SARS-CoV-2 infection serology validation of different methods: Usefulness of IgA in the early phase of infection.

Massimo Pieri1, Marco Ciotti2, Nicoletta Carlozzi3, Maria Loredana Frassanito3, Arianna Meloni3, Alessandro Cistera3, Giordano Turchetti3, Silvia Niscola3, Giuseppe Labate3, Graziella Calugi3, Sergio Bernardini4.   

Abstract

BACKGROUND AND AIMS: A novel coronavirus (SARS-CoV-2) was isolated from the respiratory samples of patients with pneumonia as showed by the sequence analysis of the virus genomes obtained in Wuhan, China. The antibody response to SARS-CoV-2 is not well understood yet, but the availability of sensitive and specific serological assays will be crucial for the early diagnosis of infection, for epidemiological studies and for defining the presence of neutralizing antibodies in response to a possible vaccine.
MATERIALS AND METHODS: We tested and compared the performances of one chemiluminescent immunoassay (CLIA), two enzyme-linked immunosorbent assay (ELISA) and an electrochemiluminescence immunoassay (ECLIA).
RESULTS: The ECLIA serological assay performed best and may be a valid screening method for SARS-COV-2 infection. The IgA detected by the ELISA assay might be a more reliable and stable early serological marker than IgM. Instead, IgGs, as expected, showed stable level after 10 days from symptoms onset.
CONCLUSION: The ECLIA method could be used as screening test, considering both the excellent performance and the cost per single test; while ELISA assay for IgG and IgA, which are present at a higher level than IgM and last longer, might be used as confirmatory test.
Copyright © 2020 Elsevier B.V. All rights reserved.

Entities:  

Keywords:  CLIA assay; Covid-19 serological test; ECLIA assay; ELISA assay; SARS-CoV-2

Mesh:

Substances:

Year:  2020        PMID: 33002475      PMCID: PMC7834570          DOI: 10.1016/j.cca.2020.09.033

Source DB:  PubMed          Journal:  Clin Chim Acta        ISSN: 0009-8981            Impact factor:   3.786


Introduction

In the late December 2019, an outbreak of pneumonia of unknown origin was reported in Wuhan, Hubei province, China. A novel coronavirus was isolated from the respiratory samples of patients with pneumonia as showed by the sequence analysis of the virus genomes obtained [1], [2]. The novel coronavirus was first named 2019-nCoV, and later SARS-CoV-2 because of its capacity to cause a severe acute respiratory syndrome (SARS) resembling that caused by SARS-CoV in 2002/2003. Infection is mild in the majority of the cases, but in some individuals, generally elderly and with comorbidities, the virus causes an atypical interstitial pneumonia progressing to acute lung injury and acute respiratory distress syndrome (ARDS) that requires respiratory support [3]. The disease, known as COVID-19 (https://www.who.int) spread rapidly all over the world, and on 11 March 2020, the WHO declared COVID-19 a pandemic. As of 16 August 2020, we have 21 294 845 COVID-19 confirmed cases and 761 779 deaths worldwide [4]. Reverse-transcriptase real-time PCR (rRT-PCR) is the method of choice for detecting SARS-CoV-2 infection and confirm suspected cases [5]. rRT-PCR is carried out on nasalpharyngeal swabs, throat swabs, bronchoalveolar lavage fluid, bronchoaspirate and sputum. Despite the high sensitive of the real-time PCR tests, sometimes samples from the upper respiratory tract may result negative even in the presence of radiological findings of pneumonia [6]. These negative results may have different explanations: i, the viral load in the upper respiratory tract is low compared to the lower respiratory tract; ii, the viral load may vary during the course of COVID-19 disease; iii.; low quality of the collected sample; iv, technical reasons linked to the assay used. In these cases, serological assays may help in making diagnosis. The antibody response to SARS-CoV-2 is not well understood yet, but the availability of sensitive and specific serological assays will be crucial for the early diagnosis of infection, for epidemiological studies (diffusion of the virus among the population, identification of asymptomatic carriers), for defining the presence of neutralizing antibodies in response to a possible vaccine. A recent publication showed how detection of antibody response in combination with RNA testing improved the sensitivity of etiological diagnosis of COVID-19 in the first week of illness [7]. Serological assays will play a key role also in the reopening of the economic activities, when it will be important to identify not only that people who were not exposed to SARS-CoV-2 and then can return to work, but also asymptomatic carriers that will be quarantined avoiding the spread of the virus in the work place and new infections [8]. Anyway, keeping in mind the limitations of the currently available serological assays and the possibility of cross-reactions with SARS-CoV (which shares 82% nucleotide identity with SARS-CoV-2 [9] and other coronaviruses, real-time PCR is nowadays the most effective diagnostic test for COVID-19 diagnosis. In this work, we tested and compared the performances of two enzyme-linked immunosorbent assays (ELISA), one chemiluminescent immunoassay (CLIA) and an electrochemiluminescence immunoassay (ECLIA). Of the latter ones, one detects IgM and IgG antibodies (SNIBE) while the other measures also IgA (Roche Diagnostics).

