Comparing oscillometric noninvasive and invasive intra-arterial blood pressure monitoring in term neonates under general anesthesia: A retrospective study.

Tasuku Fujii1, Kimitoshi Nishiwaki2.   

Abstract

BACKGROUND: Oscillometric noninvasive blood pressure and/or invasive intra-arterial blood pressure are commonly used to measure the systolic, diastolic, and mean components of blood pressure. Agreement between the two methods has been reported in adults, children, and infants, but rarely in neonates, especially under general anesthesia. AIMS: This retrospective study compared the agreement of each measured blood pressure value (oscillometric noninvasive or invasive intra-arterial blood pressure monitoring) in term neonates under general anesthesia.
METHODS: Data were collected from neonates born at ≥36 weeks of gestation whose body weight was ≥2500 g and who underwent abdominal or noncardiac thoracic surgery with both oscillometric noninvasive and invasive intra-arterial blood pressure measurements from January 2015 to March 2020. The primary outcome was the agreement of systolic, diastolic, and mean blood pressure values between the two methods using Bland-Altman analysis.
RESULTS: Paired blood pressure measurements (n = 1193) from 67 cases were compared. In Bland-Altman analysis, bias (standard deviation), 95% limits of agreement, and percentage error were -9.3 (8.4), -26.1-7.6, and 26.9% for systolic; 1.6 (6.5), -11.3-14.6, and 38.7% for diastolic; and -1.3 (5.8), -13.0-10.3, and 26.9% for mean blood pressure, respectively. During low blood pressure (intra-arterial mean blood pressure ≤30 mm Hg), the biases (standard deviation) of systolic, diastolic, and mean blood pressure were -11.4 (5.7), -0.7 (3.7), and -5.1 (4.2), whereas during high blood pressure (intra-arterial mean blood pressure ≥60 mm Hg), the values were 0.1 (9.7), 5.6 (9.4), and 6.4 (7.4), respectively.
CONCLUSIONS: Based on the bias and percentage error, the mean blood pressure exhibited the most acceptable agreement between oscillometric noninvasive and invasive intra-arterial blood pressure monitoring in term neonates under general anesthesia. However, during hypertension or hypotension, there was a large discrepancy between the two methods.
© 2020 John Wiley & Sons Ltd.

Entities:  

Keywords:  arterial blood pressure; hemodynamics; intraoperative care; neonate; oscillometric noninvasive blood pressure

Mesh:

Year:  2020        PMID: 32959496     DOI: 10.1111/pan.14020

Source DB:  PubMed          Journal:  Paediatr Anaesth        ISSN: 1155-5645            Impact factor:   2.556


  1 in total

1.  Blood pressure monitoring following kidney transplantation in children: a comparison of invasive and noninvasive measurements using Doppler as a benchmark technique.

Authors:  Eytan Kaplan; Gili Kadmon; Elhanan Nahum; Hadas Alfandary; Orly Haskin; Avichai Weissbach
Journal:  Pediatr Nephrol       Date:  2022-08-01       Impact factor: 3.651

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.