| Literature DB >> 33331811 |
Cécile Emeraud1,2,3,4, Seher Yilmaz1, Nicolas Fortineau1, Gaëlle Cuzon1, Laurent Dortet2,1,4,3.
Abstract
Introduction. Blood culture (BC) remains the gold standard for the diagnosis of bloodstream infection. Clinical microbiology laboratories must ensure the quality of their BC process from receipt to definitive results.Aim. In this study, we followed the evolution of different quality indicators for BCs over the first year of implementation of the BacT/Alert Virtuo system in a French hospital.Methodology. In our laboratory, we instituted regular monitoring of several quality indicators to track (i) delays in sample registration, (ii) delays in loading BC bottles in our incubating system (BacT/Alert Virtuo) after registration, (iii) the volume of blood in bottles and (iv) the contamination rates.Results. For 53 892 BC bottles loaded in the BacT/Alert Virtuo from 23 January to 31 December 2019, the delays in sample registration, loading and unloading were respectively 3.5 h±0.016, 44 min±0.209 and 5.8 h±0.0727. Intriguingly, the automated process performed by the BacT/Alert Virtuo system to check the blood volume in bottles was only performed for 60 % of the loaded bottles. Among these, 30 % contained the recommended volume of blood (between 7 and 13 ml). Finally, the contamination rate was found to be 27.2 % for samples at our institution.Conclusions. The delays in sample registration, loading and unloading were found to be acceptable, even though they could be improved by ensuring a continuous service during the night duty period. Furthermore, the percentage of volumes measured is insufficient and must be improved and the majority of bottles do not contain the recommended blood volume.Entities:
Keywords: BacT/AlertVirtuo; accreditation; blood culture; quality indicators
Mesh:
Year: 2020 PMID: 33331811 PMCID: PMC8346724 DOI: 10.1099/jmm.0.001300
Source DB: PubMed Journal: J Med Microbiol ISSN: 0022-2615 Impact factor: 2.472
Monitoring of quality indicators for blood cultures for 2019
|
No. of bottles |
Delay (mean) |
No. of positives |
Virtuo false-positivity rate | ||||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
|
Total |
FA |
FN |
PF |
Reception |
Loading in Virtuo |
Unloading in Virtuo |
No. of false positives* |
No. of false positives*/ total (%) |
No. of false positives*/ total positives (%) |
Measured bottles (FN, FA) |
Low volumes <7 ml |
High volumes >13 ml |
Conforming volumes† (among measured) |
Non-conforming volumes† (among measured) |
Conforming volumes † (total) |
Non-conforming volumes † (total) | |||
|
|
|
1436 |
688 |
686 |
62 |
4.3 h |
40 min |
7.35 h |
148 |
0 |
625 |
419 |
35 |
171 |
454 |
171 |
454 | ||
|
% |
|
|
|
|
|
|
|
|
|
|
|
|
| ||||||
|
|
|
5159 |
2485 |
2475 |
199 |
3.7 h |
43 min |
5.5 h |
524 |
2 |
3277 |
2117 |
280 |
880 |
2397 |
880 |
2397 | ||
|
|
|
|
|
10 |
|
|
|
|
|
|
|
|
| ||||||
|
|
|
5300 |
2528 |
2529 |
243 |
3.3 h |
45 min |
5.8 h |
576 |
3 |
3367 |
2212 |
246 |
909 |
2458 |
909 |
2458 | ||
|
% |
|
|
|
11 |
|
|
|
|
|
|
|
|
| ||||||
|
|
|
5009 |
2390 |
2388 |
231 |
3.6 h |
49 min |
5.8 h |
538 |
4 |
3006 |
2007 |
244 |
755 |
2251 |
755 |
2251 | ||
|
% |
|
|
|
11 |
|
|
|
|
|
|
|
|
| ||||||
|
|
|
4827 |
2310 |
2307 |
210 |
3.9 h |
47 min |
6.0 h |
432 |
3 |
2827 |
1759 |
284 |
784 |
2043 |
784 |
2043 | ||
|
% |
|
|
|
9 |
|
|
|
|
|
|
|
|
| ||||||
|
|
|
4843 |
2304 |
2299 |
240 |
3.9 h |
47 min |
5.3 h |
542 |
2 |
2891 |
1631 |
317 |
943 |
1948 |
943 |
1948 | ||
|
% |
|
|
|
11 |
|
|
|
|
|
|
|
|
| ||||||
|
|
|
4739 |
2270 |
2270 |
199 |
3.1 h |
39 min |
5.8 h |
556 |
0 |
2426 |
1494 |
210 |
722 |
1704 |
722 |
1704 | ||
|
% |
|
|
|
12 |
|
|
|
|
|
|
|
|
| ||||||
|
|
|
3881 |
1827 |
1820 |
234 |
2.7 h |
37 min |
5.8 h |
474 |
0 |
2223 |
1351 |
190 |
682 |
1541 |
682 |
1541 | ||
|
% |
|
|
|
12 |
|
|
|
|
|
|
|
|
| ||||||
|
|
|
4367 |
2067 |
2071 |
229 |
3.1 h |
44 min |
4.2 h |
457 |
5 |
|
|
|
|
|
|
| ||
|
% |
|
|
|
10 |
|
|
|
|
|
|
|
|
| ||||||
|
|
|
4926 |
2327 |
2326 |
273 |
3.6 h |
43 min |
6.0 h |
528 |
9 |
3416 |
2096 |
354 |
966 |
2450 |
966 |
2450 | ||
|
% |
|
|
|
11 |
|
|
|
|
|
|
|
|
| ||||||
|
|
|
4837 |
2288 |
2289 |
260 |
3.8 h |
48 min |
6.3 h |
493 |
2 |
3321 |
1991 |
354 |
976 |
2345 |
976 |
2345 | ||
|
% |
|
|
|
10 |
|
|
|
|
|
|
|
|
| ||||||
|
|
|
4568 |
2144 |
2144 |
280 |
3.6 h |
44 min |
6.4 h |
504 |
5 |
2954 |
1760 |
337 |
857 |
2097 |
857 |
2097 | ||
|
% |
|
|
|
11 |
|
|
|
|
|
|
|
|
| ||||||
|
|
|
53 892 |
25 628 |
25 604 |
|
3.5 h |
44 min |
5.8 h |
5772 |
35 |
33 060 |
20 506 |
3121 |
9433 |
23 627 |
9433 |
23 627 | ||
|
% |
|
|
|
11 |
|
|
|
|
|
|
|
|
| ||||||
∗ False positive, bottles with positive signal with BacT/Alert Virtuo and negative culture.
†Conforming volumes, volumes between 7 and 13 ml.
nd:,not determined; FA, aerobic bottles; FN, anaerobic bottles; PF, paediatric bottles.
Fig. 1.Quality indicators for blood culture bottles using the BacT/Alert Virtuo over 1 year: (a) delay from sampling to registration (=routing delay), (b) delay from sample registration to loading (=loading delay), (c) delay to unloading positive bottle (=unloading delay) and (d) percentage and distribution of measured blood volumes in bottles.
Fig. 2.Blood culture contamination over 1 year: (a) identification of germs grown in positive blood culture samples and (b) rate and identification of contaminants in positive blood cultures samples.