| Literature DB >> 33158126 |
Beatrice Casini1, Benedetta Tuvo1, Fabrizio Maggi1,2, Giuliana Del Magro3, Alessandro Ribechini4, Anna Laura Costa1, Michele Totaro1, Angelo Baggiani1, Giulia Gemignani5, Gaetano Privitera1.
Abstract
Microbiological surveillance carried out in order to verify the effectiveness of endoscope reprocessing does not include the research of viruses, although endoscopes may be associated with the transmission of viral infections. This paper reports the experience of the University Hospital of Pisa in managing the risk from severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) during an endoscopy. A review of the reprocessing procedure was conducted to assess whether improvement actions were needed. To verify the reprocessing efficacy, a virological analysis was conducted both before and after the procedure. Five bronchoscopes and 11 digestive endoscopes (6 gastroscopes and 5 colonoscopes) were sampled. The liquid samples were subjected to concentration through the use of the Macrosep Advance Centrifugal Devices (PALL Life Sciences, Port Washington, NY, USA) and subsequently analyzed using the cobas® SARS-CoV-2 Test (Roche Diagnostics, Basel, Switzerland), together with eSwab 490 CE COPAN swabs (COPAN, Brescia, Italy), which were used to sample surfaces. In accordance with the first ordinance regarding the coronavirus disease 2019 (COVID-19) emergency issued by the Tuscany Region in March 2020, a procedure dedicated to the management of the COVID-19 emergency in endoscopic practices was prepared, including the reprocessing of endoscopes. The virological analysis carried out on samples collected from endoscopes after reprocessing gave negative results, as well as on samples collected on the endoscopy column surfaces and the two washer-disinfectors that were dedicated to COVID-19 patients. The improvement in endoscope reprocessing implemented during the COVID-19 emergency was effective in ensuring the absence of SARS-CoV-2, thus reducing the risk of infections after an endoscopy on COVID-19 patients.Entities:
Keywords: COVID-19; SARS-CoV-2; endoscopy; reprocessing; virological investigation
Mesh:
Year: 2020 PMID: 33158126 PMCID: PMC7662711 DOI: 10.3390/ijerph17218142
Source DB: PubMed Journal: Int J Environ Res Public Health ISSN: 1660-4601 Impact factor: 3.390
Figure 1Number and type of endoscopy performed from January to October 2020.
Sampling dates and results of the selected endoscopes.
| Brand | Type of Endoscope | Age | Date of | Date of Reprocessing | Date of | Sites of | SARS-CoV-2 RNA |
|---|---|---|---|---|---|---|---|
| Olympus | Gastroscope | 13 | 24.04.20 | 24.04.20 | 24.04.20 | eSwab COPAN 1 | Negative |
| GIF800 | eSwab COPAN 2 | Negative | |||||
| GIF800 | |||||||
| Olympus | Gastroscope | 13 | 30.04.20 | 30.04.20 | 30.04.20 | eSwab COPAN 1 | Negative |
| GIF800 | eSwab COPAN 2 | Negative | |||||
| Liquid | Negative | ||||||
| Pentax | Colonscope | 8 | 23.04.20 | 23.04.20 | 24.04.20 | eSwab COPAN 1 | Negative |
| EC878 | eSwab COPAN 2 | Negative | |||||
| Liquid | Negative | ||||||
| Olympus | Colonscope | 15 | 24.04.20 | 24.04.20 | 27.04.20 | eSwab COPAN 1 | Negative |
| CF BIC 094 | eSwab COPAN 2 | Negative | |||||
| Liquid | Negative | ||||||
| Olympus | Gastroscope | 18 | 24.04.20 | 24.04.20 | 27.04.20 | eSwab COPAN 1 | Negative |
| GIF BIC 238 | eSwab COPAN 2 | Negative | |||||
| Liquid | Negative | ||||||
| Pentax | Gastroscope | 8 | 29.04.20 | 29.04.20 | 30.04.20 | eSwab COPAN 1 | Negative |
| EG 040 | eSwab COPAN 2 | Negative | |||||
| Liquid | Negative | ||||||
| FujiFilm | Gastroscope | 1 | 05.05.20 | 05.05.20 | 06.05.20 | eSwab COPAN 1 | Negative |
| FG098 | eSwab COPAN 2 | Negative | |||||
| Liquid | Negative | ||||||
| Pentax | Colonscope | 14 | 06.05.20 | 06.05.20 | 07.05.20 | eSwab COPAN 1 | Negative |
| EC333 | eSwab COPAN 2 | Negative | |||||
| Liquid | Negative | ||||||
| Pentax | Colonscope | 14 | 06.05.20 | 06.05.20 | 07.05.20 | eSwab COPAN 1 | Negative |
| EC 306 | eSwab COPAN 2 | Negative | |||||
| Liquid | Negative | ||||||
| Olympus | Colonscope | 5 | 21.05.20 | 21.05.20 | 22.05.20 | eSwab COPAN 1 | Negative |
| GIF241 | eSwab COPAN 2 | Negative | |||||
| Liquid | Negative | ||||||
| Olympus | Broncoscope | 11 | 24.04.20 | 24.04.20 | 27.04.20 | eSwab COPAN 1 | Negative |
| BF1T180 | eSwab COPAN 2 | Negative | |||||
| Liquid | Negative | ||||||
| Olympus | Broncoscope | 11 | 29.04.20 | 29.04.20 | 30.04.20 | eSwab COPAN 1 | Negative |
| BF1T180 | eSwab COPAN 2 | Negative | |||||
| Liquid | Negative | ||||||
| Olympus | Broncoscope | 11 | 05.05.20 | 05.05.20 | 05.05.20 | eSwab COPAN 1 | Negative |
| BF1T180 | eSwab COPAN 2 | Negative | |||||
| Liquid | Negative | ||||||
| Olympus | Broncoscope | 9 | 20.05.20 | 20.05.20 | 22.05.20 | eSwab COPAN 1 | Negative |
| BF1T180 | eSwab COPAN 2 | Negative | |||||
| Liquid | Negative |
Note: eSwab COPAN 1—external surface of endoscope; eSwab COPAN 2—valve port of the biopsy channel; Liquid—solution used for irrigation of the internal channels of the endoscope.