Performance of the Prototype NovioSense Noninvasive Biosensor for Tear Glucose in Type 1 Diabetes.

Petronella Geelhoed-Duijvestijn1, Dovile Vegelyte2, Alicja Kownacka2, Nicoleta Anton3, Maurits Joosse1, Christopher Wilson2.   

Abstract

INTRODUCTION: We recently published the results of a pilot study measuring glucose in tear fluid. We now show the results of an additional 24 patients.
METHODS: Twenty-four subjects were recruited from Haaglanden Diabetes Centre. The patients reported in a fasting state and were given a meal with half the usual dose of insulin during the test. The device was applied under the lower eyelid. Glucose levels from capillary blood and interstitial fluid with a flash glucose measurement device were recorded every 15 minutes; the current from the tear glucose sensor was recorded continuously. The eye surface and tolerability were regularly checked. A calibration algorithm to convert tear glucose to blood values was built using a neural network regression model and validated.
RESULTS: No adverse events were attributed to the sensor coil placed under the lower eyelid. The mean absolute relative difference for the 24-patient subset was 16.7 (after 6 hours total time in the eye). The median absolute relative difference was 13.3. Compared to published data from Abbott (15.7 on day 1), the present device is comparable to Libre, considering that the device was allowed only one hour of equilibration time before the measurements were made.
CONCLUSION: The NovioSense Tear glucose sensor measures blood glucose values with an acceptable accuracy and may become a good alternative to invasive devices.

Entities:  

Keywords:  glucose monitoring; glucose sensor; tear glucose; type 1 diabetes

Mesh:

Substances:

Year:  2020        PMID: 33095035      PMCID: PMC8655288          DOI: 10.1177/1932296820964844

Source DB:  PubMed          Journal:  J Diabetes Sci Technol        ISSN: 1932-2968


  18 in total

1.  A flexible and wearable biosensor for tear glucose measurement.

Authors:  Shigehito Iguchi; Hiroyuki Kudo; Takao Saito; Mitsuhiro Ogawa; Hirokazu Saito; Kimio Otsuka; Akio Funakubo; Kohji Mitsubayashi
Journal:  Biomed Microdevices       Date:  2007-08       Impact factor: 2.838

2.  Eyeglasses-based tear biosensing system: Non-invasive detection of alcohol, vitamins and glucose.

Authors:  Juliane R Sempionatto; Laís Canniatti Brazaca; Laura García-Carmona; Gulcin Bolat; Alan S Campbell; Aida Martin; Guangda Tang; Rushabh Shah; Rupesh K Mishra; Jayoung Kim; Valtencir Zucolotto; Alberto Escarpa; Joseph Wang
Journal:  Biosens Bioelectron       Date:  2019-05-07       Impact factor: 10.618

3.  The effect of intensive treatment of diabetes on the development and progression of long-term complications in insulin-dependent diabetes mellitus.

Authors:  D M Nathan; S Genuth; J Lachin; P Cleary; O Crofford; M Davis; L Rand; C Siebert
Journal:  N Engl J Med       Date:  1993-09-30       Impact factor: 91.245

Review 4.  Clinical review 2: The "metabolic memory": is more than just tight glucose control necessary to prevent diabetic complications?

Authors:  Antonio Ceriello; Michael A Ihnat; Jessica E Thorpe
Journal:  J Clin Endocrinol Metab       Date:  2008-12-09       Impact factor: 5.958

Review 5.  Tear glucose analysis for the noninvasive detection and monitoring of diabetes mellitus.

