Residual breast tissue after robot-assisted nipple sparing mastectomy.

Ko Un Park1, Gary H Tozbikian2, David Ferry3, Allan Tsung4, Mathew Chetta5, Steven Schulz6, Roman Skoracki7.   

Abstract

INTRODUCTION: While the long-term oncologic safety of robot-assisted nipple sparing mastectomy (RNSM) remains to be elucidated, histologically detected residual breast tissue (RBT) can be a surrogate for oncologically sound mastectomy. The objective of this study is to determine the presence of RBT after RNSM.
METHODS: Between August 2019-January 2020, we completed 5 cadaveric RNSMs. Full thickness biopsies from the mastectomy skin flap were obtained from predefined locations radially around the mastectomy skin envelop and nipple areolar complex to histologically evaluate for RBT.
RESULTS: The first case was not technically feasible due to inability to obtain adequate insufflation. Five mastectomy flaps were analyzable. The average mastectomy flap thickness was 2.3 mm (range 2-3 mm) and the average specimen weight was 382.72 g (range 146.9-558.3 g). Of 70 total biopsies, RBT was detected in 11 (15.7%) biopsies. Most common location for RBT was in the nipple-areolar complex, with no RBT detected from the peripheral skin flaps.
CONCLUSIONS: In this cadaveric study, RNSM is feasible leaving minimal RBT on the mastectomy flap. The most common location for RBT is in the periareolar location consistent with previous published findings after open NSM. Clinical studies are underway to evaluate the safety of RNSM.
Copyright © 2020 The Authors. Published by Elsevier Ltd.. All rights reserved.

Entities:  

Keywords:  Mastectomy; Nipple sparing mastectomy; Robot-assisted mastectomy; Robot-assisted nipple sparing mastectomy

Year:  2020        PMID: 33310481      PMCID: PMC7726449          DOI: 10.1016/j.breast.2020.11.022

Source DB:  PubMed          Journal:  Breast        ISSN: 0960-9776            Impact factor:   4.380


Brief title: robot mastectomy

Financial and conflict of interest disclosure: The authors have no relevant financial disclosure. The authors declare that there are no competing interests.

Funding

The 2019 Intramural Research Program Award (46050–502730).

Author contribution

Study conception and design: Park, Skoracki. Acquisition of data: Park, Tozbikian, Ferry, Tsung, Chetta, Schulz, Skoracki. Analysis and interpretation of data: Park, Tozbikian. Drafting of manuscript: Park, Skoracki. Critical revision: Park, Tozbikian, Ferry, Tsung, Chetta, Schulz, Skoracki. Synopsis: Robot-assisted nipple sparing mastectomy (RNSM) is a surgical advancement for women undergoing removal of the breast for a variety of reasons. Utilizing cadavers we aimed to demonstrate procedure feasibility in completing an oncologically sound mastectomy by performing histologic analysis of the mastectomy flap after RNSM.

Introduction

With the advances in breast reconstruction after mastectomy for the treatment of breast diseases including breast cancer, surgical techniques have evolved to preserve the skin flaps and nipple areolar complex (NAC) to give better aesthetic outcome without compromising oncologic outcome [1,2]. The preservation of the nipple and areola complex during mastectomy can provide psychological and emotional benefits, and improve patient satisfaction [3,4]. Nipple sparing mastectomy (NSM) can provide major psychosocial benefits for patients but is technically demanding and has variable outcomes [[5], [6], [7]]. Total mammary glandular excision in NSM can be technically challenging, due to small size of the incision and poor visualization of the dissection plane. To help overcome the challenges in open NSM, use of the robot (da Vinci Surgical System, Intuitive Surgical, Sunnyvale, CA) for robot-assisted nipple sparing mastectomy (RNSM) is an emerging operation [[8], [9], [10], [11], [12], [13], [14], [15], [16], [17], [18], [19], [20]]. There are limited number of centers throughout the world currently performing RNSM, thus the procedure is currently still under development and optimization. The use of the da Vinci surgical system is not FDA approved for use in breast surgery in the United States [21]. To study the technical feasibility and safety of RNSM, we performed a series of cadaveric studies using the robot. Our goal was to master the techniques of RNSM and to assess the potential oncologic safety by evaluating the mastectomy skin flaps for histologically detected residual breast tissue (RBT). We hypothesized that RNSM will be technically feasible, leaving minimal residual breast tissue on the mastectomy skip flap. The technique of RNSM is described and current literature is reviewed in detail in the discussion.

