The Effect of Irradiance on the Degree of Conversion and Volumetric Polymerization Shrinkage of Different Bulk-Fill Resin-Based Composites: An In Vitro Study.

Abrar N Bin Nooh1, Hend Al Nahedh1, Mohammad AlRefeai1, Fahad Alkhudhairy1.   

Abstract

OBJECTIVE: The influence of different light-emitting diode (LED) curing light intensities on the degree of conversion (DC) and volumetric polymerization shrinkage (VPS) of bulk-fill resin-based composite (RBC) restorative materials was evaluated.
MATERIALS AND METHODS: Twenty-four specimens of each RBC material (Filtek one bulk-fill posterior, Reveal HD Bulk, Tetric N-Ceram, and Filtek Z350) were prepared. The RBCs were shaped in molds and cured using an LED curing light unit at high-intensity (1,200 mW/cm2) for 20 seconds and low-intensity (650 mW/cm2) for 40 seconds Fourier-transform infrared (FTIR) spectroscopy was used to determine the DC and microcomputed tomography was used to evaluate VPS. Data were analyzed using one- and two-way ANOVA, independent t-test, and Tukey's and Scheffe's post hoc multiple comparison tests.
RESULTS: With high-intensity curing light, Reveal HD showed the highest DC (85.689 ± 6.811%) and Tetric N-Ceram the lowest (52.60 ± 9.38%). There was no statistical difference in VPS when using high- or low-intensity curing light. The highest VPS was observed for Reveal HD (2.834-3.193%); there was no statistical difference (p > 0.05) among the other RBCs.
CONCLUSION: Curing light intensities do not significantly influence the VPS of RBC materials. Reveal HD bulk cured with high-intensity light had the highest DC. European Journal of Dentistry. This is an open access article published by Thieme under the terms of the Creative Commons Attribution-NonDerivative-NonCommercial-License, permitting copying and reproduction so long as the original work is given appropriate credit. Contents may not be used for commercial purposes, or adapted, remixed, transformed or built upon. (https://creativecommons.org/licenses/by-nc-nd/4.0/).

Entities:  

Year:  2020        PMID: 33321544      PMCID: PMC8184305          DOI: 10.1055/s-0040-1721236

