Phatthranit Phattharapornjaroen1, Suwitchaya Surapornpaiboon1, Phanorn Chalermdamrichai1, Yuwares Sittichanbuncha1, Kittisak Sawanyawisuth2.
Abstract
INTRODUCTION: Adequate chest compression is crucial for cardiopulmonary resuscitation (CPR). There are several chest compression monitoring devices with different costs. This study aimed to evaluate the agreement rate of Improved Quality of Cardiopulmonary Resuscitation meter (IQ-CPR meter) and automated external defibrillator (AED) in chest compression quality monitoring.Entities:
Keywords: Quality indicators; cardiopulmonary resuscitation; health care; heart arrest; simulation training
Year: 2020 PMID: 33134972 PMCID: PMC7587996
Source DB: PubMed Journal: Arch Acad Emerg Med ISSN: 2645-4904
Figure 1Improved Quality of Cardiopulmonary Resuscitation meter (IQ-CPR meter) components
Cohen's Kappa coefficients for chest compression depth between AED and IQ-CPR meter in various populations
|
|
|
|---|---|
| All patients | 67.25% (66.02%-68.46%) |
| Male | 65.63% (64.15%-67.08%) |
| Female | 71.21% (68.98%-73.38%) |
| EP/residents | 71.20% (68.84%-73.48%) |
| Medical students, clinical years | 63.15% (60.62%-65.62%) |
| Paramedic students | 80.39% (78.31%-82.36%) |
| EMS personnel | 51.54% (48.75%-54.32%) |
| Chest compression experience < 3 years | 71.92% (70.27%-73.52%) |
| Chest compression experience > 3 years | 62.24% (60.41%-64.05%) |
CI: confidence interval; EP: electrophysiologist; EMS: emergency medical services; AED: automated external defibrillator; IQ-CPR meter: Improved Quality of Cardiopulmonary Resuscitation meter.
Sensitivity and specificity of the IQ-CPR meter compared with AED for chest compression depth
|
|
|
|
|---|---|---|
| No light (depth < 5 cm) | 32.23 (30.30-34.21) | 84.82 (83.62-85.96) |
| Green light (depth 5-6 cm) | 84.54 (83.32-85.70) | 31.35 (29.55-33.37) |
| Red light (depth > 6 cm) | 1.26 (0.03-6.85) | 99.96 (99.88-100.00) |
CI: confidence interval; AED: automated external defibrillator; IQ-CPR meter: Improved Quality of Cardiopulmonary Resuscitation meter