Comparison of internal target volumes defined by three-dimensional, four-dimensional, and cone-beam computed tomography images of a motion phantom.

Yu Wang1, Tao Liu1,2, Huiqin Chen3, Penggang Bai4, Qiong Zhan1,2, Xiaohua Liang1,2.   

Abstract

BACKGROUND: To explore the variations of the gross tumor volume (GTV) from different three-dimensional computed tomography (3DCT) scan modes and the consistency of internal target volume (ITV) between different 3DCT, cone-beam computed tomography (CBCT) and four-dimensional computed tomography (4DCT) scan, a study using a motion phantom simulating sinusoidal movement was conducted.
METHODS: For three 3DCT scan modes: the GTV was contoured, and ITVI was generated on the basis of GTV with a 10-mm margin while ITVII and ITVIII with a 0-mm margin on the motion direction. ITVCBCT and ITVMIP were contoured on the images of CBCT and maximum-intensity projection (MIP) reconstructed 4DCT images. The centroid position shifts of ITVs were analyzed. The volume consistency between ITVI, ITVII, ITVIII and ITVMIP were evaluated by calculating the Dice similarity coefficient (Dsc) and the value of Δ Volume (ΔV). Furthermore, the 3DCT and CBCT images from 12 NSCLC patients were retrospectively collected, then the Dsc and ΔV were calculated.
RESULTS: The mean ± standard deviation of centroid position of ITVI, ITVII and ITVIII were 2.3±4.7, 2.6±4.0, and 1.0±1.4 mm, respectively. The mean ± standard deviation of Dsc between ITVI, ITVII, ITVIII and ITVMIP were 0.78±0.77, 0.86±0.1, and 0.94±0.05, respectively. The ΔV of ITVI, ITVII, ITVIII were 29.67%, 17.22%, and 6.46%, respectively. The ITV from CBCT showed a deduction rate of 3.1-9.3% compared to 4DCT. For the patients, the mean Dsc andΔV between ITVI and ITVCBCT were 0.50 and 60.76%.
CONCLUSIONS: The GTV acquired from 3DCT scan mode I possessed great deviation of centroid position and target volume. ITV on the basis of this GTV was significantly larger than ITVMIP. A good similarity was showed between ITVIII and ITVMIP, 4DCT is still a golden standard for the ITV delineation, but in the absence of 4DCT, image from 3DCT scan mode III and KV-CBCT may be considered for ITV delineation with caution. 2020 Annals of Translational Medicine. All rights reserved.

Entities:  

Keywords:  Non-small cell lung cancer (NSCLC); computed tomography; gross tumor volume (GTV); internal gross tumor volume; motion phantom

Year:  2020        PMID: 33313233      PMCID: PMC7729323          DOI: 10.21037/atm-20-6246

Source DB:  PubMed          Journal:  Ann Transl Med        ISSN: 2305-5839


Introduction

High precision radiation therapy (HPRT) allows a high dose of radiation to be delivered to the target volume while reducing the dose to organs at risk. However, the tumor motion induced by respiration during positioning and treatment delivery introduces a major challenge. In patients with non-small cell lung cancer (NSCLC) who undergo HPRT, accurate target delineation is of crucial importance. Three-dimensional computed tomography (3DCT) is widely used in the delineation of target volume. However, high speed 3DCT with axial or helical mode always underestimates the tumor motion and distorts the reconstructed images of the thorax and abdomen (1). According to The International Commission on Radiation Units and Measurements Report 62 (ICRU62), to ensure dose coverage for mobile tumors, the internal target volume (ITV) includes the range of target motion during respiration on the basis of CTV (2). According to population-based motion information and guidelines, in conventional 3DCT-based treatment planning, the ITV is generated by drawing a margin on the basis of the gross tumor volume (GTV). However, the ITV contoured by this method is always larger than the real ITV, resulting in an overdose to normal tissue and geometric miss (3). Li et al. reported that the median degree of inclusion (DI) of planning target volume (PTV)3DCT in PTV4DCT was 51.60%, and the median DI of PTV4DCT in PTV3DCT was 99.94% for middle and lower lung cancer with free breathing (4). Four-dimensional CT (4DCT), which allows data of full respiratory movement to be obtained, is widely used to assess respiration-induced tumor motion. After the 4DCT images have been acquired, a maximum-intensity projection (MIP) image is usually reconstructed to efficiently delineate the target volume (5,6). The ITV generated from MIP images of 4DCT scans has been widely considered to be the golden standard for the definition of individual targets (7-9). However, because the individual ITVs acquired in 4DCT simulation are not always consistent with the real ITVs in radiation delivery, the accuracy of this technique has been questioned. Moreover, under the conditions of irregular breathing and lesion motion amplitude >2 cm, the accuracy of ITVs obtained with 4DCT is reduced further (10-13). A cone-beam CT (CBCT) scanner mounted on the gantry of an accelerator with a scan time of approximately 60 seconds during image acquisition can be used to obtain respiration information. Previous studies have shown a good volume consistency between CBCT and 4DCT (14,15). The visualized target volume in CBCT can be considered as an individual ITV which is similar to that of 4DCT. In the absence of 4DCT, ITVCBCT scans may act as a suitable substitute for ITV. Though the 4DCT has been the golden technique to get individual ITV for SBRT of isolated lung cancer. It is difficult to abandon the isotropic margin accounting for the tumor motion in 3DCT based treatment planning. So, to investigate the individual ITV with different CT techniques and choose the optimal one or a combination of them is still a thing of great sense to do. This study aimed to explore the volumetric and positional variations of GTV from different 3DCT scanning modes, and to compare the consistency of ITV between different 3DCT scan modes and CBCT and 4DCT by using a motion phantom and data from patients with NSCLC. We present the following article in accordance with the MDAR reporting checklist (available at http://dx.doi.org/10.21037/atm-20-6246).

