| Literature DB >> 32909039 |
Nada A Ahmed1,2, E H Basheir1, A B Farah1, T S Mohammedzein1,3, I I Suliman1,4.
Abstract
This study aimed to calculate patient radiation doses for adults during the seven most commonly performed conventional X-ray procedures, and to propose national diagnostic reference levels (DRLs). A representative sample of patients from 29 hospitals was included. The entrance surface air kerma (ESAK) was calculated by measuring X-ray tube output and the corresponding technical and exposure factors for each patient. Third-quartile values of the mean ESAK distributions were proposed as DRL values. The DRLs in mGy were as follows: 0.6 for chest postero-anterior (PA), 3.5 for skull AP, 1.7 for skull lateral (LAT), 2.7 for abdominal, 2.6 for pelvic AP, 3.7 for lumbar spine AP and 8 for lumbar spine LAT. Compared with literature, the maximum percentages increase were in chest PA (329%) and skull AP (187%). Since the suggested DRL for chest PA was higher than literature values, dose optimization and a review of its value is recommended.Entities:
Mesh:
Year: 2020 PMID: 32909039 DOI: 10.1093/rpd/ncaa114