| Literature DB >> 33225269 |
Smaranda Buduru1, Silvia Balhuc1, Alexandru Ciumasu1, Andreea Kui1, Cosmin Ciobanu2, Oana Almasan1, Manuela Manziuc1, Marius Negucioiu1.
Abstract
BACKGROUND AND AIMS: The objective of this study was to evaluate the accuracy of the CADIAX Compact 2 axiograph in confirming the presumptive clinical diagnosis of temporomandibular dysfunction, established according to the muscular-articular clinical examination.Entities:
Keywords: axiography; disk displacement; temporomandibular disorder
Year: 2020 PMID: 33225269 PMCID: PMC7664731 DOI: 10.15386/mpr-1754
Source DB: PubMed Journal: Med Pharm Rep ISSN: 2602-0807
Figure 1Axiograph application technique.
Figure 2The graphs in Disc Displacement with reduction (DDwR) on the right side.
Figure 3The graphs in Disc Displacement without reduction (DDwoR) on the right side.
Figure 4Distribution of subjects based on clinical and paraclinical diagnosis.
Figure 5Distribution of subjects according to the clinical diagnosis.
Figure 6Distribution of subjects according to the CADIAX Compact 2 axiograph diagnosis.
Contingency table of the subjects included in the study.
| Clinical diagnosis | |||
|---|---|---|---|
| CADIAX diagnosis | Negative | Positive | Total |
| Negative | 17.00 | 0.00 | 17.00 |
| Row% | 1.00 | 0.00 | 1.00 |
| Col% | 0.74 | 0.00 | 0.34 |
| Positive | 6.00 | 27.00 | 33.00 |
| Row% | 0.18 | 0.82 | 1.00 |
| Col% | 0.26 | 1.00 | 0.66 |
| TOTAL | 23.00 | 27.00 | 50.00 |
| Row% | 0.46 | 0.54 | 1.00 |
| Col% | 1.00 | 1.00 | 1.00 |
Extrinsic and intrinsic indicators (PPR - Positive probability ratio; NPR - Negative probability ratio; PPV - Positive predictive value; NPV - Negative predictive value).
| Sensitivity | 1.00 | 100% |
|---|---|---|
| 0.74 | 74% | |
| 0.88 | 88% | |
| 3.83 | ||
| 0.00 | ||
| 0.82 | 82% | |
| 1.00 | 100% |