Ankita Reddy1,2,3, Irene Bosch1,3,4, Nol Salcedo1, Bobby Brooke Herrera1,5, Helena de Puig2,6, Carlos F Narváez7, Diana María Caicedo-Borrero8, Ivette Lorenzana9, Leda Parham9, Kimberly García9, Marcela Mercado10, Angélica María Rico Turca11, Luis A Villar-Centeno12, Margarita Gélvez-Ramírez12, Natalia Andrea Gómez Ríos12, Megan Hiley2, Dawlyn García2, Michael S Diamond13, Lee Gehrke1,2,14.
Abstract
Since its 2013 emergence in the Americas, Chikungunya virus (CHIKV) has posed a serious threat to public health. Early and accurate diagnosis of the disease, though currently lacking in clinics, is integral to enable timely care and epidemiological response. We developed a dual detection system: a CHIKV antigen E1/E2-based enzyme-linked immunosorbent assay (ELISA) and a lateral flow test using high-affinity anti-CHIKV antibodies. The ELISA was validated with 100 PCR-tested acute Chikungunya fever samples from Honduras. The assay had an overall sensitivity and specificity of 51% and 96.67%, respectively, with accuracy reaching 95.45% sensitivity and 92.03% specificity at a cycle threshold (Ct) cutoff of 22. As the Ct value decreased from 35 to 22, the ELISA sensitivity increased. We then developed and validated two lateral flow tests using independent antibody pairs. The sensitivity and specificity reached 100% for both lateral flow tests using 39 samples from Colombia and Honduras at Ct cutoffs of 20 and 27, respectively. For both lateral flow tests, sensitivity decreased as the Ct increased after 27. Because CHIKV E1/E2 are exposed in the virion surfaces in serum during the acute infection phase, these sensitive and specific assays demonstrate opportunities for early detection of this emerging human pathogen.Entities:
Keywords: Chikungunya fever; E1/E2 antigen detection; ELISA; Latin America; acute phase diagnosis; alphavirus; lateral flow; rapid diagnosis
Mesh:
Substances:
Year: 2020 PMID: 32882998 PMCID: PMC7552019 DOI: 10.3390/v12090971
Source DB: PubMed Journal: Viruses ISSN: 1999-4915 Impact factor: 5.048
Performance of CHIKV E1/E2 ELISA Test. Area under curve (AUC) of ROC curve, 95% confidence interval (95% CI), optimal OD450 cutoff, sensitivity (%), specificity (%), and sample counts were calculated for incremental PCR Ct cutoffs for the E1/E2 ELISA test using antibody Combination A (48, 155).
| CHIKV Combination A ELISA Receiver Operator Characteristic (ROC) Analysis ( | ||||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Ct Cutoff | 20 | 21 | 22 | 23 | 24 | 25 | 26 | 27 | 28 | 29 | 30 | 31 | 32 | 33 | 34 | 35 |
| AUC | 0.86 | 0.91 | 0.94 | 0.92 | 0.89 | 0.86 | 0.83 | 0.81 | 0.79 | 0.77 | 0.72 | 0.71 | 0.70 | 0.65 | 0.61 | 0.61 |
| 95% CI | 0.74–0.97 | 0.84–0.99 | 0.89–1.00 | 0.86–0.98 | 0.80–0.97 | 0.77–0.95 | 0.74–0.93 | 0.72–0.91 | 0.70–0.88 | 0.68–0.86 | 0.63–0.82 | 0.62–0.81 | 0.61–0.79 | 0.55–0.74 | 0.52–0.71 | 0.52–0.70 |
| OD450 Cutoff | 0.22 | 0.22 | 0.22 | 0.22 | 0.22 | 0.22 | 0.22 | 0.22 | 0.22 | 0.22 | 0.22 | 0.22 | 0.22 | 0.22 | 0.22 | 0.22 |
| Sensitivity (%) | 90.00 | 94.12 | 95.45 | 87.5 | 81.48 | 79.31 | 72.73 | 69.44 | 62.79 | 58.00 | 53.57 | 52.54 | 50.82 | 45.59 | 42.47 | 41.33 |
| Specificity (%) | 84.67 | 88.81 | 92.03 | 87.5 | 92.48 | 93.13 | 93.7 | 94.35 | 95.73 | 96.36 | 97.12 | 98.02 | 97.98 | 97.83 | 97.7 | 97.65 |
| N Total Positive | 150 | 143 | 138 | 136 | 133 | 131 | 127 | 124 | 117 | 110 | 104 | 101 | 99 | 92 | 87 | 85 |
| N Total Negative | 10 | 17 | 22 | 24 | 27 | 29 | 33 | 36 | 43 | 50 | 56 | 59 | 61 | 68 | 73 | 75 |
Figure 1Chikungunya virus (CHIKV) E1/E2-based enzyme-linked immunosorbent assay (E1/E2 ELISA) and Lateral Flow Test compared to qPCR. The OD450 or signal intensity of samples tested through ELISA (A) and lateral flow (B–D), respectively, are shown in relation to the PCR Ct values. The tests used either antibody Combination A (48 and 155) or Combination B (4 and 340). The p values are <0.0001 (A), <0.0001 (B), <0.0002 (C), <0.0182 (D).
