| Literature DB >> 33160374 |
HaiYang Wang1, JunQi Liu1, YiFei Pi1, Qi Liu2, Yang Mi3, XiangXiang Yang1, YueXin Guo4, RuiTai Fan5.
Abstract
OBJECTIVE: To assess the effects of various treatment planning parameters to identify the optimal gap distance for precise two-segment total body irradiation (TBI) using helical tomotherapy (HT) with fixed jaw mode. METHODS AND MATERIALS: Data of a treatment plan for 8 acute leukemia patients (height range: 109-130 cm) were analyzed. All patients underwent total-body computed tomography (CT) with 5-mm slice thickness. A lead wire, placed at 10 cm above the patella, was used to mark the boundary between the two segments. Target volumes and organs at risk were delineated using a Varian Eclipse 10.0 physician's workstation. Different distances between the lead wire and the boundary of the two targets were used. CT images were transferred to the HT workstation to design the treatment plans, by adjusting parameters, including the field width (FW; 2.5 cm, and 5 cm), pitch (0.287 and 0.430), modulation factor (1.8). The plans were superimposed to analyze the dose distributions in the overlap region when varying target gap distances, FWs, pitches to determine the optimal combinations.Entities:
Keywords: Field width; Gap distance; Helical tomotherapy; Pitch; Total body irradiation
Mesh:
Year: 2020 PMID: 33160374 PMCID: PMC7648982 DOI: 10.1186/s13014-020-01698-x
Source DB: PubMed Journal: Radiat Oncol ISSN: 1748-717X Impact factor: 3.481
Fig. 1Schematic illustration of total body irradiation in the overlap region. The lead wire was placed at a point 10 cm above the patella. The gap distance contained the distance between the superior field and lead wire (h1) and another distance between the inferior field and lead wire (h2)
Patient characteristics
| Patient no. | Age (years) | Gender | Diagnosis | Body height (cm) | Body weight (kg) |
|---|---|---|---|---|---|
| 1 | 8 | M | All | 125.0 | 29 |
| 2 | 6 | M | AML | 109 | 19.1 |
| 3 | 10 | F | All recurrence | 122 | 30.4 |
| 4 | 7 | M | ALL | 114.5 | 22.7 |
| 5 | 11 | M | AML | 128.5 | 34.1 |
| 6 | 8 | F | AML | 116 | 24.8 |
| 7 | 10 | M | ALL | 130 | 33 |
| 8 | 9 | M | ALL | 127.5 | 28.5 |
Mean dose, maximum dose, minimum dose, and heterogeneous index (HI) of the overlap region (= h1 + h2, and h1 = h2)
| Other parameters | Target gap distance | |||||||
|---|---|---|---|---|---|---|---|---|
| 1 cm | 2 cm | 3 cm | 4 cm | 5 cm | 6 cm | |||
| Pitch = 0.287 | FW = 2.5 cm | Mean dose (Gy) | 15.04 ± 0.08 | 13.21 ± 0.13 | 11.39 ± 0.13 | 9.58 ± 0.16 | 7.79 ± 0.13 | 6.00 ± 0.13 |
| Max dose (Gy) | 18.25 ± 0.20 | 13.74 ± 0.17 | 12.85 ± 0.15 | 12.78 ± 0.21 | 12.65 ± 0.13 | 11.69 ± 0.16 | ||
| Min dose (Gy) | 12.72 ± 0.11 | 12.67 ± 0.09 | 9.03 ± 0.19 | 4.68 ± 0.24 | 2.13 ± 0.27 | 1.44 ± 0.15 | ||
| HI | 1.43 ± 0.20 | 1.08 ± 0.17 | 1.42 ± 0.19 | 2.73 ± 0.24 | 5.94 ± 0.27 | 8.12 ± 0.16 | ||
| FW = 5 cm | Mean dose (Gy) | 19.53 ± 0.27 | 17.98 ± 0.42 | 16.34 ± 0.13 | 14.52 ± 0.26 | 12.73 ± 0.25 | 10.99 ± 0.19 | |
| Max dose (Gy) | 22.10 ± 0.35 | 19.82 ± 0.53 | 17.55 ± 0.15 | 15.13 ± 0.30 | 13.02 ± 0.21 | 12.51 ± 0.19 | ||
| Min dose (Gy) | 16.32 ± 0.23 | 15.09 ± 0.24 | 13.98 ± 0.15 | 13.10 ± 0.06 | 12.52 ± 0.28 | 10.37 ± 0.17 | ||
| HI | 1.35 ± 0.35 | 1.31 ± 0.53 | 1.26 ± 0.15 | 1.15 ± 0.3 | 1.04 ± 0.28 | 1.21 ± 0.19 | ||
| Pitch = 0.43 | FW = 2.5 cm | Mean dose (Gy) | 14.96 ± 0.10 | 13.21 ± 0.16 | 11.45 ± 0.15 | 9.52 ± 0.17 | 7.92 ± 0.22 | 6.02 ± 0.19 |
| Max dose (Gy) | 18.43 ± 0.33 | 13.72 ± 0.20 | 12.65 ± 0.23 | 12.59 ± 0.22 | 12.64 ± 0.18 | 11.55 ± 0.26 | ||
| Min dose (Gy) | 12.70 ± 0.10 | 12.63 ± 0.07 | 9.22 ± 0.28 | 4.69 ± 0.18 | 2.26 ± 0.26 | 1.37 ± 0.23 | ||
| HI | 1.45 ± 0.33 | 1.09 ± 0.2 | 1.37 ± 0.28 | 2.68 ± 0.22 | 5.59 ± 0.26 | 8.43 ± 0.26 | ||
| FW = 5 cm | Mean dose (Gy) | 19.32 ± 0.31 | 17.93 ± 0.18 | 16.27 ± 0.26 | 14.55 ± 0.32 | 12.72 ± 0.30 | 10.95 ± 0.27 | |
| Max dose (Gy) | 21.89 ± 0.46 | 19.78 ± 0.25 | 17.45 ± 0.36 | 15.16 ± 0.39 | 13.04 ± 0.26 | 12.49 ± 0.17 | ||
| Min dose (Gy) | 16.17 ± 0.21 | 15.00 ± 0.24 | 13.89 ± 0.18 | 13.07 ± 0.18 | 12.48 ± 0.32 | 10.26 ± 0.38 | ||
| HI | 1.35 ± 0.46 | 1.32 ± 0.25 | 1.26 ± 0.36 | 1.16 ± 0.39 | 1.04 ± 0.32 | 1.22 ± 0.38 | ||
Fig. 2The mean dose, the maximum dose, and the minimum dose of the overlap region varied with different target gap distances. The field width (FW) is 5 cm and pitch are 0.43/0.287. *prescription dose
Fig. 3The mean dose, the maximum dose, and the minimum dose of the overlap region varied with the different target gap distances of 2.5 cm. Pitch is 0.43/0.287. *prescription dose
Fig. 4Comparison of heterogeneous index (HI) with different field widths (FW) and pitch in the overlap region for the choice of the optimal target gap distance