Methods

This study was approved by the Ethics Committee of Fujian Cancer Hospital, and written informed consent was obtained from all patients. All procedures performed in this study involving human participants were in accordance with the Declaration of Helsinki (as revised in 2013).

Respiratory phantom study

Motion phantom

The QUASARTM Programmable Motion Phantom (Modus Medical Devices Inc., London, ON, Canada) with a cubic target simulating sinusoidal motion, was chosen, and is shown in . Three motion amplitudes (A) were set (5, 10, and 15 mm), and three motion frequencies (F) were set (20, 15, and 10 times/min). The theoretical ITV can be calculated based on the given diameter, motion, and magnitude of the motion target.
Figure 1

The motion phantom filled with air and a built-in polystyrene cube.

The motion phantom filled with air and a built-in polystyrene cube.

Patient characteristics

The 3DCT and CBCT images of 12 NSCLC patients who had been treated with intensity-modified radiation therapy (IMRT) between 2011 January and 2014 December were retrospectively analyzed. The patients had a median age of 65 (range, 60–78) years old. All of the patients were pathologically confirmed and were considered unsuitable for surgery. The patient and tumor characteristics are shown in .
Table 1

Patient and tumor characteristics

Characteristics
Gender
   Male8
   Female4
Age (years)
   Range60–78
   Median65
Location
   Upper lobe7
   Lower lobe5
T stage
   T14
   T25
   T33
Histology
   Squamous cell carcinoma5
   Adenocarcinoma7

Acquisition of CT images

3DCT scanning of the phantom

The 3DCT images were acquired with a Philips Brilliance CT Big Bore scanner (Philips Medical Systems, Inc., Cleveland, OH, USA). The phantom was placed on the scanning bed and used to simulate the movement of a lung tumor in the cranial–caudal (CC) direction. Three scan modes were set as follows: (I) pitch =1.06, rotation time =1.0 s; (II) pitch =0.813, rotation time =1.5 s; and (III) pitch =0.313, rotation time =1.5 s. The 3DCT scanning parameters were as follows: a 2-mm slice thickness, 120 kV scan voltage, and 398 mA electric current. The DICOM data of the moving phantom were reconstructed with a 1-mm slice thickness and transferred to the Pinnacle treatment planning system (Version 9.2).

4DCT scanning of the phantom

For the 4DCT images, the motion parameters of the phantom were set as mentioned above. The scanning process involved a 50-second scanning time and a slice thickness of 2 mm. The respiratory signal was recorded by tracking the trajectory of the infrared markers placed on the phantom with the Real-time Position Management™ System (RPM, Varian Medical Systems, Palo Alto, CA, USA). The 4DCT images were sorted into 10 bins according to the phase of the breathing signal. MIPs of the 4DCT data sets were then generated and transferred to the Pinnacle treatment planning system (Version 9.2) to contour the ITV4DCT.

CBCT scanning of the phantom

Cone-beam CTs were acquired using a Varian® Trilogy linear accelerator fitted with an On-Board Imager® (OBI) with system version 1.4.13.0 (Varian Medical Systems, PaloAlto, CA). The same setup, frequency settings, and slice thickness were used for CBCT scanning. The setting parameters were as follows: Super mode; collimator, S20, filter, F0; frame number, 1,320; frame clockwise rotation range of the rack, −180° to +180°. The obtained images were reconstructed with a 1-mm slice thickness and transferred to the Pinnacle treatment planning system to contour the ITVCBCT.

