Supraclavicular versus infraclavicular approach in inserting totally implantable central venous access for cancer therapy: A comparative retrospective study.

Amine Souadka1, Hajar Essangri1, Imad Boualaoui1, Abdelilah Ghannam2, Amine Benkabbou1, Laila Amrani1, Raouf Mohsine1, Mohammed Anass Majbar1.   

Abstract

INTRODUCTION: The insertion of an implantable central venous access is performed according to a variety of approaches which allow the access to the subclavian vein, yet the supraclavicular technique has been underused and never compared to the other methods. The aim of this study was to testify on the efficacy and safety of the subclavian puncture without ultrasound guidance « Yoffa » in comparison with the classical infraclavicular approach (ICA).
MATERIAL AND METHODS: This is a retrospective study with prospective data collection on patients followed at the national oncology institute for cancer, in the period extending from May 1st 2017 to August 31st 2017. All patients had a totally implantable central venous access device inserted by the same surgeon AS for chemotherapy administration and demographic characteristics, as well as procedure details were examined. The primary outcomes were the intraoperative complications, while the secondary outcomes represented immediate postoperative and mid-term complications (at 15 months of follow up). Outcomes were compared between techniques by means of non parametric tests and the Fischer test.
RESULTS: Our study included 135 patients with 70 patients undergoing the subclavian technique, while 65 were subject to the infraclavicular approach. Both groups had no statistically significant demographic characteristics. The number of vein puncture attempts exceeding once, the accidental artery puncture and operative time were more significant in the ICA group; (39,6 vs 17,6 p = 0,01) (9.2% vs 0; p = 0,01) and (27± 13 vs 23± 8min, p = 0.045) respectively. There was no statistically significant difference in the immediate and midterm complication rate between the two methods 1(1,4) vs 2 (3) p = 0.5.
CONCLUSION: In case of unavailability of ultrasonographic guidance, the use of the supra-clavicular landmarks approach is linked to higher success rates and less arterial punctures, thereby proving to be a safe and reliable approach.

Entities:  

Mesh:

Year:  2020        PMID: 33232361      PMCID: PMC7685501          DOI: 10.1371/journal.pone.0242727

Source DB:  PubMed          Journal:  PLoS One        ISSN: 1932-6203            Impact factor:   3.240


Introduction

The first placement of a totally implantable central venous access device (TIVAD) was performed in 1982 using the cephalic vein and ever since, TIVADs are standardly used in the oncology field for the administration of chemotherapeutic agents, intravenous hydration, parenteral nutrition and blood withdrawal. This tool has proven to provide constant and reliable parenteral access and reduce to some level the vascular complications of cancer chemotherapy, as well as improve the overall quality of life of these patients [1]. Yet many aspects of TIVADs use are still a matter of discussion, namely the choice of device, vein and technique of insertion. The most frequently used central venous catheterization approaches vary between the internal jugular used for resuscitation and the subclavian vein which is often adopted during surgical procedures. The access to the subclavian vein can either be possible through the infra-clavicular or the supra-clavicular approach, with the latter being less frequently used. This supraclavicular technique for central vein puncture, first described In 1965 by Yoffa [2], was mainly used in emergency resuscitation by anaesthesiologist as a quicker way to central line catheterization, allowing higher first attempt success rate and accessibility during CPR, tube thoracostomy and active surgery [3, 4]. In addition, the lack of resources such as ultrasonographic (US) guidance—which has been standardized in TIVAD and central line access insertion—has put the blind approach relying on landmark localization for vein puncture on the front line as well as prompted investigating these methods [5-7]. We wanted to examine the potential benefits of the supraclavicular puncture in terms of efficacy, safety and intra interventional pain perception. The aim of this study is to compare the supraclavicular puncture to the classical infra-clavicular approach, both without ultrasound guidance and demonstrate the efficacy and safety of this underused technique.

Methods

Patients

This is a retrospective analysis of a prospective database conducted at the National Institute of Oncology (NIO) of Rabat, Morocco in the period from May 1st 2017 to August 31st 2017. All patients with the indication of systemic chemotherapy and TIVAD insertion by a single senior surgeon (A.S) were eligible and admitted in a day-hospital setting. Only Patients meeting the following criteria were included: patients aged > 18 year old, with an Eastern Cooperative Oncology Group (ECOG) performance status of 0 to 2, able to bear a lying position for at least 45 minutes and diagnosed with mainly digestive or gynecologic cancer. Patients with superior vena cava syndrome, active infection, coagulopathy (defined as a platelet count <50 000/l and/or a prothrombin rate < 50%), life expectancy <6 months or the inability to sign the written informed consent were excluded.

Implantable ports and access to central vein

Patients underwent the insertion of a single type of silastic port with a silicone membrane and connected to an 8F silicone rubber catheter (Bard Port, Bard Inc., Salt Lake City, UT) using a percutaneous landmark access approach to the subclavian vein by a single surgeon skilled in both techniques. Both approaches are standard procedures in our institution. Patients were divided in 2 groups, depending on the TIVAD implantation technique and period in which they underwent the procedure. Group A patients received TIVAD through a supraclavicular approach and group B through an infraclavicular approach in the first and second half of our study period respectively (Table 1). The right side was attempted whenever possible. At the 3rd failed attempt, the surgeon would try the other puncture site and in case this second route was also unsuccessful, patients were excluded from the analysis.
Table 1

Demographics of patients.

