Literature DB >> 33314853

[Research of Methods to Reduce Alarm Fatigue of Monitoring System].

Mengxing Liu1,2, Zehui Sun2, Wenyu Ye2, Sanchao Liu2, Xianliang He2, Cheng Wang2, Ye Li1.   

Abstract

OBJECTIVE: In order to solve alarm fatigue, the algorithm optimization strategies were researched to reduce false and worthless alarms.
METHODS: A four-lead arrhythmia analysis algorithm, a multiparameter fusion analysis algorithm, an intelligent threshold reminder, a refractory period delay technique were proposed and tested with collected 28 679 alarms in multi-center study.
RESULTS: The sampling survey indicate that the 80.8% of arrhythmia false alarms were reduced by the four-lead analysis, the 55.9% of arrhythmia and pulse false alarms were reduced by the multi-parameter fusion analysis, the 28.0% and 29.8% of clinical worthless alarms were reduced by the intelligent threshold and refractory period delay techniques respectively. Finally, the total quantity of alarms decreased to 12 724.
CONCLUSIONS: To increase the dimensionality of parametric analysis and control the alarm limits and delay time are conducive to reduce alarm fatigue in intensive care units.

Entities:  

Keywords:  alarm fatigue; critical care; intelligent threshold; multi-parameter fusion; nursing management

Year:  2020        PMID: 33314853     DOI: 10.3969/j.issn.1671-7104.2020.06.003

Source DB:  PubMed          Journal:  Zhongguo Yi Liao Qi Xie Za Zhi        ISSN: 1671-7104


  1 in total

1.  The Current Situation and Influencing Factors of the Alarm Fatigue of Nurses' Medical Equipment in the Intensive Care Unit Based on Intelligent Medical Care.

Authors:  Yunxia Zhao; Minlin Wan; Huisong Liu; Mei Ma
Journal:  J Healthc Eng       Date:  2021-07-03       Impact factor: 2.682

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.