To optimize sensitivity, there has been an increasing interest in the miniaturization of NMR detectors. In our lab, a stripline NMR detector has been developed, which provides high resolution and is scalable to a large range of sample volumes. These features make it an ideal detector for hyphenated techniques. In this manuscript, we demonstrate a stripline probe, which is designed for combining supercritical fluid chromatography (SFC) experiments with NMR. It features a novel stripline chip, designed to reduce the signal from the contact pads, which results in an improved lineshape. An external lock circuit provides stability over time to perform signal averaging or multidimensional experiments. As proof of concept, we demonstrate the SFC-NMR technique with this stripline probe using a mixture of cholesterol and cholestanol, which is relevant for studying cerebrotendinous xanthomatosis. Additionally, this probe makes it possible to record high-resolution spectra of samples with a high spin density. This means that it is possible to directly observe shifts due to the nuclear demagnetizing field in the "homomolecular" case, which is challenging using conventional probes due to broadening effects from radiation damping.
To optimize sensitivity, there has been an increasing interest in the miniaturization of NMR detectors. In our lab, a stripline NMR detector has been developed, which provides high resolution and is scalable to a large range of sample volumes. These features make it an ideal detector for hyphenated techniques. In this manuscript, we demonstrate a stripline probe, which is designed for combining supercritical fluid chromatography (SFC) experiments with NMR. It features a novel stripline chip, designed to reduce the signal from the contact pads, which results in an improved lineshape. An external lock circuit provides stability over time to perform signal averaging or multidimensional experiments. As proof of concept, we demonstrate the SFC-NMR technique with this stripline probe using a mixture of cholesterol and cholestanol, which is relevant for studying cerebrotendinous xanthomatosis. Additionally, this probe makes it possible to record high-resolution spectra of samples with a high spin density. This means that it is possible to directly observe shifts due to the nuclear demagnetizing field in the "homomolecular" case, which is challenging using conventional probes due to broadening effects from radiation damping.
In recent years, there
has been an increasing interest in miniaturizing
detectors for nuclear magnetic resonance (NMR) spectroscopy.[1] This is because smaller coils can provide a higher
sensitivity per spin for mass-limited samples. For example, for a
flat circular coil, the sensitivity per sample volume is inversely
proportional to its diameter, assuming the resistance of the coil
does not change. Decreasing the coil diameter by a factor two doubles
the signal-to-noise ratio of the NMR spectrum. Although the sensitivity
increases, the number of spins that will fit inside the detected volume
decreases. The size of the detector should therefore be matched to
the available sample volume for optimal sensitivity. For mass- or
volume-limited samples, such as body fluids, it is worthwhile to use
a miniaturized detector. The sensitivity of the probe not only depends
on its size but also on the coil design. Most of the research has
been focused on developing planar helical coils and tightly wound
solenoid microcoils.[2−5] An alternative design is the planar microslot waveguide probe.[6]However, with these types of microcoils,
the resolution is often
significantly reduced due to the close proximity of the microcoil
structure to the sample. In our lab, a stripline detector has been
developed,[7] which solves this issue by
using a planar structure oriented along the main magnetic field. The
stripline NMR detector can record spectra with high sensitivity of
samples contained within fused-silica capillaries. Also, the stripline
can easily be designed for a wide range of sample volumes. This makes
it straightforward to hyphenate this type of NMR detector to other
techniques such as chromatography and capillary electrophoresis.[8] An interesting development is the hyphenation
of supercritical fluid chromatography (SFC) with stripline NMR.[9,10]In this manuscript, we demonstrate a novel NMR probe design,
which
is optimized for on-line SFC-NMR experiments. It features an improved
stripline design, which provides a further increase in resolution,
and an external lock circuit to provide stability over time for longer
experiments. An interesting application of SFC-NMR is the separation
and identification of lipids. In the work by Oostendorp et al.,[11] the analysis and quantification of lipids in
body fluids are performed using 1H NMR without separation
in a regular NMR setup. However, when too many lipids are present
or when some are present at high concentration, identifying and quantifying
the lipids can be difficult in these crowded NMR spectra. SFC-1H NMR could resolve this problem since the lipids are separated
before NMR analysis, thus solving the issue of a spectral overlap.
We demonstrate the use of this novel probe design in a SFC-NMR setup
for the separation and structural analysis of a mixture of cholesterol
and cholestanol. This can be relevant for analyzing the blood plasma
of patients suffering from cerebrotendinous xanthomatosis (CTX), which
shows an accumulation of cholestanol due to a defect in a gene coding
for the enzyme sterol 27-hydroxylase.