Group A (YOFFA)Group B (ICA)P value
N = 70N = 65
Mean age (years) ± DS54 ± 1651 ± 160.427
Gender N (%)0.9
    • Female34 (38.6)31 (47.7)
    • Male36 (51.4)34 (52.3)
ASA0,88
139 (55)37 (57)
2–331 (45)28 (43)
Mean BMI (Kg/m2)26 ± 3,224,6 ± 2,70,67
Side of puncture, N (%)0.512
    • Right60 (86)53 (81)
    • Left10 (14)12 (19)
Neoplastic Disease N (%)0.56
    • Digestive45 (64.3)40 (61.5)
    • Gynecologic18 (25.7)21 (32.3)
    • Others7 (10)4 (4.2)
Chemotherapy administration before TIVAD insertion, N (%)0,06
Yes24 (34,2)13 (20)
No46 (65,8)52 (80)

Technique

Supraclavicular approach

Patients are placed in a Trendelenburg’s position with slight shoulders extension and the procedure is performed under local anesthesia (20 cc of lidocaine 1%), following the previously described technique by David YOFFA [2]. The key success factor is the correct identification of the clavisternomastoid angle landmark formed by the junction of the lateral head of the sternocleidomastoid muscle and the clavicle using active rising of the patient’s head. The site of needle insertion is 0.5 to 1cm lateral to the clavicular head of the sternocleidomastoid muscle and 1 cm posterior to the clavicle. The needle is oriented at a 45° angle to the sagittal and transverse planes and a 15° angle below the coronal plane with the contralateral nipple as a target for directing the introducer [8]. Subsequently to the vein puncture, a guidewire is inserted through the needle which is withdrawn and replaced by a ‘peel-away sheath’ introducer the time of the completion of the second step of the procedure by the surgeon.

Infraclavicular approach

Following the same positioning and initial steps, the landmark of needle insertion for this approach is 1 cm lateral to the middle third of the clavicle with a 10 to 15° incline from the coronal plane to the direction of the sternal notch or the contralateral acromioclavicular articulation. In order to decrease malposition risk of the guidewire in the ipsilateral Internal jugular vein, a sterile finger may be placed in the corresponding supraclavicular fossa.

Port, catheter insertion and control

A surgically dissected space is prepared for port placement, while being connected to the catheter. A first surgical time (t1) is marked by skin incision and insertion of a tunneller from the dissected space to the venipuncture site to ensure a subcutaneous path for the catheter, as well as a pocket for the venous access device. The second surgical time (t2), is marked by the advancement of the ‘peel-away sheath’ dilator into the vein. Following the removal of the dilator, the catheter is inserted through the vein and the port is then flushed by 20cc of saline after checking for blood reflux. A chest X-ray was routinely performed following the catheter placement in order to visualize the location at the lower or middle part of the superior vena cava and rule out a pneumothorax. Antibiotic prophylaxis was not routinely administered.

Endpoints

Primary outcomes were the intraoperative result defined by a rate of successful first attempt, the number of accidental artery puncture, the occurrence of an intraoperative incident that led to change in technique (ex: local Hematoma), difficulties identifying the landmarks as well as the total operative time and pain perception. During the procedure each patient was asked to individually rank the intra-interventional pain perception on a visual analogue scale (1 = no pain and 10 = maximum imaginable pain) at (t1) and (t2). Secondary outcomes were surgery related complications either categorized as early complications, defined by the occurrence of infection, a hematoma or pneumothorax in the first week following the procedure and prior to the first use of the device, or late complication after assessment in March 2018 such as infection, thrombotic obstruction, fracture embolism (Fat emboli) or patient death leading to the non-use or removal of the device.

Statistical evaluation

Continuous variables were presented with a mean value ± SD or a median with interquartile range and categorical variables were expressed with frequencies and percentages. For the pain perception evaluation, a Wilcoxon t-test was used. Categorical variables and the comparison of the two groups was performed using the χ2 test or Fisher's exact test when appropriate. A p value of <0.05 was considered statistically significant. All statistical analysis was performed using SPSS software (SPSS 13.0; SPSS Inc, Chicago, IL).

Results and discussion

Results

During this study period, one hundred and thirty five patients underwent the successful placement of a TIVAD prior to chemotherapy with seventy (51.9%) patients in the group A (using the supraclavicular approach) versus sixty-five (48.1%) patients in the group B (using the infraclavicular approach). There were no significant differences between the two groups concerning mean age, gender, American Society of Anesthesiologists (ASA) score, Body mass index (BMI), neoplastic disease and the side of puncture. Further demographic characteristics are shown in Table 1. Regarding intraoperative outcomes, the first attempt success rate was 82.4% (56/70) vs. 61% (39/65) for group A and B respectively(p = 0.01). Group A had a null rate of accidental artery punctures compared to group B 6/65 (9.2%) with p = 0.01. The procedure was significantly longer in duration for group B (27± 13 vs 23± 8min, p = 0.045) and the surgeon had to switch the techniques in 9 cases vs 2 cases in group B and A respectively. This is mainly due to a failure at the 3rd attempt in 5/65 of cases from group B compared to 1/70 cases in group A. Table 2.
Table 2

Outcomes comparison between the Yoffa and the ICA technique groups.