Methods

Under The Ohio State University institutional Ethics Committee approval (in lieu of Institutional Review Board) between August 2019–January 2020, 6 RNSM cadaveric cases were performed in three female fresh cadavers by a single surgeon (KUP). Two of the cadaveric surgeries were performed with the daVinci Si system (Intuitive Surgical, Sunnyvale, CA) and one was performed with daVinci Xi system. Robotic surgery certified breast surgical oncologist (KP) at the console performed all the surgeries with an experienced robotic certified first assistant (DF) at the bedside. After breast removal, the mastectomy flap was measured with a caliper to assess the thickness of the flap.

Variations in surgical technique

The cadaver was placed in the supine position with the ipsilateral arm to the operative site positioned out of the field. Prior to the start of the procedure, the midline, inframammary fold, extent of breast tissue, anterior axillary line, and horizontal axillary crease were marked and used to guide the position of the linear incision along the anterior axillary line. Next, the subcutaneous dissection was performed using scissors and electrocautery as far as the instruments would reach under direct visualization to create a ‘working space’. Cadaver # 1 and #3 were performed using the daVinci Si® system and cadaver #2 was performed using the daVinci Xi® system. For the first cadaveric case, we used the entire linear incision to directly place the three robot ports adjacent to each other and secured the ports in place using suture and surgical skin stapler as previously described between ports [22]. Insufflation was difficult to achieve using this technique. Therefore, for the 2nd and 3rd cadaver cases, we modified the procedure and used the single port system (GelPOINT Mini, Applied Medical, Rancho Santa Margarita, CA). The single port system was inserted into the incision and the robot ports were placed into GelSeal Cap connected to an insufflator to keep the pressure at 8 mm Hg. The robot was positioned at the cadaver’s head and docked. The 30-degree camera was placed in the middle port. Subcutaneous dissection was performed in the lateral to medial direction using the monopolar-curved scissors and bipolar grasping forceps for traction and exposure. The bedside assistant guided the surgeon at the console on the extent of the dissection as outlined by the skin markings prior to docking the robotic system. Using a similar technique of sharp and cautery dissection, the gland was separated from the pectoralis major muscle. The specimen was removed from the anterior axillary incision by manual gentle traction using the “waving flag technique” (move the gland back and forth gently until it is removed). If it is not feasible to remove the entire gland easily, the incision was extended to assure removal of the intact specimen.

Assessment of residual breast tissue

Prior to skin closure, approximately 1 cm [2] full thickness biopsies from the mastectomy skin flap were obtained using fourteen predefined locations radially around the mastectomy skin envelop and nipple areolar complex (Supplementary Fig. 1) as previously published [23]. Locations L, M, N, and O correspond to the area behind the NAC. The tissue was formalin-fixed, paraffin-embedded and stained with hematoxylin and eosin (H&E) as per standard institutional protocol. The tissue was assessed histologically by a board certified breast pathologist (GT) for presence of RBT.

Data availability

The authors declare that the data supporting the findings of this study are available within the paper.

Results

Operative characteristics

Fresh non-formalin fixed female cadavers without previous history of breast cancer were used for the study. The cadavers were elderly and none had evidence of prior breast surgery (Table 1). There were no skin injuries (ie. button-hole) related to use of the robot (Table 2).
Table 1

Characteristics of cadavers.

Cadaver #1Cadaver #2Cadaver #3
Age767185
RaceWhiteBlackWhite
Cause of deathCOPDRetroperitoneal neoplasmDementia/Alzheimers
Estimated bra sizeABC
Ptosis of breast (grade)333
Body Mass Index32.1821.2827.45
Table 2

Operative characteristics of robot-assisted nipple sparing mastectomies.

Case 1Case 2Case 3Case 4Case 5Case 6
Case completed (yes/no)NoYesYesYesYesYes
Mastectomy flap thickness (mm)n/a2.02.02.53.02.0
Skin injury (yes/no)n/aNoNoNoNoNo
Weight of mastectomy specimen (grams)n/a146.9378.9289.2558.3540.3
Console time (minutes)n/a46708024775
Characteristics of cadavers. Operative characteristics of robot-assisted nipple sparing mastectomies. Console time was measured from the time of docking the robot to undocking the system. The console time ranged from 46 to 247 min (Table 2). Case #5 was significantly longer as mid-case the robot was undocked to trial a different type of single-port system, GelPOINT Path Transanal access platform (Applied Medical, Rancho Santa Margarita, CA). Although this helped in acquiring a complete seal so that the CO2 did not leak at the skin-port junction, the self-retaining sleeve was too rigid and narrow, resulting in instrument collision.