Source DB:  PubMed          Journal:  Eur J Dent


Introduction

Dental curing light units used to polymerize restorative materials revolutionized modern dentistry. Different light sources (e.g., light emitting diodes [LEDs] and tungsten halogen) are available for photopolymerization in dentistry applications that require light sources that are safe, efficient, promote fast curing, and compatible with dental materials. Resin-based composite (RBC) materials are now widely used as restorative materials due to increasing esthetic demands. The RBCs are composed of organic resin monomers, inorganic fillers, photoinitiators, and silane coupling agents to bind the fillers to the matrix. Bisphenol A-glycidyl methacrylate (Bis-GMA) has historically been used as the base monomer in commercial dental composites, though other dimethacrylates (e.g., triethylene glycol dimethacrylate [TEGDMA]) have been added due to its high viscosity. 1 Photoinitiators in RBCs trigger polymerization, with the most common photoinitiator system being camphorquinone (CQ), which is accelerated by a tertiary amine. Some commercial formulations include alternative photoinitiators (e.g., Ivocerin in Tetric N-Ceram) that are more color stable and less yellow than CQ. 1 2 Attempts by manufacturers to improve RBCs have simplified and facilitated rapid placement of large restorations, known as bulk-fill RBCs. These bulk-fill composites provide sufficient polymerization with a 4-mm bulk placement in a single layer. In addition, manufacturers employ a variety of methods to increase the depth of cure (DOC) of bulk-fill RBCs, such as increasing the filler particles size, reducing the filler content, and using an additional photoinitiator. 2 The degree of conversion (DC) is an important indication of the mechanical performance of RBCs, as adequate polymerization results in enhanced mechanical and physical properties. 3 4 During photopolymerization, monomers are transformed into complex polymer chains, although not all monomers will be converted into polymer structures, resulting in some unreacted monomers remaining. The polymerization process begins with light absorption in a specific wavelength range by the RBC; the activation process and reaction occurs with the aliphatic amine present in the composite to produce free radicals. The carboncarbon double bonds present in the monomers are converted into single bonds upon polymer chain formation during the process of polymerization, and this percentage of carboncarbon double bond to single bond conversion is referred to as the DC (%). 5 Multiple techniques are used to measure the DC, including indirect methods such as microhardness testing using Knoop or Vickers indenters, 6 or by employing a scraping method. However, it has been noted that these techniques overestimate the DOC values, and they also suffer from being difficult to standardize. 7 8 Vibrational spectroscopy is a direct method for measuring unreacted carboncarbon double bond and converted single bond percentages in cured materials. While Fourier-transform infrared (FTIR) spectroscopy based on light absorption or Raman spectroscopy based on light scattering can be used, 9 FTIR spectroscopy is the most common technique employed, as it provides reliable results when measuring the DC. 10 Volumetric polymerization shrinkage (VPS) is one major disadvantage of RBC restorations. Carbon monomers form polymer chains during light irradiation, resulting in 2 to 3 vol% shrinkage. 11 This VPS can lead to inner stresses generated at the resin-tooth interface and cause pain, loss of marginal sealing due to interfacial gaps, and microleakage, which may evolve secondary caries and lead to restoration failure. 12 Many factors can lead to VPS of RBC materials, such as filler load, filler particles, monomer system, photoinitiators, and light-curing units. The resin matrix composition and filler content determine the extent of VPS. 13 Several methods and devices have been utilized to measure polymerization shrinkage in terms of linear and volumetric shrinkage and cuspal displacements. Indirect techniques such as finite element analysis, microleakage assessment, and three-dimensional (3D) micro-computed tomography (μ-CT) have been used. The μ-CT method is a nondestructive technique that results in a 3D image 14 that has been effectively employed in the assessment of real volumetric shrinkage of RBCs. 15 With the development of light-curing devices, dentists can cure composites faster by increasing the light intensity to reduce the exposure time, as justified by the Bunsen–Roscoe law (BRL) of reciprocity, 16 although there are many debates in the literature concerning the validity of BRL. Currently, it is generally agreed that RBC materials require exposure to the radiation of particular wavelengths for a specific duration of time to yield acceptable results. 17 The aim of this study was to evaluate the effects of different LED curing light intensities on the DC and VPS of bulk-fill RBC restorative materials. The hypothesis tested was that no statistically significant effect on the DC and VPS of different bulk-fill RBCs would occur from applying curing light at different intensities.

Materials and Methods

Specimen Preparation

A total of 96 specimens consisting of 24 disc-shaped specimens of each Filtek one bulk-fill posterior, Reveal HD Bulk, Tetric N-Ceram, and Filtek Z350 (control group) RBC material were fabricated, and shade A2 was selected for all the composites except for Tetric N-Ceram, in which shade IVA was used (the material compositions and manufacturers are listed in Table 1 ). Each set of specimens was divided into four groups: two groups were used to measure the DC and the other two groups were used to measure the VPS. Curing with a Bluephase N light-curing unit was investigated at two intensities: high-intensity output (1,200 mW/cm 2 ) for 20 seconds and low-intensity output (650 mW/cm 2 ) for 40 seconds. The power intensity was measured using a dental Bluephase radiometer (Ivoclar Vivadent). For each group, six specimens were used to measure DC and VPS at each curing light intensity.
Table 1