Group A (Yoffa)Group B (ICA)P
Success at first attempt, N (%)56 (82.4)39 (61)0.01
Accidental artery puncture, N (%)0 (0)6 (9.2)0.01
Mean operative time (±SD)23±827± 130.042
Success rate, N (%)68 (97.25)56 (86.2)0.019
Cause of changing site of puncture,
N (%)2 (2.85)9 (13.8)0.019
    • Hematoma14
    • Failure at the 3rd attempt15
Median Pain perception rate in scale from 1 to 10 (quartiles)
At (T1)1 (0–3)0 (0–2)0.001
At (T2)1 (1–4)3 (1–5)0.001
As regards pain perception, patients who chose the supraclavicular approach (groupe A) reported a more significant pain sensation at the time of the tunneller insertion from the dissected space to the venipuncture site (t1). Opposingly, the same group described less pain during the insertion of the dilator over the guidewire and into the vein. No TIVAD insertion related death was observed, thus all patients completed at least 15 months of follow up. Table 3 is a representation of both early and late postoperative outcomes for the two techniques. Only one early infection occurred in group A and one device defunctioning in group B, which required the interruption of the TIVAD use during the antibiotic treatment period for the first case and the device location change to the contralateral side in the other. No case of thrombosis was described after the 15 months follow up period, however, a pinch off syndrome was observed in group B leading to a late change to the second technique.
Table 3

Early and late complications (follow up 15 months) of surgical insertion of TIVAD.

Groupe AGroupe B
Early complications
Infection1 (1.4)0
dysfonction01(1.5)
Late complications
Thrombosis00
Pinch Off syndrom01 (1.5)

Discussion

This study showed that the supraclavicular approach to TIVAD implantation is more successful (97.2% vs 86.2), shorter in duration (23 vs 27 min) and has a higher rate of success at first attempt with less arterial puncture incidents than the infraclavicular approach. This approach offers many advantages compared to the infraclavicular puncture technique due to its easily localizable landmarks even in obese patients, the shorter needle insertion depth (smaller skin vein distance), the larger target area, the straighter route to the vena cava and reduced risk of lung, pleural or arterial injuries [9-12]. This explains the significantly higher success at first attempt rate of 86% and the global success rate of 97.2%, alongside the very few early and late complications. Only three comparative studies inspected the supraclavicular and infraclavicular approaches, with success rates ranging from 84.5% to 93.3% and 80% to 87% for both approaches respectively. The rates of complications were heterogeneous depending on the definitions [3, 13, 14]. Very few data in the literature addresses patient pain perspective in this type of procedure with only one study demonstrating the jugular catheter placement to be significantly less painful than the infra-clavicular insertion using the subjective pain scale [15]. Our study categorized pain assessment according to the two different painful moment of the procedure: t1 being the time of tunneller insertion ensuring the subcutaneous path of the catheter, which was not significantly different in the two groups despite the longer distance from the dissected space to the vein puncture in the supraclavicular approach. Furthermore, the second surgical time t2, corresponding to the ‘peel-away sheath’ dilator insertion was significantly more painful in the subclavian approach which could probably be due to the proximity of the pleura and lung, as well as the length of the distance from the skin to the vein [9]. Although the use of ultrasound guidance (US) represents a significant advance in central line placement, following this approach in the subclavian vein puncture often includes some additional risks [16]. Although US guided procedures could increase the anatomical perception of the performers [17, 18], the false sense of security, routinely reliance on this tool and the complication risk resulting from the effect on healthcare professionals’ landmark-based techniques could also be argued [19]. For this reason, landmark based approaches remain a skill physicians need to have in their armamentarium, which applies to TIVAD insertion as well. Limitations of this study are the retrospective aspect, absence of randomization and the fact that the results of this study were directly dependent on the senior surgeon's experience and skill. Therefore, caution is required when interpreting and applying our results to procedures where novices or physicians, especially at the beginning of their learning curve, are attempting to perform successfully in both approaches. An additional limitation is the relatively small number of patients included, however, this will allow us to decide the required sample size for future studies addressing this technique.