Authors:  Justin T Baca; David N Finegold; Sanford A Asher
Journal:  Ocul Surf       Date:  2007-10       Impact factor: 5.033

6.  Time lag of glucose from intravascular to interstitial compartment in humans.

Authors:  Ananda Basu; Simmi Dube; Michael Slama; Isabel Errazuriz; Jose Carlos Amezcua; Yogish C Kudva; Thomas Peyser; Rickey E Carter; Claudio Cobelli; Rita Basu
Journal:  Diabetes       Date:  2013-09-05       Impact factor: 9.461

7.  The Performance and Usability of a Factory-Calibrated Flash Glucose Monitoring System.

Authors:  Timothy Bailey; Bruce W Bode; Mark P Christiansen; Leslie J Klaff; Shridhara Alva
Journal:  Diabetes Technol Ther       Date:  2015-07-14       Impact factor: 6.118

8.  Adherence of self-monitoring of blood glucose in persons with type 1 diabetes in Sweden.

Authors:  Peter Moström; Elsa Ahlén; Henrik Imberg; Per-Olof Hansson; Marcus Lind
Journal:  BMJ Open Diabetes Res Care       Date:  2017-04-06

9.  Noninvasive Self-diagnostic Device for Tear Collection and Glucose Measurement.

Authors:  Seung Ho Lee; Yong Chan Cho; Young Bin Choy
Journal:  Sci Rep       Date:  2019-03-18       Impact factor: 4.379

10.  Clinical Evidence for Use of a Noninvasive Biosensor for Tear Glucose as an Alternative to Painful Finger-Prick for Diabetes Management Utilizing a Biopolymer Coating.

Authors:  Alicja E Kownacka; Dovile Vegelyte; Maurits Joosse; Nicoleta Anton; B Jelle Toebes; Jan Lauko; Irene Buzzacchera; Katarzyna Lipinska; Daniela A Wilson; Nel Geelhoed-Duijvestijn; Christopher J Wilson
Journal:  Biomacromolecules       Date:  2018-10-25       Impact factor: 6.988

View more
  6 in total

Review 1.  Sense and Learn: Recent Advances in Wearable Sensing and Machine Learning for Blood Glucose Monitoring and Trend-Detection.

Authors:  Ahmad Yaser Alhaddad; Hussein Aly; Hoda Gad; Abdulaziz Al-Ali; Kishor Kumar Sadasivuni; John-John Cabibihan; Rayaz A Malik
Journal:  Front Bioeng Biotechnol       Date:  2022-05-12

Review 2.  Small-volume detection: platform developments for clinically-relevant applications.

Authors:  Wei-Hsuan Sung; Yu-Ting Tsao; Ching-Ju Shen; Chia-Ying Tsai; Chao-Min Cheng
Journal:  J Nanobiotechnology       Date:  2021-04-21       Impact factor: 10.435

3.  Quantification of tear glucose levels and their correlation with blood glucose levels in dogs.

Authors:  Eunji Lee; Seonmi Kang; Jaeho Shim; Dajeong Jeong; Youngseok Jeong; Junyeong Ahn; Kangmoon Seo
Journal:  Vet Med Sci       Date:  2022-03-19

Review 4.  Products for Monitoring Glucose Levels in the Human Body With Noninvasive Optical, Noninvasive Fluid Sampling, or Minimally Invasive Technologies.

Authors:  Trisha Shang; Jennifer Y Zhang; Andreas Thomas; Mark A Arnold; Beatrice N Vetter; Lutz Heinemann; David C Klonoff
Journal:  J Diabetes Sci Technol       Date:  2021-06-13

Review 5.  Commercial and Scientific Solutions for Blood Glucose Monitoring-A Review.

Authors:  Yirui Xue; Angelika S Thalmayer; Samuel Zeising; Georg Fischer; Maximilian Lübke
Journal:  Sensors (Basel)       Date:  2022-01-06       Impact factor: 3.576

Review 6.  Exhaled Breath Analysis for Diabetes Diagnosis and Monitoring: Relevance, Challenges and Possibilities.

Authors:  Kaushiki Dixit; Somayeh Fardindoost; Adithya Ravishankara; Nishat Tasnim; Mina Hoorfar
Journal:  Biosensors (Basel)       Date:  2021-11-25
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.