| Literature DB >> 32900908 |
Christian A Taschner1, Christian Paul Stracke2, Franziska Dorn3, Krzysztof Bartosz Kadziolka4, Kornelia Kreiser5, László Solymosi6, Mirko Pham6, Jan Hendrik Buhk7, Bernd Turowski8, Wolfgang Reith9, Samer Elsheikh10, Stephan Meckel10, Hendrik Janssen11,12, Alexander Hammer11, Oliver Beuing13, Olav Jansen14, Horst Urbach10, Michael Knauth15, Carolin Jenkner16, René Chapot2.
Abstract
BACKGROUND: Flow diverters (FD) are used regularly for the endovascular treatment of unruptured intracranial aneurysms. We aimed to assess the safety and effectiveness of the Derivo embolization device (DED) with respect to long-term clinical and angiographic outcomes.Entities:
Keywords: aneurysm; flow diverter
Mesh:
Year: 2020 PMID: 32900908 PMCID: PMC8142444 DOI: 10.1136/neurintsurg-2020-016303
Source DB: PubMed Journal: J Neurointerv Surg ISSN: 1759-8478 Impact factor: 5.836
Patient data and characteristics of aneurysms treated with the Derivo embolization device
|
| 96 |
| Aneurysms | 96 |
| Women (%) | 71/96 (74%) |
| Age (years, mean±SD) | 54±12.0 |
|
| |
| Asymptomatic | 61 (64%) |
| SAH from treated aneurysm >60 days | 1 (1%) |
| Recurrent aneurysm after coiling, SAC, or failed clipping | 14 (15%) |
|
| |
| mRS 0 | 72 (75%) |
| mRS 1 | 24 (25%) |
|
| |
| Mean±SD | 14.2±16.9 |
| <5 mm | 14 (15%) |
| 5–9.9 mm | 36 (37%) |
| 10–20 mm | 33 (34%) |
| >20 mm | 13 (14%) |
|
| |
| Mean±SD | 7.7±9.6 |
| Aneurysm partially thrombosed | |
| yes | 18 (19%) |
|
| |
| ICA | 82 (86%) |
| ACA | 1 (1%) |
| MCA | 1 (1%) |
| VA | 8 (8%) |
| BA | 4 (4%) |
|
| |
| Wide-neck saccular | 77 (80%) |
| Fusiform/dissecting | 19 (20%) |
ACA, anterior cerebral artery; BA, basilar artery; ICA, internal carotid artery; MCA, middle cerebral artery; mRS, modified Rankin Scale score; SAC, stent-assisted coiling; SAH, subarachnoid hemorrhage; SD, SD deviation; VA, vertebral artery.
Treatment characteristics
| Dual antiplatelet therapy | 96 (100%) |
| Number of DED implanted, mean (range) | 1.2 (1–3) |
| Additional coiling | 47 (49%) |
| PTA after DED placement | 18 (19%) |
| Procedural time, mean±SD (range) | 89±43 min (20-235) |
| Entire neck covered * | 94 (98%) |
| Covered side branches * | 89 (93%) |
| Complete wall apposition * | 80 (84%) |
Procedural time=groin puncture to final DSA; SD=SD deviation.
*Core-laboratory adjudicated data.
DED, Derivo embolization device; PTA, percutaneous transluminal angioplasty.
Periprocedural complications
| Flow diverter could not be delivered through microcatheter | 2 (2%) |
| Flow diverter did not open | 3 (3%) |
| Flow diverter twisted | 1 (1%) |
| Flow diverter displaced(proximally, distally) | 6 (6%) [4, 2] |
| Fish mouthing of the DED *(proximal, distal) | 11 (11.5%) [9, 2] |
| ICA dissection | 1 (1%) |
| Parent vessel occlusion(thrombus, DED related) | 4 (4%) [1, 3] |
| Multiple embolic infarcts | 1 (1%) |
| Symptomatic intracranial hemorrhage | 1 (1%) |
| Femoral artery pseudoaneurysm, groin hematoma, retroperitoneal hematoma | 4 (4%) |
*Eight cases signaled by core laboratory, three cases reported by enrolling center; ICA=internal carotid artery.
DED, Derivo embolization device.
Figure 1Scores on the modified Rankin Scale (n=96) distribution of scores at baseline, discharge, and last available follow-up (14.8±5.2 months) in the trial population.