Conclusions

Nowadays, central venous access is performed under ultrasound guidance, which is not always possible in resource-restricted environments. In this case, the supraclavicular line is an attractive approach that appears to be safe and possibly easier to perform, with easier localizable landmarks and less misplacement than more frequently used lines and. As such, the supraclavicular approach could challenge more traditional techniques and more studies investigating this alternative are required. (SAV) Click here for additional data file. 1 Jun 2020 Submitted filename: PLOS one comments .docx Click here for additional data file. 28 Aug 2020 PONE-D-20-14056 Supraclavicular versus infraclavicular approach in inserting totally implantable central venous access for cancer therapy : a comparative study on safety, efficacy and intra interventional pain perception PLOS ONE Dear Dr. Souadka, Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process. Please address the reviewers' concerns and revise the manuscriopt accordingly. Please submit your revised manuscript by Oct 12 2020 11:59PM. If you will need more time than this to complete your revisions, please reply to this message or contact the journal office at plosone@plos.org. When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file. Please include the following items when submitting your revised manuscript: A rebuttal letter that responds to each point raised by the academic editor and reviewer(s). You should upload this letter as a separate file labeled 'Response to Reviewers'. A marked-up copy of your manuscript that highlights changes made to the original version. You should upload this as a separate file labeled 'Revised Manuscript with Track Changes'. An unmarked version of your revised paper without tracked changes. You should upload this as a separate file labeled 'Manuscript'. If you would like to make changes to your financial disclosure, please include your updated statement in your cover letter. Guidelines for resubmitting your figure files are available below the reviewer comments at the end of this letter. If applicable, we recommend that you deposit your laboratory protocols in protocols.io to enhance the reproducibility of your results. Protocols.io assigns your protocol its own identifier (DOI) so that it can be cited independently in the future. For instructions see: http://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols We look forward to receiving your revised manuscript. Kind regards, Robert Jeenchen Chen, MD, MPH Academic Editor PLOS ONE Journal Requirements: When submitting your revision, we need you to address these additional requirements. 1. Please ensure that your manuscript meets PLOS ONE's style requirements, including those for file naming. The PLOS ONE style templates can be found at https://journals.plos.org/plosone/s/file?id=wjVg/PLOSOne_formatting_sample_main_body.pdf and https://journals.plos.org/plosone/s/file?id=ba62/PLOSOne_formatting_sample_title_authors_affiliations.pdf 2.In your Data Availability statement, you have not specified where the minimal data set underlying the results described in your manuscript can be found. PLOS defines a study's minimal data set as the underlying data used to reach the conclusions drawn in the manuscript and any additional data required to replicate the reported study findings in their entirety. All PLOS journals require that the minimal data set be made fully available. For more information about our data policy, please see http://journals.plos.org/plosone/s/data-availability. Upon re-submitting your revised manuscript, please upload your study’s minimal underlying data set as either Supporting Information files or to a stable, public repository and include the relevant URLs, DOIs, or accession numbers within your revised cover letter. For a list of acceptable repositories, please see http://journals.plos.org/plosone/s/data-availability#loc-recommended-repositories. Any potentially identifying patient information must be fully anonymized. Important: If there are ethical or legal restrictions to sharing your data publicly, please explain these restrictions in detail. Please see our guidelines for more information on what we consider unacceptable restrictions to publicly sharing data: http://journals.plos.org/plosone/s/data-availability#loc-unacceptable-data-access-restrictions. Note that it is not acceptable for the authors to be the sole named individuals responsible for ensuring data access. We will update your Data Availability statement to reflect the information you provide in your cover letter. 3.Thank you for stating the following financial disclosure: [No]. At this time, please address the following queries: Please clarify the sources of funding (financial or material support) for your study. List the grants or organizations that supported your study, including funding received from your institution. State what role the funders took in the study. If the funders had no role in your study, please state: “The funders had no role in study design, data collection and analysis, decision to publish, or preparation of the manuscript.” If any authors received a salary from any of your funders, please state which authors and which funders. If you did not receive any funding for this study, please state: “The authors received no specific funding for this work.” Please include your amended statements within your cover letter; we will change the online submission form on your behalf. 4.Thank you for stating the following in your Competing Interests section: [No]. Please complete your Competing Interests on the online submission form to state any Competing Interests. If you have no competing interests, please state "The authors have declared that no competing interests exist.", as detailed online in our guide for authors at http://journals.plos.org/plosone/s/submit-now This information should be included in your cover letter; we will change the online submission form on your behalf. Please know it is PLOS ONE policy for corresponding authors to declare, on behalf of all authors, all potential competing interests for the purposes of transparency. PLOS defines a competing interest as anything that interferes with, or could reasonably be perceived as interfering with, the full and objective presentation, peer review, editorial decision-making, or publication of research or non-research articles submitted to one of the journals. Competing interests can be financial or non-financial, professional, or personal. Competing interests can arise in relationship to an organization or another person. Please follow this link to our website for more details on competing interests: http://journals.plos.org/plosone/s/competing-interests [Note: HTML markup is below. Please do not edit.] Reviewers' comments: Reviewer's Responses to Questions Comments to the Author 1. Is the manuscript technically sound, and do the data support the conclusions? The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented. Reviewer #1: Yes Reviewer #2: Yes ********** 2. Has the statistical analysis been performed appropriately and rigorously? Reviewer #1: Yes Reviewer #2: Yes ********** 3. Have the authors made all data underlying the findings in their manuscript fully available? The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified. Reviewer #1: No Reviewer #2: No ********** 4. Is the manuscript presented in an intelligible fashion and written in standard English? PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here. Reviewer #1: Yes Reviewer #2: Yes ********** 5. Review Comments to the Author Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters) Reviewer #1: This is a study comparing the supraclavicular and infraclavicular approaches for central venous access and for placing implantable venous access device. Both approaches were applied using landmark technique (without ultrasound guidance). The main objective was to compare these two techniques