3DCT and CBCT scanning of the patients

The 12 NSCLC patients were immobilized in the supine position with the arms raised above the head using vacuum bags. The 3DCT scanning was performed during free breathing using a Philips Brilliance CT Big Bore scanner (Philips Medical Systems, Inc., Cleveland, OH, USA). Images were reconstructed with a 2-mm slice thickness, and transferred to the Pinnacle treatment planning system. For the initial radiation treatment, CBCT scanning was used to verify the tumor position. The CBCT images were acquired by rotation from −180° to +180° and transmitted to the Pinnacle treatment planning system after reconstruction with a slice thickness of 2 mm.

GTV contouring and ITV generation

The GTV, ITV3DCT, ITVCBCT, and ITVMIP were contoured by the same oncologist. The GTV was contoured under the same lung window conditions for each patient, with a window width of 1,600 Hu and a window level of −600 Hu. The ITVCBCT was contoured with a window width of 400 Hu and a window level of 40 Hu. The volume difference was calculated using the Dice similarity coefficient (Dsc) (16) and volume difference (ΔV). Dsc represents the degree of overlap between the two volumes, with the Dsc value approaching 1 as the agreement improves and if the two volumes do not overlap at all, the Dsc will be zero. The Dsc is used to evaluate the match index of the two volumes and is defined as follows: ΔV illustrates how the A is located within the B and is described as follows:

Statistical analysis

All statistical analyses were performed using the SPSS17.0 package (SPSS Inc, Chicago, IL, USA). The Wilcoxon test was used to compare the tumor volumes and the central position gained by different scanning modes. Differences were considered to be significant if the P value was less than 0.05.

Results

Motion phantom study

Geometrical variations of GTV from 3DCT scanning

GTV was contoured on 162 series of 3DCT images, and the length of the GTV in the CC direction along with the target motion was analyzed. The median length of the GTV was 38 mm (range, 18–59 mm). The characteristics of GTV length are shown in .
Figure 2

Length of the gross tumor volume (GTV) of the motion phantom acquired from three 3DCT scanning modes. The diameter of the cubic target was 30 mm. The median length of the GTV was 38 mm (range, 18–59 mm).

Figure 3

The gross tumor volume (GTV) variation under 3DCT scan mode I. The red outline depicts the GTV. (A) shows the cubic target within the motion phantom at conditions of rest; (B), (C), (D), and (E) show the length deviation of GTV under the same moving conditions: A =15 mm, F =15 times/min.

Length of the gross tumor volume (GTV) of the motion phantom acquired from three 3DCT scanning modes. The diameter of the cubic target was 30 mm. The median length of the GTV was 38 mm (range, 18–59 mm). The gross tumor volume (GTV) variation under 3DCT scan mode I. The red outline depicts the GTV. (A) shows the cubic target within the motion phantom at conditions of rest; (B), (C), (D), and (E) show the length deviation of GTV under the same moving conditions: A =15 mm, F =15 times/min.

Centroid position deviation derived from the three 3DCT scanning modes

The position of the center of the mass (COM) in the CC direction was analyzed. The mobility vector in the x, y direction was 0, and the vector in the z direction was the amplitude of the phantom. show the deviation of the target center. As more respiration information was incorporated into scanning mode III, the mean deviation of COM derived was significantly smaller than that of scanning modes I and II (P<0.001). The variation in the centroid position of scan mode I, II and III were 2.3±4.7 mm (range, −10–16 mm), 2.6±4.0 mm (range, −5.5–11.5 mm), and 1.0±1.4 mm (range, −2–5 mm), respectively.
Figure 4

Central position deviations of gross tumor volume from different 3DCT scanning modes were 2.3±4.7 mm (range, −10–16 mm), 2.6±4.0 mm (range, −5.5–11.5 mm), and 1.0±1.4 mm (range, −2–5 mm), respectively (P<0.000).

Figure 5

The centroid position deviation becoming stronger as the motion vector increases under the three 3DCT scanning modes. (A,B,C) The central deviation under scanning mode I with three amplitudes (A =5, 10, and 15 mm). Mean ± SD: 2.8±3.6 (0–16.0 mm), 4.8±2.8 (0–9.0 mm), and 4.4±3.4 (0–10.5 mm), respectively (P=0.037). (D,E,F) The central deviation under scanning mode II with three amplitudes. Mean ± SD: 1.4±1.3 (0–4.5 mm), 4.2±2.2 (1–9.0 mm), and 5.7±3.0 (1–11.5 mm), respectively (P=0.000). (G,H,I) The central position under scanning mode III had high reproducibility, with a central deviation of 0.6±0.8 (0–3.0 mm), 1.8±1.3 (0.5–5.0 mm), and 1.2±1.1 (0–3.5 mm), respectively (P=0.03).