Materials and methods

Patients and serum specimens

Serum samples were collected from rRT-PCR-diagnosed SARS-CoV-2 positive (n = 40) and negative patients (n = 40) from “Tor Vergata” University Covid-Hospital of Rome. The study was conducted in accordance with the guidelines of the local Ethic Committee (approval number: R.S.44.20) and the Helsinki Declaration, as revised in 2013. Sera were collected by centrifugation at 2500 g for 10 min, within 1 h from collection. All serum samples were collected from 1 to 50 days after nasopharyngeal swab. Positive patients have been divided into three groups: early infection time (1–10 days after nasopharyngeal swab); late infection time (11–45 days after nasopharyngeal swab) and COVID convalescent patients (>45 days after a positive nasopharyngeal swab; discharged home).

Reverse real time-PCR (rRT-PCR)

Nasopharyngeal swabs were tested for SARS-CoV-2 infection with Seegene Allplex™2019-nCoV assay (Seegene, Seoul, South Korea), according to the manufacturer’s protocol. Automated RNA extraction and PCR setup were carried out using Seegene NIMBUS, an automatic liquid handling workstation. rRT-PCR was run on a CFX96TMDx platform (Bio-Rad Laboratories, Inc., CA, USA) and subsequently interpreted by Seegene’s Viewer Software. The Seegene Allplex™2019-nCoV assay identifies the virus by multiplex real-time PCR targeting three viral genes (E, RdRP and N), thus complying with international validated testing protocols [10].

Serological tests

Automated ROCHE Electrochemiluminescence Immunoassay (ECLIA) for anti-SARS-CoV-2 IgG, IgM or IgA detection.

The Roche Elecsys Anti-SARS-CoV-2 (Roche Diagnostics, Monza, Italy) is an immunoassay for the in vitro qualitative detection of antibodies (including IgG, IgM and IgA) to SARS-CoV-2 in human serum and plasma. The assay uses a recombinant protein representing the nucleocapsid (N) antigen for the determination of antibodies against SARS-CoV-2. The result is given as cut-off index (COI): positive (COI ≥ 1.0) or negative (COI < 1.0). The test was performed on Cobas 6000 immunoassay analyzer according to the manufacturer’s instructions.

SNIBE Chemiluminescence Immunoassay (CLIA)

The SNIBE 2019-nCoV IgG and 2019-nCoV IgM assays are an indirect chemiluminescence immunoassay from SNIBE (SNIBE Diagnostics, Shenzhen, China). The assay uses magnetic microbeads coated with 2019-nCoV recombinant antigen (nucleocapsid protein, NP and Spike Protein, SP) and anti-human IgG and IgM antibodies labeled with N‐(aminobutyl)‐N‐(ethylisoluminol) (ABEI) that form complexes. The light signal measured by a photomultiplier as relative light units (RLUs), is proportional to the concentration of 2019-nCoV IgG and IgM presented in the sample. The cut-off value of SARS-CoV-2 IgM/IgG positive was >1 AU/ml. The test was performed on Maglumi 600 according to the manufacturer’s instructions.

Euroimmun Enzyme-Linked Immunosorbent Assay (ELISA)

Immunoenzymatic assays for SARS-COV-2 IgG and IgA antibodies determination in human plasma and serum performed on the fully automated EURO LabWorkstation (Euroimmun, Lübeck, Germany). Each kit contains microplate strips with 8 break-off reagent wells coated with recombinant structural protein of SARS-CoV-2 (nucleocapsid protein, NP). In the first reaction step, diluted patient samples are incubated in the wells. In the case of positive samples, specific antibodies will bind to the antigens. To detect the bound antibodies, a second incubation is carried out using an enzyme-labelled anti-human IgA or IgG (enzyme conjugate) catalysing a colour reaction. Results are evaluated semi-quantitatively by calculation of a ratio of the extinction of the control or patient sample over the extinction of the calibrator. The ratio is interpreted as follows: <0.8 negative; ≥0.8 to <1.0 borderline; ≥1.1 positive.

