Literature DB >> 33018767

Temporally Interfering TMS: Focal and Dynamic Stimulation Location.

Majid Memarian Sorkhabi, Karen Wendt, Timothy Denison.   

Abstract

In this study, we present a temporal interference (TI) concept to achieve focal and steerable stimulation in the targeted brain area through transcranial magnetic stimulation (TMS). This method works by inducing two high-frequency electric fields with a slight frequency difference via two independent coils. The intrinsic nonlinear nature of the nerve membrane, which acts as a low-pass filter, does not allow the nerve to engage at high frequencies. Instead, neurons at the intersection of two electric fields can follow the frequency difference of the two fields. For 3D MRI-derived head models, the finite element method is used to compute the electric field induced by the time-varying magnetic field along with the electric field penetration depth and the activated volume for the specific coil parameters. A deeper stimulation with an acceptable spatial spread can be obtained by controlling the intersection of the fields by finding the optimal position and orientation of the two coils. Moreover, by changing the voltage ratio of the coils, and not their mechanical orientation, the intended area can be dynamically driven. The computational results show that the TI technique is an efficient approach to resolve the electric field depth-focality trade-off, which can be a reasonable alternative to complex coil designs. The system proposed in this paper shows a great promise for a more dynamic and focused magnetic stimulation.

Entities:  

Mesh:

Year:  2020        PMID: 33018767      PMCID: PMC7610609          DOI: 10.1109/EMBC44109.2020.9176249

Source DB:  PubMed          Journal:  Annu Int Conf IEEE Eng Med Biol Soc        ISSN: 2375-7477


  13 in total

Review 1.  Resonance, oscillation and the intrinsic frequency preferences of neurons.

Authors:  B Hutcheon; Y Yarom
Journal:  Trends Neurosci       Date:  2000-05       Impact factor: 13.837

2.  Developing a more focal magnetic stimulator. Part I: Some basic principles.

Authors:  D Cohen; B N Cuffin
Journal:  J Clin Neurophysiol       Date:  1991-01       Impact factor: 2.177

3.  Why image-guided navigation becomes essential in the practice of transcranial magnetic stimulation.

Authors:  J-P Lefaucheur
Journal:  Neurophysiol Clin       Date:  2009-11-24       Impact factor: 3.734

Review 4.  Fundamentals of transcranial electric and magnetic stimulation dose: definition, selection, and reporting practices.

Authors:  Angel V Peterchev; Timothy A Wagner; Pedro C Miranda; Michael A Nitsche; Walter Paulus; Sarah H Lisanby; Alvaro Pascual-Leone; Marom Bikson
Journal:  Brain Stimul       Date:  2011-11-01       Impact factor: 8.955

5.  Design of transcranial magnetic stimulation coils with optimal trade-off between depth, focality, and energy.

Authors:  Luis J Gomez; Stefan M Goetz; Angel V Peterchev
Journal:  J Neural Eng       Date:  2018-06-01       Impact factor: 5.379

6.  STIMULUS: Noninvasive Dynamic Patterns of Neurostimulation Using Spatio-Temporal Interference.

Authors:  Jiaming Cao; Pulkit Grover
Journal:  IEEE Trans Biomed Eng       Date:  2019-05-30       Impact factor: 4.538

7.  Non-invasive magnetic stimulation of human motor cortex.

Authors:  A T Barker; R Jalinous; I L Freeston
Journal:  Lancet       Date:  1985-05-11       Impact factor: 79.321

8.  Electric field depth-focality tradeoff in transcranial magnetic stimulation: simulation comparison of 50 coil designs.

Authors:  Zhi-De Deng; Sarah H Lisanby; Angel V Peterchev
Journal:  Brain Stimul       Date:  2012-03-21       Impact factor: 8.955

Review 9.  Safety, ethical considerations, and application guidelines for the use of transcranial magnetic stimulation in clinical practice and research.

Authors:  Simone Rossi; Mark Hallett; Paolo M Rossini; Alvaro Pascual-Leone
Journal:  Clin Neurophysiol       Date:  2009-10-14       Impact factor: 3.708

10.  Theoretical Analysis of Transcranial Magneto-Acoustical Stimulation with Hodgkin-Huxley Neuron Model.

Authors:  Yi Yuan; Yudong Chen; Xiaoli Li
Journal:  Front Comput Neurosci       Date:  2016-04-19       Impact factor: 2.380

View more
  2 in total

1.  Physiological Artifacts and the Implications for Brain-Machine-Interface Design.

Authors:  Majid Memarian Sorkhabi; Moaad Benjaber; Peter Brown; Timothy Denison
Journal:  Conf Proc IEEE Int Conf Syst Man Cybern       Date:  2020-10

Review 2.  Shaping and Focusing Magnetic Field in the Human Body: State-of-the Art and Promising Technologies.

Authors:  Sabrina Rotundo; Danilo Brizi; Alessandra Flori; Giulio Giovannetti; Luca Menichetti; Agostino Monorchio
Journal:  Sensors (Basel)       Date:  2022-07-08       Impact factor: 3.847

  2 in total