Cyrille Haddar1, Paul O Verhoeven2, Thomas Bourlet3, Bruno Pozzetto4, Sylvie Pillet5.
Abstract
BACKGROUND: Facing the emergence of a new RNA virus, clinical laboratories are often helpless in the case of a shortage of reagents recommended by Reference Centres.Entities:
Keywords: Coronavirus; Emerging virus; Laboratory-developed assay; RNA virus; RT-qPCR; SARS-CoV-2
Mesh:
Substances:
Year: 2020 PMID: 33099260 PMCID: PMC7476897 DOI: 10.1016/j.jcv.2020.104636
Source DB: PubMed Journal: J Clin Virol ISSN: 1386-6532 Impact factor: 3.168
Overall results comparing the 5 open one step RT-qPCR reagents to the reference assay.
| Complete denomination (commercialisation) | Abbreviation used in this study | True positive samples (n) | True negative samples (n) | False negative samples (n) | Concordance (%) | Kappa coefficient (CI 95 %) |
|---|---|---|---|---|---|---|
| SuperScript™ III Platinum™ One-Step qRT-PCR kit (Invitrogen, Cergy-Pontoise) | SIII (reference assay) | 51 | 3 | 0 | – | – |
| TaqMan Fast Virus 1-Step Master Mix (Applied Biosystems) | Fast Virus | 50 | 3 | 1 | 98.15 | 0.848 (0.584−1.111) |
| Luna® Universal Probe One-Step RT-qPCR Kit (New England Biolabs, Evry) | Luna | 49 | 3 | 2 | 96.30 | 0.731 (0.474−0.988) |
| GoTaq® Probe 1- Step RT-qPCR System (Promega France, Charbonnières-les-Bains) | GoTaq | 51 | 3 | 0 | 100 | 1 (0.733−1.267) |
| LightCycler® Multiplex RNA Virus Master (Roche, Meylan) | LC MM | 49 | 3 | 2 | 96.30 | 0.731 (0.474−0.988) |
| One-step PrimeScript RT-PCR kit (Takara, Ozyme, Saint-Cyr-L’École) | PrimeScript | 50 | 3 | 1 | 98.15 | 0.848 (0.584−1.111) |
CI: Confidence interval.
The reverse transcription (RT) step was done during 5 min at 48 °C and after inactivation of the enzyme during 20 s at 95 °C, the cycling program was of 5 s at 95 °C and 30 s at 58 °C during 50 cycles.
Same RT and cycling programs than the reference assay.
Fig. 1Scatter plots of correlation between the CT values of the IP4 region of the viral RNA dependant RNA polymerase gene determined using 5 alternative open one step RT-qPCR reagents by analysis with the 48 extracts exhibiting a CT value <35 with the reference assay (SuperScript™ III Platinum™ One-Step qRT-PCR kit Invitrogen). The abbreviations of all reagents tested are listed in the Table 1. A to E- Logistic regression. Each point represents the CT value measured by the tested reagent in comparison to that of the reference assay (SIII) of the same extract. The Pearson coefficients with their 95 % Confidence interval are drawn on each figure. F to J- Bland-Altman plot. Each point represents the difference observed between the results of the two methods against their mean. A and F- Comparison of the Fast Virus to the SIII. B and G- Comparison of the Luna to the SIII. C and H- Comparison of the LC MM to the SIII. D and I- Comparison of the GoTaq to the SIII. E and J- Comparison of the PrimeScript to the SIII.