Central position deviations of gross tumor volume from different 3DCT scanning modes were 2.3±4.7 mm (range, −10–16 mm), 2.6±4.0 mm (range, −5.5–11.5 mm), and 1.0±1.4 mm (range, −2–5 mm), respectively (P<0.000). The centroid position deviation becoming stronger as the motion vector increases under the three 3DCT scanning modes. (A,B,C) The central deviation under scanning mode I with three amplitudes (A =5, 10, and 15 mm). Mean ± SD: 2.8±3.6 (0–16.0 mm), 4.8±2.8 (0–9.0 mm), and 4.4±3.4 (0–10.5 mm), respectively (P=0.037). (D,E,F) The central deviation under scanning mode II with three amplitudes. Mean ± SD: 1.4±1.3 (0–4.5 mm), 4.2±2.2 (1–9.0 mm), and 5.7±3.0 (1–11.5 mm), respectively (P=0.000). (G,H,I) The central position under scanning mode III had high reproducibility, with a central deviation of 0.6±0.8 (0–3.0 mm), 1.8±1.3 (0.5–5.0 mm), and 1.2±1.1 (0–3.5 mm), respectively (P=0.03).

The length of the ITVCBCT compared to that of ITVMIP

ITVCBCT was positively correlated with the amplitude of the phantom. The ITV length from CBCT showed a deduction rate of 3.1–9.3% compared to that of 4DCT. The MIP derived from 4DCT with a clear visualization of the target without motion artifacts is taken as the golden standard for estimating the target motion. For the three amplitudes (A=5 mm, 10 mm, 15 mm), the mean ratio of ITV from CBCT to 4DCT was 0.97, 0.91, and 0.91, respectively. shows the characteristics of the ITVCBCT.
Table 2

The length of the ITVCBCT compared to that of the ITVMIP

Amplitude (mm)MinimumCBCT (mm)MaximumCBCT (mm)MeanCBCT (mm)MeanMIP (mm)Ratio of length (CBCT/4DCT)Volume deduction of CBCT
5363935360.973.1%
10474943470.919.3%
15535750550.918.9%

ITVCBCT refers to the internal gross tumor volume of cone-beam computed tomography (CT); ITVMIP refers to the internal gross tumor volume of four-dimensional CT; MinCBCT refers to the minimum length of ITV from cone-beam CT; MaxCBCT refers to the maximum length of ITV from cone-beam CT; MeanCBCT refers to the mean length of ITV from cone-beam CT; MeanMIP refers to the mean length of ITV from four-dimensional CT.

ITVCBCT refers to the internal gross tumor volume of cone-beam computed tomography (CT); ITVMIP refers to the internal gross tumor volume of four-dimensional CT; MinCBCT refers to the minimum length of ITV from cone-beam CT; MaxCBCT refers to the maximum length of ITV from cone-beam CT; MeanCBCT refers to the mean length of ITV from cone-beam CT; MeanMIP refers to the mean length of ITV from four-dimensional CT.

Comparison of the target length between ITV3DCT, ITVCBCT, and ITVMIP

Compared to those with scanning modes I and II, the ITV derived from scanning mode III showed a good consistency with ITVMIP and ITVCBCT, with a better DSC and smaller ΔV. The Dsc between ITVI, ITVII, ITVIII, and ITVMIP were 0.78±0.77 (range, 0.45–0.91), 0.86±0.1 (range, 0.57–0.99), and 0.94±0.05 (range, 0.77–1.0), respectively. The ΔV between ITVI, ITVII, ITVIII and ITVMIP was 29.67% (range, 0–63.89%), 17.22% (range, 0–57.45%), 6.46% (range, 0–23.4%), respectively (). The mean ratio of ITVI, ITVII, ITVIII to ITVCBCT were 1.21, 0.91, and 1.06, respectively, and the mean ratios to ITVMIP were 1.16, 0.85, and 0.95, respectively. ITVI with a margin of 10-mm on CC direction and ITVII with a margin of 0-mm on the basis of GTV showed poor consistency with ITVCBCT and ITVMIP (P<0.05). ITVIII without a margin showed good consistency with ITVMIP and ITVCBCT (P=0.044 and P=0.082, respectively). The data are shown in .
Figure 6

The Dsc and ΔV of ITV from three 3D-CT scan modes. (A) The Dsc between ITVI, ITVII, ITVIII and ITVMIP were 0.78±0.77 (range: 0.45–0.91), 0.86±0.1 (range: 0.57–0.99), 0.94±0.05 (range: 0.77–1.0), respectively. (B) The ΔV between ITVI, ITVII, ITVIII and ITVMIP were 29.67% (range: 0–63.89%), 17.22% (range: 0–57.45%), 6.46% (range: 0–23.4%).