Experimental Section
Split-Contact
Stripline
Although the stripline detector
has a good homogeneity compared to other microcoil designs, the resonances
often exhibited a broadening at the base. This is caused by the signal
originating from the sample located on the contact pads (the top and
bottom parts of the chip shown in Figure a). The magnetic field inhomogeneity in these
areas is caused by the electrical contacts and the casing, which holds
the stripline. This means that the magnetic field at the center of
the stripline and the contact pads cannot be simultaneously homogenized
using shims. To reduce this effect and thus improve the resolution,
we have developed a new stripline design where the part of the contact
pad directly below the capillary is disconnected from the rest of
the contact pad. This reduces the current density near the capillary
and thus decreases the RF field strength and sensitivity for the sample
located on the contact pads. The design is called the split-contact
stripline and is shown in Figure b. To study the behavior of the split-contact stripline,
we have used a stripline chip, which is 25 mm long and 10 mm wide.
The sensitive region of the strip has a width of 500 μm, separated
by gaps of 250 μm, and the tapered regions have 45° angles.
Figure 1
(a) Pictures
of the regular stripline and (b) the new split-contact
design. Both designs use ground planes (not shown) to confine the
RF field. More information on the stripline design is provided in
the Supplementary Information.
(a) Pictures
of the regular stripline and (b) the new split-contact
design. Both designs use ground planes (not shown) to confine the
RF field. More information on the stripline design is provided in
the Supplementary Information.The expected difference in B1-field
strengths between the regular and split-contact stripline was simulated
(COMSOL Multiphysics). The details of this simulation are provided
in the Supplementary Information. In Figure , the simulated B1-field strengths (at 600 MHz) are shown as
a function of the position along the stripline z.
At each position along z, the distribution in B1-fields is determined in a cylindrical volume
at the position of the sample. In blue, the B1-profile of a regular stripline is shown and in red the B1-profile of the split-contact stripline. At
the narrow region of the stripline, there is a distribution in B1-field strengths over the capillary. In this
region, both profiles are nearly identical; this means that the split
contacts hardly influence the main sensitive region of the stripline.
On the other hand, outside of the narrow sensitive region, the B1-field drops more rapidly in the split-contact
case and is approximately 4 times lower compared to the regular stripline.
This greatly reduces the excitation and detection of signals originating
from sample located outside the narrow region of the stripline.
Figure 2
Simulation
of the distribution in B1-field strengths
in the sample as a function of position along the
stripline for a regular stripline (blue contours) and a split-contact
stripline (red contours). The black line shows the full width of the
strip.
Simulation
of the distribution in B1-field strengths
in the sample as a function of position along the
stripline for a regular stripline (blue contours) and a split-contact
stripline (red contours). The black line shows the full width of the
strip.
External Lock
To compensate for drifting of the magnet,
most commercial NMR probes contain a deuterium lock channel. In these
cases, the deuterium resonance of the deuterated solvent is used to
correct for variations in the magnetic field. However, in SFC-NMR,
the samples are not always mixed with deuterated solvent. This means
that internal locking can only be used in a limit number of cases.
For this reason, we have implemented an external lock into the probe
design. The external lock circuit consists of a coil wound around
a glass tube filled with deuterated solvent, as can be seen in Figure S9, which is part of a resonant circuit
that matches the deuterium frequency (92 MHz). In this case, deuterated
acetone was used as a lock sample, as its resonance is nearly temperature
independent. The lock circuit in this probe design is completely independent
of the NMR circuits.
SFC-NMR
The SFC-NMR setup used for
this research has
been introduced in a previous publication.[9] It consists of a Waters Acquity UPC[2] instrument,
which is coupled to a Varian VNMRS spectrometer operating at 600 MHz
Larmor frequency (14.1T). Selection and concentration of the fraction
of interest for NMR analysis proceed as follows; During chromatography,
when a peak of interest passes the UV detector, a valve is switched,
letting the sample including mobile phase into a purged 100 μL
sample loop. When the valve switches back, the sample expands to near
atmospheric pressure through a fixed restriction in the flowline.
