| Literature DB >> 33178769 |
Giray Bozkaya1, Ali Rıza Sisman2.
Abstract
BACKGROUND: The results of cardiac troponin I (cTnI) methods used in the diagnosis of acute myocardial infarction (AMI) are highly variable. In this study, it was aimed to compare the analytical performance of the Mindray CL-series TnI method with the Beckman Coulter-Access II AccuTnI+3 method.Entities:
Keywords: Analytical performance; bias; cardiac troponin I (cTnI); limit of quantification (LoQ); method comparison; precision
Year: 2020 PMID: 33178769 PMCID: PMC7607087 DOI: 10.21037/atm-20-6104
Source DB: PubMed Journal: Ann Transl Med ISSN: 2305-5839
The verification data of precision and bias study
| PI claims or user estimates | Mindray CL 1000i | Beckman Coulter Access II | |||||
|---|---|---|---|---|---|---|---|
| Level 1 | Level 2 | Level 3 | Level 1 | Level 2 | Level 3 | ||
| PI claims | |||||||
| Mean (ng/mL) | 0.410 | * | 34.873 | 0.070 | 0.910 | 13.970 | |
| Repeatability (CV%) | 5.15 | * | 4.15 | 5 | 2 | 2 | |
| WL imprecision (CV%) | 9.06 | * | 7.11 | 7 | 5 | 5 | |
| User estimates | |||||||
| N | 25 | 25 | 25 | 25 | 25 | 25 | |
| Repeatability (CV%) | 1.86 | 1.75 | 1.08 | 2.58 | 3.13 | 2.00 | |
| UVL for repeatability imprecision (CV%) | 6.90 | 6.90 | 5.56 | 6.70 | 2.68 | 2.68 | |
| WL imprecision (CV%) | 3.36 | 5.27 | 5.82 | 12.01 | 8.60 | 9.13 | |
| UVL for WL imprecision (CV%) | 14.13 | 14.13 | 11.09 | 10.29 | 8.30 | 8.30 | |
| Target value of QC material (ng/mL) | 0.78 | 2.52 | 27.1 | 0.22 | 0.88 | 12.1 | |
| Overall mean, n=25 (ng/mL) | 0.775 | 2.486 | 27.023 | 0.234 | 0.935 | 11.437 | |
| Verification interval for bias (ng/mL) | 0.74–0.82 | 2.30–2.74 | 24.36–29.84 | 0.17–0.27 | 0.75–1.02 | 10.29–13.91 | |
| Bias % | −0.64 | 0.04 | −0.28 | 6.36 | 6.25 | −5.48 | |
| Total analytical error % | 6.18 | 8.74 | 9.88 | 26.19 | 20.44 | 20.55 | |
| Sigma metric for TEa =27.91% | 8.12 | 5.29 | 4.75 | 1.8 | 2.52 | 2.46 | |
| Sigma metric for TEa =10% | 0.56 | 1.89 | 1.67 | 0.31 | 0.44 | 0.5 | |
*, the manufacturer has no claim for level 2. PI, package insert; WL, within laboratory; UVL, upper verification limit; TEa, total allowable error.
Figure 1There was an average of 17.5% difference between the two methods between 0.04–10 ng/mL measuring range. As observed in the Bland-Altman difference plot, the Mindray CL-series TnI method gave higher results compared to Beckman Coulter Access II AccuTnI+3. It was found this difference exceeded 100% in low concentrations and 30% in high concentrations.
Figure 2In the Passing-Bablok regression analysis, the relationship between the two methods was found as Mindray CL-1000i TnI = Beckman Coulter Access II AccuTnI+3×1.208+0.00110.
Figure 3Mindray CL-series TnI method had a LoQ of approximately 0.0085 ng/mL under 10%CV.
Figure 4Beckman Coulter Access II AccuTnI+3 method had a LoQ of approximately 0.0366 ng/mL under 10%CV.