Table 3

Comparison of ITV3DCT with ITVCBCT and ITVMIP for the phantom

Amplitude (mm)ITVI (CC)ITVII (CC)ITVIII (CC)ITVCBCT (CC)ITVMIP (CC)Mean ITVI/ITVCBCTMean ITVII/ITVCBCTMean ITVIII/ITVCBCTMean ITVI/ITVMIPMean ITVII/ITVMIPMean ITVIII/ITVMIP
539.6–53.1 (46.0±4.6)19.8–36.9 (28.9±5.7)28.8–34.2 (32.7±1.4)31.5–32.9 (32.2±0.39)31.9–33.3 (32.4±0.43)1.430.921.061.420.891.01
1036–62.1 (48.5±9.5)18–44.1 (34.4±7.9)32.4–42.3 (38.9±3.0)35.6–40.5 (39.9±0.44)41.4–43.2 (42.3±0.46)1.220.891.051.150.810.92
1534.2–71.1 (46.0±14.9)32.4–53.1 (41.8±5.6)39.6–48.6 (45.4±2.3)45.9–48.2 (46.8±0.62)48.9–50.4 (49.5±0.38)0.990.921.050.930.840.92
Overall34.2–71.1 (46.8±10.4)18–53.1 (35±8.3)28.8–48.6 (39.0±5.7)31.5–48.2 (39.6±6)31.9–50.4 (41.4±7)1.210.911.061.160.850.95
Z−3.79−2.03−2.06−2.4−3.57−1.749
P<0.0000.0410.0440.0160.0000.082

ITV3DCT refers to the internal gross tumor volume of three-dimensional computed tomography (CT). ITVMIP refers to the internal gross tumor volume of four-dimensional CT; ITVI refers to the internal gross tumor volume of 3DCT scan mode I, ITVII refers to the internal gross tumor volume of 3DCT scan mode II, ITVIII refers to the internal gross tumor volume of 3DCT scan mode III.

The Dsc and ΔV of ITV from three 3D-CT scan modes. (A) The Dsc between ITVI, ITVII, ITVIII and ITVMIP were 0.78±0.77 (range: 0.45–0.91), 0.86±0.1 (range: 0.57–0.99), 0.94±0.05 (range: 0.77–1.0), respectively. (B) The ΔV between ITVI, ITVII, ITVIII and ITVMIP were 29.67% (range: 0–63.89%), 17.22% (range: 0–57.45%), 6.46% (range: 0–23.4%). ITV3DCT refers to the internal gross tumor volume of three-dimensional computed tomography (CT). ITVMIP refers to the internal gross tumor volume of four-dimensional CT; ITVI refers to the internal gross tumor volume of 3DCT scan mode I, ITVII refers to the internal gross tumor volume of 3DCT scan mode II, ITVIII refers to the internal gross tumor volume of 3DCT scan mode III.

Patients

CBCT and 3DCT were performed in 12 patients with lung cancer who had previously received IMRT. The ITVCBCT and ITV3DCT [with a 1-cm extension on the CC direction, and a 0.5-cm extension on the left-right (LR) and superior-inferior (SI) directions] were compared, and the volume difference between them was statistically significant (P=0.002). The difference between ITVCBCT and ITV3DCT was expressed by V, and the mean value of V was 35.4 cm3. The mean Dsc and ΔV between ITV3DCT and ITVCBCT were 0.50 and 60.76%, respectively. Data was shown in and .
Table 4

Comparison of ITV10 mm and ITVCBCT in lung cancer patients with 3DCT scanning mode I

NumberGTV3DCT (CC)ITV10 mm (CC)ITVCBCT (CC)Volume difference (CC)
16.125.55.719.8
23.621.86.115.7
312.740.216.723.5
495.9212.3119.592.8
56.529.711.018.7
645.5106.353.652.7
71.310.61.69.0
821.366.228.238
951.0122.760.262.5
104.325.25.220.0
1111.432.815.017.8
1223.179.925.854.1
Mean35.4

ITV10 mm: the internal gross tumor volume (ITV) of lung cancer with a 10-mm margin on the cranial–caudal (CC) direction, 5-mm on the left-right (LR) and superior-inferior (SI) directions on the basis of gross tumor volume (GTV). ITVCBCT: The ITV of lung cancer from cone-beam computed tomography (CBCT).