The pressure drop causes the CO2 to evaporate and the compound
of interest to stay behind on the walls of the tube. This selection
procedure can be performed multiple times to collect the sample from
multiple chromatographic runs. After selecting the peak, a water flow
is switched on, which moves through the loop toward the NMR probe
in a PEEK tubing with an inner diameter of 0.254 mm at a flow rate
of 0.1 mL min–1. After the tubing is completely
filled with water, a plug of strong solvent, in this case, 15 μL
of CDCl3 (Cambridge Isotope Laboratories, 99.8 deuterated
+0.05 v/v TMS), is introduced into the water flow to elute the sample.
This plug flows toward the middle of the NMR stripline detector at
which point the flow is stopped. Stopping the flow enables the acquisition
of multiple scans to improve the signal-to-noise ratio. After the
experiment, the valve returns to its original position and the sample
flows back from the NMR probe to the waste. More details on the chromatography
used in the SFC-NMR experiments are provided in the Supplementary Information.For the NMR experiments,
a home-built probe with the split-contact stripline as described above
was used. The stripline chip used for SFC-NMR is of a larger design
(dimensions are provided in the Supplementary Information) to accommodate a capillary with an outer diameter
of 700 μm and an inner diameter of 500 μm, which means
it has an active detection volume of ∼1.6 μL. The SFC-NMR
spectra and reference cholesterol spectrum were acquired with a relaxation
delay of 10 s and an acquisition time of 1 s. For the SFC-NMR spectra,
5250 scans were recorded (16 h and 2 min) and 4500 scans for the reference
cholesterol spectrum (13 h and 48 min, 25.9 mM). All spectra were
recorded with a receiver bandwidth of 8 kHz. The spectra are referenced
to TMS. The spectra were recorded in blocks of five scans. Before
the summation of the spectra, the maxima of the TMS peaks was aligned
to compensate for any remaining drift in the spectra over time, and
the spectra were baseline corrected. All spectra were processed using
ssNake.[12]
Results
The split-contact stripline
design was tested by comparing the RF-field strengths with a regular
stripline design. For both designs, a chip was fabricated with a constriction
length of 6.0 mm and a width of 500 μm. Using a capillary (OD
350 μm, ID 250 μm) with a 1 mm plug of water in FC-40,
the power levels where calibrated such that the constricted part of
the stripline chip produced an RF field of approximately 250 kHz.
Afterward, the water sample was repositioned to the contact pad, and
a nutation experiment was performed to determine the B1-field produced at the contact pad. The positions of
the sample are shown in Figure . The repositioning of the capillary did not influence the
tuning and matching of the probe. The results of these experiments
are shown in Figure . These B1-fields correspond well to
those predicted by the simulations. The regular stripline produced
a B1-field of approximately 21.5 kHz (simulated:
23.0 kHz) at the contact pad, whereas the split-contact stripline
produced a B1-field of only 6.0 kHz (simulated:
6.1 kHz). As can be seen from the figures, the sensitivity at the
contact pads is also significantly lower for the split-contact design,
as is expected from the reciprocity principle.[13] The reduced B1-field strength
also means that when a pulse with a flip angle of 90° is given
at the center of the stripline; the spins on the split-contacts experience
a pulse with a flip-angle of 1.7° (compared to a flip-angle of
7.8° for a conventional stripline).
Figure 3
Positions of a sample
plug (water in FC-40). The position in blue
is used to calibrate the RF field. The position, shown in red, is
used to determine the RF-field at the contact-pads.
Figure 4
(a) Simulation and (b) experiment of the nutation frequency of
a 1 mm plug of water in FC-40. The plug is positioned at the contact
pads of a regular stripline (blue) and a split-contact stripline (orange).
Positions of a sample
plug (water in FC-40). The position in blue
is used to calibrate the RF field. The position, shown in red, is
used to determine the RF-field at the contact-pads.(a) Simulation and (b) experiment of the nutation frequency of
a 1 mm plug of water in FC-40. The plug is positioned at the contact
pads of a regular stripline (blue) and a split-contact stripline (orange).The reduced sensitivity of the contacts of the
stripline results
in an improvement of the resolution. This can be seen in Figure where the CH3 resonance of toluene is compared between a spectrum obtained
using regular and split-contact striplines. The FWHM improves from
5.5 Hz to about 3 Hz; however, the biggest improvement can be seen
in the base of the peak where most of the signal originates from the
contact pads. The split-contact spectrum is still slightly asymmetric,
which is most likely caused by local field disturbances, which are
hard to correct using a regular shim system. An example of a source
of these local field disturbances is the magnetic impurities in the
“seed layer” used in the manufacturing of the stripline
chip. One way to compensate for these local field disturbances would
be to implement a Shim-on-Chip system.[14]
Figure 5
CH3 resonance of toluene recorded using a regular (blue)
and a split-contact stripline.
