| Literature DB >> 33083654 |
Andrew R Barsky1, Michele M Kim1, Alexandra Dreyfuss1, Lei Dong1, James M Metz1, Shibu Anamalayil1, Taoran Li1, Neil K Taunk1.
Abstract
PURPOSE: Radiation therapy (RT) is commonly used in the treatment of gynecologic cancers. Intensity-modulated RT (IMRT) has been shown to reduce gastrointestinal toxicity compared with 2-dimensional and 3-dimensional RT modalities. We report the initial clinical experience using IMRT for gynecologic cancers with a novel 6MV flattening filter free O-ring linear accelerator (6X-FFF ORL). METHODS AND MATERIALS: We retrospectively identified consecutive women with uterine or cervical cancer who received pelvic RT on Halcyon (Varian Medical Systems, Palo Alto, CA), a novel 6X-FFF ORL. We report their clinicopathologic data, RT details, early disease-control outcomes, acute toxicities, dose-volume histogram data, couch corrections, and treatment times.Entities:
Year: 2020 PMID: 33083654 PMCID: PMC7557142 DOI: 10.1016/j.adro.2020.01.004
Source DB: PubMed Journal: Adv Radiat Oncol ISSN: 2452-1094
Figure 1Imaging modality comparison. Representative images from planning computed tomography (CT), 6MV flattening filter free O-ring linear accelerator (6X-FFF ORL) MV cone beam CT (CBCT), and 6X-FFF ORL KV CBCT.
Details of radiation therapy course
| Variable | Value (%) |
|---|---|
| RT position | |
| Supine | 17 (100) |
| RT field | |
| Whole pelvis | 15 (88) |
| Extended field | 2 (12) |
| Delivered dose (Gy) | |
| Median | 50.4 |
| Range | 19.8-2555.0 |
| No. of fractions | |
| Median | 25 |
| Range | 11-28 |
| Bladder filling | |
| Full and empty | 16 (94) |
| Comfortable | 1 (6) |
| IGRT | |
| MV CBCT | 6 (35) |
| kV CBCT | 11 (65) |
| Brachytherapy boost | |
| Yes | 6 (35) |
| No | 11 (65) |
| Boost modality | |
| High dose rate | 6 (100) |
| Boost dose (Gy) | |
| Median | 18.5 |
| Range | 10-27.5 |
| Boost no. of fractions | |
| Median | 3 |
| Range | 2-5 |
| Systemic therapy | |
| Concurrent only | 3 (18) |
| Sequential only | 6 (35) |
| Concurrent and sequential | 2 (12) |
| None | 6 (35) |
Abbreviations: CBCT = cone beam computed tomography; IGRT = image guided radiation therapy; RT = radiation therapy.
One patient was planned to receive 50.4 Gy in 28 fractions but did not complete her course owing to social difficulties. She received 19.8 Gy in 11 fractions, with concurrent cisplatin.
Patients were simulated with both full and empty bladders and treated with attempted full bladders. One patient was unable to tolerate bladder filling and was simulated and treated with a comfortably full bladder.
Brachytherapy boosts are not delivered on the 6MV flattening filter free O-ring linear accelerator (6X-FFF ORL).
Dosimetric parameters of targets and OARs
| Variable | Value (%) |
|---|---|
| PTV | |
| V97 (median, range) | 98.9 (95.8-100) |
| V 110 | 0 (0-0.3) |
| Rectum | |
| V40 | 59.2 (16.2-98.9) |
| Bowel bag | |
| V40 | 14.4 (1.7-35.3) |
| Large bowel | |
| Maximum (Gy) | 53.2 (44.9-56.7) |
| Bladder | |
| V45 | 27.1 (0.2-52.3) |
| Pelvic bones | |
| V10 | 96.7 (90-99.5) |
| V25 | 71.5 (45.5-84.3) |
| Left pelvic bones | |
| V10 | 95.5 (85.4-99.2) |
| V25 | 66.5 (44.6-81.8) |
| Right pelvic bones | |
| V10 | 95.9 (87.1-99.5) |
| V25 | 64.4 (40.7-78.2) |
| Sacrum | |
| V10 | 100 (93-100) |
| V25 | 96.7 (47.9-100) |
| Duodenum | |
| D0.03 mL (Gy) | 47.8 (44-51.7) |
| V55 (mL) | 0 (0-0) |
| Kidneys | |
| V18 | 17.9 (13.2-22.5) |
| Spinal cord | |
| Maximum (Gy) | 23 (18.4-27.7) |
Abbreviations: OAR = organs-at-risk; PTV = planning target volume.
Only extended-field radiation therapy patients were evaluated.
Figure 2Couch corrections after initial setup based on image-guided radiation therapy (IGRT). Average 3-dimensional vector couch correction between skin-marker alignment and IGRT for all patients, for all fractions of 6MV flattening filter free O-ring linear accelerator treatment.
Figure 3Treatment time. Average treatment time (from start of imaging to beam off), for 8 patients, for all standard fractions of 6MV flattening filter free O-ring linear accelerator treatment. The average treatment times for nonstandard fractions are separately displayed.
Figure 4Total room time. Average total room time (time from leaving gowned waiting area to return), for 8 patients, for all standard fractions of 6MV flattening filter free O-ring linear accelerator treatment. The average total room times for nonstandard fractions are separately displayed.