Accuracy of a chairside intraoral scanner compared with a laboratory scanner for the completely edentulous maxilla: An in vitro 3-dimensional comparative analysis.

Fernando Zarone1, Gennaro Ruggiero2, Marco Ferrari3, Francesco Mangano4, Tim Joda5, Roberto Sorrentino6.   

Abstract

STATEMENT OF PROBLEM: Intraoral scanners are promising options for removable prosthodontics. However, analog aids, including occlusion rims, are still used, as a completely digital workflow is challenging and scientific evidence on the topic is scarce.
PURPOSE: The purpose of this in vitro study was to assess and compare the trueness and precision of scans obtained from a reference typodont of a completely edentulous maxilla by using an intraoral scanner (TRIOS 3 Pod; 3Shape A/S) with scans obtained by using a laboratory scanner (DScan 3; EGS S.R.L.) from both Type IV stone casts and polysulfide impressions.
MATERIAL AND METHODS: The polyurethane resin reference typodont was replicated from a clinical cast and was scanned with a metrological machine to obtain a reference scan. Ten digital casts were obtained by applying standardized scanning strategies to the reference typodont with the intraoral scanner. A device was created to make 10 consistent polysulfide impressions, and a scan of each impression was made with the laboratory scanner and then digitally reversed to obtain 10 digital reversed casts. Ten Type IV stone casts were poured and then scanned with the laboratory scanner to obtain 10 digital extraoral scanner casts. The scans in standard tessellation language (STL) format were imported into a dedicated software program, and the trueness and precision were calculated in μm. In addition to descriptive statistics (confidence interval 95%), 1-way ANOVA followed by the Bonferroni test or the Kruskal-Wallis and the Dunn tests were used to analyze differences among groups (α=.05).
RESULTS: The trueness values (95% confidence interval) were digital intraoral scanner cast=48.7 (37.8-59.5), digital reversed cast=249.9 (121.3-378.5), and digital extraoral scanner cast=308.8 (186.6-430.9); significant differences were detected between digital intraoral scanner cast and digital reversed cast (P<.001) and between digital IOS casts and digital extraoral scanner cast (P<.001). The precision values (95% confidence interval) were digital intraoral scanner cast=46.7 (29.7-63.7), digital reversed cast=271.2 (94.6-447.8), and digital extraoral scanner cast=341.4 (175.5-507.3); significant differences were detected between digital intraoral scanner cast and digital reversed cast (P=.003) and between digital intraoral scanner cast and digital extraoral scanner cast (P=.001).
CONCLUSIONS: Directly scanning a solid typodont of a completely edentulous maxilla with the intraoral scanner produced better trueness and precision than scanning the polysulfide impressions or the stone casts with a laboratory scanner.
Copyright © 2020 Editorial Council for the Journal of Prosthetic Dentistry. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2020        PMID: 33289647     DOI: 10.1016/j.prosdent.2020.07.018

Source DB:  PubMed          Journal:  J Prosthet Dent        ISSN: 0022-3913            Impact factor:   3.426


  2 in total

1.  Accuracy of Mini-Implant Placement Using a Computer-Aided Designed Surgical Guide, with Information of Intraoral Scan and the Use of a Cone-Beam CT.

Authors:  Georgios Vasoglou; Ioanna Stefanidaki; Konstantinos Apostolopoulos; Evmorfia Fotakidou; Michail Vasoglou
Journal:  Dent J (Basel)       Date:  2022-06-08

2.  Marginal and internal fit and intaglio surface trueness of interim crowns fabricated from tooth preparation of four finish line locations.

Authors:  Keunbada Son; Young-Tak Son; Ji-Min Lee; Kyu-Bok Lee
Journal:  Sci Rep       Date:  2021-07-06       Impact factor: 4.379

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.