CH3 resonance of toluene recorded using a regular (blue)
and a split-contact stripline.To demonstrate the stability of the external
lock with the stripline detector, we have recorded a series of single-pulse
NMR spectra over time. In Figure S10, the
drift of the CH3 resonance of methanol is shown for a period
up to 35 h. What can be observed is that without a lock, the magnetic
field takes nearly 10 h to stabilize completely after the probe has
been inserted into the magnet. Furthermore, there are sharp features
visible at around 5, 15, and 30 h after the experiment was started.
These are caused by the usage of the Bitter magnets in the High Field
Magnet Laboratory (HFML) building about 100 m away from the NMR lab.
To correct for these effects and for the drifting of the magnet, an
external lock was implemented in the stripline probe. In this case,
the external lock circuit contains a sample of deuterated acetone.
Spectra of a sample of acetone were recorded for nearly 30 h. With
the external lock, the initial drift after inserting the probe is
corrected, as well as the influences of the magnets of the HFML. Some
small variations in the resonance frequency can still be seen, with
a maximum deviation of about 1 Hz. These variations are most likely
caused by slight differences between the sample and lock sample.
Demagnetizing Field
An interesting effect that can
be directly observed with this stripline probe is the shift of a resonance
due to the nuclear demagnetizing field (long-range dipole–dipole
interactions). The nonlinear effects of the nuclear demagnetizing
field have been known for a long time, in particular, in the context
of paramagnetic impurities,[15,16] and the multiple echoes
caused by the demagnetizing field have been observed experimentally
in different materials.[17]In this
experiment, we placed a capillary with an inner diameter of 50 μm
filled with H2O in the stripline probe. The narrow diameter
of the capillary minimizes radiation damping and lowers the B1-gradient over the sample. Simple one-pulse
NMR experiments were performed on this sample with varying pulse widths.
The results are presented in Figure , which shows that there is a difference in the resonance
frequency after a small flip angle compared to a near 180° pulse.
This difference in frequency is about 2.3 Hz (3.8 ppb).
Figure 6
Spectra of
H2O as a function of flip angle showing the
shifting resonance. The black ellipse shows the path that the peak
maximum/minimum makes. The width of the ellipse is 2.3 Hz.
Spectra of
H2O as a function of flip angle showing the
shifting resonance. The black ellipse shows the path that the peak
maximum/minimum makes. The width of the ellipse is 2.3 Hz.Due to the small diameter of the sample (i.e., 50 μm),
the B1 homogeneity is improved, which
means that
the resonance shift can be monitored for a number of cycles, as demonstrated
in Figure a. In this
figure, the black line approximately follows the resonance frequency.
This line has an oscillation frequency (87 kHz) and a decay rate (Gaussian)
identical to that of the signal amplitude (the orange line of Figure b) and only has a
90° phase difference. The resonance frequency also slowly shifts
to lower frequency during the experiment, this is most likely caused
by a slow increase in the sample temperature.