in terms of procedural success and performance time. The authors found a better first pass success rate and shorter procedure time for supraclavicular approach compared to infraclavicular method. Plase find below my comments: Major strengths of the study: Comparison of one popular central access technique to a disregarded technique which may be a reliable alternative for placing TIVAD. Major weaknesses of the study and comments for authors: Ultrasound guidance is a standardized method for central venous access as the authors mentioned. In major institutions where TIVADs are placed, ultrasound guidance is essential for those procedures. The patients receiving the two procedures were not randomized which may lead to selection bias (as this was a retrospective cohort study). -Please mention type of the study method in the title. Introduction: The hypothesis and objective of the study should be explained in detail in the introduction part. Methods: -The sample size may not be sufficient to determine rarely seen early and late complications. -In the sentence: “Group A patients received TIVAD through a supraclavicular approach and group B through an infraclavicular approach in the first and second half of our study period respectively (see details below).” Which details below where? A table? Insertion techniques? Results -Duration of catheter placement was mentioned in discussion part but there is no data in the results. Discussion -Personally, I do not agree with the authors about the presumption that ultrasound guided catheter placement may lead to deskilling. Although this is arguable, I believe that us guided procedures increase anatomicaly perception of the performers regarding the anatomic structures as they see them every time they perform the procedure (see anatomic variations as well). I would like to declare that I have no conflict of interest. Reviewer #2: Thank you for the chance to read your paper - it has highlighted a technique for access of which I was not previously aware. The key limitations are already described by yourselves, and it is a shame as they would be 'easily' addressed by a longer study period and true randomisation rather than by time period. Major comments: Abstract: 1. "yet the supraclavicular technique has been underused and never compared to the other method". No true - you yourselves reference several published comparisons! 2. (1,4% vs 7% p=0,21). This is a large difference without significance. Ranges should be included with the data to understand why this is so. In the results table the figure is 1.4 vs 1.5. Typo? Introduction: The text needs clarity. The techniques you are talking about need to be made very clear in the introduction, as at the moment it is a little confusing for me as a reader. For example: Be clear in the abstract and the introduction that you are just assessing access to the subclavian vein. "Accordingly, the most frequent central venous catheterization approaches vary between the subclavian which is often adopted during surgical procedures and the internal jugular used for resuscitation. The use of the supraclavicular approach on the other hand is less usual. In fact, the supraclavicular technique for central vein puncture was first described In 1965 by Yoffa [2]." The phrasing of this paragraph makes it seem like you are discussing subclavian access to supraclavicular IJV access, due to structure. If unfamiliar with Yoffa's technique (which I was before reading your paper, so thank you for the opportunity to improve my knowledge on the topic), then it is unclear that your study is actually comparing the supra and infra clavicular approach to subclavian vein access alone. This could be clarified with a short section describing the two techniques you are comparing clearly. The Plos-one readership is generalist. Methods: 1. The time period for collection is short - By your results, the volume of cases in your centre is high. The numbers you have included exceed those in other literature. However your analysis would be even more powerful with greater numbers. 2. The study is not randomised. This is a major limitation, and could be resolved easily. I base this on the methodology stating the team used one technique first for a month, then simply switched over. Why not randomised? 3. The follow up for secondary outcomes is over a year earlier than the study period. Is this a typo? "Secondary outcomes were surgery related complications either categorized as early complications, defined by the occurrence of infection, a hematoma or pneumothorax in the first week following the procedure and prior to the first use of the device, or late complication after assessment in march 2016 such as infection," Results: You start by using the term subclavicular, whereas you've used infraclavicular throughout the text. Discussion: This statement is not accurate - reference 17 specifically identifies benefits of US guidance for subclavian access from a recent meta-analysis, although not significant. "US guidance is associated with frequent risk of posterior vessel wall penetration as well as more lateral puncture sites as in the landmark puncture technique which may lead to pleural injury." Minor comments: 1. Abstract: ICA is used without prior reference to the phrase it represents The complications being studied are the primary outcome (I assume?). Please include information in the abstract. Author needs to confirm - was the jugular approach also performed without US? Some minor spelling mistakes ********** 6. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files. If you choose “no”, your identity will remain anonymous but your review may still be made public. Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy. Reviewer #1: No Reviewer #2: Yes: RA Benson [NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files.] While revising your submission, please upload your figure files to the Preflight Analysis and Conversion Engine (PACE) digital diagnostic tool, https://pacev2.apexcovantage.com/. PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email PLOS at figures@plos.org. Please note that Supporting Information files do not need this step. 18 Sep 2020 Dear editors We would like to thank you for your feedback and inform you that all comments were followed. Firstly, we would like to inform you that we are submitting our minimal underlying data set as an additional supplementary file according to PLOS journals requirement. All potentially identifiable patient information was fully anonymized. Secondly, the responses to reviewer comments are the following: Reviewer #1: 1- Please mention the type of the study method in the title. The required change has been made. 2- Introduction: The hypothesis and objective of the study should be explained in detail in the introduction part. The required change has been made. 3- Methods: The sample size may not be sufficient to determine rarely seen early and late complications. Indeed the sample size is a limitation of our study which was mentioned in our limitations paragraph in the discussion. However, a future study with a bigger sample of patients should take place. 4- In the sentence: “Group A patients received TIVAD through a supraclavicular approach and group B through an infraclavicular approach in the first and second half of our study period respectively (see details below).” Which details below where? A table? Insertion techniques? The mention refers to the Table 1 in the results section and was rectified. 