ImmunoDiagnostics Enzyme-Linked Immunosorbent Assay (ELISA)

Immunoenzymatic assays for the determination of SARS-COV-2 IgM and IgG antibodies (Immunodiagnostic; ImmunoDiagnostics Limited Hong Kong Science Park, Sha Tin, Hong Kong) in human plasma and serum was performed on the fully automated Alisei Q.S., according to the manufacturer’s instructions. ImmunoDiagnostics SARS-CoV-2 NP IgG ELISA kit is a two-step incubation immunoassay. Recombinant nucleocapsid protein (NP) of SARS-CoV-2 pre-coated onto the polystyrene microwell strips is specifically recognized by anti-NP antibodies present in human serum or plasma specimens. A detection solution containing HRP-conjugated antihuman IgM/IgG is added for 1hr incubation, wherein HRP-conjugated anti-human IgM/IgG binds to the IgM/IgG class antibodies previously bound to NP protein on the plate. After removal of nonspecific binding, a substrate solution containing HRP is added resulting in the formation of a blue color. Color reaction is stopped by 2 M H2SO4, transforming the blue color to yellow signal, which is quantified by an absorbance microplate reader at 450 nm. The color intensity is proportional to the amount of anti-NP IgG antibodies captured inside the wells. Ratio is considered negative for all the values <0.9 COI (Cut-Off Index); equivocal for all the values between 0.9 and 1.1 COI; positive for all the values >1.1 COI. Table 1 summarizes the technical features of the aforementioned assays.
Table 1

Technical features of the serological assays tested in the study.

ManufacturerMethodQualitative/SemiquantitativeTarget AntigenAntibody measuredSensitivitySpecificity
ROCHEECLIAQualitativeNucleocapsid ProteinIgA/IgM/IgG65.5–100%*99.8%
SNIBECLIASemiquantitativeSpike and Nucleocapsid ProteinIgM/IgG91.2% (IgG)78.6% (IgM)97.3% (IgG)97.5% (IgM)
EUROIMMUNEELISASemiquantitativeNucleocapsid ProteinIgA/IgG33.3–80% (IgG)*50–100% (IgA)*98.5% (IgG)92.5% (IgA)
IMMUNODIAGNOSTICELISASemiquantitativeNucleocapsid ProteinIgM/IgG92.5% (IgG)88.2% (IgM)93.3% (IgG)92% (IgM)

Sensitivity range determined after PCR confirmation at different time points.

Technical features of the serological assays tested in the study. Sensitivity range determined after PCR confirmation at different time points.

Internal Quality Control (IQC) and External Quality Assessment (EQA)

All assays underwent to an Internal Quality Control (IQC), and precisions were expressed as the coefficient of variations (CV %) calculated as the standard deviation divided by the mean value, for each of the studied parameters. The External Quality Assessment (EQA) was performed with the Austrian Association for Quality Assurance and Standardization of Medical and Diagnostic Tests (ÖQUASTA).

Statistical analysis

Specificity and sensitivity were calculated by Receiver Operating Characteristic Curves (ROC Curve). All data were analyzed using Med Calc Ver.18.2.18 (MedCalc Software Ltd, Ostend, Belgium). The investigator was blinded to the group allocation during the experiment.

Results

Specificity and sensitivity of the serological assays

Sensitivities and specificities obtained with the immunoassays using ROC curve are summarized in Table 2 and Table 3 . The Roche combined antibody test is a screening test that provides a qualitative result. The Roche test showed excellent results with 99% sensitivity and 100% specificity considering all samples and the different groups of patients, Table 2.
Table 2

Sensitivities and specificities obtained with ROC curve for Roche ECLIA.

CONTROL NEGATIVE GROUP N = 40Total (1–50 days) N = 40Early infection time (1–10 days) N = 16Late infection time (11–45 days) N = 16Covid patients after discharge (>45) N = 8
TESTECLIA ROCHEECLIA ROCHEECLIA ROCHEECLIA ROCHE
Sensitivity (%)9910010099
Specificity (%)100100100100
Kit Cut-off>1 COI>1 COI>1 COI>1 COI
ROC curve (AUC)1111
Table 3

Sensitivities and specificities of the tested methods during the period of observation (1–50 days) based on the ROC curve.