Residual breast tissue

The average mastectomy flap thickness was 2.3 mm (range 2–3 mm) and the average specimen weight was 382.72 g (range 146.9–558.3 g) (Table 2). Of the 70 total biopsies, RBT was detected in 11 (15.7%) biopsies. All of the RBT were detected in tissue behind the NAC: in case 2 RBT was detected in location M,N, and O; in case 3 RBT in N, in case 4 in L,M,N, and O; in case 5 in N; and in case 6 in L, and N (Fig. 1). There were no RBT detected outside the NAC in the mastectomy skin flaps.
Fig. 1

Representative slide of preiareolar skin, subcutaneous tissue, and residual breast parenchyma stained with hemoxylin and eosin. Arrow indicates periareolar skin and asterisk indicates terminal ductal lobular unit. Photo is taken with 40× magnification.

Representative slide of preiareolar skin, subcutaneous tissue, and residual breast parenchyma stained with hemoxylin and eosin. Arrow indicates periareolar skin and asterisk indicates terminal ductal lobular unit. Photo is taken with 40× magnification.

Discussion

In this cadaveric study, we demonstrate the technical reproducibility of robot-assisted nipple sparing (RNSM) and low residual breast tissue (RBT) after RNSM supporting our initial hypothesis. We detected at least one RBT-positive biopsy in all five RNSM but none were detected outside of the NAC. Although the cadavers in the present study were elderly with significant degree of ptosis and hence, not the typically patient who undergoes NSM, we were able to successfully complete five cadaveric RNSM procedures.

Residual breast tissue after mastectomy

Technical feasibility and reproducibility of RNSM have been demonstrated but previous studies did not evaluate for presence of RBT after mastectomy. In this study, we demonstrate that there were no histologically detectable RBT in the mastectomy flap outside of the NAC. The presence of residual breast tissue in the NAC is not surprising in this study. Presence of RBT after open NSM is not uncommon and is most common in tissue behind the NAC [23,24]. In a study using MRI to evaluate for RBT after mastectomy, the overall frequency of RBT was 77.8% and when the NAC region was excluded, the frequency of RBT was 51.6% in NSMs [24]. Papassotiropoulos et al. reported RBT was histologically detected in 18–28% of central biopsies around the NAC in open NSM cases [23]. This phenomenon may be due to lack of Cooper’s ligaments under the NAC, leading to more difficult precise dissection in the dermis-fibroglandular tissue junction [25]. More importantly, the incidence of local recurrence after open NSM is low despite the presence of RBT [2,[26], [27], [28]]. Nipple areolar complex recurrence after open NSM ranges from 0 to 7% in most recent series [25]. While the exact clinical relevance of RBT after mastectomy remains elusive, it is important to note that previous studies evaluating local recurrence and NAC recurrence demonstrate that tumor biology is a stronger risk factor than surgical technique.

Emergence of robotic-assisted nipple sparing mastectomy

Surgeons experience greater physical symptoms such as neck and lower back pain, mental strain, and fatigue from performing open NSM compared to skin sparing mastectomy [6,29]. Thus, a more ergonomically sound technique with greater visualization is needed to improve surgeon working conditions and to improve the ease of the operation during open NSM. Robot assisted surgery facilitates minimally invasive surgery through 3D vision with magnification, improved lighting, and seven degrees of motion of the instrument tip to allow for proper dissection of the surgical plane. Despite wide use in other surgeries, the use of the robot for breast surgery has been limited until recently. One of the earliest reports of RNSM is from Dr.Antonio Toesca [30]. In this initial report of 3 consecutive successful cases, the authors report the surgical technique in detail. Through a 2.5 cm extra-mammary axillary line incision a subcutaneous flap is first developed. A single port access system (Access Transformer OCTO; Seoul, Korea) was used to dock the robotic arms and to keep the robotic arm elbows from clashing. CO2 insufflation created the necessary working space and retraction on the mastectomy flap. The dissection was performed using Maryland Bipolar Forceps and monopolar cautery. Subsequently, few additional publications have described the use of RNSM (Table 3) [[8], [9], [10], [11], [12], [13], [14], [15], [16], [17], [18], [19], [20]]. Reported complications include nipple areolar complex necrosis, mastectomy flap necrosis, temporary skin blistering, hematoma, seroma, infection, loss of implant from infection, delayed axillary wound healing, transient brachial plexus neurapraxia (due to arm positioning), transient common peronal nerve neurapraxia (due to patient positioning during surgery). Thus far, no local recurrences have been reported with a mean follow-up of 20 months, although longer follow up studies are pending [31]. The learning curve for RNSM ranged from 8 to 13 cases using the cumulative sum plot method [16,20].
Table 3

Summary of published results on robot-assisted nipple sparing mastectomy as of December 2019 (case reports excluded).

