Analysis of sputum/tracheal aspirate and nasopharyngeal samples for SARS-CoV-2 detection by laboratory-developed test and Panther Fusion system.

Phyu M Thwe1, Ping Ren2.   

Abstract

In this study, 127 sputum/tracheal aspirate specimens were evaluated by a laboratory-developed real-time RT-PCR method and Fusion SARS-CoV-2 assay. These specimens were collected from the patients who have nasopharyngeal swab (NPS) samples being used for SARS-CoV-2 detection previously or simultaneously. The overall agreement was 96% between the lower respiratory tract (LRT) and NPS samples, suggesting that LRT specimens could be an option for patients who develop a productive cough or those receiving invasive mechanical ventilation.
Copyright © 2020 Elsevier Inc. All rights reserved.

Entities:  

Keywords:  COVID-19; Lower respiratory tract; Nasopharyngeal swab sample; SARS-CoV-2; Sputum; Tracheal aspirate

Year:  2020        PMID: 33190093      PMCID: PMC7525323          DOI: 10.1016/j.diagmicrobio.2020.115228

Source DB:  PubMed          Journal:  Diagn Microbiol Infect Dis        ISSN: 0732-8893            Impact factor:   2.803


Introduction

Diagnostic testing of severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) from nasopharyngeal (NPS) and oropharyngeal (OP) swabs were the first FDA approved sample types to be used on laboratory-developed tests (LDT) or commercial platforms. The Centers for Disease Control and Prevention (CDC) stated that lower respiratory tract (LRT) samples, such as sputum, are permissible for patients with a productive cough or under special circumstances (e.g., invasive mechanical ventilation) (https://www.cdc.gov/coronavirus/2019-ncov/lab/guidelines-clinical-specimens.html). Testing sputum can also be helpful for patients under investigation (PUI), whose NPS swabs are negative for SARS-CoV-2. Previous studies have shown that SARS-CoV-2 RNA could be detected in non-NP sample types, such as sputum, bronchoalveolar (BAL) fluid, and saliva (Wang et al., 2020a; Mohammadi et al., 2020; Wang et al., 2020b). Here, we examined the potential utilization of sputum/tracheal aspirate (collectively as sputum) samples for diagnosis of coronavirus disease 2019 (COVID-19) by real-time RT-PCR using LDT and Fusion SARS-CoV-2 assay (Hologic, San Diego, CA). The LDT had previously been established in our laboratory for NPS in March 2020. The Fusion assay was modified from the Emergency Use Authorization (EUA) approved protocol by using dithiothreitol treated sputum after 1:1 water dilution.

Material and methods

One hundred and twenty-seven retrospective sputum samples from 103 in-patients were collected. NPS samples from these 103 patients were tested for initial diagnosis of COVID-19, whereas sputum samples were collected at the time of NPS collection and/or during hospitalization. Out of 103 patients, 73 sputum samples were from 48 NPS-positive patients whereas 54 were from 54 NPS-negative patients (Table 1 ). Twenty-five out of 73 positive sputa were from the repeated collection. Sixteen of 48 NPS-positive patients had repeated sputa, collected at least twice during hospitalization. A majority of positive NPS (37) were tested by ID NOW COVID-19 (Abbott, Abbott Park, IL, USA), 5 by Panther Fusion® SARS-CoV-2 Assay (Hologic, San Diego, CA, USA), 2 by Abbott RealTime SARS-CoV-2 (Abbott Park, IL, USA), and 1 by Xpert® Xpress SARS-CoV-2 (Cepheid, Sunnyvale, CA, USA). All negative NPS were tested at least once by one of the RT-PCR platforms or tested multiple times by ID NOW COVID-19 assay.
Table 1

List of positive NPS and sputum samples with LDT and Fusion SARS-CoV-2 testing results.