Table 5

The Dice similarity coefficients (Dsc) and ΔV of ITVCBCT and ITV10 mm of 12 patients

NumberDSCΔV (100%)
10.353677.65
20.432472.02
30.574358.46
40.455943.71
50.540562.96
60.670449.58
70.262194.33
80.591657.40
90.683450.93
100.325540.08
110.624754.27
120.488267.71
Mean0.500260.76

Dsc, dice similarity coefficient. ΔV represents the delta volume of the two targets.

Figure 7

The Dsc and ΔV of the 12 lung cancer patients. (A) Showed the Dsc distribution of the 12 patients with lung cancer, the mean Dsc was only 0.50. (B) Showed the ΔV between ITV3D-CT and ITVCBCT of 12 patients with lung cancer, the mean ΔV was up to 60.7%.

ITV10 mm: the internal gross tumor volume (ITV) of lung cancer with a 10-mm margin on the cranial–caudal (CC) direction, 5-mm on the left-right (LR) and superior-inferior (SI) directions on the basis of gross tumor volume (GTV). ITVCBCT: The ITV of lung cancer from cone-beam computed tomography (CBCT). Dsc, dice similarity coefficient. ΔV represents the delta volume of the two targets. The Dsc and ΔV of the 12 lung cancer patients. (A) Showed the Dsc distribution of the 12 patients with lung cancer, the mean Dsc was only 0.50. (B) Showed the ΔV between ITV3D-CT and ITVCBCT of 12 patients with lung cancer, the mean ΔV was up to 60.7%.

