Metal artifacts of hip arthroplasty implants at 1.5-T and 3.0-T: a closer look into the B1 effects.

Iman Khodarahmi1, John Kirsch2, Gregory Chang1, Jan Fritz3.   

Abstract

OBJECTIVE: To evaluate the effect of circular polarization (CP) and elliptical polarization (EP) of the B1 field on metal implant-induced artifacts of titanium (Ti) and cobalt-chromium (CoCr) hip arthroplasty implants at 1.5-T and 3.0-T field strengths.
MATERIAL AND METHODS: In vitro Ti and CoCr total hip arthroplasty implants were evaluated using high transmit and receive bandwidth turbo spin echo (HBW-TSE) and slice encoding for metal artifact correction (SEMAC) metal artifact reduction techniques. Each technique was implemented at 1.5-T, which only allows for CP of B1 field as the system default, as well as 3.0-T, which permitted CP and EP. Manual segmentation quantified the size of the metal artifacts at the level of the acetabular cup, femoral neck, and femoral shaft.
RESULTS: In the acetabular cup and femoral neck, 1.5-T CP achieved smaller artifact sizes than 3.0-T CP (28-29% on HBW-TSE, p = 0.002-0.005; 17-34% on SEMAC, p = 0.019-0.102) and 3.0-T EP (25-28% on HBW-TSE, p = 0.010-0.011; 14-36% on SEMAC, p = 0.058-0.135) techniques. In the femoral stem region, 3.0-T EP achieved more efficient artifact suppression than 3.0-T CP (HBW-TSE 44-45%, p < 0.001-0.022; SEMAC 76-104%, p < 0.001-0.022) and 1.5-T CP (HBW-TSE 76-96%, p < 0.001-0.003; SEMAC 138-173%, p = 0.003-0.005) techniques.
CONCLUSION: Despite slightly superior metal reduction ability of the 1.5-T in the region of the acetabular cup and prosthesis neck, 3.0-T MRI of hip arthroplasty implants using elliptically polarized RF pulses may overall be more effective in reducing metal artifacts than the current standard 1.5-T MRI techniques, which by default implements circularly polarized RF pulses.

Entities:  

Keywords:  3.0-T; Hip arthroplasty; MRI; Metal artifact; Polarization; Radiofrequency pulse

Mesh:

Substances:

Year:  2020        PMID: 32918566     DOI: 10.1007/s00256-020-03597-4

Source DB:  PubMed          Journal:  Skeletal Radiol        ISSN: 0364-2348            Impact factor:   2.199


  2 in total

Review 1.  MRI-guided sacroiliac joint injections in children and adults: current practice and future developments.

Authors:  Danoob Dalili; Amanda Isaac; Jan Fritz
Journal:  Skeletal Radiol       Date:  2022-08-25       Impact factor: 2.128

2.  New-Generation Low-Field Magnetic Resonance Imaging of Hip Arthroplasty Implants Using Slice Encoding for Metal Artifact Correction: First In Vitro Experience at 0.55 T and Comparison With 1.5 T.

Authors:  Iman Khodarahmi; Inge M Brinkmann; Dana J Lin; Mary Bruno; Patricia M Johnson; Florian Knoll; Mahesh B Keerthivasan; Hersh Chandarana; Jan Fritz
Journal:  Invest Radiol       Date:  2022-03-04       Impact factor: 10.065

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.