CONTROL NEGATIVE GROUP N = 40Total (1–50 days) N = 40Early infection time (1–10 days) N = 16Late infection time (11–45 days) N = 16Covid patients after discharge (>45) N = 8
TESTCLIA_SNIBE
CLIA_SNIBE
CLIA_SNIBE
CLIA_SNIBE
IgMIgGIgMIgGIgMIgGIgMIgG
Sensitivity (%)759187.5881001005083
Specificity (%)10010075100100100100100
Kit Cut-off>1.1 COI>1.1 COI>1.1 COI>1.1 COI>1.1 COI>1.1 COI>1.1 COI>1.1 COI
Area under the ROC curve (AUC); 95% Confidence interval0.9210.9880.950.9680.94210.8210.988



TESTEuroimmune ELISA
Euroimmune ELISA
Euroimmune ELISA
Euroimmune ELISA
IgA
IgG
IgA
IgG
IgA
IgG
IgA
IgG
Sensitivity (%)87.59791.394949990100
Specificity (%)94100941009410094100
Kit Cut-off>1.1 COI>1.1 COI>1.1 COI>1.1 COI>1.1 COI>1.1 COI>1.1 COI>1.1 COI
Area under the ROC curve (AUC); 95% Confidence interval0.9870.9870.9770.96710.99410.9921



TESTImmunoDiagnostic ELISA
ImmunoDiagnostic ELISA
ImmunoDiagnostic ELISA
ImmunoDiagnostic ELISA
IgM
IgG
IgM
IgG
IgM
IgG
IgM
IgG
Sensitivity (%)75919486971008887.5
Specificity (%)100100100100979997100
Kit Cut-off>1.1 COI>1.1 COI>1.1 COI>1.1 COI>1.1 COI>1.1 COI>1.1 COI>1.1 COI
Area under the ROC curve (AUC); 95% Confidence interval0.9210.9880.9940.987110.9960.963
Sensitivities and specificities obtained with ROC curve for Roche ECLIA. Sensitivities and specificities of the tested methods during the period of observation (1–50 days) based on the ROC curve. The automated SNIBE chemiluminescence test on Maglumi instrument is a semi-quantitative test that reveals the presence of IgG and IgM antibodies against SARS-CoV-2. The system showed 100% specificity and 91% sensitivity for IgG antibodies, and a specificity and sensitivity for IgM antibodies of 100% and 75%, respectively (Table 3). Finally, we tested on automated systems two ELISA assays that measure IgA, IgM and IgG antibodies against NP. The Euroimmune ELISA measures IgA and IgG antibodies against NP. The sensitivity of IgG Euroimmune ELISA at ≤10 days was 94%, at 11–45 days was 99%; and at >45 days was 100%. The specificity was 100% in all groups. The sensitivity of IgA at ≤10 days was 91.3%, at 11–45 days 94%, and 90% at >45 days. Specificity was 94% in all groups, Table 3. The sensitivity of the Immunodiagnostic ELISA for IgG was 86% at ≤10 days, 100% at 11–45 days, and 87.5% >45 days. The specificity of IgG at ≤10 days was 100%, at 11–45 days 99%, and 100% at >45 days. The sensitivity for IgM at ≤10 days was 94%, at 11–45 days 97%, and was 88% at >45 days. Specificity was 100% at ≤10 days, 97% at 11–45 days, and 97% at >45 days, Table 3. Overall, sensitivity and specificity for IgG was similar for both CLIA and ELISA methods. Considering all patients, the sensitivity is similar with all methods, except for group >45 days from the first nasopharyngeal swab where a higher sensitivity was observed with ELISA methods (87.5–100% vs 83%). Among the control samples with negative RT-PCR, no false positives were observed with IgG ELISA and CLIA methods; while some false positives were observed with IgA and IgM ELISA. Fig. 1 shows the kinetic results for the 40 study patients at different days, from 1 to 50 days after the first positive RT-PCR, divided into time categories for IgA, IgM and IgG. The graph shows average values and corresponding standard error for IgA, IgM and IgG immunoglobulins at each infection time. IgA, IgM and IgG were detected since the first day after positive RT-PCR results, although differences were observed during the prefixed timeline (1–50 days). In particular, IgA and IgG immunoglobulins increased sharply up to 20 days, while IgM showed an unstable increase within the same time laps. After 20 days from the first positive swab, IgA levels decreased progressively but were still detectable until 50 days; instead, IgMs were barely detectable at the same time point. Finally, IgG maintained stable levels with absorbance values clearly above the cut-off level.
Fig. 1

Kinetcs of IgA (A), IgM (B) and IgG (C) from the symptoms onset in the 40 study patients. Different time frames are shown. Mean and standard deviations are plotted. Dashed-dotted line represent the cut-off value.