Composition of the materials tested as provided by the manufacturer

MaterialManufacturerComposition
Abbreviations: AUDMA, aromatic urethane dimethacrylate; BisGMA, bisphenol A-glycidyl methacrylate; UDMA, urethane dimethacrylate.
Filtek one bulk-fill posterior3M ESPEMonomer matrix: AUDMA, addition-fragmentation monomer (dynamic stress-relieving monomer), 1,12-diaminododecane (dimethacrylate), and UDMAFillers: a combination of a nonagglomerated/nonaggregated 20 nm silica filler, a nonagglomerated/nonaggregated 4 to 11 nm zirconia filler, an aggregated zirconia/silica cluster filler (comprised of 20 nm silica and 4 to 11 nm zirconia particles), and an ytterbium trifluoride filler consisting of agglomerate 100 nm particles. The inorganic filler loading is ~76.5 wt% (58.4 vol%)
Reveal HD Bulk FillBISCOMonomer matrix: UDMA and BisGMAFiller: ytterbium fluoride
Tetric N-CeramIvoclar Vivadent,Schaan, LiechtensteinMonomer matrix: dimethacrylates.Filler: barium glass, ytterbium trifluoride pre-polymer, and mixed oxides. Filler content is 75 to 77 wt% (53–55 vol%)
Filtek Z350 XT (control)3M ESPEMonomer matrix: Bis-GMA, UDMA, TEGDMA, and bis-EMA.Filler: Combination of nonagglomerated/nonaggregated 20 nm silica filler, nonagglomerated/nonaggregated 4 to 11 nmzirconia filler, and aggregated zirconia/silica cluster filler (comprised of 20 nm silica and 4–11 nm zirconia particles). The inorganic filler loading is ~72.5 wt% (55.6 vol%)
A special custom Teflon mold for the VPS samples and a two-part brass mold for DC samples (10 mm diameter and 4 mm depth) were used for Filtek one bulk-fill posterior, Reveal HD Bulk, and Tetric N-Ceram measurements, and different size molds (10 mm diameter, 2 mm depth) were used for the Filtek Z350 control group. After the materials were placed in the molds, a clear Mylar strip (Mylar Uni-strip, Caulk/Dentsply, Milford, Delaware, United States) and a 1-mm thick glass plate were secured over each mold to flatten the surface and then gently pressed to remove excess material on the mold.

Degree of Conversion

Twelve samples of each RBC were prepared for DC measurements. The mold was placed on a dark nonreflective surface for measurements before and after polymerization. The absorbance and transmission peaks were obtained using the reflectance mode of the FTIR spectrometer (Thermo Scientific, NICOLET iS10, United States). The DC (%) were determined from the ratio of the absorbance intensities of the aliphatic C=C peak (1,638 cm -1 ) against the internal reference aromatic C=C peak (1,608 cm -1 ). The DC was determined according to the following equation:

Micro-Computed Tomography Analysis

Twelve samples were prepared from each composite. The mold was positioned inside the μ-CT chamber, and each sample was scanned for 1 hour pre- and postpolymerization using a high-resolution 3D X-ray microscope (Bruker SkyScan 1172, Kontich, Belgium). The μ-CT projection images were acquired with 88 kV voltage, 114 µA anode current, 316 milliseconds exposure time, 15.89 µm image pixel size, Al + Cu filter, 0.4 rotation step for 360 degrees, frame averaging of four for improved signal-to-noise ratio, and eight random movements to minimize ring artifacts. A flat-field correction was performed before the scanning procedure to account for variations in camera pixel sensitivity. The μ-CT data were evaluated with DataViewer software (version 1.5.6.2, Bruker SkyScan, Kontich, Belgium) after image reconstruction. The pre- and postpolymerization scans were analyzed using the 3D analysis tool to give a volume output value using CTAn software (Bruker Skyscan, Kontich, Belgium). Finally, CTVol software (version 2.3.2.0, Bruker SkyScan, Kontich, Belgium) was used for 3D visualization and production of color-coded images of the samples.

Statistical Analysis

Data are expressed as the mean and standard deviation determined using SPSS Statistics software (version 22.0; IBM SPSS, Armonk, New York, United States). The results were first tested for normality using the Shapiro–Wilk test. The results were then analyzed using two-way analysis of variance (ANOVA) followed by independent t -test and one-way ANOVA. Comparisons between the individual groups were made with Tukey’s or Scheffe’s post hoc test to determine the significant differences between the materials and curing light intensities. All tests were performed at a significance level of p <0.05.

