Literature DB >> 33282524

Flexible A-scan rate MHz-OCT: efficient computational downscaling by coherent averaging.

Tom Pfeiffer1,2, Madita Göb1, Wolfgang Draxinger1, Sebastian Karpf1, Jan Philip Kolb1,3, Robert Huber1.   

Abstract

In order to realize adjustable A-scan rates of fast optical coherence tomography (OCT) systems, we investigate averaging of OCT image data acquired with a MHz-OCT system based on a Fourier Domain Mode Locked (FDML) laser. Increased system sensitivity and image quality can be achieved with the same system at the cost of lower imaging speed. Effectively, the A-scan rate can be reduced in software by a freely selectable factor. We demonstrate a detailed technical layout of the strategies necessary to achieve efficient coherent averaging. Since there are many new challenges specific to coherent averaging in swept source MHz-OCT, we analyze them point by point and describe the appropriate solutions. We prove that coherent averaging is possible at MHz OCT-speed without special interferometer designs or digital phase stabilization. We find, that in our system up to ∼100x coherent averaging is possible while achieving a sensitivity increase close to the ideal values. This corresponds to a speed reduction from 3.3 MHz to 33 kHz and a sensitivity gain of 20 dB. We show an imaging comparison between coherent and magnitude averaging of a human finger knuckle joint in vivo with 121 dB sensitivity for the coherent case. Further, the benefits of computational downscaling in low sensitivity MHz-OCT systems are analyzed.
© 2020 Optical Society of America under the terms of the OSA Open Access Publishing Agreement.

Entities:  

Year:  2020        PMID: 33282524      PMCID: PMC7687947          DOI: 10.1364/BOE.402477

Source DB:  PubMed          Journal:  Biomed Opt Express        ISSN: 2156-7085            Impact factor:   3.732


  28 in total

1.  High-speed wavelength-swept semiconductor laser with a polygon-scanner-based wavelength filter.

Authors:  S H Yun; C Boudoux; G J Tearney; B E Bouma
Journal:  Opt Lett       Date:  2003-10-15       Impact factor: 3.776

2.  Multi-megahertz OCT: High quality 3D imaging at 20 million A-scans and 4.5 GVoxels per second.

Authors:  Wolfgang Wieser; Benjamin R Biedermann; Thomas Klein; Christoph M Eigenwillig; Robert Huber
Journal:  Opt Express       Date:  2010-07-05       Impact factor: 3.894

3.  Sensitivity advantage of swept source and Fourier domain optical coherence tomography.

Authors:  Michael Choma; Marinko Sarunic; Changhuei Yang; Joseph Izatt
Journal:  Opt Express       Date:  2003-09-08       Impact factor: 3.894

4.  Motion artifacts in optical coherence tomography with frequency-domain ranging.

Authors:  S H Yun; G Tearney; J de Boer; B Bouma
Journal:  Opt Express       Date:  2004-06-28       Impact factor: 3.894

5.  Intravascular optical coherence tomography imaging at 3200 frames per second.

Authors:  Tianshi Wang; Wolfgang Wieser; Geert Springeling; Robert Beurskens; Charles T Lancee; Tom Pfeiffer; Antonius F W van der Steen; Robert Huber; Gijs van Soest
Journal:  Opt Lett       Date:  2013-05-15       Impact factor: 3.776

6.  Ultrahigh-speed non-invasive widefield angiography.

Authors:  Cedric Blatter; Thomas Klein; Branislav Grajciar; Tilman Schmoll; Wolfgang Wieser; Raphael Andre; Robert Huber; Rainer A Leitgeb
Journal:  J Biomed Opt       Date:  2012-07       Impact factor: 3.170

7.  Compensation of spectral and RF errors in swept-source OCT for high extinction complex demodulation.

Authors:  Meena Siddiqui; Serhat Tozburun; Ellen Ziyi Zhang; Benjamin J Vakoc
Journal:  Opt Express       Date:  2015-03-09       Impact factor: 3.894

8.  Strategies to improve phase-stability of ultrafast swept source optical coherence tomography for single shot imaging of transient mechanical waves at 16 kHz frame rate.

Authors:  Shaozhen Song; Wei Wei; Bao-Yu Hsieh; Ivan Pelivanov; Tueng T Shen; Matthew O'Donnell; Ruikang K Wang
Journal:  Appl Phys Lett       Date:  2016-05-10       Impact factor: 3.791

9.  Phase-stable swept source OCT angiography in human skin using an akinetic source.

Authors:  Zhe Chen; Mengyang Liu; Michael Minneman; Laurin Ginner; Erich Hoover; Harald Sattmann; Marco Bonesi; Wolfgang Drexler; Rainer A Leitgeb
Journal:  Biomed Opt Express       Date:  2016-07-12       Impact factor: 3.732

10.  Spectral domain optical coherence tomography of multi-MHz A-scan rates at 1310 nm range and real-time 4D-display up to 41 volumes/second.

Authors:  Dong-Hak Choi; Hideaki Hiro-Oka; Kimiya Shimizu; Kohji Ohbayashi
Journal:  Biomed Opt Express       Date:  2012-11-01       Impact factor: 3.732

View more
  2 in total

1.  Continuous spectral zooming for in vivo live 4D-OCT with MHz A-scan rates and long coherence.

Authors:  Madita Göb; Tom Pfeiffer; Wolfgang Draxinger; Simon Lotz; Jan Philip Kolb; Robert Huber
Journal:  Biomed Opt Express       Date:  2022-01-13       Impact factor: 3.732

2.  Resolution-enhanced OCT and expanded framework of information capacity and resolution in coherent imaging.

Authors:  Nichaluk Leartprapun; Steven G Adie
Journal:  Sci Rep       Date:  2021-10-15       Impact factor: 4.379

  2 in total