Subject #Specimen #Assay used for initial NPS tested positiveSputum collection days post NPSSputum resultsLDT RT-PCR results (Ct value)
Fusion results (Ct value)
EN2ORF8RNase P (IC)ORF1ab
11Abbott m20000Pos#32.0734.2934.1324.61NT@
22ID NOW0Pos16.1917.0217.0736.3414.3
33ID NOW0Pos17.5318.0417.9324.86NT
44ID NOW0Pos20.0820.8120.9219.9NT
55ID NOW0Pos23.3824.8325.1319.61NT
66ID NOW0Pos24.2225.526.6820.75NT
77ID NOW0Pos25.1326.0526.620.3NT
88ID NOW0Pos26.8328.2427.3831.31NT
99ID NOW0Pos29.2731.2431.1624.04NT
1010ID NOW0Pos29.6230.2129.3229.19NT
1111ID NOW0Pos30.2231.3931.527.23NT
1212FUSION0Pos24.1525.3224.8132.4NT
1313ID NOW1Pos22.1422.3722.0429.56NT
1414ID NOW1Pos22.3424.3322.8625.621.2
1515ID NOW1Pos25.927.2326.6629.33NT
1616ID NOW1Pos29.7131.231.1727.328.5
1717ID NOW1Pos38.6738.9627.4438.5
1818FUSION1Pos33.1834.834.4919.3534.1
1919NA^3Pos27.4629.0928.8922.16NT
2020ID NOW5Pos15.2116.216.320.6815
2121ID NOW5Pos20.5521.3921.7626.2618.8
2222ID NOW5Pos25.6927.1126.1720.33NT
2323ID NOW5Pos27.529.128.9323.426.8
2424FUSION6Pos25.6625.9325.1829.23NT
2525ID NOW7Pos27.7228.0528.524.11NT
2626ID NOW7Pos29.6430.1230.5728.21NT
2727ID NOW7Pos32.6533.9134.3922.95NT
2828ID NOW8Pos30.1431.4930.6527.04NT
2929ID NOW8Pos33.4733.9234.1624.51NT
3030ID NOW11Pos26.4727.8828.0429.07NT
3131ID NOW11Pos29.3530.3330.120.76NT
3232FUSION15Pos36.7237.2536.4426.28NT
3333ID NOW0Pos20.2423.7722.1719.37NT
342Pos21.3222.7422.6824.39NT
3435ID NOW0Pos23.5625.9625.0228.12NT
366Pos36.7235.3735.4724.05NT
3537FUSION4Pos27.3628.628.1625.327.4
388Pos32.1133.1132.3124.5329.8
3639ID NOW5Pos21.3722.221.3926.22NT
4010Pos27.5628.7127.9925.59NT
3741Abbott m20009Pos29.9331.2531.2924.03NT
4213Pos36.7438.6737.3921.16NT
3843ID NOW14Pos36.8737.2236.5125.2134.7
4417Pos39.0240.1522.01NT
3945ID NOW14Pos36.3636.4436.4722.934.5
4620Neg25.25Neg&
4047ID NOW14Pos36.8239.0538.423.0435.1
4821Neg23.57Neg
4149ID NOW20Pos32.9233.532.4322.56NT
5023Pos36.0437.7737.0725.9NT
4251NA30Neg24.31Neg
5232Pos39.3338.0627.06
4353ID NOW0Pos35.4536.7736.7528.1538.1
545Neg26.48Neg
558Pos39.7641.6625.71NT
4456Xpert0Pos24.9526.1625.6228.4223
579Pos27.3228.3228.0120.0327.1
5813Pos32.434.2533.3919.7932.4
4559ID NOW12Pos36.2238.4239.0120.4138.1
6017Pos38.7540.0522.63NT
6118Pos39.9123.11NT
4662ID NOW3Pos23.9224.6125.1220.4523.6
639Pos32.3433.3733.8820.7632
6419Pos39.3442.0439.5324.5438.5
6527Neg22.94Neg
4766LDT8Pos22.7524.2424.2721.92NT
6721Pos35.9837.2937.5424.3535.6
6826Pos36.4637.2937.523.31NT
6932Pos41.4840.8923.15NT
4870ID NOW9Pos28.9929.0528.6724.86NT
7110Pos29.3930.3230.1420.39NT
7210Pos32.3533.0732.1825.62NT
7314Pos34.3234.6333.7225.42NT

^NA: Not available.

