Utilization of a 3D printed dental splint for registration during electromagnetically navigated mandibular surgery.

S G Brouwer de Koning1, J J M Riksen2, T P Ter Braak2, M J A van Alphen1, F van der Heijden2, W H Schreuder1, L H E Karssemakers1, M B Karakullukcu1, R L P van Veen3.   

Abstract

PURPOSE: A dental splint was developed for non-invasive rigid point-based registration in electromagnetically (EM) navigated mandibular surgery. Navigational accuracies of the dental splint were compared with the common approach, that is, using screws as landmarks.
METHODS: A dental splint that includes reference registration notches was 3D printed. Different sets of three points were used for rigid point-based registration on a mandibular phantom: notches on the dental splint only, screws on the mandible, contralateral screws (the side of the mandible where the sensor is not fixated) and a combination of screws on the mandible and notches on the dental splint. The accuracy of each registration method was calculated using 45 notches at one side of the mandible and expressed as the target registration error (TRE).
RESULTS: Average TREs of 0.83 mm (range 0.7-1.39 mm), 1.28 mm (1.03-1.7 mm), 2.62 mm (1.91-4.0 mm), and 1.34 mm (1.30-1.39 mm) were found, respectively, for point-based registration based on the splint only, screws on the mandible, screws on the contralateral side only, and screws combined with the splint.
CONCLUSION: For dentate patients, rigid point-based registration performs best utilizing a dental splint with notches. The dental splint is easy to implement in the surgical, and navigational, workflow, and the notches can be pinpointed and designated on the CT scan with high accuracy. For edentate patients, screws can be used for rigid point-based registration. However, a new design of the screws is recommended to improve the accuracy of designation on the CT scan.

Entities:  

Keywords:  Dental splint; Electromagnetic surgical navigation; Mandibular surgery; Rigid point-based registration; Target registration error

Mesh:

Year:  2020        PMID: 33067757     DOI: 10.1007/s11548-020-02271-3

Source DB:  PubMed          Journal:  Int J Comput Assist Radiol Surg        ISSN: 1861-6410            Impact factor:   2.924


  1 in total

1.  Electromagnetic surgical navigation in patients undergoing mandibular surgery.

Authors:  S G Brouwer de Koning; F Geldof; R L P van Veen; M J A van Alphen; L H E Karssemakers; J Nijkamp; W H Schreuder; T J M Ruers; M B Karakullukcu
Journal:  Sci Rep       Date:  2021-02-25       Impact factor: 4.379

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.