Distortion of the Intracranial Pressure Waveform by Extraventricular Drainage System.

Daniel Teichmann, James C Lynch, Thomas Heldt.   

Abstract

OBJECTIVE: To investigate whether intracranial pressure (ICP) waveform measurements obtained from extraventricular drainage (EVD) systems are suitable for the calculation of intracranial elastance (ICE) or cerebrovascular pressure autoregulation (PAR) indices.
METHODS: The transfer characteristic of an EVD system is investigated by its step and frequency responses with focus on the low frequency (LF) range from 0.02 to 0.065 Hz (important in PAR) and the location of the system's first resonance frequency (important for ICE). The effects of opening the distal end of the EVD for drainage of cerebrospinal fluid and the presence of trapped air bubbles are also investigated.
RESULTS: The EVD system exhibits a first resonant frequency below 4 Hz, resulting in significant distortion of the measured ICP waveform. The frequency response in the LF range only remains flat when the EVD is closed. Opening the drain results in drops in magnitude and phase along the entire frequency range above DC. Air bubbles close to the EVD catheter tip affect the LF range while an air bubble close to the pressure transducer further decreases the first resonant frequency. Tests with actual ICP waveforms confirmed EVD-induced waveform distortions that can lead to erroneous ICE estimation.
CONCLUSION: EVD-based ICP measurements distort the waveform morphology. PAR indices based on LF information are only valid if the EVD is closed. EVD-based ICE estimation is to be avoided. SIGNIFICANCE: ICP waveform analyses to derive information about ICE and PAR should be critically questioned if only EVD derived ICP signals are at hand.

Entities:  

Year:  2021        PMID: 33156777     DOI: 10.1109/TBME.2020.3036283

Source DB:  PubMed          Journal:  IEEE Trans Biomed Eng        ISSN: 0018-9294            Impact factor:   4.538


  2 in total

1.  Predictive Values for Time from Transducer Stopcock Closure to Accurate Intracranial Pressure Reading.

Authors:  Corrie Earthman; Kathrina Siaron; Jennifer Wilson; DaiWai M Olson
Journal:  Neurocrit Care       Date:  2022-08-24       Impact factor: 3.532

2.  Waveform Morphology as a Surrogate for ICP Monitoring: A Comparison Between an Invasive and a Noninvasive Method.

Authors:  Fabiano Moulin de Moraes; Eva Rocha; Felipe Chaves Duarte Barros; Flávio Geraldo Rezende Freitas; Maramelia Miranda; Raul Alberto Valiente; João Brainer Clares de Andrade; Feres Eduardo Aparecido Chaddad Neto; Gisele Sampaio Silva
Journal:  Neurocrit Care       Date:  2022-03-24       Impact factor: 3.532

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.