Discussion

Physiologic motions, especially respiratory movement, are known to influence the precision of radiation procedures in the treatment of thoracic tumors. The images obtained by conventional 3DCT only represent the random shapes and volumes during the period of respiration and cannot truly reflect all of the necessary respiratory movement information. Slow scanning methods include slow spiral CT and CBCT, are more reproducible and can better capture tumor movement as a result of the larger CTV and a higher accuracy rate (17,18). The ITV from 4DCT is a fusion of the GTVs contoured from the 10 respiratory phases during scanning and is considered to be the golden standard for ITV contouring (7-9). In order to validate that the image from conventional 3DCT scan was the instant shape and volume of a phase from respiratory cycle that comes randomly, we applied three different 3DCT scanning modes by adjusting the pitch and rotation time to scan the phantom in our study. We found out the variation of the target volume and central position were significantly correlated with the scan mode. For the cubic target with a diameter of 30 mm, the obtained length of the GTV varied greatly from 18–59 mm. The ITV became larger as the motion amplitude increasing for scan mode II and III, while this relationship can’t be detected by scan mode I. The mean and standard deviation of the centroid shift decreased as the scan time being longer for ITVI, ITVII and ITVIII with value of 2.3±4.7, 2.6±4.0, and 1.0±1.4 mm, respectively. In this study, the motion cycle of the phantom was 3–6 seconds for three motion amplitudes. For scanning mode I with a pitch value of 0.813 and a rotation time of 1.0 seconds, the acquisition time was shorter than a motion cycle, leading to the great variation in centroid position and volume of GTV. While the scanning mode III with a smaller pitch value of 0.313 and a longer rotation time of 1.5 seconds resulted in a longer image acquisition time than the motion cycle. As more respiration information was incorporated into scanning mode III, the mean deviation of the central position and the target volume derived was significantly smaller than that of scanning modes I and II (P<0.001). We also found out that the deviation of central position became significantly greater as the motion amplitude increased (P<0.05) under same scanning mode. This result indicated that the tumor motion may contribute to increasing the target volume variation, because a tumor with larger motion amplitude is likely to produce more residual motion artifacts. Similar results were also reported by Chen et al. (1), who demonstrated that the volume and position of the targets obtained by conventional spiral CT scanning were obviously different and the obtained GTV was distorted after 3D reconstruction. Li (4) also evaluated the centroid shifts of PTVs derived from 3DCT and demonstrated the uncertainty of a target derived from 3DCT. In our study, we found the GTV3DCT from scan mode I, II underestimate the MIP volume by 28%, 15% respectively and they were statistically significant with P value of 0.000. This result was similar to the previous study from Lingaiah et al. (15), who found that the GTV by helical mode scanning underestimate the MIP volume by 13% and 21% respectively and was statistically significant with a P value of 0.00. In addition, our study showed the GTV3DCT from scan mode III underestimate the MIP volume only by 5% and this difference was no statistically significant (P=0.082). Thus, scan mode III could effectively decrease the uncertainty and variation from the variation of respiration. This certainly due to the smaller pitch value and longer rotation time of scan mode III which guarantee much more time to acquire much more target motion information. By analyzing the Dsc and ΔV between targets, the difference in ITV obtained through different imaging techniques can be more clearly established. The Dsc between targets was directly influenced by the shift of the target central position. In our study, we evaluated the Dsc between ITV3DCT, and ITVCBCT and ITVMIP. For the 3DCT scan mode I and II, the ITV obtained was much larger than the ITVCBCT or ITVMIP for the phantom. A good similarity between ITVIII and ITVMIP with a mean Dsc and ΔV of 0.94 and 6.46% was showed in our study. The mean ratio of ITVIII to ITVCBCT and ITVMIP was 1.06 and 0.95, respectively. The present data suggest that scanning mode III, with a pitch of 0.313 and rotation of 1.5 seconds, could obtain considerably more respiration information and had a good consistency with 4DCT. The ITVs from scan mode III may be considered as individual ITVs for IMRT in lung cancer. The data of 12 patients with lung cancer who had previously received radiotherapy were also analyzed. The defined ITV3DCT with a margin of 10 mm in the CC direction and 5 mm in the SI and LR directions showed an increasing volume of 35.27 cm3 and a mean size ratio of 3.1 compared to ITVCBCT. Poor consistency was revealed between the ITVCBCT and ITV3DCT, with a mean DSC of 0.50 and ΔV of 60%. These results support those of some previous studies (3,19,20). For instance, Li et al. investigated thirty-one patients with NSCLC underwent 3DCT and CBCT scans of the thorax during free breathing and found the mean Dsc between the two image modalities was 0.66 and the ratio of ITV3DCT to ITVCBCT was 2.33. For the 12 patients undergo 3DCT scan mode I, the corresponding ITV derived was much larger than the ITVCBCT, which resulted in good coverage of tumor motion while carrying a high possibility of unnecessary irradiation to normal tissue. CBCT images may suffer from projection deduction due to the interplay between the gantry rotation and the motion target. The volume from CBCT possessed great reproducibility for a certain amplitude in this study, this suggested that the image projection deduction may have no correlation with the amplitude of motion. In our study, the ITV obtained from CBCT showed a volume deduction of 3.1–9.3% compared to the ITV obtained from the MIP image of 4DCT. This result was similar to those of some reports; for instance, Wang et al. (14) reported a volume difference of less than 8%. Our study showed that the volume ratio of CBCT to 4DCT was 0.93. Lingaiah et al. (15) reported that CBCT was well correlated with the 4DCT volume, with a ratio of 1.02. Although the ITVCBCT was slightly smaller than the ITV4DCT, it contained considerably more respiratory information due to its longer scan time. Previous studies (10,21) have reported that under sinusoidal breathing conditions, CBCT and 4DCT can realistically reflect the actual range of tumor respiratory motion; however, when the respiratory amplitude is too large, or the respiratory curve is irregular, the ITVs obtained by CBCT and 4DCT are often partially deleted, making them smaller than that of the theoretical ITV. Under such conditions, the volumes of CBCT and 4DCT can be reduced by 20–30% and the length of the ITV can be reduced by 7–11 mm. Our study also evaluated the difference in ITV gained from different CT scans and came to a similar conclusion. In this article, we verified the randomness of GTV from 3DCT scanning and investigate the deviation of ITV from different 3DCT scan modes and different CT scan techniques. There is still much work to do to improve the techniques of 3DCT and 4DCT and to explore a optimized combination of these techniques. Our study has some limitations that need to be addressed. First, the study used only three sinusoidal phantom motions, which were constrained to the CC direction, and the data of the patients involved were collected and analyzed retrospectively. In addition, though we set strict criteria for collecting the patients’ data, the 12 number of the patients was also limited. These situations are certainly ideal; however, it may not be the case in a clinical setting. Further investigations involving 3D target motion and non-sinusoidal motion patterns as well as a larger amount of real patients’ breathing patterns between 3DCT scanning mode III, CBCT, and 4DCT would be worthwhile.