IC: internal control.

Pos: Positive.

Neg: Negative.

NT: Not tested.

List of positive NPS and sputum samples with LDT and Fusion SARS-CoV-2 testing results. ^NA: Not available. IC: internal control. Pos: Positive. Neg: Negative. NT: Not tested. All 127 sputum samples were evaluated by the LDT. Our LDT method utilized QIAamp Viral RNA Mini kit (Qiagen, Germantown, MD, USA) for nucleic acid extraction and Reliance One-Step Multiplex Supermix (Bio-Rad, Hercules, CA, USA) for RT-PCR reaction. The assay targeted SARS-CoV-2 E, N2, and ORF8 genes with the Ct cutoff at 42. The human housekeeping gene RNase P was used as an internal control. RT-PCR was performed on the CFX96 Touch Real-Time PCR Detection System (Bio-Rad, Hercules, CA, USA). Briefly, sputum was diluted with molecular grade water in a 1:1 ratio as mentioned in Branche et al. (Branche et al., 2014), and the mixture was vortexed and settled for 15–30 min; 140 μL of diluted samples were processed for the downstream RNA extraction and RT-PCR reactions. Due to the limitation in Fusion reagents, only 50 specimens (27 positives and 23 negatives) out of 127 were tested by Fusion. The diluted (1:1) sputum samples from above were mixed with an equal volume of freshly prepared 10 mM dithiothreitol (DTT) dissolved in sterile PBS (pH = 7.2), and the mixture was incubated at room temperature with intermittent mixing until mucus was liquefied (up to 30 min), according to the CDC's sputum processing protocol (https://www.cdc.gov/coronavirus/2019-ncov/downloads/processing-sputum-specimens.pdf). The liquified sample mix was then transferred to the Fusion lysis tube per manufacturer's instruction (Hologic Panther Fusion® SARS-CoV-2, 2020). The limit-of-detection (LoD) analyses for both methods were performed by spiking the whole genome of SARS-CoV-2 viral RNA, provided by The World Reference Center for Emerging Viruses and Arboviruses (WRCEVA) at the University of Texas Medical Branch (UTMB). Twenty replicates were tested at the estimated lowest detection levels to achieve minimal 19 positive results, based on the FDA guidelines. Statistical analysis was performed using Medcalc stats (https://www.medcalc.org/calc/diagnostic_test.php). Cohen's kappa coefficient (κ) was also calculated using the Graphpad Prism QuickCalcs website (https://www.graphpad.com/quickcalcs/) as a measure of the overall agreement, with values representing levels of agreement that are categorized as almost perfect (0.81–1.00), substantial agreement (0.61–0.80), moderate agreement (0.41–0.60), fair agreement (0.21–0.40), and slight agreement (0.00–0.20) (Landis and Koch, 1977).

