Pulmonary ridge coverage and device-related thrombosis after left atrial appendage occlusion.

Xavier Freixa1, Pedro Cepas-Guillen, Eduardo Flores-Umanzor, Ander Regueiro, Laura Sanchis, Andrea Fernandez-Valledor, Salvatore Brugaletta, Maria José Carretero, Barbara Vidal, Mónica Masotti, Victoria Martin-Yuste, Mercè Roqué, Marta Sitges, Manel Sabaté.   

Abstract

AIMS: The aim of this study was to evaluate the impact of pulmonary ridge (PR) coverage on both clinical and imaging follow-up outcomes in patients undergoing left atrial appendage occlusion (LAAO). METHODS AND
RESULTS: The study included consecutive patients with non-valvular atrial fibrillation who underwent LAAO with disc and lobe devices. Patients were classified into two groups according to the PR coverage. A total of 147 patients were included. Among these, the PR was covered in 109 (74%) and uncovered in 38 (26%). Successful implantation was achieved in 98.6%. No differences in procedural outcomes were observed between the groups. The rate of procedural major adverse events was 3% (only major bleedings and/or vascular access complications). No device embolisation, cardiac tamponade or in-hospital mortality was observed. After a mean follow-up of 1.77±2.2 years, the annualised ischaemic stroke and major bleeding rate was 1.3%/year and 6.5%/year, respectively, without differences between groups. At follow-up, patients with a covered PR presented a lower incidence of device-related thrombosis (DRT) (1%) than those with an uncovered PR (27%); p<0.001. In multivariable analysis, the presence of PR coverage emerged as an independent predictor of DRT.
CONCLUSIONS: Pulmonary ridge coverage was associated with a lower incidence of DRT after LAAO. Procedural and follow-up clinical outcomes did not differ between covered PR and uncovered PR patients.

Entities:  

Mesh:

Year:  2021        PMID: 33164895     DOI: 10.4244/EIJ-D-20-00886

Source DB:  PubMed          Journal:  EuroIntervention        ISSN: 1774-024X            Impact factor:   6.534


  1 in total

1.  Clinical and echocardiographic risk factors for device-related thrombus after left atrial appendage closure: an analysis from the multicenter EUROC-DRT registry.

Authors:  Vivian Vij; Kerstin Piayda; Dominik Nelles; Steffen Gloekler; Roberto Galea; Monika Fürholz; Bernhard Meier; Marco Valgimigli; Gilles O'Hara; Dabit Arzamendi; Victor Agudelo; Lluis Asmarats; Xavier Freixa; Eduardo Flores-Umanzor; Ole De Backer; Lars Sondergaard; Luis Nombela-Franco; Angela McInerney; Kasper Korsholm; Jens Erik Nielsen-Kudsk; Shazia Afzal; Tobias Zeus; Felix Operhalski; Boris Schmidt; Gilles Montalescot; Paul Guedeney; Xavier Iriart; Noelie Miton; Jacqueline Saw; Thomas Gilhofer; Laurent Fauchier; Egzon Veliqi; Felix Meincke; Nils Petri; Peter Nordbeck; Dmitrii Ognerubov; Evgeny Merkulov; Ignacio Cruz-González; Rocio Gonzalez-Ferreiro; Deepak L Bhatt; Alessandra Laricchia; Antonio Mangieri; Heyder Omran; Jan Wilko Schrickel; Josep Rodes-Cabau; Horst Sievert; Georg Nickenig; Alexander Sedaghat
Journal:  Clin Res Cardiol       Date:  2022-07-18       Impact factor: 6.138

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.