Figure 2Performance of CHIKV E1/E2 ELISA Test. Receiver operating characteristic (ROC) curve analysis of PCR-tested patient samples from Honduras. Incremental Ct values are used to characterize positive and negative samples, to which ELISA performance is compared. Test performance is demonstrated in terms of true positive rate (%) versus false positive rate (%).
Figure 3Performance of CHIKV Lateral Flow Test. Receiver operating characteristic (ROC) curve analysis of PCR-tested patient samples. The lateral flow test was constructed with Combination A antibodies using 19 samples from Honduras (A), Combination B antibodies using 19 samples from Honduras (B), and Combination B antibodies using 10 samples from Colombia (C). Incremental Ct values are used to characterize positive and negative samples, to which lateral flow test performance is compared. Test performance is demonstrated in terms of true positive rate (%) versus false positive rate (%).
Performance of CHIKV E1/E2 Lateral Flow Test. Area under curve (AUC) of ROC curve, 95% confidence interval (95% CI), signal intensity cutoff, sensitivity (%), specificity (%), and sample counts were calculated for incremental PCR Ct cutoffs for the E1/E2 lateral flow test using antibody Combination A (48 and 155) in Honduras (A) and antibody Combination B (4 and 340) in Honduras (B) and Colombia (C).
| A. | |||||
|---|---|---|---|---|---|
| CHIKV Combination A Lateral Flow Receiver Operator Characteristic (ROC) Analysis-Honduras ( | |||||
| Ct Cutoff | 20 | 21–24 | 25–26 | 27 | 28 |
| AUC | 0.96 | 0.95 | 0.94 | 1.00 | 0.81 |
| 95% CI | 0.88–1.00 | 0.85–1.00 | 0.83–1.00 | 1.00–1.00 | 0.61–1.000 |
| Lateral Flow Signal Intensity Cutoff | 0.82 | 0.75 | 0.67 | 0.57 | 0.57 |
| Sensitivity (%) | 100 | 100 | 100 | 100 | 62.50 |
| Specificity (%) | 92.31 | 91.67 | 90.91 | 100 | 100 |
| N Total Positive | 13 | 12 | 11 | 9 | 3 |
| N Total Negative | 6 | 7 | 8 | 10 | 16 |
| B. | |||||
| CHIKV Combination B Lateral Flow Receiver Operator Characteristic (ROC) Analysis-Honduras ( | |||||
| Ct Cutoff | 20 | 21–26 | 27 | 28 | |
| AUC | 0.94 | 0.93 | 1.00 | 0.81 | |
| 95% CI | 0.82–1.00 | 0.82–1.00 | 1.00–1.00 | 0.61–1.00 | |
| Lateral Flow Signal Intensity Cutoff | 0.72 | 0.53 | 0.53 | 0.53 | |
| Sensitivity (%) | 83.33 | 100 | 100 | 62.50 | |
| Specificity (%) | 92.31 | 75.00 | 100 | 100 | |
| N Total Positive | 13 | 12 | 9 | 3 | |
| N Total Negative | 6 | 7 | 10 | 16 | |
| C. | |||||
| CHIKV Combination B Lateral Flow Receiver Operator Characteristic (ROC) Analysis-Colombia ( | |||||
| Ct Cutoff | 18–19 | 20 | 21 | ||
| AUC | 0.86 | 1.00 | 0.78 | ||
| 95% CI | 0.60–1.00 | 1.00–1.00 | 0.51–1.00 | ||
| Lateral Flow Signal Intensity Cutoff | 0.63 | 0.59 | 0.56 | ||
| Sensitivity (%) | 100 | 100 | 77.78 | ||
| Specificity (%) | 85.71 | 100 | 100 | ||
| N Total Positive | 7 | 4 | 1 | ||
| N Total Negative | 3 | 6 | 9 | ||