Kinetcs of IgA (A), IgM (B) and IgG (C) from the symptoms onset in the 40 study patients. Different time frames are shown. Mean and standard deviations are plotted. Dashed-dotted line represent the cut-off value.

Internal and external quality assessment

Due to limited reagent availability, imprecision was determined using positive and negative Internal Quality Control (IQC) only between run for a period of one month. Positive IQC generated the following coefficient of variations (CV %): 7.08% for ECLIA method; 6.29% (IgM) and 5.33% (IgG) for CLIA method. The two ELISA methods obtained 7.8% (IgG), 6.9% (IgA) and 5.1% (IgG), 5.8% (IgM) for Euroimmune ELISA and Immunodiagnostic ELISA respectively. Negative IQC showed a precision similar to that claimed by the manufacturer. The External Quality Assessment (EQA) for the identification of anti-SARS-CoV-2 IgG/IgM/IgA antibodies showed a perfect match with the results generated by the serological assays under evaluation.

Discussion

Several serological assays are available on the market for the diagnosis of SARS-CoV-2 infection [7]; therefore, assessment of their analytical performance by using clinical specimens is of critical importance. In this study, we evaluated the specificity and sensitivity of two ELISA assays (Euroimmun SARS-COV-2 IgG and IgA; ImmunoDiagnostics SARS-COV-2 NP IgM and IgG)) and two chemiluminescent enzyme assays (Roche Elecsys Anti-SARS-CoV-2; SNIBE 2019-nCoV IgG) for the detection of SARS-CoV-2 antibodies. Roche assay measures total immunoglobulins directed towards a recombinant NP of SARS-CoV-2. ELISA assays detect IgA, IgM and IgG to NP-antigen [11], [12]. Differences in the performance of single assays were observed that might be partially explained by differences in the targeted antigen. For instance, the spike (S) glycoprotein is densely glycosylated, with 66 N-linked glycosylation sites per trimer [8], but only few of them are targeted by neutralizing antibodies. The currently very low seroprevalence of SARS-CoV-2 in most regions of the world explains the low positive predictive values of serological testing. This can be somewhat improved by selecting the groups to be tested. The variable performance of the assays evaluated in this study highlights the need for laboratories to select carefully the best performing assays in order to optimize SARS-CoV-2 serodiagnostics. Our results demonstrated that the IgG testing is useful for the clinical diagnosis of SARS-CoV-2, with optimal specificity and sensitivity observed with both CLIA and ELISA methods, and particularly in patients tested over 2 weeks from the first positive nasopharyngeal swab. In the humoral immune response, IgM and IgA antibodies are generally produced earlier than IgG isotypes. Our data shows significant difference in the detection rate of IgM and IgA antibodies with IgA detected earlier than IgM suggesting that IgA might be more useful than IgM for early diagnosis of SARS-CoV-2 infection. The kinetics of IgM and IgG evaluated during a time course showed that both IgM and IgG rapidly increased after the onset of symptoms. Considering the cut-offs suggested by the manufacturer (COI > 1.0), the immunoglobulin rise appears already significant from the first days after symptoms onset. These findings are in agreement with those recently reported by other works [8], [13], [14], [15]. This study also confirms that simultaneous measurement of IgM/IgA and IgG can be useful, especially in the early phase of infection. In our opinion, it is really important to find a trustable marker able to detect an early infection in order to provide a risk scale of upcoming seroconversion during pandemic, trace contacts and activate serological surveillance to identify those who already came into contact with virus [8].

Conclusion

Among the platforms assessed in this study, Roche serological assay performed best, and may be a valid screening method for SARS-COV-2 infection. The IgA detected by the Euroimmune assay might be a more reliable and stable early serological marker than IgM. Instead, IgGs, as expected, showed stable level after 10 days from symptoms onset. Taken together, if a reflex test could be set in the laboratory, the ECLIA method could be used as screening test, considering both the excellent performance and the cost per single test; while the ELISA assay that detects IgG and IgA, which are present at a higher level than IgM and last longer, might be used as confirmatory test.

