F H M van Zelst1,2, S G J van Meerten1,2, P J M van Bentum1, A P M Kentgens1. 1. Institute for Molecules and Materials (IMM) , Radboud University , Nijmegen 6525 AJ , The Netherlands. 2. TA-COAST , Science Park 904 , Amsterdam 1098 XH , The Netherlands.
Abstract
By coupling supercritical fluid chromatography (SFC) and nuclear magnetic resonance (NMR) in-line, a powerful analytical method arises that enables chemically specific analysis of a broad range of complex mixtures. However, during chromatography, the compounds are diluted in the mobile phase, in this case supercritical CO2 (scCO2), often resulting in concentrations that are too low to be detected by NMR spectroscopy or at least requiring excessive signal averaging. We present a hyphenated SFC-NMR setup with an integrated approach for concentrating samples in-line, which are diluted in scCO2 during chromatography. This in-line concentration is achieved by controlled in-line expansion of the scCO2. As a proof of concept four isomers of vitamin E (tocopherol) were isolated by SFC, concentrated in-line by expanding CO2 from 120 to 50 bar, and finally shuttled to the NMR spectrometer fitted with a dedicated probehead for spectroscopic characterization of microfluidic samples. The abundant isomers were readily detected, supporting the viability of SFC-NMR as a powerful analytical tool.
By coupling supercritical fluid chromatography (SFC) and nuclear magnetic resonance (NMR) in-line, a powerful analytical method arises that enables chemically specific analysis of a broad range of complex mixtures. However, during chromatography, the compounds are diluted in the mobile phase, in this case supercritical CO2 (scCO2), often resulting in concentrations that are too low to be detected by NMR spectroscopy or at least requiring excessive signal averaging. We present a hyphenated SFC-NMR setup with an integrated approach for concentrating samples in-line, which are diluted in scCO2 during chromatography. This in-line concentration is achieved by controlled in-line expansion of the scCO2. As a proof of concept four isomers of vitamin E (tocopherol) were isolated by SFC, concentrated in-line by expanding CO2 from 120 to 50 bar, and finally shuttled to the NMR spectrometer fitted with a dedicated probehead for spectroscopic characterization of microfluidic samples. The abundant isomers were readily detected, supporting the viability of SFC-NMR as a powerful analytical tool.
Analyzing
complex mixtures,
which are often only available in small amounts, is one of the challenges
in analytical chemistry. To analyze these mixtures, it is necessary
to separate the molecular components in the mixture prior to spectroscopic
characterization. Common hyphenated chromatographic techniques couple
the separation by liquid chromatography (LC) or gas chromatography
(GC) to detection by a photodiode array (PDA) detector or mass spectrometer
(MS).[1] Although PDA or MS detection is
often used, the structural information that can be gained from these
detection techniques is limited compared to nuclear magnetic resonance
spectroscopy (NMR).[2] The main advantages
of using NMR as a detector is that it is quantitative and provides
chemical resolution. It is therefore the preferred technique for obtaining
detailed molecular structures without prior knowledge of the molecule.
Many examples of separation techniques hyphenated with NMR have appeared
over the years,[3,4] such as capillary HPLC-NMR[5,6] having the advantage of low solvent volumes, GC-NMR[7] for volatile compounds, and many examples of (HP)LC-NMR,[8] often with parallel MS detection.[9]As an alternative to LC or GC, supercritical fluid
chromatography
(SFC) can be used for chromatographic separation. In SFC, supercritical
carbon dioxide (scCO2) is used as the mobile phase. Due
to the higher diffusivity and lower viscosity of supercritical fluids,
ten times higher flow rates are achieved in SFC than in normal phase
LC. Another advantage of using SFC instead of LC when hyphenated with
NMR is the mobile phase. In LC-NMR solvent suppression techniques
are necessary to reduce signal coming from the mobile phase. In SFC-NMR
however, the mobile phase, scCO2, does not give a 1H NMR signal, so “cleaner” spectra are obtained.
Furthermore, scCO2 is a “green” and low-cost
solvent, unlike the organic solvents normally employed in LC. By adding
small amounts of other solvents to the scCO2, known as
modifiers or cosolvents, the polarity of the mobile phase can be optimized
so that a broad range of compounds can be separated. A modifier gradient
in CO2 is often used to speed up the chromatography, eluting
the more polar compounds with an increasingly polar mobile phase.[4] The downside of using these modifiers is that
their signals will show up in the NMR spectra of the sample of interest.