5- Results : Duration of catheter placement was mentioned in discussion part but there is no data in the results. The duration of catheter placement was already mentioned in paragraph 2 from the results section as well as the results section, Table 2, under the Mean operative time. 6- Discussion : Personally, I do not agree with the authors about the presumption that ultrasound guided catheter placement may lead to deskilling. Although this is arguable, I believe that us guided procedures increase anatomically perception of the performers regarding the anatomic structures as they see them every time they perform the procedure (see anatomic variations as well). This section was modified. Reviewer #2: 1- Abstract: "yet the supraclavicular technique has been underused and never compared to the other method". No true - you yourselves reference several published comparisons! indeed the different approaches in central line placement have been previously discussed, however, this is the first comparison of both techniques in port catheter implantation. 2- Abstract: (1,4% vs 7% p=0,21). This is a large difference without significance. Ranges should be included with the data to understand why this is so. In the results table the figure is 1.4 vs 1.5. Typo? This was a typo which we rectified. 3- The text needs clarity. The techniques you are talking about need to be made very clear in the introduction, as at the moment it is a little confusing for me as a reader. For example: Be clear in the abstract and the introduction that you are just assessing access to the subclavian vein. "Accordingly, the most frequent central venous catheterization approaches vary between the subclavian which is often adopted during surgical procedures and the internal jugular used for resuscitation. The use of the supraclavicular approach on the other hand is less usual. In fact, the supraclavicular technique for central vein puncture was first described In 1965 by Yoffa [2]." The phrasing of this paragraph makes it seem like you are discussing subclavian access to supraclavicular IJV access, due to structure. If unfamiliar with Yoffa's technique (which I was before reading your paper, so thank you for the opportunity to improve my knowledge on the topic), then it is unclear that your study is actually comparing the supra and infra clavicular approach to subclavian vein access alone. This could be clarified with a short section describing the two techniques you are comparing clearly. The Plos-one readership is generalist. The required change has been made to the abstract and introduction. 4- Methods: The time period for collection is short - By your results, the volume of cases in your centre is high. The numbers you have included exceed those in other literature. However your analysis would be even more powerful with greater numbers. We agree that a bigger sample would benefit and support this technique better. This retrospective study aimed to present a preliminary scientific basis for future randomized trials since no previous paper describes this approach for catheter implantation. 5- Methods: The study is not randomised. This is a major limitation, and could be resolved easily. I base this on the methodology stating the team used one technique first for a month, then simply switched over. Why not randomised? We agree with your opinion, however, it is difficult to receive ethical approval for a randomized trial on a technique which was never presented in a scientific publication. As such, we reported this series to present this approach for the first time in literature. 6- Methods: . The follow up for secondary outcomes is over a year earlier than the study period. Is this a typo? This was a typo which we rectified. 7- Results: You start by using the term subclavicular, whereas you've used infraclavicular throughout the text. The required change has been made and the same term used throughout the manuscript. 8- Discussion: This statement is not accurate - reference 17 specifically identifies benefits of US guidance for subclavian access from a recent meta-analysis, although not significant. "US guidance is associated with frequent risk of posterior vessel wall penetration as well as more lateral puncture sites as in the landmark puncture technique which may lead to pleural injury." This statement was removed from the manuscript. 9- Minor comments : Abstract: ICA is used without prior reference to the phrase it represents ; We referenced the abbreviation as requested. 10- The complications being studied are the primary outcome (I assume?). Please include information in the abstract. We specified our primary outcomes in the abstract as requested. 11- Author needs to confirm - was the jugular approach also performed without US? We would like to confirm that the jugular approach was also performed without US 12- Some minor spelling mistakes The manuscript grammar and spelling were revised. Submitted filename: Yoffa comments.docx Click here for additional data file. 20 Oct 2020 PONE-D-20-14056R1 Supraclavicular versus infraclavicular approach in inserting totally implantable central venous access for cancer therapy : a comparative study on safety, efficacy and intra interventional pain perception PLOS ONE Dear Dr. Souadka, Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process. Please revise the Title to be concise and informative. Please address the reviewers' concerns and revise accordingly. Please submit your revised manuscript by Dec 04 2020 11:59PM. If you will need more time than this to complete your revisions, please reply to this message or contact the journal office at plosone@plos.org. When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file. Please include the following items when submitting your revised manuscript: A rebuttal letter that responds to each point raised by the academic editor and reviewer(s). You should upload this letter as a separate file labeled 'Response to Reviewers'. A marked-up copy of your manuscript that highlights changes made to the original version. You should upload this as a separate file labeled 'Revised Manuscript with Track Changes'. An unmarked version of your revised paper without tracked changes. You should upload this as a separate file labeled 'Manuscript'. If you would like to make changes to your financial disclosure, please include your updated statement in your cover letter. Guidelines for resubmitting your figure files are available below the reviewer comments at the end of this letter. If applicable, we recommend that you deposit your laboratory protocols in protocols.io to enhance the reproducibility of your results. Protocols.io assigns your protocol its own identifier (DOI) so that it can be cited independently in the future. For instructions see: http://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols We look forward to receiving your revised manuscript. Kind regards, Robert Jeenchen Chen, MD, MPH Academic Editor PLOS ONE [Note: HTML markup is below. Please do not edit.] Reviewers' comments: Reviewer's Responses to Questions Comments to the Author 1. If the authors have adequately addressed your comments raised in a previous round of review and you feel that this manuscript is now acceptable for publication, you may indicate that here to bypass the “Comments to the Author” section, enter your conflict of interest statement in the “Confidential to Editor” section, and submit your "Accept" recommendation. Reviewer #1: (No Response) Reviewer #3: (No Response) ********** 2. Is the manuscript technically sound, and do the data support the conclusions? The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented. Reviewer #1: Partly Reviewer #3: (No Response) ********** 3. Has the statistical analysis been performed appropriately and rigorously? Reviewer #1: No Reviewer #3: (No Response) ********** 4. Have the authors made all data underlying the findings in their manuscript fully available? The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified. Reviewer #1: Yes Reviewer #3: (No Response) ********** 5. Is the manuscript presented in an intelligible fashion and written in standard English? PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here. Reviewer #1: Yes Reviewer #3: (No Response) ********** 6. Review Comments to the Author Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters) Reviewer #1: The authors are appreciated for their effort however, the required revisions were insufficiently done. Non-randomized characteristics of the study is a flaw. Ultrasound guidance is a standard for central venous access which is considered mandatory for increasing procedural success and reducing complications. Larger sample size is needed for assessing rare early and late complications. Reviewer #3: (No Response) ********** 7. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files. If you choose “no”, your identity will remain anonymous but your review may still be made public. Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy. Reviewer #1: No Reviewer #3: No [NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files.] While revising your submission, please upload your figure files to the Preflight Analysis and Conversion Engine (PACE) digital diagnostic tool, https://pacev2.apexcovantage.com/. PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email PLOS at figures@plos.org. Please note that Supporting Information files do not need this step. 27 Oct 2020 Dear editors, Firstly, we would like to thank you for your feedback and constructive criticism. As requested, we provided a shorter title for our manuscript. As regards the reviewer comment : “ Reviewer #1: The authors are appreciated for their effort however, the required revisions were insufficiently done. Non-randomized characteristics of the study is a flaw. Ultrasound guidance is a standard for central venous access which is considered mandatory for increasing procedural success and reducing complications. Larger sample size is needed for assessing rare early and late complications “ As previously mentioned, we did acknowledge the limitation originating from the non randomized character of the study in the limitations section as well as specified the need to consider our results with caution : “Limitations of this study are the retrospective aspect, absence of randomization and the fact that the results of this study were directly dependent on the senior surgeon's experience and skill. Therefore, caution is required when applying our results” We also acknowledge the sample size limitation, and the need for a larger sample in our future investigations regarding this subject. In fact, we believe this study provides us with guidance as we added to our limitation section. “ An additional limitation is the relatively small number of patients included, however, this will allow us to decide the required sample size for future studies addressing this technique.” On the other hand, we agree that ultrasound guidance is the standard of use for central venous access as mentioned in our conclusion. However, we made modifications to our conclusion to further emphasize on ultrasound guidance prioritization. In fact, central line placement is a procedure conducted by radiologists, intensive care specialists and surgeons with the latter using landmark techniques more frequently. This is particularly more frequent in low and middle income contexts where the availability of ultrasounds devices is not always possible, due to resource restrictions. In this case and only in the absence of ultrasound guidance is the yoffa approach preferred to other techniques. Submitted filename: Response to comments ; yoffa.docx Click here for additional data file. 9 Nov 2020 Supraclavicular versus infraclavicular approach in inserting totally implantable central venous access for cancer therapy : a comparative retrospective study PONE-D-20-14056R2 Dear Dr. Souadka, We’re pleased to inform you that your manuscript has been judged scientifically suitable for publication and will be formally accepted for publication once it meets all outstanding technical requirements. Within one week, you’ll receive an e-mail detailing the required amendments. When these have been addressed, you’ll receive a formal acceptance letter and your manuscript will be scheduled for publication. An invoice for payment will follow shortly after the formal acceptance. To ensure an efficient process, please log into Editorial Manager at http://www.editorialmanager.com/pone/, click the 'Update My Information' link at the top of the page, and double check that your user information is up-to-date. If you have any billing related questions, please contact our Author Billing department directly at authorbilling@plos.org. If your institution or institutions have a press office, please notify them about your upcoming paper to help maximize its impact. If they’ll be preparing press materials, please inform our press team as soon as possible -- no later than 48 hours after receiving the formal acceptance. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information, please contact onepress@plos.org. Kind regards, Robert Jeenchen Chen, MD, MPH Academic Editor PLOS ONE Additional Editor Comments (optional): Reviewers' comments: Reviewer's Responses to Questions Comments to the Author 1. If the authors have adequately addressed your comments raised in a previous round of review and you feel that this manuscript is now acceptable for publication, you may indicate that here to bypass the “Comments to the Author” section, enter your conflict of interest statement in the “Confidential to Editor” section, and submit your "Accept" recommendation. Reviewer #4: All comments have been addressed Reviewer #5: All comments have been addressed ********** 2. Is the manuscript technically sound, and do the data support the conclusions? The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented. Reviewer #4: Partly Reviewer #5: Yes ********** 3. Has the statistical analysis been performed appropriately and rigorously? Reviewer #4: Yes Reviewer #5: I Don't Know ********** 4. Have the authors made all data underlying the findings in their manuscript fully available? The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified. Reviewer #4: Yes Reviewer #5: Yes ********** 5. Is the manuscript presented in an intelligible fashion and written in standard English? PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here. Reviewer #4: Yes Reviewer #5: Yes ********** 6. Review Comments to the Author Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters) Reviewer #4: The paper is well revised. It deserves a wide scale reders exposure. Although the sample size is small but still it provides valuble information about TIVAP. Reviewer #5: I read the article of the authors titled "Supraclavicular versus infraclavicular approach in inserting totally implantable central venous access for cancer therapy : a comparative retrospective study" with interest. When I examined the answers given to the previous criticisms, I saw that the necessary corrections were made. ********** 7. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files. If you choose “no”, your identity will remain anonymous but your review may still be made public. Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy. Reviewer #4: Yes: Aram Baram Reviewer #5: No Submitted filename: Comments to the authors.docx Click here for additional data file. 11 Nov 2020 PONE-D-20-14056R2 Supraclavicular versus infraclavicular approach in inserting totally implantable central venous access for cancer therapy : a comparative retrospective study Dear Dr. Souadka: I'm pleased to inform you that your manuscript has been deemed suitable for publication in PLOS ONE. Congratulations! Your manuscript is now with our production department. If your institution or institutions have a press office, please let them know about your upcoming paper now to help maximize its impact. If they'll be preparing press materials, please inform our press team within the next 48 hours. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information please contact onepress@plos.org. If we can help with anything else, please email us at plosone@plos.org. Thank you for submitting your work to PLOS ONE and supporting open access. Kind regards, PLOS ONE Editorial Office Staff on behalf of Dr. Robert Jeenchen Chen Academic Editor PLOS ONE
  19 in total