AuthorYear of publicationNumber of casesMastectomy flap or nipple areolar complex necrosisOverall complication rateComplication requiring reoperation
Toesca [8]201729000
Sarfati [12]20186304.8%3.2%
Lai [16]2018395.1%30.8%2.6%
Park [20]20191225%25%0
Houvenaghel [17]2019273.7%40.7%Not reported
Toesca [31]2019941.1%22.3%4.3%
Summary of published results on robot-assisted nipple sparing mastectomy as of December 2019 (case reports excluded).

Future studies in robotic mastectomy

Despite these early successes, the safety of robotic oncologic surgery has been called to question by the FDA and some of this may be due to inferior outcomes in minimally invasive surgery compared to open radical hysterectomy for cervical cancer (Laparoscopic Approach to Cervical Cancer [LACC] Trial) [32]. It is important to note that in this study there were no randomization between laparoscopic versus robotic approach and of the minimally invasive surgery cases only 15.6% were performed with robotic-assistance (N = 45). There are speculations that the use of uterine or cervical manipulators and carbon dioxide pneumoperitonium promoted tumor recurrence [33]. However, other abdominal robotic surgeries that require carbon dioxide such as minimally invasive pancreatectomy and proctectomy report comparable rates of overall survival compared to open surgery [34,35]. In response to the safety concerns, our institution has received FDA approval of an Investigational Device Exemption (IDE) to initiate a clinical study to evaluate the safety of RNSM (clinicaltrials.gov NCT04537312). This type of research is necessary to demonstrate the safety and feasibility of RNSM before this technique can be incorporated into routine practice. Currently, there are numerous ongoing trials outside of the United States evaluating outcomes after RNSM [[36], [37], [38], [39], [40]], including a randomized study comparing open to robot-assisted NSM [41]. These ongoing international studies may help elucidate the long-term oncologic safety of RNSM. Furthermore, future studies assessing the mastectomy skin flap viability may demonstrate low rates of flap complication and thus improved success of immediate breast reconstruction. Although current published number of cases are small, there is a trend for low skin flap necrosis [8,12,16,17,31]. While the same anatomic planes are adhered to during RNSM compared to open, the skin retraction with CO2 insufflation rather than physical retraction may be less traumatic through the duration of the surgery. Critical evaluation of the mastectomy skin flap in future studies is still needed.

Conclusion

Herein, we describe the first study evaluating the presence of RBT after robotic mastectomy. RNSM was technically feasible and presence of RBT was comparable to that reported in previous published literature on open NSM. The most common location of RBT, as expected, was centrally in the tissue behind the NAC. Further studies are underway to evaluate the safety and long-term oncologic outcomes of RNSM.
  3 in total

1.  Prospective pilot study protocol evaluating the safety and feasibility of robot-assisted nipple-sparing mastectomy (RNSM).

Authors:  Ko Un Park; Sandy Lee; Angela Sarna; Matthew Chetta; Steven Schulz; Doreen Agnese; Valerie Grignol; William Carson; Roman J Skoracki
Journal:  BMJ Open       Date:  2021-11-15       Impact factor: 2.692

2.  Novel technique for endoscopic-assisted nipple-sparing mastectomy and immediate breast reconstruction with endoscopic-assisted latissimus dorsi muscle flap harvest through a single axillary incision: a retrospective cohort study of comparing endoscopic and open surgery.

Authors:  Juanjuan Qiu; Nan Wen; Yanyan Xie; Yu Feng; Faqing Liang; Qing Lv; Zhenggui Du
Journal:  Gland Surg       Date:  2022-08

3.  Safe Technical Innovation: Development and Implementation of a Robotic Breast Operation Program.

Authors:  D Brock Hewitt; Ko Un Park
Journal:  Ann Surg Open       Date:  2022-06-24
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.