The amount of modifier is however much smaller (typically a few percent)
than in LC. Their signal can be suppressed by switching to deuterated
solvents after chromatography, which will be described in more detail
later in this paper. If SFC and NMR can be coupled, then a powerful
tool for analyzing a broad range of complex mixtures becomes available.Although SFC-NMR has been reported earlier as an analysis technique[10−12] the limitations in the sensitivity of NMR make the technique less
popular, especially for nanoscale sample volumes. To compensate for
the low sensitivity, large amounts of sample are needed to obtain
a good signal. However, this leads to overloading of the chromatographic
column, thus reducing the chromatographic resolution.[13] To solve this issue, the sensitivity of NMR for small sample
volumes must be improved and the sample should be concentrated before
analysis in the NMR spectrometer.Several innovations have been
made to enhance the sensitivity in
NMR for small sample volumes. By miniaturizing the detection coil,
smaller sample volumes can be measured. Research has been focused
on developing tightly wound solenoid coils and planar helical microcoils.[14−17] An alternative approach was the development of planar microslot
waveguide probes.[18] The challenge in miniaturization
is to get the highest sensitivity while maintaining an uncompromised
high resolution.[19] Over the past few years
our group has introduced on-chip stripline detectors for high-resolution
NMR of mass-limited samples.[20] The conventional
coil geometry used for detection has been replaced by flat stripline
RF inductors. The flat geometry of the stripline can be used for in-flow
detection, by passing a capillary over the stripline. Stripline probes
can achieve a high resolution and a single scan sensitivity of 0.1
nanomole 1H spins per square-root Hz receiver bandwidth
and of a few nanomoles for 13C for sample volumes from
100 nL to 1 mL.[20,21]To improve sensitivity
further, the sample can be concentrated
before analysis. In the work by Tayler et al. this was done by evaporating
the sample off-line before detecting it with NMR.[22] The authors first separated a mixture of four isomers of
vitamin E (tocopherol) by SFC. Each isomer (Figure S1b of the Supporting Information, SI) was then collected separately in
a vial 10 times, the excess solvent was removed and each isomer was
redissolved in 0.5 μL methanol-d4. This small amount
of sample was then inserted into a capillary, and NMR spectra were
acquired during 25 min for each isomer, averaging over 500 scans with
a relaxation delay of 2 s. Implementing an in-line concentration step
between chromatography and spectroscopy would greatly reduce the analysis
time, and make automated and more quantitative analysis possible by
reducing losses in the sample collection and preparation steps. In
this study, the possibility to concentrate samples, which are diluted
in scCO2 during chromatography, is investigated.In the high density liquid regime of CO2, above the
critical temperature and pressure of 31.9 °C and 73 bar respectively,[13] molecules can be dissolved into CO2. By changing the pressure and or temperature of a supercritical
fluid, its density changes and thereby its solvating ability. When
supercritical CO2 expands and becomes a gas, the density
decreases with decreasing pressure and the sample will not stay dissolved.
This principle is used for implementing an in-line concentration step
to convert the off-line SFC-NMR setup into an in-line analysis technique.
Experimental
Section
Set-Up
In our initial SFC-NMR setup, in which SFC was
coupled directly to the NMR probe through a flow control system, the
pressure of the SFC system and the pressure of the tubing going to
the NMR probe was controlled by the same backpressure regulator at
120 bar. However, the concentration of the sample in the plug was
too low to detect by NMR spectroscopy within a reasonable amount of
time. Therefore, the system was extended with an in-line concentrator
(indicated in green in Figure a). In this setup, the tubing going to the NMR probe is connected
to a different manual backpressure regulator with an operating range
between 20 and 103 bar (Vici Jour). This backpressure regulator allows
us to lower the pressure of a sample fraction which was selected after
SFC. Lowering the pressure leads to expansion of the plug in the water
flow. An extra stainless steel tube, with an inner diameter of 2.159
mm and length of ∼20 cm, is added to the system before the
NMR probehead to allow for separation of the phases (scCO2, gas phase CO2 and sample), as shown in Figure . The PEEK tubing going from
the expansion tube to the bottom of the NMR probe has an inner diameter
of 0.254 mm. All other PEEK tubing is 0.508 mm wide. The fused silica
capillary (Polymicro Technologies) in the stripline probe has an inner
diameter of 250 μm and outer diamer of 360 μm. To select
the desired fraction coming from the SFC instrument and to stop the
flow in the NMR probehead, two 6-port binary-position valves (Vici
Valco, Inc.) can be switched, namely valve 1 and valve 2 in Figure b, respectively.
