Literature DB >> 33261734

Patient-specific computational fluid dynamics of femoro-popliteal stent-graft thrombosis.

Michele Conti1, Anna Ferrarini2, Alice Finotello3, Giancarlo Salsano4, Ferdinando Auricchio2, Domenico Palombo5, Giovanni Spinella5, Bianca Pane5.   

Abstract

Intra-stent thrombosis is one of the major failure modes of popliteal aneurysm endovascular repair, especially when the diseased arterial segment is long and requires overlapping stent-grafts having different nominal diameters in order to accommodate the native arterial tapering. However, the interplay between stent sizing, post-operative arterial tortuosity, luminal diameter, local hemodynamics, and thrombosis onset is not elucidated, yet. In the present study, a popliteal aneurysm was treated with endovascular deployment of two overlapped stent-grafts, showing intra-stent thrombosis at one-year follow-up examination. Patient-specific computational fluid-dynamics analyses including straight- and bent-leg position were performed. The computational fluid-dynamics analysis showed that the overlapping of the stent-grafts induces a severe discontinuity of lumen, dividing the stented artery in two regions: the proximal part, affected by thrombosis, is characterized by larger diameter, low tortuosity, low flow velocity, low helicity, and low wall shear stress; the distal part presents higher tortuosity and smaller lumen diameter promoting higher flow velocity, higher helicity, and higher wall shear stress. Moreover, leg bending induces an overall increase of arterial tortuosity and reduces flow velocity promoting furtherly the luminal area exposed to low wall shear stress.
Copyright © 2020. Published by Elsevier Ltd.

Entities:  

Keywords:  Endovascular treatment; Femoro-popliteal segment; Helicity; Intra-stent thrombosis; Limb flexion; Medical image analysis; Peripheral stenting; Popliteal artery aneurysm

Year:  2020        PMID: 33261734     DOI: 10.1016/j.medengphy.2020.10.011

Source DB:  PubMed          Journal:  Med Eng Phys        ISSN: 1350-4533            Impact factor:   2.242


  2 in total

1.  A Computational Framework Examining the Mechanical Behaviour of Bare and Polymer-Covered Self-Expanding Laser-Cut Stents.

Authors:  Ciara G McKenna; Ted J Vaughan
Journal:  Cardiovasc Eng Technol       Date:  2021-11-30       Impact factor: 2.305

2.  Hemodynamic Parameters Predict In-stent Thrombosis After Multibranched Endovascular Repair of Complex Abdominal Aortic Aneurysms: A Retrospective Study of Branched Stent-Graft Thrombosis.

Authors:  Ming-Yuan Liu; Yang Jiao; Junjun Liu; Simeng Zhang; Wei Li
Journal:  Front Cardiovasc Med       Date:  2021-04-23