| Literature DB >> 32906745 |
Yong Woo Song1, Min Young Kim2, Min Hee Chung2, Young Kwon Yang2, Jin Chul Park2.
Abstract
In South Korea, the gradual increase in particulate matter generation has received significant attention from central and local governments. Exhaust gas, which contains nitrogen oxides (NOx), is one of the main sources of particulate matter. In this study, the reduction of NOx using a coating material mixed with a titanium dioxide (TiO2) photocatalyst was demonstrated. The NOx reduction performance of the TiO2 photocatalyst-infused coating was evaluated by applying the ISO 22197-1: 2007 standard. Subsequently, the performance was evaluated by changing the NO gas concentration and ultraviolet (UV)-A irradiance under standard experimental conditions. It was determined that NOx reduction can be achieved even if the NO gas concentration and UV-A irradiance are lower than those under the standard conditions when the TiO2 photocatalyst-infused coating was used. This study revealed that NOx reduction can be realized through TiO2 photocatalyst-infused coating in winter or cloudy days with a low solar altitude. It was also confirmed that compared with the UV-A irradiance, the NO gas concentration has a greater effect on the NOx reduction performance of the TiO2 photocatalyst-infused coating. These findings can be used to evaluate a variety of construction materials with TiO2 photocatalysts in the future.Entities:
Keywords: nitrogen oxides (NOx); particulate matter; reduction test; secondary source; titanium dioxide (TiO2)
Mesh:
Substances:
Year: 2020 PMID: 32906745 PMCID: PMC7571102 DOI: 10.3390/molecules25184087
Source DB: PubMed Journal: Molecules ISSN: 1420-3049 Impact factor: 4.411
TiO2 coating composition.
| Contained Chemicals | Proportion |
|---|---|
| Titanium dioxide (anatase) | 1.75% |
| Silicone compound | 5.6% |
| Ethanol | 41.6% |
| Water | 51.0% |
| Other | 0.05% |
Field emission scanning electronic microscope (FESEM) and energy dispersive spectroscopy (EDS) measuring equipment.
| Classification | Contents |
|---|---|
| Model | SIGMA 500 (Carl Ziess) |
| Detector | SE 2 |
| EDS detector | X-MaxN50 (Oxford) |
| Acceleration voltage | 18.0 kV |
| Working distance | 8.5 mm |
| Magnification | 500× to 1000× |
| Time-resolution | 0.8 nm |
STUB (Sample Holder) composition.
| Element | wt% |
|---|---|
| C | 12.84 |
| O | 3.04 |
| Al | 79.34 |
| Cu | 4.79 |
| Total | 100.00 |
Figure 1STUB (Sample Holder) field emission scanning electronic microscope (FESEM) image.
TiO2 coating composition.
| Element | wt% |
|---|---|
| C | 3.47 |
| O | 65.09 |
| Al | 10.39 |
| Ti | 21.05 |
| Total | 100.00 |
Figure 2TiO2 coating FESEM and energy dispersive spectroscopy (EDS) mapping image.
Figure 3Test chamber diagram.
Specifications of measuring equipment.
|
|
| |||
|
|
|
|
| |
| Temperature | −20 to +80 °C | ±3% of the leading value ± 0.25 °C | 0.1 °C | |
| Relative humidity | 0–100% RH | Accuracy: ±1.8% RH Calibration | 0.1% RH | |
| Velocity | 0.00–5.00 m/s | ±3% of the leading value ± 0.05 m/s | 0.01 m/s | |
|
|
| |||
| Range | Automatic | 0–20 ppm |
| |
| USEPA approval | 0.0–0.5 ppm | |||
| TUVEN certified | Less than NO (0–1000 ppm), | |||
| Accuracy/precision | Precision | 0.4 ppb or 0.5% of reading (the lesser of the two) | ||
| Linearity | ±1% of the total scale | |||
| Reaction time | 90% in 15 s | |||
| Sample flow rate | 0.3 slpm (total flow rate of 0.6 slpm for the NO and NOx flow path) | |||
ISO standard condition test values.