Results

A significant effect associated with curing light intensity (F = 6.09, p < 0.01) and type of RBC (F = 22.05, p < 0.000) was observed when determining the DC ( Table 2 for the two-way ANOVA test results). However, the interaction of RBC and curing light intensity was not significant (F = 1.07, p < 0.37) (one-way ANOVA results of the materials for each light-cure intensity are listed in Table 3 ). When applying high-intensity curing light, the DC of Reveal HD Bulk was the highest (85.689 ± 6.811%), followed by Filtek One (74.92 ± 3.58%), Filtek Z350 (69.15 ± 10.86%), and Tetric N-Ceram (55.561 ± 4.898%). There was a significant difference between one of the material F-test with p <0.00. To identify the material, we did the multiple comparison test (Tukey) which showed that Reveal HD Bulk with 85.689 ± 6.811% DC was more significant than Filtek Z350 with 69.152 ± 10.865% DC at p <0.003 and Tetric N-Ceram with 55.561 ± 4.898% DC at p <0.00. Tetric N-Ceram exhibited significantly less DC than Filtek one bulk fill posterior at p <0.001 and Filtek Z350 at p <0.017.
Table 2

Two-way analysis of variance

SourceType III sum of squaresdfMean squareF-ratio p -Value
Abbreviations: RBC, resin-based composite.
Curing light intensity384.4391384.4396.0990.018
RBC4170.51731390.17222.054<0.000
Curing light intensity × RBC202.679367.5601.0720.372
Error2521.3744063.034
Total232512.48848
Corrected total7279.00847
Table 3

One-way analysis of variance and multiple comparison test (Tukey)

Curing light intensityRBC materialMean (%)SD F test p -Value 95% confidence interval for meanMultiple comparison test (Tukey)
Lower boundUpper boundFiltek Z350Filtek OneTetricN-CeramRevealHD
Abbreviations: RBC, resin-based composite; SD, standard deviation.
HighFiltek Z35069.15210.865<0.00057.75080.5531
Filtek One74.9223.58771.15778.6860.5091
Tetric N-Ceram55.5614.89850.42160.7020.0170.0011
Reveal HD85.6896.81178.54192.8360.0030.0700.0001
LowFiltek Z35065.74613.077<0.00252.02379.4691
Filtek One71.4185.50265.64477.1910.6771
Tetric N-Ceram52.6019.38542.75262.4490.0720.0061
Reveal HD72.9193.69469.04276.7960.4980.9900.0031
Application of low-intensity curing light resulted in Reveal HD bulk fill again having the highest DC (72.91 ± 3.69%) among all the RBCs, followed by Filtek one bulk fill posterior (71.41 ± 5.50%), Filtek Z350 (65.74 ± 13.07%), and Tetric N-Ceram (52.60 ± 9.38%). Tetric N-Ceram was significantly different from Reveal HD bulk fill at p < 0.003 and Filtek One at p <0.006. The effect of curing light intensity for each material is listed in Table 4 . It is clear that the only material that exhibits a significant difference when using different curing light intensities is Reveal HD bulk fill. With high-intensity curing light, Reveal HD exhibited a greater DC (85.689 ± 6.811%) compared with when low-intensity curing light was applied (72.919 ± 3.694%; p < 0.002).
Table 4

Descriptive t -test analysis

RBC materialCuring light intensityMean (%)SD95% confidence interval for mean t -test p -Value
Lower boundUpper bound
Abbreviations: RBC, resin-based composite; SD, standard deviation.
Filtek Z350High69.15210.86557.75080.5530.634
Low65.74613.07752.02379.469
Filtek OneHigh74.9223.58771.15778.6860.220
Low71.4185.50265.64477.191
Tetric N-CeramHigh55.5614.89850.42160.7020.509
Low52.6019.38542.75262.449
Reveal HDHigh85.6896.81178.54192.8360.002
Low72.9193.69469.04276.796

Volumetric Polymerization Shrinkage

VPS was investigated next to determine whether curing light intensity had a significant effect. Two-way ANOVA ( Table 5 ) indicates a significant effect associated with the RBCs (F = 16.41, p < 0.000), whereas the curing light intensity (F = 0.001, p < 0.97) and the interaction of the curing light with RBCs (F = 1.80, p < 0.16) is not significant for VPS.
Table 5

analysis of variance related to volumetric polymerization shrinkage

SourceType III sum of squaresdfMean squareF-ratio p -Value
Abbreviation: RBC, resin-based composite.
Light cure intensity0.00010.0000.0010.972
RBC4.94531.64816.412<0.000
Light cure intensity × RBC0.54330.1811.8040.162
Error4.018400.100
Total308.30348
Corrected total9.50647
The VPS of RBC materials was compared at low and high intensity; the results of which are listed in Table 6 . When using the high-intensity curing light, Reveal HD Bulk (3.19 ± 0.59%) had the highest VPS among all the materials, followed by Tetric N-Ceram (2.42 ± 0.24%), Filtek Z350 (2.25 ± 0.18%), and Filtek one bulk fill posterior (2.09 ± 0.17%).
Table 6

