium implants. Our study supports use of PyroCarbon in humeral head hemiarthroplasty as a viable alternative to conventional metal hemiarthroplasty. Further investigation of PyroCarbon performance in clinical settings is warranted.
© 2018 The British Elbow & Shoulder Society.

Entities:  

Keywords:  PyroCarbon; cobalt chromium alloy; glenoid erosion; in vitro wear testing; shoulder hemiarthroplasty

Year:  2018        PMID: 33343712      PMCID: PMC7726179          DOI: 10.1177/1758573218796837

Source DB:  PubMed          Journal:  Shoulder Elbow        ISSN: 1758-5732


  61 in total

1.  Capitellar erosion caused by a metal radial head prosthesis. A case report.

Authors:  Roger P Van Riet; Francis Van Glabbeek; Olivier Verborgt; Jan Gielen
Journal:  J Bone Joint Surg Am       Date:  2004-05       Impact factor: 5.284

2.  Conformational and adsorptive characteristics of albumin affect interfacial protein boundary lubrication: from experimental to molecular dynamics simulation approaches.

Authors:  Hsu-Wei Fang; Man-Ching Hsieh; Huei-Ting Huang; Cheng-Yen Tsai; Min-Hui Chang
Journal:  Colloids Surf B Biointerfaces       Date:  2008-10-19       Impact factor: 5.268

3.  Bone compressive strength: the influence of density and strain rate.

Authors:  D R Carter; W C Hayes
Journal:  Science       Date:  1976-12-10       Impact factor: 47.728

4.  Surface-active phospholipid as the lubricating component of lubricin.

Authors:  I M Schwarz; B A Hills
Journal:  Br J Rheumatol       Date:  1998-01

5.  Oligolamellar nature of the articular surface.

Authors:  B A Hills
Journal:  J Rheumatol       Date:  1990-03       Impact factor: 4.666

6.  Radial head replacement with the MoPyC pyrocarbon prosthesis.

Authors:  Ioannis K Sarris; Margaritis J Kyrkos; Nikiforos N Galanis; Kyriakos A Papavasiliou; Fares E Sayegh; George A Kapetanos
Journal:  J Shoulder Elbow Surg       Date:  2012-02-17       Impact factor: 3.019

7.  Surface replacement arthroplasty for glenohumeral arthropathy in patients aged younger than fifty years: results after a minimum ten-year follow-up.

Authors:  Ofer Levy; Oren Tsvieli; Julia Merchant; Lora Young; Alberto Trimarchi; Rupen Dattani; Ruben Abraham; Stephen A Copeland; Ali Narvani; Ehud Atoun
Journal:  J Shoulder Elbow Surg       Date:  2015-01-16       Impact factor: 3.019

8.  Glenoid component failure in total shoulder arthroplasty.

Authors:  Frederick A Matsen; Jeremiah Clinton; Joseph Lynch; Alexander Bertelsen; Michael L Richardson
Journal:  J Bone Joint Surg Am       Date:  2008-04       Impact factor: 5.284

9.  Cementless surface replacement arthroplasty of the shoulder with biologic resurfacing of the glenoid.

Authors:  Keng Thiam Lee; Simon Bell; John Salmon
Journal:  J Shoulder Elbow Surg       Date:  2009-03-27       Impact factor: 3.019

10.  Hemiarthroplasty versus total shoulder prosthesis: results of cemented glenoid components.

Authors:  M Pfahler; F Jena; L Neyton; F Sirveaux; D Molé
Journal:  J Shoulder Elbow Surg       Date:  2006 Mar-Apr       Impact factor: 3.019

View more
  4 in total

1.  Survivorship of anatomic press-fit short-stem radial head replacement with a pyrocarbon bearing.

Authors:  Aparna I Viswanath; Adam C Watts
Journal:  Shoulder Elbow       Date:  2021-06-17

Review 2.  Innovations in Shoulder Arthroplasty.

Authors:  Nels Leafblad; Elise Asghar; Robert Z Tashjian
Journal:  J Clin Med       Date:  2022-05-16       Impact factor: 4.964

3.  Pyrocarbon Humeral Head in a Shoulder Hemiarthroplasty: Preliminary Results at 3 Years Follow-Up and Review of the Current Literature.

Authors:  Eva Campos-Pereira; Luís Henrique-Barros; Rui Claro
Journal:  Case Rep Orthop       Date:  2021-04-14

4.  Midterm results of pyrocarbon interposition shoulder arthroplasty: good outcomes after posttraumatic osteonecrosis without malunion of the tuberosities.

Authors:  Jérôme Garret; Arnaud Godenèche; Pascal Boileau; Daniel Molé; Mikael Etzner; Luc Favard; Christophe Lévigne; François Sirveaux; Gilles Walch
Journal:  JSES Int       Date:  2022-06-23
  4 in total