Contribution

MP, MC, GC and SB designed the study. MP, ML and GC was responsible for data collection and management. CG, MP, NC enrolled patients. AM, AC, SN, GL and GT performed assays. MP was responsible for biostatistics analyses. MP, MC and GC were responsible for interpretation of data. MP and MC prepared the tables and figures. MP, MC, GC and SB were drafting the manuscript. All authors contributed to revision of the manuscript, and approved it for submission.

Funding

This research did not receive any specific grant from funding agencies in the public, commercial, or not-for-profit sectors.

Declaration of Competing Interest

The authors declare that they have no known competing financial interests or personal relationships that could have appeared to influence the work reported in this paper.
  14 in total

1.  Analytical performances of a chemiluminescence immunoassay for SARS-CoV-2 IgM/IgG and antibody kinetics.

Authors:  Andrea Padoan; Chiara Cosma; Laura Sciacovelli; Diego Faggian; Mario Plebani
Journal:  Clin Chem Lab Med       Date:  2020-06-25       Impact factor: 3.694

2.  Correlation of Chest CT and RT-PCR Testing for Coronavirus Disease 2019 (COVID-19) in China: A Report of 1014 Cases.

Authors:  Tao Ai; Zhenlu Yang; Hongyan Hou; Chenao Zhan; Chong Chen; Wenzhi Lv; Qian Tao; Ziyong Sun; Liming Xia
Journal:  Radiology       Date:  2020-02-26       Impact factor: 11.105

3.  SARS-CoV-2 infection serology: a useful tool to overcome lockdown?

Authors:  Marzia Nuccetelli; Massimo Pieri; Sandro Grelli; Marco Ciotti; Roberto Miano; Massimo Andreoni; Sergio Bernardini
Journal:  Cell Death Discov       Date:  2020-05-26

4.  Clinical Performance of the Roche SARS-CoV-2 Serologic Assay.

Authors:  Mei San Tang; Karl G Hock; Nicole M Logsdon; Jennifer E Hayes; Ann M Gronowski; Neil W Anderson; Christopher W Farnsworth
Journal:  Clin Chem       Date:  2020-08-01       Impact factor: 8.327

5.  The outbreak of COVID-19: An overview.

Authors:  Yi-Chi Wu; Ching-Sung Chen; Yu-Jiun Chan
Journal:  J Chin Med Assoc       Date:  2020-03       Impact factor: 2.743

6.  Acute stress, behavioural symptoms and mood states among school-age children with attention-deficit/hyperactive disorder during the COVID-19 outbreak.

Authors:  Jinsong Zhang; Lan Shuai; Hui Yu; Zhouye Wang; Meihui Qiu; Lu Lu; Xuan Cao; Weipin Xia; Yuanyuan Wang; Runsen Chen
Journal:  Asian J Psychiatr       Date:  2020-04-09

7.  Detection of 2019 novel coronavirus (2019-nCoV) by real-time RT-PCR.

Authors:  Victor M Corman; Olfert Landt; Marco Kaiser; Richard Molenkamp; Adam Meijer; Daniel Kw Chu; Tobias Bleicker; Sebastian Brünink; Julia Schneider; Marie Luisa Schmidt; Daphne Gjc Mulders; Bart L Haagmans; Bas van der Veer; Sharon van den Brink; Lisa Wijsman; Gabriel Goderski; Jean-Louis Romette; Joanna Ellis; Maria Zambon; Malik Peiris; Herman Goossens; Chantal Reusken; Marion Pg Koopmans; Christian Drosten
Journal:  Euro Surveill       Date:  2020-01

8.  Genomic characterization of the 2019 novel human-pathogenic coronavirus isolated from a patient with atypical pneumonia after visiting Wuhan.