One-way analysis of variance and multiple comparison test (Tukey)

Curing light intensityRBC materialMean (%)SD F-test p -value 95% confidence interval for meanMultiple comparison test (Tukey)
Lower boundUpper boundFiltek Z350Filtek OneTetricN-CeramReveal HD
Abbreviations: RBC, resin-based composite; SD, standard deviation.
High intensityFiltek Z3502.2540.185<0.0002.0592.4491
Filtek One2.0980.1761.9132.2820.8631Tetric N-Ceram2.429
0.2422.1752.6820.8190.3741Reveal HD3.1930.5992.565
3.8210.0010.0000.0061Low intensityFiltek Z3502.1960.352<0.0082.075
2.8131Filtek One2.1960.2191.966
2.4260.4461Tetric N-Ceram2.5120.3172.180
2.8440.9750.2461Reveal HD2.8340.2222.601
3.0680.1130.0050.2331Abbreviations: RBC, resin-based composite; SD, standard deviation.
There was a significant difference between the materials (F-test p < 0.00) under high-intensity irradiation. The multiple comparison test showed that Reveal HD bulk fill had a significantly higher VPS than Tetric N-Ceram ( p < 0.006), Filtek Z350 ( p < 0.001), and Filtek One Bulk Fill ( p < 0.000). Comparison of RBCs subjected to low-intensity curing light gave similar results, with Reveal HD again exhibiting the highest VPS (2.83 ± 0.22), followed by Tetric N-Ceram (2.51 ± 0.31), Filtek Z350 (2.44 ± 0.35), and Filtek One Bulk Fill (2.19 ± 0.21). There was no significant different between the materials, except for Reveal HD and Filtek one bulk fill ( p < 0.005). The effect of curing light intensity on the VPS for each RBC analyzed with the t -test is displayed in Table 7 . The curing light intensity had no significant effect on the VPS for all of the tested materials. Postpolymerization 3D images of the materials are shown in Fig. 1 .
Table 7

Descriptive t -test analysis

RBC materialCuring light intensityMean (%)SD95% confidence interval for mean t -test p -Value
Lower boundUpper bound
Abbreviations: RBC, resin-based composite; SD, standard deviation.
Filtek Z350High2.2540.1852.0592.4490.269
Low2.4440.3522.0752.813
Filtek OneHigh2.0980.1761.9132.2820.414
Low2.1960.2191.9662.426
Tetric N-CeramHigh2.4290.2422.1752.6820.619
Low2.5120.3172.1802.844
Reveal HDHigh3.1930.5992.5653.8210.100
Low2.8340.2222.6013.068
Fig. 1

Images of three-dimensional reconstructions of composite restorations. 1: high-intensity curing light; 2: low-intensity curing light; A: Filtek One Bulk Fill Posterior; B: Filtek Z350; C: Tetric N-Ceram; D: Reveal HD Bulk.

Images of three-dimensional reconstructions of composite restorations. 1: high-intensity curing light; 2: low-intensity curing light; A: Filtek One Bulk Fill Posterior; B: Filtek Z350; C: Tetric N-Ceram; D: Reveal HD Bulk.

