Procedure for optimal implementation of automatic tube potential selection in pediatric CT to reduce radiation dose and improve workflow.

Jacinta E Browne1, Michael R Bruesewitz1, Vrieze Thomas1, Kristen B Thomas1, Nathan C Hull1, Cynthia H McCollough1, Lifeng Yu1.   

Abstract

It is important to employ radiation dose reduction techniques in pediatric computed tomography (CT) to reduce potential risks of radiation-induced malignancy. Automatic tube potential (kV) selection tools have been developed and become available on many CT scanners, which select the optimum kV based on the patient size and clinical task to improve the radiation dose efficiency. However, its use in pediatric CT has been mostly empirical, following manufacturer's default recommendation without solid demonstration for quality improvement. This study aimed to implement an automatic tube potential tool (CAREkV, Siemens Healthcare) into routine pediatric CT practice, using the "Plan-Do-Study-Act" quality improvement process, in place of an existing kV/mAs technique chart. The design of this quality improvement project involved Plan-Do-Study-Act stages. Plan and Do stages identified the criteria for optimal automatic kV selection; a range of phantoms representing typical pediatric groups were scanned on a dual-source 128-slice scanner using a fast-pitch scanning mode. The identified CAREkV settings were implemented into the CT protocol and evaluated after a 6-month period. In the Study stage, an objective evaluation of the image metrics and radiation dose for two similar patient cohorts using CAREkV and the technique-chart, respectively, were compared. The kV selected, image quality and radiation dose determined by CAREkV were comparable to those obtained while using the technique-chart. The CAREkV was successfully implemented into our pediatric abdominopelvic CT practice. By utilizing the "PDSA" process optimal image quality and radiation dose reduction were achieved with an automatic kV selection tool to improve CT workflow.
© 2020 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, Inc. on behalf of American Association of Physicists in Medicine.

Entities:  

Keywords:  automatic tube potential; image quality; pediatric abdominopelvic CT; radiation dose reduction; ‘Plan-Do-Study-Act’ quality improvement

Mesh:

Year:  2020        PMID: 33338314      PMCID: PMC7882104          DOI: 10.1002/acm2.13098

Source DB:  PubMed          Journal:  J Appl Clin Med Phys        ISSN: 1526-9914            Impact factor:   2.102


  21 in total

1.  Dose reduction in pediatric CT: a rational approach.

Authors:  John M Boone; Estella M Geraghty; J Anthony Seibert; Sandra L Wootton-Gorges
Journal:  Radiology       Date:  2003-08       Impact factor: 11.105

2.  Radiation dose reduction in computed tomography: techniques and future perspective.

Authors:  Lifeng Yu; Xin Liu; Shuai Leng; James M Kofler; Juan C Ramirez-Giraldo; Mingliang Qu; Jodie Christner; Joel G Fletcher; Cynthia H McCollough
Journal:  Imaging Med       Date:  2009-10

3.  A three-dimensional statistical approach to improved image quality for multislice helical CT.

Authors:  Jean-Baptiste Thibault; Ken D Sauer; Charles A Bouman; Jiang Hsieh
Journal:  Med Phys       Date:  2007-11       Impact factor: 4.071

4.  Application- and patient size-dependent optimization of x-ray spectra for CT.

Authors:  Willi A Kalender; Paul Deak; Markus Kellermeier; Marcel van Straten; Sabrina V Vollmar
Journal:  Med Phys       Date:  2009-03       Impact factor: 4.071

5.  Automatic selection of tube potential for radiation dose reduction in CT: a general strategy.

Authors:  Lifeng Yu; Hua Li; Joel G Fletcher; Cynthia H McCollough
Journal:  Med Phys       Date:  2010-01       Impact factor: 4.071

Review 6.  Optimal tube potential for radiation dose reduction in pediatric CT: principles, clinical implementations, and pitfalls.

Authors:  Lifeng Yu; Michael R Bruesewitz; Kristen B Thomas; Joel G Fletcher; James M Kofler; Cynthia H McCollough
Journal:  Radiographics       Date:  2011 May-Jun       Impact factor: 5.333

Review 7.  A review of patient dose and optimisation methods in adult and paediatric CT scanning.

Authors:  E Dougeni; K Faulkner; G Panayiotakis
Journal:  Eur J Radiol       Date:  2011-06-17       Impact factor: 3.528

Review 8.  Strategies for CT radiation dose optimization.

Authors:  Mannudeep K Kalra; Michael M Maher; Thomas L Toth; Leena M Hamberg; Michael A Blake; Jo-Anne Shepard; Sanjay Saini
Journal:  Radiology       Date:  2004-01-22       Impact factor: 11.105

9.  Radiation Dose Reduction in Pediatric Body CT Using Iterative Reconstruction and a Novel Image-Based Denoising Method.

Authors:  Lifeng Yu; Joel G Fletcher; Maria Shiung; Kristen B Thomas; Jane M Matsumoto; Shannon N Zingula; Cynthia H McCollough
Journal:  AJR Am J Roentgenol       Date:  2015-11       Impact factor: 3.959

10.  Evaluation of AAPM Reports 204 and 220: Estimation of effective diameter, water-equivalent diameter, and ellipticity ratios for chest, abdomen, pelvis, and head CT scans.

Authors:  Christiane S Burton; Timothy P Szczykutowicz
Journal:  J Appl Clin Med Phys       Date:  2017-11-27       Impact factor: 2.102