Gore-Tex implant extrusion following revision medialization thyroplasty: Case report and review of the literature.

Ariel Frost1, Natasha Mirza1.   

Abstract

OBJECTIVE: To review current literature and experience with Gore-TexⓇ implant extrusion following medialization thyroplasty, as well as to report the unique case of Gore-TexⓇ implant extrusion following revision medialization thyroplasty.
METHODS: Review of existing literature and description of personal experience with unique case of Gore-TexⓇ implant extrusion following revision medialization thyroplasty.
RESULTS: Review of existing literature found no prior reported cases of Gore-TexⓇ implant extrusion following revision medialization thyroplasty. Risk factors for implant extrusion include the pressure of the implant on insertion and the inability to secure the implant. Cases of implant extrusion can be managed operatively via an endoscopic or via an external open approach.
CONCLUSION: This is the first reported case of Gore-TexⓇ implant extrusion following revision medialization thyroplasty. Careful consideration should be given in revision medialization thyroplasty as additional implant material may cause increased pressure, a risk factor for implant extrusion.
© 2020 The Authors.

Entities:  

Keywords:  Complications; Gore-TexⓇ; Implant extrusion; Laryngoplasty; Medialization thyroplasty; Revision

Year:  2020        PMID: 32983585      PMCID: PMC7502009          DOI: 10.1016/j.wjorl.2020.08.001

Source DB:  PubMed          Journal:  World J Otorhinolaryngol Head Neck Surg        ISSN: 2095-8811


Introduction

Medialization thyroplasty (MT) is the most widely used laryngeal framework surgery technique available to treat unilateral vocal fold paralysis (UVFP), enabling improvement in voice and swallowing function, and preventing life-threatening aspiration events. While other surgical techniques such as injection laryngoplasty, arytenoid adduction (AA), and laryngeal innervation are used to treat UVFP, MT has the unique advantage of intra-operative manipulation to optimize vocal quality. , A variety of laryngeal implant materials are available for MT, including Gore-TexⓇ, SilasticⓇ, and hydroxyapatite. A 2010 survey on laryngeal framework surgery by Young et al revealed a downward trend in the use of SilasticⓇ (76% previously, 24% at the time of survey), and an upward trend in the use of Gore-TexⓇ (36% previously, 64% at the time of survey). However, all types of laryngeal implants carry the risk of extrusion, with an overall rate of 0.8%. , Here, we describe a case of delayed implant extrusion following a revision Gore-TexⓇ MT. While implant extrusion following MT is a recognized complication, this case uniquely occurred in both a delayed fashion and following revision Gore-TexⓇ MT (5 years following the patient's original Gore-TexⓇ MT, and 4 years following the patient's revision Gore-TexⓇ MT). This is the first case in the literature of MT implant extrusion following a revision MT.

Case

The patient is a 59-year-old female with history of non-small cell lung cancer, status post left pneumonectomy in 2013, complicated by left vocal cord paralysis, for which she underwent left Gore-TexⓇ MT in 2014 and revision Gore-TexⓇ MT in 2015. She presented to the outpatient clinic with 1 week of sudden onset throat pain and hoarseness. Nasopharyngolaryngoscopy (NPL) revealed Gore-TexⓇ implant extrusion laryngeally at the level of the false vocal cord, partially obstructing the airway. Imaging was deferred and the patient proceeded emergently to the operating room for direct laryngoscopy and endoscopic implant removal. In the operating room, the extruding Gore-TexⓇ implant was removed (Fig. 1 , Fig. 2 ), revealing the second Gore-TexⓇ implant, which was also extruding and therefore also removed (Fig. 3 , Fig. 4 ). Following removal of the Gore-TexⓇ implants, granulation tissue was seen in the left ventricle, which was excised and injected with Decadron (Fig. 5 , Fig. 6 ). The patient tolerated the procedure well without complication and was discharged home the same day.
Fig. 1

Laryngeal Gore-TexⓇ implant extrusion.

Fig. 2

Endoscopic removal of laryngeal Gore-TexⓇ implant extrusion.

Fig. 3

2nd laryngeal Gore-TexⓇ implant extrusion.

Fig. 4

Endoscopic removal of 2nd laryngeal Gore-TexⓇ implant extrusion.

Fig. 5

Granulation tissue in the left ventricle.

Fig. 6

Endoscopic excision of granulation tissue in the left ventricle.

