Kyle DeGruy1, Katherine Klein2, Zilma Rey1, Patricia Hall1, Andrea Kim1, Heather Alexander1.
Abstract
BACKGROUND: Proficiency testing (PT) is part of a comprehensive quality assurance programme, which is critical to ensuring patients receive accurate and reliable diagnostic testing. Implementation of the Cepheid Xpert® MTB/RIF assay to aid in the diagnosis of tuberculosis has expanded rapidly in recent years; however, PT material for Xpert MTB/RIF is not readily available in many resource-limited settings.Entities:
Keywords: DTS; EQA; Xpert MTB/RIF; dried tube specimen; external quality assessment; proficiency testing
Year: 2020 PMID: 33102169 PMCID: PMC7564946 DOI: 10.4102/ajlm.v9i1.1166
Source DB: PubMed Journal: Afr J Lab Med ISSN: 2225-2002
FIGURE 1Overview of dried tube specimens for Xpert MTB/RIF preparation procedure, 2013–2015, United States Centers for Disease Control and Prevention (Atlanta, Georgia, United States).
FIGURE 2Testing dried tube specimens using Xpert MTB/RIF, 2013–2015, United States Centers for Disease Control and Prevention (Atlanta, Georgia, United States).
Accuracy of dried tube specimen Xpert MTB/RIF test results collected during panel validation, 2013–2015, United States Centers for Disease Control and Prevention (Atlanta, Georgia, United States).
| PT Panel | Concordant | Discordant | RIF Indeterminate | Error | Total | |||||
|---|---|---|---|---|---|---|---|---|---|---|
| % | % | % | % | % | ||||||
| 2013-A | 49 | 98.0 | 0 | 0.0 | 0 | 0.0 | 1 | 2.0 | 50 | 100 |
| 2013-B | 81 | 90.0 | 1 | 1.1 | 4 | 4.4 | 4 | 4.4 | 90 | 100 |
| 2013-C | 121 | 96.8 | 1 | 0.8 | 3 | 2.4 | 0 | 0.0 | 125 | 100 |
| 2013-D | 122 | 97.6 | 0 | 0.0 | 1 | 0.8 | 2 | 1.6 | 125 | 100 |
| 2014-A | 120 | 96.0 | 0 | 0.0 | 0 | 0.0 | 5 | 4.0 | 125 | 100 |
| 2015-A | 241 | 96.4 | 0 | 0.0 | 0 | 0.0 | 9 | 3.6 | 250 | 100 |
| 2015-B | 244 | 97.6 | 1 | 0.4 | 0 | 0.0 | 5 | 2.0 | 250 | 100 |
| 2015-C | 271 | 98.5 | 0 | 0.0 | 0 | 0.0 | 4 | 1.5 | 275 | 100 |
PT, proficiency testing; RIF, rifampicin.
, Exact proportions are provided; totals may not add up to 100% because of unrepresented fractions.
Errors received during validation of dried tube specimens for 2013–2015 proficiency testing panels, 2013–2015, United States Centers for Disease Control and Prevention (Atlanta, Georgia, United States).
| Error code | Error description | Number detected | |
|---|---|---|---|
| % | |||
| 2005 | Motion of the syringe drive was not detected. | 3 | 10 |
| 2014 | The digital temperature reading of n for Thermistor A/Thermistor B/Ambient Thermistor/Optic Thermistor was not within the acceptable range of m1 to m2. | 1 | 3.3 |
| 2037 | Cartridge integrity test failed at valve position n. | 1 | 3.3 |
| 5006/5007 | X probe check failed. | 24 | 80 |
| 5011 | Signal loss detected in the amplification curve for analyte [x]. | 1 | 3.3 |
, Error descriptions are from GeneXpert Dx 4.0 Operator Manual; 300-7607. Rev2010.
, Exact proportions are provided; totals may not add up to 100% because of unrepresented fractions.
FIGURE 3Standard deviations of Probe A and specimen processing control cycle threshold values for dried tube specimens containing Mycobacterium tuberculosis complex tested during panel validation, 2013–2015, United States Centers for Disease Control and Prevention (Atlanta, Georgia, United States).