1.  Before we go too far: ultrasound-guided central catheter placement.

Authors:  Alexander B Levitov; Sameh Aziz; Anthony D Slonim
Journal:  Crit Care Med       Date:  2009-08       Impact factor: 7.598

2.  Subclavian vein portacath placement - the Yoffa supraclavicular technique.

Authors:  A Ricci; A Majbar
Journal:  J Visc Surg       Date:  2011-04-11       Impact factor: 2.043

3.  Subclavian venipuncture: a review.

Authors:  R J Defalque
Journal:  Anesth Analg       Date:  1968 Nov-Dec       Impact factor: 5.108

4.  A comparison of the supraclavicular approach and the infraclavicular approach for subclavian vein catheterization.

Authors:  S Sterner; D W Plummer; J Clinton; E Ruiz
Journal:  Ann Emerg Med       Date:  1986-04       Impact factor: 5.721

5.  Supraclavicular subclavian venepuncture and catheterisation.

Authors:  D Yoffa
Journal:  Lancet       Date:  1965-09-25       Impact factor: 79.321

6.  Insertion of totally implantable central venous access devices by surgeons.

Authors:  Hyeonjun An; Chun-Geun Ryu; Eun-Joo Jung; Hyun Jong Kang; Jin Hee Paik; Jung-Hyun Yang; Dae-Yong Hwang
Journal:  Ann Coloproctol       Date:  2015-04-30

7.  A modified approach to supraclavicular subclavian vein catheter placement: the pocket approach.

Authors:  Julie Gorchynski; Worth W Everett; Eleni Pentheroudakis
Journal:  Cal J Emerg Med       Date:  2004-07

8.  A retrospective clinical study: complications of totally implanted central venous access ports.

Authors:  June Pill Seok; Young Jin Kim; Hyun Min Cho; Han Young Ryu; Wan Jin Hwang; Tae Yun Sung
Journal:  Korean J Thorac Cardiovasc Surg       Date:  2014-02-05

9.  Subclavian vein catheterization during cardiopulmonary resuscitation. A prospective comparison of the supraclavicular and infraclavicular percutaneous approaches.

Authors:  S Dronen; B Thompson; R Nowak; M Tomlanovich
Journal:  JAMA       Date:  1982-06-18       Impact factor: 56.272

10.  Comparison between ultrasound-guided supraclavicular and infraclavicular approaches for subclavian venous catheterization in children--a randomized trial.

Authors:  H-J Byon; G-W Lee; J-H Lee; Y-H Park; H-S Kim; C-S Kim; J-T Kim
Journal:  Br J Anaesth       Date:  2013-06-10       Impact factor: 9.166

View more
  2 in total

Review 1.  Success Rate and Complications of the Supraclavicular Approach for Central Venous Access: A Systematic Review.

Authors:  Atif Nazir; Khurram Niazi; Syed Muhammad Jawad Zaidi; Muhammad Ali; Saeed Maqsood; Jahanzeb Malik; Mehwish Kaneez; Amin Mehmoodi
Journal:  Cureus       Date:  2022-04-03

Review 2.  Forty years after the first totally implantable venous access device (TIVAD) implant: the pure surgical cut-down technique only avoids immediate complications that can be fatal.

Authors:  Adriana Toro; Elena Schembari; Emanuele Gaspare Fontana; Salomone Di Saverio; Isidoro Di Carlo
Journal:  Langenbecks Arch Surg       Date:  2021-06-09       Impact factor: 3.445

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.