| Literature DB >> 33278855 |
Haitham Saeed1, Hasnaa Osama1, Yasmin M Madney1, Hadeer S Harb1, Mona A Abdelrahman1, Carsten Ehrhardt2, Mohamed E A Abdelrahim1.
Abstract
BACKGROUND: The new coronavirus (SARS-COV-2) that emerged at the end of 2019 was stated in China and infected millions of people around the world, with the highest spread rate amongst humans compared with other coronaviruses. This paper aimed to review and analyse the published studies about COVID-19 diagnosis, prevention, and treatment.Entities:
Mesh:
Year: 2021 PMID: 33278855 PMCID: PMC7883281 DOI: 10.1111/ijcp.13886
Source DB: PubMed Journal: Int J Clin Pract ISSN: 1368-5031 Impact factor: 3.149
Target level of oxygen saturation for different COVID‐19 patients
| Condition | Target level |
|---|---|
| Adult and non‐pregnant patient | SpO2 ≥ 90% |
| Pregnant patient | SpO2 ≥ 92%‐95% |
| Children | SpO2 ≥ 90% |
| Children or adults with dyspnea, severe respiratory distress, apnoea, coma, shock, and central cyanosis | SpO2 ≥ 94% |
Factors affecting aerosol delivery during mechanical ventilation and recommended settings , ,
| Factor | Recommendations |
|---|---|
| Nebulizer type | It is preferred to use VMN instead of traditional jet nebulizer (JN) as the VMN drug chamber is completely separated from the exhaled air pathway. |
| If available use VMN for aerosol delivery during mechanical ventilation. | |
| Nebulizer position | Placing the VMN prior to the humidifier could improve the efficiency of the aerosol therapy and further reduce retrograde contamination from the infected subject. |
| JN (if used) should be placed close to the ventilator. | |
| Position of pMDI attached to spacer | pMDI + spacer should be placed at least 15 cm away from the endotracheal tube. |
| Nebulizer fill volume | Add at least 2 ml of the vehicle to the drug for JN. |
| Adding saline to the nebulized drug solution would increase the delivered fraction of the inhalable drug. | |
| Increasing the fill volume of VMN is not important. | |
| Residual dead volume | The dead volume that remains in the nebulizer chamber affects the amounts of aerosol that could be delivered to the patient. Also, this amount of drug could be contaminated with microorganisms in case of jet nebulizers. |
| In case of using JN, it should be disinfected and discarded carefully after end of nebulization. | |
| VMNs are recommended if available. | |
| Filters | It is recommended to Attach a HEPA filter to the expiratory limb of the ventilator to reduce the exposure to the fugitive aerosol and avoid the transmission of contaminated droplets through the ventilator circuit. |