Figure 7
(a) Shift and (b) amplitude
of the proton resonance as a function
of RF-pulse width. The black and orange lines have a frequency of
87 kHz (90° out of phase) and a Gaussian decay. The black line
has an amplitude of 1.2 Hz
(a) Shift and (b) amplitude
of the proton resonance as a function
of RF-pulse width. The black and orange lines have a frequency of
87 kHz (90° out of phase) and a Gaussian decay. The black line
has an amplitude of 1.2 HzThe effect of this demagnetizing field can be observed in this
stripline setup because the sample is locked using an external lock
circuit. In regular liquid-state NMR, it is common to lock on the
deuterated solvent in the sample. This means that the demagnetizing
field shift is not directly observed as shown here since the lock
signal will be affected by the demagnetizing field also. It can also
result in anomalous crosspeaks, for example, in COSY experiments.[18]It is interesting to note that the magnitude
(in Hz) of the shift
in Figure cannot
be explained by the classical picture of bulk magnetization created
by the sum of the individual proton spins. The shift between 0°
and 180° in the classical picture is predicted to be 1.6 Hz,[15,18] whereas we observe 2.3 Hz. In the quantum mechanical framework,
the shift is described by long-range dipole–dipole interactions,
which are not averaged by molecular motion.[19] In this case, the predicted shift is times larger for “like”
spins
or so-called “homomolecular” spins.[20] This is because the B-term in the dipolar Hamiltonian contributes
to the total shift for these types of spins.Direct detection
of this kind of effect can only be done conveniently
in a microliter setting. Recording spectra of samples with high spin
densities in regular liquid NMR probes is usually complicated by other
nonlinear effects, such as radiation damping. With the stripline probe,
it is straightforward to reduce these effects so that the pulse width
effect on the resonance position can be observed with high accuracy.The stripline described above was optimized
to analyze microliter (μL) sample volumes, thus matching the
volumes coming from chromatographic techniques such as SFC. Therefore,
we implemented the detector in our hyphenated SFC-NMR system. The
design specifications of the stripline chip are provided in the Supporting Information. Due to high solubility
of lipids in supercritical CO2, which has similar solvent
strength as hexane, SFC is used more often in the separation of different
classes of lipids in the field of lipidomics.[21] Here, we look at the SFC separation of cholesterol, one of the most
known lipids, and cholestanol, which has a similar structure to cholesterol,
differing in only one double bond (Figure ). For the separation of these compounds,
a 1 μL mixture of 28 mM cholestanol and 43 mM cholesterol in
chloroform was injected into the SFC. The optimal separation conditions
were obtained with a Torus Diol column from Waters, using a gradient
of 2–5% methanol in CO2 over 3 min, the result of
which can be found in the Supporting information. Although the two molecules are very similar in structure, they
can still be separated well under these conditions by SFC.
Figure 8
SFC-1H NMR spectra of the first chromatographic peak
(green) and the second chromatographic peak (orange) compared to reference
cholestanol 1H NMR spectrum in CDCl3. The molecular
structures of cholesterol and cholestanol are shown on the left.
SFC-1H NMR spectra of the first chromatographic peak
(green) and the second chromatographic peak (orange) compared to reference
cholestanol1H NMR spectrum in CDCl3. The molecular
structures of cholesterol and cholestanol are shown on the left.To test if the hyphenation of SFC-NMR could be
a feasible analytic
technique in the field of lipidomics, the split-contact stripline
detector, optimized for μL volumes, was coupled in-line to the
SFC apparatus. After SFC separation, both lipids were each collected
six times in separate runs, before eluting them to the middle of the
stripline NMR detector, in a plug of 15 μL of CDCl3. There, the flow was stopped, allowing for the acquisition of multiple
scans, to obtain a higher signal-to-noise ratio. To obtain a reference
spectrum of cholesterol, instead of selecting the sample from the
SFC, a 15 μL plug of 26 mM cholesterol in CDCl3 was
injected directly into the flow of water toward the NMR spectrometer.
The resulting spectra are shown in Figure . By comparing the SFC-1H NMR
spectra with reference spectra, it is clear that the first chromatographic
peak corresponds to cholestanol and the second peak corresponds to
cholesterol. By using the split-contact stripline detector, the resolution
and peak shape is improved (compared to using a regular stripline)
and peak splittings due to J-couplings can be observed, e.g., around
3.55 ppm. The peak at 5.3 ppm, corresponding to the proton on the
ring in cholesterol, which is absent in the cholestanol spectrum,
confirms the selection of the correct molecules. Also, some shifts
are observed for other peaks, reflecting the differences in the molecular
structure. A mixture of very similar molecules can thus be separated
and characterized by SFC-NMR. If the spectra are examined closely,
a small impurity of cholesterol can however be observed in the cholestanol
spectrum, probably due to a slight overlap of the peaks during chromatography.