Results

We demonstrated a 100% negative agreement for NPS-sputum negative pairs (54/54) and a 93% (95%CI: 85–98%) positive agreement for NPS-sputum positive pairs (68/73) by LDT (Table 2 ). The overall agreements between NPS and sputum pairs were 96% (95% CI: 91–99%) and 90% (95%CI: 78–96%) for LDT and Fusion, respectively. Due to a limited supply of Fusion reagents, the sample size for Fusion was smaller than LDT, which likely resulted in a slightly lower agreement than LDT. The ĸ values were 0.92 (95% CI: 85.2–98.9%) and 0.80 (95% CI: 64–96%) for LDT and Fusion, respectively, indicating that the agreements were almost perfect to substantial. The results from all samples tested on Fusion correlated with LDT. Five discrepant samples (i.e., NP positive but sputum negative) (Samples# 46, 48, 51, 52, and 65) tested by LDT were repeated on Fusion and resulted as negative, which was consistent with the LDT results. Among those discrepant sample sets, 2 were from 2 different patients (#51 and 54) and were most likely poor quality since the repeat testing of recollected sputum within 2 days (#52 and 55) resulted in positive. The remaining discrepant samples (#46, 48, and 65) were from 3 different patients (Subject #39, 40, and 46), who had at least two sputum specimens collected throughout their hospitalization.
Table 2

Sputum vs. NPS samples results by LDT and Fusion.

Sputum analyses byNasopharyngeal swab samples
PositiveNegative
LDTPositive680
Negative554
FusionPositive220
Negative523
Sputum vs. NPS samples results by LDT and Fusion. The LoD were determined to be 10,000 and 3,000 copies/mL for LDT and Fusion method on sputum samples, respectively.

Discussion

We demonstrated that sputum is one of the alternative specimen types for SARS-CoV-2 diagnosis in the hospitalized patient population. The discrepancy (i.e., positive NPS and negative sputum) could be attributed to i) poor quality of sputum and ii) sample collection at the disease recovery stage. One of the potential issues with sputum is the presence of mucus which poses difficulty during sample preparation, particularly on automated platforms. For Fusion SARS-CoV-2 assay, the manufacturer discourages the mucus part of LRT samples being transferred into Fusion lysis tubes. Despite the 1:1 dilution of sputum in sterile water (leftover samples from LDT performance), the persistent viscosity deterred successful handling by the Panther Fusion instrument. We have overcome these issues, by applying the CDC guidelines for sputum preparation and successfully demonstrated that sputum can be validated on any platform, whether it is a manual or an automated instrumentation system. Most clinical sites utilize two consecutive negative NPS as an indication of infection clearance. However, some patients who develop a productive cough at the later course of the disease may still be positive with SARS-CoV-2 RNA in the sputum despite the NPS became negative, a finding described in Chen et al. (Chen et al., 2020). Wolfel et al. (Wölfel et al., n.d.) have demonstrated the presence of infectious viral particles in LRT, the detection of SARS-CoV-2 RNA in sputum, particularly those with low to moderate Ct values, likely indicate the possibility of persistent infection. Therefore, testing of sputum may be necessary for a certain patient population indicated below. A testing algorithm for sputum may need to be developed with the support of infectious disease clinicians. We believe that our findings are beneficial in cases of i) hospitalized individuals with endotracheal intubation before the NPS samples could be taken for COVID-19 evaluation (endotracheal aspiration may be indicated in this situation); ii) patients with a traumatic fracture to the facial/nasal area or anatomic anomaly; iii) symptomatic patients who have a productive cough with negative NPS results. One pitfall of sputum collection is the generation of aerosols; therefore, induced sputum is not recommended by the CDC. However, endotracheal sputum aspirates or self-collected sputum (with proper instructions) in a defined patient population would provide several advantages over NPS, such as the discomfort associated with NPS sampling.

Conclusions

To our knowledge, this is the first study analyzing the use of sputum on LDT and Fusion platforms. Overall, we have demonstrated that both platforms are validated and can be utilized for the diagnosis of COVID-19 from sputum samples in symptomatic patients.

Funding

This project did not receive any funding support from any agencies in the public, commercial, or not-for-profit sectors.

Conflict of interest

All authors have no conflict of interest.

Ethical approval

The project is approved by the UTMB Institutional Review Board (IRB) under an expedited review process with IRB # 20–0143.

Informed consent

Not applicable.

CRediT author statement.

