| Literature DB >> 33060791 |
Moritz Späth1,2, Martin Hohmann3,4, Clemens Roider3, Benjamin Lengenfelder3,4, Florian Stelzle3,5,4, Stefan Wirtz6, Florian Klämpfl3,4.
Abstract
Due to significant advantages, the trend in the field of medical technology is moving towards minimally or even non-invasive examination methods. In this respect, optical methods offer inherent benefits, as does diffuse reflectance imaging (DRI). The present study attempts to prove the suitability of DRI-when implemented alongside a suitable setup and data evaluation algorithm-to derive information from anatomically correctly scaled human capillaries (diameter: [Formula: see text], length: [Formula: see text]) by conducting extensive Monte-Carlo simulations and by verifying the findings through laboratory experiments. As a result, the method of shifted position-diffuse reflectance imaging (SP-DRI) is established by which average signal modulations of up to 5% could be generated with an illumination wavelength of [Formula: see text] and a core diameter of the illumination fiber of [Formula: see text]. No reference image is needed for this technique. The present study reveals that the diffuse reflectance data in combination with the SP-DRI normalization are suitable to localize human capillaries within turbid media.Entities:
Mesh:
Year: 2020 PMID: 33060791 PMCID: PMC7567838 DOI: 10.1038/s41598-020-74447-5
Source DB: PubMed Journal: Sci Rep ISSN: 2045-2322 Impact factor: 4.379
Skin layers considered in the MC simulations.
| Tissue layer | Thickness along | Vasculature (diameter) |
|---|---|---|
| Stratum corneum | 20 | – |
| Epidermis | 80 | – |
| Papillary dermis | 150 | Loop ( |
| Upper blood net dermis | 80 | Horizontal cylinder ( |
| Reticular dermis | 1500 | – |
| Deep blood net dermis | 80 | – |
| Subcutaneous tissue | 6000 | – |
Figure 1(a) Illustration of the vasculature considered in the MC simulation, consisting of the capillary loops and the superficial vascular plexus. Most of the capillary loops are aligned along the y–z plane, while four of them are aligned along the x–z plane. The coordinate system relates to the complete simulation volume. The spacing (along the x axis) between two branches of the superficial vascular plexus is 30 px ( ). (b) Detailed view of a vasculature branch with its dimensions. The dimensioning starts at the centers of the cylinders.
Optical properties of the skin layers and the microvasculature considered in the MC simulation[18–22].
| Element | 414 nm | 424 nm | 540 nm | |||||
|---|---|---|---|---|---|---|---|---|
| Stratum corneum | 1.57 | 50.00 | 1.46 | 50.00 | 1.00 | 50.00 | 0.90 | 1.53 |
| Epidermis | 3.44 | 14.60 | 3.19 | 13.96 | 1.81 | 7.84 | 0.85 | 1.34 |
| Papillary dermis | 0.86 | 13.50 | 0.80 | 13.36 | 0.45 | 11.09 | 0.80 | 1.40 |
| Upper blood net dermis | 1.23 | 13.50 | 1.14 | 13.36 | 0.65 | 11.09 | 0.90 | 1.39 |
| Reticular dermis | 0.74 | 13.50 | 0.68 | 13.36 | 0.39 | 11.09 | 0.76 | 1.40 |
| Deep blood net dermis | 1.47 | 13.50 | 1.37 | 13.36 | 0.78 | 11.09 | 0.95 | 1.39 |
| Subcutaneous tissue | 0.69 | 7.00 | 0.64 | 7.00 | 0.36 | 7.00 | 0.80 | 1.44 |
| Microvasculature | 283.00 | 8.00 | 203.17 | 8.00 | 28.75 | 8.00 | 0.96 | 1.36 |
Figure 2(a) Schematic of the experimental setup consisting of an optical phantom with a cantilever incorporated to imitate a capillary loop, a traversable illumination by means of a glass fiber and a microscope setup to image the diffuse reflection from the phantom surface on a camera chip. By way of illustration, the cantilever is turned by 90 degree in the sketch. (b) Photograph of the setup. (c) Transmitted light microscope image of the cantilever used in the optical phantom. To optimally mimic a capillary loop, material was removed from the cantilever over a large area by ultrashort pulse laser ablation.
Figure 3(a) Result for the SP-DRI method. The default vasculature is encountered and the positions of the vasculature and the capillary loops are indicated by the dashed lines (the capillary loops that are actually aligned along the y axis are marked in white, compare Fig. 1a. The position of the illumination fiber is also shown. (b) Logarithmic signal for the raw data set 1 which was used for the SP-DRI normalization in (a). The position of the cross-sectional plane is also shown. (c) Graph of the cross-sectional plane at px for the raw data in (b).
Figure 4Results for the examination of the specificity with a difference in the number of capillary loops and vasculature branches. The positions of the vasculature and the capillary loops are indicated by the dashed lines (the capillary loops that are actually aligned along the y axis are marked in white). The position of the illumination fiber (size not according to scale) and of the cross-sectional plane is also shown. (a) SP-DRI signal for the setup with only three capillary loops at each of the vasculature branches. (b) SP-DRI signal for the setup with only two vasculature branches. (c) Graph of the cross-sectional plane at px to compare the two validation setups with the default setup as shown in Fig. 3.
Figure 5Results for the examination of the ability of two-point discrimination with a random distribution of the vasculature. (a) Map with the positions of the four capillary loops indicated by the dashed lines. The capillary loops that are actually aligned along the y axis are marked in white. The positions of the illumination fiber and of the cross-sectional planes are also shown. (b) Cross-section at the two relevant x positions.
Figure 6Results of the laboratory experiment for the proof of concept of the SP-DRI normalization method. (a) Image of the result, where the position of the cantilever structure inside the optical phantom is clearly visible. The position of the illumination fiber is not shown here as it is outside of the range illustrated on the map. (b) A cross-section through the image from (a) at the position px.