| Classification | Value |
|---|---|
| UV-A irradiance | 10 W/m2 |
| NO gas concentration | 1.00 ppm |
| Temperature | 25 ± 2.5 ℃ |
| Relative humidity | 50% |
| Test time/measurement interval | 3 h/1 min |
| Number of experiments | 3 times |
Changes in condition test value.
| Classification | UV-A Irradiance (W/m2) | |||||||
|---|---|---|---|---|---|---|---|---|
| 2.50 | 5.00 | 7.50 | 10.0 | |||||
|
| CASE 1 | 0.25 | CASE 5 | 0.25 | CASE 9 | 0.25 | CASE 13 | 0.25 |
| CASE 2 | 0.50 | CASE 6 | 0.50 | CASE 10 | 0.50 | CASE 14 | 0.50 | |
| CASE 3 | 0.75 | CASE 7 | 0.75 | CASE 11 | 0.75 | CASE 15 | 0.75 | |
| CASE 4 | 1.00 | CASE 8 | 1.00 | CASE 12 | 1.00 | CASE 16 | 1.00 | |
ISO standard condition test result.
| Classification | Start Concentration | End Concentration | Reduction Rate (Reduction Amount) |
|---|---|---|---|
|
| 1.015 ppm | 0.653 ppm | 35.67% (12.06 µmol/10 cm2·3 h) |
|
| 0.000 ppm | 0.143 ppm | 3.11 µmol/10 cm2·3 h (generated) |
|
| 1.015 ppm | 0.796 ppm | 21.58% (8.95 µmol/10 cm2·3 h) |
Figure 4NOx reduction results for the ISO standard condition.
Test for changes in the condition that resulted in NO reduction.
|
|
| |||
|
| 0.25 ppm | 0.50 ppm | 0.75 ppm | |
| 2.50 W/m2 | 0.119 ppm | 0.179 ppm | 0.189 ppm | |
| 5.00 W/m2 | 0.108 ppm | 0.136 ppm | 0.234 ppm | |
| 7.50 W/m2 | 0.113 ppm | 0.201 ppm | 0.327 ppm | |
NO2 generation results based on NO concentration and UV-A irradiance.
|
|
| |||
|
| 0.25 ppm | 0.50 ppm | 0.75 ppm | |
| 2.50 W/m2 | 0.053 ppm | 0.105 ppm | 0.091 ppm | |
| 5.00 W/m2 | 0.028 ppm | 0.060 ppm | 0.088 ppm | |
| 7.50 W/m2 | 0.024 ppm | 0.076 ppm | 0.094 ppm | |
Test results of changes in condition to achieve NOx reduction.
|
|
| |||
|
| 0.25 ppm | 0.50 ppm | 0.75 ppm | |
| 2.50 W/m2 | 0.067 ppm | 0.074 ppm | 0.098 ppm | |
| 5.00 W/m2 | 0.080 ppm | 0.076 ppm | 0.121 ppm | |
| 7.50 W/m2 | 0.089 ppm | 0.126 ppm | 0.202 ppm | |
Figure 5Test results of changes in condition to achieve NOx reduction. NOx reduction according to change in the irradiance based on a concentration of (a) 0.25 ppm, (b) 0.50 ppm, and (c) 0.75 ppm. NOx reduction according to the change in the reference concentration at a light rate of (d) 2.50 W/cm2, (e) 5.00 W/cm2, and (f) 7.50 W/cm2.
NOx reduction amount.
|
| 0.25 | 0.50 | 0.75 | 1.00 | |
|
|
| ||||
| 2.50 |
|
| 4.32 | 5.41 | |
| 5.00 | 2.99 | 3.23 | 5.34 | 5.45 | |
| 7.50 | 3.24 | 5.05 | 8.18 | 7.99 | |
| 10.0 | 4.45 | 6.08 | 8.07 | 8.95 | |
Figure 6Results of NOx reduction rate.
Figure 7Total NOx reduction test results.