Phyu M. Thwe: Resources, Investigation, Formal analysis, Writing – original draft preparation and editing. Ping Ren: Conceptualization, Resources, Formal analysis, Project administration, Writing – Reviewing and Editing.
  8 in total

Review 1.  Biological characteristics and biomarkers of novel SARS-CoV-2 facilitated rapid development and implementation of diagnostic tools and surveillance measures.

Authors:  Gajanan Sampatrao Ghodake; Surendra Krushna Shinde; Avinash Ashok Kadam; Rijuta Ganesh Saratale; Ganesh Dattatraya Saratale; Asad Syed; Abdallah M Elgorban; Najat Marraiki; Dae-Young Kim
Journal:  Biosens Bioelectron       Date:  2021-01-04       Impact factor: 10.618

2.  Evaluation of SARS-CoV-2 antibody persistence and viral spread in stool: a long-term care experience before COVID-19 vaccination.

Authors:  Oya Ozlem Eren-Kutsoylu; Ozgur Appak; Arzu Nazli-Zeka; Gokcen Omeroglu-Simsek; Nil Tekin; Basak Bayram; Ayca Arzu Sayiner; Nesim Tanglay; Vildan Avkan-Oguz
Journal:  Ir J Med Sci       Date:  2022-07-12       Impact factor: 2.089

3.  Comparative Diagnostic Accuracy of the STANDARD M10 Assay for the Molecular Diagnosis of SARS-CoV-2 in the Point-of-Care and Critical Care Settings.

Authors:  Alexander Domnich; Andrea Orsi; Carlo-Simone Trombetta; Elisabetta Costa; Giulia Guarona; Miriana Lucente; Valentina Ricucci; Bianca Bruzzone; Giancarlo Icardi
Journal:  J Clin Med       Date:  2022-04-27       Impact factor: 4.964

4.  Discordant SARS-CoV-2 Detection in the Nasopharynx Versus Trachea for Patients With Tracheostomies.

Authors:  Joshua D Smith; Jason A Correll; Jennifer L Stein; Robbi A Kupfer; Norman D Hogikyan; Robert J Morrison; Andrew P Stein
Journal:  Laryngoscope       Date:  2021-05-14       Impact factor: 2.970

Review 5.  Updating the use of nano-biosensors as promising devices for the diagnosis of coronavirus family members: A systematic review.

Authors:  Adriano Aquino; Vania Margaret Flosi Paschoalin; Leticia Louize Gonçalves Tessaro; Paulo A Raymundo-Pereira; Carlos Adam Conte-Junior
Journal:  J Pharm Biomed Anal       Date:  2022-01-25       Impact factor: 3.935

Review 6.  Advances in the Rapid Diagnostic of Viral Respiratory Tract Infections.

Authors:  Gratiela Gradisteanu Pircalabioru; Florina Silvia Iliescu; Grigore Mihaescu; Alina Irina Cucu; Octavian Narcis Ionescu; Melania Popescu; Monica Simion; Liliana Burlibasa; Mihaela Tica; Mariana Carmen Chifiriuc; Ciprian Iliescu
Journal:  Front Cell Infect Microbiol       Date:  2022-02-10       Impact factor: 5.293

Review 7.  Different Respiratory Samples for COVID-19 Detection by Standard and Direct Quantitative RT-PCR: A Literature Review.

Authors:  Maryam Ahmadzadeh; Hossein Vahidi; Arash Mahboubi; Fariba Hajifathaliha; Leila Nematollahi; Elham Mohit
Journal:  Iran J Pharm Res       Date:  2021       Impact factor: 1.696

8.  Comparison of nasopharyngeal swab and nasopharyngeal aspiration in adults for SARS-CoV-2 identification using reverse transcription-polymerase chain reaction.

Authors:  Selin Ünsaler; Ayşe Okan; Süda Tekin; Ayşenur M Hafız; Ozan Gökler; Ozan Altuntaş
Journal:  J Med Virol       Date:  2021-08-07       Impact factor: 20.693

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.