Conclusions

Our results indicated a good correlation between the ITV got from 3DCT scan mode III and MIP volume. The ITV from CBCT showed a deduction rate of 3.1–9.3% compared to 4DCT. In cases where 4DCT scan imager is not possible, this “slow” 3DCT scan mode III or CBCT scan may be considered for determining ITV volumes, with a small additional isotropic margin for safer treatment. The GTV obtained from 3DCT scan mode I had a strong uncertainty, and the target centroid position deviated with great randomness; the ITV formed on the basis of this GTV3DCT was larger than the ITVCBCT or ITV4DCT, which may lead to geographical miss radiation of the target and excessive radiation of surrounding normal tissue. The article’s supplementary files as
  19 in total

1.  Artifacts in computed tomography scanning of moving objects.

Authors:  George T Y Chen; Jong H Kung; Kevin P Beaudette
Journal:  Semin Radiat Oncol       Date:  2004-01       Impact factor: 5.934

2.  Improving target delineation on 4-dimensional CT scans in stage I NSCLC using a deformable registration tool.

Authors:  Iris E van Dam; John R van Sörnsen de Koste; Gerard G Hanna; Rebecca Muirhead; Ben J Slotman; Suresh Senan
Journal:  Radiother Oncol       Date:  2010-07       Impact factor: 6.280

3.  Quantifying ITV instabilities arising from 4DCT: a simulation study using patient data.

Authors:  Sara St James; Pankaj Mishra; Fred Hacker; Ross I Berbeco; John H Lewis
Journal:  Phys Med Biol       Date:  2012-02-17       Impact factor: 3.609

4.  Assessment of gross tumor volume regression and motion changes during radiotherapy for non-small-cell lung cancer as measured by four-dimensional computed tomography.

Authors:  Keith R Britton; George Starkschall; Susan L Tucker; Tinsu Pan; Christopher Nelson; Joe Y Chang; James D Cox; Radhe Mohan; Ritsuko Komaki
Journal:  Int J Radiat Oncol Biol Phys       Date:  2007-03-26       Impact factor: 7.038

5.  Dosimetric comparison of liver tumour radiotherapy in all respiratory phases and in one phase using 4DCT.

Authors:  Dorota Gabryś; Roland Kulik; Krystyna Trela; Krzysztof Ślosarek
Journal:  Radiother Oncol       Date:  2011-10-03       Impact factor: 6.280

6.  Projection-data based temporal maximum attenuation computed tomography: determination of internal target volume for lung cancer against intra-fraction motion.

Authors:  Shinichiro Mori; Nobuyuki Kanematsu; Hiroshi Asakura; Masahiro Endo
Journal:  Phys Med Biol       Date:  2007-01-23       Impact factor: 3.609

7.  Comparison of the planning target volume based on three-dimensional CT and four-dimensional CT images of non-small-cell lung cancer.

Authors:  Feng Xiang Li; Jian Bin Li; Ying Jie Zhang; Tong Hai Liu; Shi Yu Tian; Min Xu; Dong Ping Shang; Chang Sheng Ma
Journal:  Radiother Oncol       Date:  2011-05-04       Impact factor: 6.280

8.  Multiple "slow" CT scans for incorporating lung tumor mobility in radiotherapy planning.

Authors:  F J Lagerwaard; J R Van Sornsen de Koste; M R Nijssen-Visser; R H Schuchhard-Schipper; S S Oei; A Munne; S Senan
Journal:  Int J Radiat Oncol Biol Phys       Date:  2001-11-15       Impact factor: 7.038

9.  The effect of respiratory motion variability and tumor size on the accuracy of average intensity projection from four-dimensional computed tomography: an investigation based on dynamic MRI.

Authors:  Jing Cai; Paul W Read; Ke Sheng
Journal:  Med Phys       Date:  2008-11       Impact factor: 4.071

10.  Comparison of internal target volumes defined on 3-dimensional, 4-dimensonal, and cone-beam CT images of non-small-cell lung cancer.

Authors:  Fengxiang Li; Jianbin Li; Zhifang Ma; Yingjie Zhang; Jun Xing; Huanpeng Qi; Dongping Shang
Journal:  Onco Targets Ther       Date:  2016-11-17       Impact factor: 4.147

View more
  2 in total

1.  Evaluation of lung tumor motion in a large sample: Target-related and clinical factors influencing tumor motion based on four-dimensional CT.

Authors:  Fengxiang Li; Yanlin Qu; Tingting Zhang; Zhen Cui; Xin Sun; Tao Zhang; Jianbin Li
Journal:  Cancer Med       Date:  2021-09-14       Impact factor: 4.452

2.  Evaluation of Lung Tumor Target Volume in a Large Sample: Target and Clinical Factors Influencing the Volume Derived From Four-Dimensional CT and Cone Beam CT.

Authors:  Fengxiang Li; Tingting Zhang; Xin Sun; Yanlin Qu; Zhen Cui; Tao Zhang; Jianbin Li
Journal:  Front Oncol       Date:  2022-01-20       Impact factor: 6.244

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.