| Literature DB >> 33015380 |
A Hunt1,2, I Hanson3, A Dunlop3, H Barnes2, L Bower1,2, J Chick3, C Cruickshank4, E Hall4, T Herbert2, R Lawes2, D McQuaid3, H McNair1,2, A Mitchell3, J Mohajer3, T Morgan1,2, U Oelfke3, G Smith2, S Nill3, R Huddart1,2, S Hafeez1,2.
Abstract
Whole bladder magnetic resonance image-guided radiotherapy using the 1.5 Telsa MR-linac is feasible. Full online adaptive planning workflow based on the anatomy seen at each fraction was performed. This was delivered within 45 min. Intra-fraction bladder filling did not compromise target coverage. Patients reported acceptable tolerance of treatment.Entities:
Keywords: Adaptive radiotherapy; MR-guided radiotherapy; MR-linac; bladder cancer
Year: 2020 PMID: 33015380 PMCID: PMC7522378 DOI: 10.1016/j.ctro.2020.09.002
Source DB: PubMed Journal: Clin Transl Radiat Oncol ISSN: 2405-6308
Patient characteristics.
| Patient 1 | Patient 2 | Patient 3 | Patient 4 | Patient 5 | |
|---|---|---|---|---|---|
| Age | 86 | 88 | 86 | 73 | 88 |
| Sex | Male | Male | Male | Female | Male |
| 90 | 70 | 80 | 80 | 80 | |
| Cancer staging | T3N0M0 bladder | T4aN0M0 bladder | T4bN1M0 bladder | T2N0Mx bladder | T2N2M0 bladder |
T2N0M0 prostate | T4N3M1 | T2N1M0 prostate | |||
Ureter | |||||
Whole bladder | Whole bladder | Whole bladder | Whole bladder | Whole bladder | |
Urethra (1.5 cm) | Prostate | Pelvic sidewall | Urethra (1.0c m) | Prostate | |
Prostate | |||||
| 106 | 134 | 175 | 87 | 334 | |
| Prescription dose | 36Gy in 6f | 36Gy in 6f | 36Gy in 6f | 30Gy in 5f | 36Gy in 6f |
| Clinical Notes | – | Long term urinary catheter in situ | – | Metastatic upper tract cancer, bladder radiotherapy given for bleeding | – |
KPS karnofsky peformance score.
Fig. 1Summary of results. Panel A: Time taken for key workflow stages. Box and whisker plot where x is mean value, horizontal line is median value, box edges are interquartile range, whisker edges are the maximum and minimum values, and circles are outliers. Panel B: Inter- and intra-patient variation in CTV as seen on MRIsession image. Panel C: Intra-fraction CTV change, note for patient 2 volume decreases occurred at some fractions likely due to presence of urinary catheter (on free flow). Panel D: Plan conformity index (CIRTOG) comparing CIRTOG of plan at time of MRIsession image compared to estimated delivered dose on MRIpost imaging. Aiming for a value approaching 1. Panel E: CTV coverage based on estimated dose delivered to the anatomy as seen at MRIverification and MRIpost time points, magnified graph on the right. Aiming for CTV coverage to be >95% (dotted line), achieved in all fractions apart from Patient 1 fraction 6.