Literature DB >> 32890572

Performance evaluation of automated BD Phoenix NMIC-500 panel for carbapenemase detection in carbapenem-resistant and carbapenem-susceptible Enterobacterales.

Hanwool Cho1, Jung Ok Kim1, Ji Eun Choi1, Howon Lee1, Woong Heo1, Young Jong Cha1, In Young Yoo1, Yeon-Joon Park2.   

Abstract

Rapid detection of carbapenemases and accurate reporting of carbapenem MICs is critical for appropriate treatment and infection control. We evaluated the BD Phoenix NMIC-500 panel for detection and classification of carbapenemases and antimicrobial susceptibility testing (AST) for carbapenems. A total of 235 isolates were tested; 47 carbapenemase-producing Enterobacterales, 52 non-carbapenemase-producing carbapenem-resistant Enterobacterales (non-CP-CRE), 136 carbapenem-susceptible Enterobacterales (CSE). The sensitivity of carbapenemase-producing organism (CPO) detection was 97.9%, the specificity was 100% for CSE but 32.7% for non-CP-CREs. All the 35 false-positive cases were non-CP-CREs; 23 out of the 35 were determined as untyped carbapenemase producer (CP), nine were mistyped as class B, and three were as class A. The detection rate/correct classification rate for class A, B, and D carbapenemase was 100%/78.6%, 100%/100%, and 80%/60%, respectively. To supplement the low specificity, it is suggested to report carbapenemase-producer (CP) positive results as "strongly suspicious for carbapenem resistance but carbapenemase production needs to be confirmed" and perform the confirmatory test. The EA and CA for ertapenem, imipenem, and meropenem was 99.1%/99.6%, 89.4%/90.6%, and 95.3%/95.7%. In conclusion, the BD Phoenix CPO detect panel provides advantage in that the carbapenemase test is automated and the results can be obtained within 6 h but the low specificity in CREs needs to be improved. In addition, accurate reporting of meropenem MICs will be helpful for clinicians to choose treatment options.
Copyright © 2020 Elsevier B.V. All rights reserved.

Entities:  

Keywords:  Antimicrobial susceptibility testing; CPO; Carbapenem-resistant Enterobacterales; Carbapenemase-producing organisms; Phoenix NMIC-500

Mesh:

Substances:

Year:  2020        PMID: 32890572     DOI: 10.1016/j.mimet.2020.106042

Source DB:  PubMed          Journal:  J Microbiol Methods        ISSN: 0167-7012            Impact factor:   2.363


  2 in total

1.  Comparative Evaluation of Two Automated ID/AST Systems and Mikrolatest Kit in Assessing the In Vitro Colistin Susceptibility of Carbapenem-Resistant Enterobacteriaceae Isolates: A Single-Center Exploratory Study from North India.

Authors:  Mohit Bhatia; Ravi Indrapal Singh; Diksha Rani; U Sasi Rekha; Ranjana Rohilla; Balram Ji Omar; Pratima Gupta
Journal:  J Pharm Bioallied Sci       Date:  2022-05-19

2.  Evaluation of the BD Phoenix CPO detect panel for prediction of Ambler class carbapenemases.

Authors:  Daniel Jonas; Sandra Reuter; Sarah Klassen; Sabine Weber; Marion Buck; Tommaso Giani; Gian Maria Rossolini; Hajo Grundmann
Journal:  Sci Rep       Date:  2021-06-23       Impact factor: 4.379