Ahmad Hasan Ahangari1, Kianoosh Torabi, Sasan Rasaei Pour, Safoura Ghodsi.   

Abstract

BACKGROUND: Mandibular movement analysis is a critical step in making the functional occlusal morphology and improving the diagnosis and treatment of temporomandibular joint disorders (TMDs). Cadiax Compact® is an electronic condylograph that claims to record the horizontal condylar inclination (HCI), Bennett angle (BA) and relative shape of the articular eminence. This study aims at assessing the accuracy of Cadiax Compact® II in its claimed abilities.
MATERIALS AND METHODS: The electronic condylograph (Cadiax Compact®II) was fitted on the fully adjustable articulator (gamma dental reference-SL). After setting of HCI and BA on the arbitrary degrees, eccentric movements were produced manually on the articulator. The Cadiax recorded these preadjusted angles and the accuracy of its recordings was assessed by comparison of the results with the preadjusted HCI, BA and color inserts as references.
RESULTS: The majority of the comparisons showed statistically significant differences between articulator settings and Cadiax recordings. However, the maximum difference was about 2.50 which seems acceptable for clinical practice.
CONCLUSION: The obtained results showed that Cadiax Compact is an accurate and reliable instrument for diagnostic purposes, yielding reproducible measurements. Despite this, Cadiax is a technically sensitive device that can preclude its routine usage.

Entities:  

Mesh:

Year:  2012        PMID: 23151700     DOI: 10.5005/jp-journals-10024-1176

Source DB:  PubMed          Journal:  J Contemp Dent Pract        ISSN: 1526-3711


  4 in total

1.  Temporomandibular dysfunction diagnosis by means of computerized axiography.

Authors:  Smaranda Buduru; Silvia Balhuc; Alexandru Ciumasu; Andreea Kui; Cosmin Ciobanu; Oana Almasan; Manuela Manziuc; Marius Negucioiu
Journal:  Med Pharm Rep       Date:  2020-10-25

2.  Diagnostic use of computerized axiography in TMJ disc displacements.

Authors:  Daniel Talmaceanu; Nicolae Bolog; Daniel Leucuta; Ioan Andrei Tig; Smaranda Buduru
Journal:  Exp Ther Med       Date:  2022-01-11       Impact factor: 2.447

3.  Comparison of selected kinematic facebows applied to mandibular tracing.

Authors:  Mieszko Wieckiewicz; Marek Zietek; Danuta Nowakowska; Wlodzimierz Wieckiewicz
Journal:  Biomed Res Int       Date:  2014-05-07       Impact factor: 3.411

4.  The contribution of computerized axiography to the functional evaluation of the temporomandibular joint: a case report.

Authors:  Alexandra Maria Botos; Anca Stefania Mesaros; Adela Ioana Zimbran
Journal:  Clujul Med       Date:  2016-07-28
  4 in total