| Literature DB >> 33109134 |
Marika Toscano1, Kam Szlachetka2, Natalie Whaley2, Loralei L Thornburg2.
Abstract
BACKGROUND: Point-of-care ultrasound (POC-US) is a diagnostic test conducted at the site of patient care with direct interpretation by the clinician, providing immediate results. POC-US for gynecologic application is not well characterized by current literature yet has the potential to increase access in limited resource settings. We compared the diagnostics of three POC-US devices for gynecologic (GYN) pathology and then performed evaluation of sensitivity and specificity of a single best POC-US device for intended use in a low resource setting.Entities:
Keywords: Low-resource setting; POCUS; Pelvic ultrasound; Point-of-care ultrasound; Sensitivity; Specificity
Year: 2020 PMID: 33109134 PMCID: PMC7590494 DOI: 10.1186/s12880-020-00518-8
Source DB: PubMed Journal: BMC Med Imaging ISSN: 1471-2342 Impact factor: 1.930
Baseline characteristics of study participants
| Characteristic | United States study participants (n = 60) | Field study participants (n = 20) | |
|---|---|---|---|
| Agea (years) | 36 (28.5–38) | 31 (24–53.5) | 0.56c |
| BMI > 30 | 23/60 (38%) | 0/20 (0%) | < 0.01 |
| Paritya | 1 (0–2) | 2 (0.5–5.5) | 0.13c |
| Race | < 0.01 | ||
| Caucasian | 40/60 (67%) | 0 (0%) | |
| Black | 17/60 (28%) | 20/20 (100%) | |
| Other | 3/60 (5%) | 0 (0%) | |
| Ethnic origin | 0.23d | ||
| Hispanic | 5/60 (8%) | 0/20 (0%) | |
| Non-Hispanic | 55/60 (92%) | 20/20 (100%) | |
| Menopausal status | 0.64d | ||
| Pre-menopausal | 48/60 (80%) | 16/20 (80%) | |
| Post-menopausal | 12/60 (20%) | 4/20 (20%) | |
| Day of menstrual cycleb | 20 ± 12 | Unknown | n/a |
| Indication for USN | < 0.01 | ||
| Pelvic Pain | 10/60 (17%) | 20/20 (100%) | |
| Abnormal uterine bleeding | 17/60 (28%) | 0/20 (0%) | |
| IUD localization | 8/60 (13%) | 0/20 (0%) | |
| Ovarian cyst | 16/60 (27%) | 0/20 (0%) | |
| Follow up of known pathology | 9/60 (15%) | 0/20 (0%) |
All data presented as n(%) unless otherwise noted
n/a not applicable
All p values calculated with chi-square test unless otherwise noted
aMedian (interquartile range)
bMean ± standard deviation
cMann–Whitney U test used to calculate p value
dFisher’s exact test used to calculate p value
Fig. 1Comparative imaging of simple ovarian cyst with a GE Vscan, b Sonosite Iviz, c Philips Lumify
Fig. 2Bar graphs comparing, a image quality score, b correlation to diagnostic ultrasound image score for three POC-US devices. Bars followed by a different letter are significantly different at alpha level of 0.05 by Fisher’s exact test. Error bars represent 95% CI
Fig. 3Agreement between measurements of three POC-US devices and diagnostic ultrasound. Absolute mean difference in measurements for a uterine volume, b endometrial thickness, c right ovarian mean diameter, d left ovarian mean diameter, e pathology of interest mean diameter shown with error bars representing 95% limits of agreement. Bars followed by asterisk indicate a significant difference in measurements between POC-US and Dx-US by t test at alpha level of 0.05
Concordance of nominal variables Lumify POC-US versus Dx-Us
| Variable | na | Categories | Cohen’s unweighted kappa coefficient | |
|---|---|---|---|---|
| Uterine position | 43 | Anteverted, retroverted, mid-position | 0.7 | < 0.0001 |
| Endometrial pathology | 37 | Yes, no | 0.9 | < 0.0001 |
| Endometrial pathology type | 12 | Thickened, fluid, IUD, none | 0.8 | < 0.0001 |
| Fibroids | 39 | Yes, no | 0.7 | < 0.0001 |
| Right ovary | 38 | Seen, not seen | 0.4 | 0.02 |
| Left ovary | 37 | Seen, not seen | 0.4 | 0.002 |
| Adnexal pathology present | 40 | Seen, not seen | 0.7 | < 0.0001 |
| Side of ovarian pathology | 14 | Left, right | 1 | < 0.0001 |
| Type of ovarian pathology | 18 | Normal, functional follicle, abnormal | 0.7 | < 0.0001 |
| Final diagnosisa | 50 | Normal, ovarian finding, endometrial finding, fibroid, cul de sac finding | 0.7 | < 0.0001 |
aSubjects with more than 1 diagnosis included for each category
Fig. 4Agreement between measurements of Lumify POC-US and Dx-US of a uterine volume, b endometrial thickness, c right ovarian mean diameter, d left ovarian mean diameter, e adnexal pathology mean diameter, f myoma mean diameter by (i) linear regression and (ii) Bland–Altman plot analysis.
Sensitivity, specificity, positive predictive value and negative predictive value of Lumify POC-US for pelvic pathology
| Variable | PPV (%) | NPV (%) | Sensitivity (%) | Specificity (%) |
|---|---|---|---|---|
| Endometrial pathology | 90.9 | 100 | 100 | 96.4 |
| Fibroids | 80 | 93.1 | 80 | 93.1 |
| Adnexal pathology | 76.5 | 82.6 | 76.5 | 82.6 |
| Overall | 81.3 | 80.4 | 74.7 | 85.7 |
| Vscan | Iviz | Lumify | |
|---|---|---|---|
| Battery charging time | 90 min | 120 min using external charger 4 h with battery in the system using USB charger | Depends on tablet device used |
| Presets | Cardiac, Abdominal Obstetrics | Depends on probe selected | Gallbladder, Abdomen, Lung, OBGYN |
| Depth Adjustment | 6–24 cm | 5.1–32 cm | 1–30 cm |
| Gain Adjustment | No, Autogain | Yes (0–100); near gain and far gain present | Yes (0–100) |
| Color Flow | Yes | Yes | Yes, fast flow and slow flow |
| M-mode | No | Yes | Yes |
| Cine | No | Yes, 256 frames | No |
| Save loop | Yes (2 s) | Yes (2, 4, 6, 10, 15, 30 and 60 s) | Yes (3–10 s) |
| Power | No | No | Yes (0 to negative 30) |
| Dynamic range | No | Yes | No |
| Measurements | Caliper distance (single per image) | Caliper distance and volume (multiple per image) | Caliper distance and volume (multiple per image |
| No additional calculation packages | Additional calculation packages | Additional calculation packages | |
| Annotatation | Voice notations | Keyboard | Keyboard |
| Can't annotate and measure on same image | Can annotate and measure on same image | ||
| Can customize labels | – | ||
| Export | USB, Email | Local PACS, Cloud, USB, Email, Printer | DICOM, network share, Local directory, USB, Email |
| Barcode scanner | No | Yes | Yes |
| Worklist | No | Yes | Yes |
| Start-up time | 20 s | < 35 s | < 5 s |
| Controls | Buttons | Touchscreen | Touchscreen |
| Ability to reopen a closed study | No | No | No |
| Structured reports | No | Yes | Yes |