Discussion

In this study, the DC and VPS of bulk-fill RBCs were evaluated with exposure to two different curing light intensities. The results of this study showed that the curing light intensity had no effect on the VPS and also had no effect on the DC, except for the Reveal HD bulk-fill RBC, in which high-intensity curing light exposure resulted in a high percentage of DC. FTIR spectroscopy was employed to determine the DC; FTIR is a reliable technique that is frequently used to obtain the DC of carboncarbon double bonds converted to single bonds. 10 Several factors can affect the DC such as the chemistry of the organic matrix, the photoinitiator used, and the filler amount, size, and distribution. 4 A very important factor that also affects the DC is the light source used, 18 photoactivation method, 19 wavelength, 20 light tip size, power density, 21 and irradiation time. 22 A significant difference in the DC was observed among the materials tested. A systematic review by Lima et al 23 evaluated the DOC in bulk-fill RBCs revealing that the polymerization of bulk-fill RBCs depends on both the radiant exposure (irradiance and time) and material composition, recommending curing light with LED irradiance ≥1,000 mW/cm 2 with 20-second exposure time to obtain acceptable DOC of bulk-fill RBCs. Reveal HD bulk-fill RBC exhibited the highest DC (72.9%) with low-intensity light and 85.6% DC with high-intensity light irradiation. This can be explained by the Reveal HD filler technology, which utilizes fillers to refract light deeper in the mass of the material, allowing for a higher DOC. The Reveal HD bulk-fill RBC was the only material affected by curing light intensity, whereby higher intensity light irradiation resulted in a higher percentage of DC. Wydra et al 24 observed similar DC behavior with a dimethacrylate RBC exposed to high and low irradiance, whereby high-intensity light resulted in a higher DC. However, Filtek one bulk-fill posterior, Filtek Z350, and Tetric N-Ceram bulk-fill RBCs were not affected by the curing light intensity, which is in agreement with other studies 25 26 which also showed that the light cure intensity did not affect the DC of bulk-fill composite resins. Hasslen et al 25 evaluated the DOC of bulk-fill composites with three different irradiance exposures (i.e., standard power mode, 1,000 mW/cm 2 ; high-power mode, 1,400 mW/cm 2 ; and Xtra power mode, 3,200 mW/cm 2 ), concluding that the curing light mode did not affect the DOC of the tested bulk-fill materials. Although an additional photoinitiator (Ivocerin) is added to the Tetric N-Ceram bulk-fill RBC, it resulted in the lowest DC, which is consistent with a previous report that also observed a low DC for this bulk-fill RBC. 27 The low DC of Tetric N-Ceram could be due to differences in the material composition. The two main monomer characteristics that affect the DC are the initial monomer viscosity and chemical structure flexibility. 28 Bis-GMA is the least flexible and most viscous monomer among the monomers in this study, which is due to the presence of rigid aromatic groups that interact and strong hydrogen bonding of the hydroxyl groups present in the monomer. UDMA possesses lower viscosity and higher flexibility compared with Bis-GMA due to weak hydrogen bonding of its amine group compared with that of the hydroxyl groups. 29 In the present study, the Reveal HD Bulk RBC had the highest VPS and highest DC compared with the rest of the tested composites. This can be explained by the direct relationship between VPS and DC. 30 Filtek one bulk-fill posterior had the least VPS, which the manufacturer claims arises from the innovative aromatic urethane dimethacrylate and addition-fragmentation monomer present in the formulation, which reduce shrinkage and relieve stress during polymerization. Filtek Z350 was used in the study as a conventional RBC to act as a control, ultimately exhibiting no significant difference in the VPS compared with the bulk-fill RBCs. This was in agreement with a study by Abbasi et al, 31 whereby no significant difference was observed between the conventional and bulk-fill RBC in terms of VPS. Another study by Rizzante et al 32 examined the VPS of bulk-fill composites using μ-CT concluding that the bulk-fill RBCs provided similar results to the conventional RBC. Curing light intensity did not affect the VPS in the current study, similar to what Sampaio et al observed, noting also that what mostly affects the VPS are the material compositions. 33 In a randomized clinical study, the effectiveness of using high-intensity LED curing light on the clinical performance of bulk-fill RBC restorations was evaluated; no difference in performance was observed between the conventionally used intensity and high-intensity light used for curing the restorations. 34 All of the tested RBCs in this study had a VPS ranging from 2.090 to 3.193%, which is an acceptable value. 35 Furthermore, the use of high-intensity curing light for 20 seconds or low-intensity curing light for 40 seconds had no significant effect on the VPS.