Laryngeal Gore-TexⓇ implant extrusion. Endoscopic removal of laryngeal Gore-TexⓇ implant extrusion. 2nd laryngeal Gore-TexⓇ implant extrusion. Endoscopic removal of 2nd laryngeal Gore-TexⓇ implant extrusion. Granulation tissue in the left ventricle. Endoscopic excision of granulation tissue in the left ventricle. At follow-up the patient was breathing comfortably without any respiratory complaints. Three months after her Gore-TexⓇ implant removal, she underwent left vocal cord injection laryngoplasty with calcium hydroxylapatite (CaHA), with subsequent vocal improvement and closure of glottic gap on NPL. Her most recent follow-up has been delayed due to the coronavirus 2019 (COVID-19) pandemic.

Discussion

Laryngeal framework surgery, including MT and AA, are common treatments for UVFP with glottic incompetence. In the 2010 survey on laryngeal framework surgery by Young et al, the overall complication rate of MT was 15%, airway compromise requiring intervention was observed in 2.2%, suboptimal voice outcome was observed in 4% (3% unchanged, 1% worse voice), and the overall revision rate was 6%. The most common revision surgery was replacement with another implant. Implant extrusion is a recognized postoperative complication (0.8%) and can occur medially into the airway (0.5%) or laterally into the paraglottic space (0.3%). Since the introduction of Gore-TexⓇ in the 1990s, documentation of Gore-TexⓇ implant extrusion has been limited to case reports and case series. A total of 13 cases of Gore-TexⓇ implant extrusion have been described in the literature so far, plus our own case (Table 1 ).6, 7, 8, 9, 10, 11, 12, 13, 14, 15, 16, 17 Implant extrusion occurred anywhere between 2 months and 10 years postoperatively, and more frequently occurred medially into the airway, with patients presenting with cough, dysphonia, and/or globus sensation.6, 7, 8, 9, 10, 11, 12
Table 1

Case summary of Gore-TexⓇ implant extrusion.

Pt No.AuthorYearSexSurgeryInner perichondriumOnsetTreatment
1Zeitels2003NRMTPreservedNRRemoved
2Laccourreye2003FMTRemoved49 monthsRemoved
3Cohen2004NRMTIncisedNRRemoved
4Cohen2004NRMTIncisedNRRemoved
5Halum2005FMTNANRRemoved
6Halum2005NRMTNANRRemoved
7Feinberg2010FMTNA6 yearsRemoved
8Sims2014FMTNA5 monthsRemoved
9Krane2016FMTPreserved5 yearsRemoved
10Morris2016FMTNA10 yearsRemoved
11Hardman2016NRMTNANRRemoved
12Watanabe2017FMT/AAPreserved2 monthsRemoved
13Kirke2018FMTIncised6 yearsPartial removal
14Frosta2020FRevision MTNA4 yearsRemoved

NR: not reported; MT: medialization thyroplasty; AA: arytenoid adduction.

Presented case.

Case summary of Gore-TexⓇ implant extrusion. NR: not reported; MT: medialization thyroplasty; AA: arytenoid adduction. Presented case. Review of the literature highlighted several factors that may cause a Gore-TexⓇ implant to extrude. Surgical factors include improper implant placement (removal/violation of the inner perichondrium, pressure of the implant during insertion) and suboptimal implant stabilization (inability to lock the implant to the thyroid cartilage lamina window). , One case reported an unusually large amount of Gore-TexⓇ needed to achieve medialization, which is another potential risk factor for implant extrusion. Postoperative risk factors include infection and risks associated with subsequent surgeries (size of endotracheal tube (ETT), traumatic intubation, thoracic surgery with double-lumen ETT, and length of surgery), which may weaken the tissue around the implant. Finally, amongst patient factors, female gender has been noted to be a risk factor due to females having a relatively smaller larynx. , The case presented here has several unique aspects. This is the only published case of Gore-TexⓇ implant extrusion following revision MT. 14 months after the patient's initial Gore-TexⓇ MT, her voice continued to be weak and breathy and she was found to have increased left vocal cord atrophy with a persistent glottic gap, for which she underwent revision MT with additional Gore-TexⓇ implant. The additional implant material may have exerted increased pressure on the laryngeal tissue medial to the implant. In addition, the inevitable vocal cord atrophy in the setting of UVFP, as well as the patient's chronic steroid use given her transplant history, may have weakened the laryngeal tissue medial to the implant. As previously mentioned, female gender has been noted to also be a risk factor. In summary, this was a female patient who underwent revision MT with placement of additional Gore-TexⓇ, which likely caused increased pressure on the laryngeal surface, which itself weakened in the setting of expected progressive vocal cord atrophy and chronic steroid use. In this case, after suboptimal vocal quality with the initial Gore-TexⓇ implant, it was believed that a 2nd Gore-TexⓇ implant would improve her vocal quality, however perhaps the initial Gore-TexⓇ implant should have been removed completely and replaced with a different implant material or an autogenous implant. Consistent with the published literature, at our institution MT implant extrusion is managed in the operating room endoscopically or via an external surgical approach. The endoscopic approach is preferred for medial implant extrusion into the airway. The external surgical approach is indicated for lateral implant extrusion. One advantage of the external approach is the ability to perform immediate reimplantation. However previous studies have suggested that reimplantation is often unnecessary following implant extrusion, as many patients retain adequate glottic closure secondary to scar tissue build up in the paraglottic space. , , More recently, ready-made MT implant systems such as the MontgomeryⓇ Thyroplasty Implant System (Boston Medical Products, MA, USA), which is made of customizable SilasticⓇ, and VoCoM® (GYRUS ACMI, TN, USA), which is made of hydroxyapatite, have been developed to decrease the time involved in implant customization and to reduce complications. In 2018 Chao et al published on a novel approach to MT with autogenous tissue using tensor fascia lata to improve outcomes and decrease complications such as extrusion or wound infection, particularly in patients that have received prior radiotherapy to the head and neck. Whether these new implant systems improve vocal outcomes and decrease complications long term, such as extrusion, remains to be determined.