This does however not influence the structural characterization of
the molecule in the spectrum since the concentration of the impurity
is very low.The SFC-NMR spectra shown here were acquired overnight
over 16
h, which is a long time for an analysis. The signal-to-noise ratio
is, however, much higher than required for the identification of these
compounds. A spectrum with a signal-to-noise ratio of 3 for the peak
around 5.3 ppm would require an experimental time of only 1 min. This
is also due to the advantage of using the stripline probe with a larger
volume. In our previous work,[9] a stripline
probe with a detection volume of only 150 nL was used. The current
stripline has a volume of ∼1.6 μL. The volume that is
selected from the SFC is 100 μL for each selected peak. To get
a higher sample concentration in the NMR detector, however, the mobile
phase is evaporated out and the sample is redissolved in 15 μL
of solvent, in this case, deuterated chloroform. The detection volume
of 1.6 μL therefore matches the sample volume much better than
in our previous probe in which only a small part of the sample was
used for detection. The current probe volume could be increased even
more, to better match the sample volume from the SFC. However, some
mixing of the sample plug with the transportation medium, water, occurs
at the edges of the plug. To prevent the water from interfering with
the spectra, it is best not to use the full sample plug for detection
but only the middle part. This μL volume stripline probe is
therefore the best match in this hyphenation of SFC with in-line NMR
detection. Although the absolute sensitivity of the stripline decreases
by scaling up the design, the increased number of spins in the active
detection volume leads to an overall gain in signal intensity. As
an estimation, to obtain a signal-to-noise ratio of 3 for the cholesterol
spectrum with the smaller volume stripline probe, 15 min of the experimental
time is required, instead of 1 min. It is therefore beneficial to
match the stripline probe volume to the sample volume obtained from
the SFC.The concentrations of cholestanol and cholesterol used
in this
SFC-NMR analysis are 28 and 43 mM, respectively, which are high compared
to the concentrations of 0.2 and 3 mM that are present in the blood
plasma of patients suffering from CTX.[11] If a signal-to-noise of 3 would be taken as the minimum for identification
of cholestanol/cholesterol, the concentrations of cholestanol and
cholesterol can go down to 0.8 and 1.2 mM, respectively. If enough
sample is available, more injections from the SFC can also be an option
to get a higher concentration for the NMR analysis. In this experiment,
the same peak was selected six times, in six separate SFC runs, after
which it was eluted in one plug of chloroform to the NMR detector.
This was done to obtain a higher concentration of the sample for detection.
For the six injections that were done in this analysis, only 12 μL
of the sample was needed. To get even higher concentrations of the
sample into the detector, while keeping the same initial sample concentration
going into the SFC, more injections can be done. For example, by selecting
the peak 12 times from the SFC, a gain of a factor 2 in concentration
is achieved. This either enables a shorter analysis time in NMR, gaining
a factor 4 in time, or this allows going to lower initial sample concentrations
of 0.4 and 0.6 mM for cholestanol and cholesterol, respectively. This
then approaches the concentrations that are present in patients suffering
from CTX, making SFC-NMR a feasible technique for detecting this biomarker.
Conclusions
In this manuscript, we presented a stripline
probe optimized for
(in-line) detection of μL sample volumes and showed a proof-of-concept
application of SFC-NMR. The stripline has an active volume of approximately
1.6 μL. A novel design minimizes the signal from the contact
pads, which results in an improved lineshape compared to the traditional
stripline. An external lock circuit provides long term stability for
multidimensional experiments and signal averaging. This high resolution
and stability make it possible to record spectra of small sample volumes
with high spin densities. This way, it was possible to directly observe
the shift due to the distant dipolar field. To our knowledge, a direct
observation of this effect for “like” spins has not
been demonstrated previously under thermal conditions due to resolution
limitations.This type of stripline probe is well suited for
hyphenated chromatography-NMR
systems. The stripline described above was optimized to analyze μL
sample volumes, thus matching the volumes coming from chromatographic
techniques such as SFC. We therefore used the optimized stripline
detector in our hyphenated SFC-NMR system.An SFC-NMR setup
has been demonstrated to separate a mixture of
cholesterol and cholestanol with subsequent identification by NMR.
These lipids are present in high concentrations in blood plasma of
patients suffering from cerebrotendinous xanthomatosis (CTX). Both
compounds can be isolated using this setup and analyzed with good
resolution. In this proof-of-principle experiment, concentrations
of cholesterol and cholestanol of 43 and 28 mM, respectively, were
used for the analysis. However, with simple adaptations, like selecting
the sample more often, this setup allows for the detection and identification
of lipids in sub mM range, thus approaching the concentrations of
3 mM cholesterol and 0.2 mM cholestanol that are present in patients
suffering from CTX. Although experiments still need to be performed
on real patient samples, this is a proof-of-principle that the SFC-NMR
setup with the optimized stripline probe can be a feasible analysis
technique for full structural characterization of lipids in body fluids.
Authors: A P M Kentgens; J Bart; P J M van Bentum; A Brinkmann; E R H van Eck; J G E Gardeniers; J W G Janssen; P Knijn; S Vasa; M H W Verkuijlen Journal: J Chem Phys Date: 2008-02-07 Impact factor: 3.488