Authors:  Jasper Fuk-Woo Chan; Kin-Hang Kok; Zheng Zhu; Hin Chu; Kelvin Kai-Wang To; Shuofeng Yuan; Kwok-Yung Yuen
Journal:  Emerg Microbes Infect       Date:  2020-01-28       Impact factor: 7.163

9.  Antibody Responses to SARS-CoV-2 in Patients With Novel Coronavirus Disease 2019.

Authors:  Juanjuan Zhao; Quan Yuan; Haiyan Wang; Wei Liu; Xuejiao Liao; Yingying Su; Xin Wang; Jing Yuan; Tingdong Li; Jinxiu Li; Shen Qian; Congming Hong; Fuxiang Wang; Yingxia Liu; Zhaoqin Wang; Qing He; Zhiyong Li; Bin He; Tianying Zhang; Yang Fu; Shengxiang Ge; Lei Liu; Jun Zhang; Ningshao Xia; Zheng Zhang
Journal:  Clin Infect Dis       Date:  2020-11-19       Impact factor: 9.079

10.  Diagnostic performances and thresholds: The key to harmonization in serological SARS-CoV-2 assays?

Authors:  Mario Plebani; Andrea Padoan; Davide Negrini; Benedetta Carpinteri; Laura Sciacovelli
Journal:  Clin Chim Acta       Date:  2020-05-30       Impact factor: 3.786

View more
  7 in total

1.  Re-evaluating positive serum samples for SARS-CoV-2-specific IgA and IgG antibodies using an in-house serological assay.

Authors:  Margherita Cacaci; Giulia Menchinelli; Rosalba Ricci; Flavio De Maio; Melinda Mariotti; Riccardo Torelli; Grazia Angela Morandotti; Francesca Bugli; Maurizio Sanguinetti; Brunella Posteraro
Journal:  Clin Microbiol Infect       Date:  2021-01-02       Impact factor: 8.067

Review 2.  SARS-CoV-2 Infection and the COVID-19 Pandemic Emergency: The Importance of Diagnostic Methods.

Authors:  Marco Ciotti; Francesca Benedetti; Davide Zella; Silvia Angeletti; Massimo Ciccozzi; Sergio Bernardini
Journal:  Chemotherapy       Date:  2021-03-19       Impact factor: 2.544

3.  Clinical validation of a second generation anti-SARS-CoV-2 IgG and IgM automated chemiluminescent immunoassay.

Authors:  Massimo Pieri; Marzia Nuccetelli; Eleonora Nicolai; Serena Sarubbi; Sandro Grelli; Sergio Bernardini
Journal:  J Med Virol       Date:  2021-01-26       Impact factor: 2.327

Review 4.  Biology of severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) and the humoral immunoresponse: a systematic review of evidence to support global policy-level actions and research.

Authors:  Shalini Nair; Xinguang Chen
Journal:  Glob Health J       Date:  2021-11-19

5.  Detection of SARS-CoV-2 antibodies in febrile patients from an endemic region of dengue and chikungunya in Peru.

Authors:  Yordi Tarazona-Castro; Lucinda Troyes-Rivera; Johanna Martins-Luna; Felipe Cabellos-Altamirano; Miguel Angel Aguilar-Luis; Hugo Carrillo-Ng; Luis J Del Valle; Sungmin Kym; Sebastian Miranda-Maravi; Wilmer Silva-Caso; Saul Levy-Blitchtein; Juana Del Valle-Mendoza
Journal:  PLoS One       Date:  2022-04-08       Impact factor: 3.240

6.  Prevalence of SARS-CoV-2 infection and dynamics of antibodies response among previously undiagnosed healthcare workers in a university hospital: A prospective cohort study.

Authors:  Sherief Musa; Shereen Abdel Alem; Khaled Amer; Tarek Elnagdy; Wael A Hassan; Mohamed A Ali; Yasmine Gaber; Hedy A Badary; Omnia Tantawi; Reham Abdelmoniem; Amr Radwan; Hanaa Yousof; Shereen Shawky; Hala Talaat; Rabab Fouad; Abdel Meguid Kassem
Journal:  J Infect Public Health       Date:  2021-06-10       Impact factor: 3.718

7.  Evaluation of S-RBD and high specificity ACE-2-binding antibodies on SARS-CoV-2 patients after six months from infection.

Authors:  Flaminia Tomassetti; Marzia Nuccetelli; Serena Sarubbi; Francesca Gisone; Marco Ciotti; Francesco Spinazzola; Cristina Ricotta; Monica Cagnoli; Monica Borgatti; Marco Iannetta; Massimo Andreoni; Graziella Calugi; Massimo Pieri; Sergio Bernardini
Journal:  Int Immunopharmacol       Date:  2021-07-27       Impact factor: 4.932

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.