Conclusions

We describe a unique case of Gore-TexⓇ implant extrusion following revision MT, in which additional Gore-TexⓇ was added in the revision surgery. The combination of the patient's second Gore-TexⓇ implant causing increased pressure on the laryngeal surface, which itself weakened in the setting of expected progressive vocal cord atrophy and chronic steroid use, may have contributed to medial implant extrusion. Implant extrusion is a recognized postoperative complication that can be managed endoscopically or via an external open approach based on medial or lateral extrusion. During MT careful attention is given to implant placement and stabilization.

Declaration of competing interest

None.
  18 in total

1.  Analysis of laryngeal framework surgery: 10-year follow-up to a national survey.

Authors:  VyVy N Young; Thomas G Zullo; Clark A Rosen
Journal:  Laryngoscope       Date:  2010-08       Impact factor: 3.325

2.  Hypopharyngeal extrusion of 2.5 feet (76 cm) of polytetrafluoroethylene (Gore-Tex): initial laser-assisted office-based removal and micropharyngeal completion.

Authors:  Steven Feinberg; Gerardo Lopez-Guerra; Steven M Zeitels
Journal:  Ann Otol Rhinol Laryngol       Date:  2010-09       Impact factor: 1.547

3.  Complications of phonosurgery: results of a national survey.

Authors:  C A Rosen
Journal:  Laryngoscope       Date:  1998-11       Impact factor: 3.325

4.  Interesting case of late Gore-Tex extrusion following medialization laryngoplasty.

Authors:  Diana N Kirke; Andrew Blitzer
Journal:  Laryngoscope       Date:  2018-11-08       Impact factor: 3.325

5.  Displacement of Residual Gore-Tex Thyroplasty Implant Presenting as a True Vocal Fold Mass.

Authors:  John R Sims; Ian J Lalich; Dale C Ekbom
Journal:  Otolaryngol Head Neck Surg       Date:  2014-01-31       Impact factor: 3.497

6.  Case of migrated Gore-Tex implant following external thyroplasty.

Authors:  Joel Hardman; Malcolm A Buchanan; Faruque Riffat
Journal:  BMJ Case Rep       Date:  2016-06-08

7.  Gore-Tex extrusion following type I thyroplasty.

Authors:  Natalie A Krane; Amanda Hu; Robert T Sataloff
Journal:  Ear Nose Throat J       Date:  2016 Apr-May       Impact factor: 1.697

8.  Medialization laryngoplasty with Gore-Tex for voice restoration secondary to glottal incompetence: indications and observations.

Authors:  Steven M Zeitels; Marcelo Mauri; Seth H Dailey
Journal:  Ann Otol Rhinol Laryngol       Date:  2003-02       Impact factor: 1.547

9.  Endolaryngeal extrusion of expanded polytetrafluoroethylene implant after medialization thyroplasty.

Authors:  Ollivier Laccourreye; Stéphane Hans
Journal:  Ann Otol Rhinol Laryngol       Date:  2003-11       Impact factor: 1.547

Review 10.  Evaluation and treatment of the unilateral paralyzed vocal fold.

Authors:  M S Benninger; R L Crumley; C N Ford; W J Gould; D G Hanson; R H Ossoff; R T Sataloff
Journal:  Otolaryngol Head Neck Surg       Date:  1994-10       Impact factor: 3.497

View more
  1 in total

Review 1.  Gore-Tex implant extrusion following revision medialization thyroplasty: Case report and review of the literature.

Authors:  Ariel Frost; Natasha Mirza
Journal:  World J Otorhinolaryngol Head Neck Surg       Date:  2020-09-19
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.