Conclusion

Different curing light intensities did not significantly influence VPS in all of the RBC materials tested. The Reveal HD bulk-fill composite cured under high-intensity light exhibited the highest DC. Light intensity had no effect on the DC for the rest of the materials. The hypothesis in this study was partially accepted, since there was no statistically significant effect on VPS when different curing light intensities were used. However, curing light intensity did affect the DC of one material, Reveal HD bulk fill, which had a higher DC when using high-intensity curing light. The results of this study can aid dentists to optimize the clinical performance of restorations. However, more longitudinal clinical studies are needed to support the use of bulk-fill RBCs in clinical practice.
  29 in total

1.  The micro-Raman spectroscopy, a useful tool to determine the degree of conversion of light-activated composite resins.

Authors:  C Pianelli; J Devaux; S Bebelman; G Leloup
Journal:  J Biomed Mater Res       Date:  1999

2.  New insight into the "depth of cure" of dimethacrylate-based dental composites.

Authors:  Julian G Leprince; Philippe Leveque; Bernard Nysten; Bernard Gallez; Jacques Devaux; Gaetane Leloup
Journal:  Dent Mater       Date:  2012-01-02       Impact factor: 5.304

3.  Comparative properties of low-shrinkage composite resins.

Authors:  John Burgess; Deniz Cakir
Journal:  Compend Contin Educ Dent       Date:  2010-05

4.  Polymerization shrinkage and contraction stress of dental resin composites.

Authors:  Cornelis J Kleverlaan; Albert J Feilzer
Journal:  Dent Mater       Date:  2005-07-22       Impact factor: 5.304

5.  Degree of conversion of bulk-fill compared to conventional resin-composites at two time intervals.

Authors:  Ruwaida Z Alshali; Nick Silikas; Julian D Satterthwaite
Journal:  Dent Mater       Date:  2013-07-08       Impact factor: 5.304

6.  Polymerization shrinkage, microhardness and depth of cure of bulk fill resin composites.

Authors:  Fabio Antonio Piola Rizzante; Jussaro Alves Duque; Marco Antônio Húngaro Duarte; Rafael Francisco Lia Mondelli; Gustavo Mendonça; Sérgio Kiyoshi Ishikiriama
Journal:  Dent Mater J       Date:  2019-03-26       Impact factor: 2.102

7.  Light curing - guidelines for practitioners - a consensus statement from the 2014 symposium on light curing in dentistry held at Dalhousie University, Halifax, Canada.

Authors:  Jean-Francois Roulet; Richard Price
Journal:  J Adhes Dent       Date:  2014-08       Impact factor: 2.359

8.  Depth of cure of resin composites: is the ISO 4049 method suitable for bulk fill materials?

Authors:  Simon Flury; Stefanie Hayoz; Anne Peutzfeldt; Jürg Hüsler; Adrian Lussi
Journal:  Dent Mater       Date:  2012-03-03       Impact factor: 5.304

9.  Bulk-fill resin-based composites: an in vitro assessment of their mechanical performance.

Authors:  N Ilie; S Bucuta; M Draenert
Journal:  Oper Dent       Date:  2013-04-09       Impact factor: 2.440

10.  Evaluation of degree of conversion and hardness of dental composites photo-activated with different light guide tips.

Authors:  Marília Regalado Galvão; Sergei Godeiro Fernandes Rabelo Caldas; Vanderlei Salvador Bagnato; Alessandra Nara de Souza Rastelli; Marcelo Ferrarezi de Andrade
Journal:  Eur J Dent       Date:  2013-01
View more
  2 in total

1.  Microleakage of Direct Restorations-Comparisonbetween Bulk-Fill and Traditional Composite Resins:Systematic Review and Meta-Analysis.

Authors:  Francesca Zotti; Edoardo Falavigna; Giorgia Capocasale; Daniele De Santis; Massimo Albanese
Journal:  Eur J Dent       Date:  2021-08-27

2.  Influence of Tip Diameter and Light Spectrum of Curing Units on the Properties of Bulk-Fill Resin Composites.

Authors:  Igor Oliveiros Cardoso; Alexandre Coelho Machado; Luísa de Oliveira Fernandes; Paulo Vinícius Soares; Luís Henrique Araújo Raposo
Journal:  Eur J Dent       Date:  2021-12-14
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.