Figure 1
(a) Schematic
representation of the SFC-NMR system. A broad expansion
tube (green) is added before the NMR spectrometer for concentrating
the sample. (b) The flow control system: After SFC, a fraction of
interest can be selected from the UV chromatogram using valve 1, flowing
into a 100 μL sample loop. The water flow pushes this plug toward
the center of the NMR stripline chip at a rate of 0.1 mL/min. By switching
valve 2, the flow can be stopped to acquire multiple scans.
(a) Schematic
representation of the SFC-NMR system. A broad expansion
tube (green) is added before the NMR spectrometer for concentrating
the sample. (b) The flow control system: After SFC, a fraction of
interest can be selected from the UV chromatogram using valve 1, flowing
into a 100 μL sample loop. The water flow pushes this plug toward
the center of the NMR stripline chip at a rate of 0.1 mL/min. By switching
valve 2, the flow can be stopped to acquire multiple scans.The flow control system of the
SFC-NMR setup can be seen in detail
in Figure b. In the
SFC machine, the sample is injected into a CO2 flow and
loaded onto the column for separation at high pressure (120 bar in
this case). This separation is followed by UV detection, after which
the sample goes through the backpressure regulator to the waste. Once
a sample of interest passes the UV detector, valve 1 can be switched.
A 100 μL loop is then filled with the sample, scCO2 and possibly a cosolvent. Once the loop is filled, the valve switches
back, thus inserting the sample plug into a flow of an incompressible
and immiscible medium, water in our case. This separate water flow
line, leading to the NMR probe, is kept at a lower pressure to expand
the scCO2 plug. After passing a wider tube, the sample
is concentrated at the trailing edge of the plug (as will be discussed
in the Results section) and flows at a rate
of 0.1 mL/min toward the center of the stripline for detection. This
sample can be detected by NMR in flow, or the flow can be stopped
by switching the second valve. In stop-flow mode, more scans can be
acquired to enhance signal intensities. The system remains under pressure
when the flow is stopped. After the NMR experiment, the sample will
flow through the backpressure regulator to the waste.
Instruments
In the SFC-NMR setup, we coupled a Waters
Acquity UPC2 instrument to a Varian VNMRS spectrometer
at 600 MHz Larmor frequency (14.1 T). A home-built stripline probe
was used,[20] with a 300 μm wide and
4 mm long chip, optimized for detecting an active volume of 150 nL.
In-flow experiments were performed with an acquisition time of 0.5
s and a relaxation delay of 0.2 s (in total 0.7 s per spectrum), while
stop-flow experiments were recorded with an acquisition time of 1
s and a relaxation delay of 5 s (for the samples separated in methanol/CO2) or 10 s (for the sample separated in toluene/CO2). All spectra were recorded with a receiver bandwidth of 10 kHz.
The spectra are referenced to the CH3 peak of toluene,
which was set to 2.1 ppm.
SFC Chromatography
A mixture of
four vitamin E (tocopherol)
isomers (Sigma-Aldrich, CAS 1406–66–2, not less than
80% β, γ and δ isomers) was diluted in toluene (Fisher
Scientific) at 100 mg/mL (0.24 M). Two μL of this sample (0.2
mg tocopherol) was injected onto a 100 mm × 1.7 μm packed
BEH SFC column (Waters) at a backpressure of 120 bar, a temperature
of 50 °C and a flow rate of 1.5 mL/min. For the separation in
toluene in CO2 (Linde Gas Benelux, food grade), a linear
solvent gradient of 20 to 30% toluene/CO2 in 4 min was
used. For the separation in methanol/CO2, a linear solvent
gradient of 2 to 5% methanol (BioSolve, SFC grade) in CO2 in 3 min was used.
Results and Discussion
Expansion of Toluene in
Supercritical CO2
Separation of the four isomers
of tocopherol, differing in the methylation
pattern of the aromatic ring, has been achieved before by SFC.[22] The same tocopherol mixture as used by Tayler
et al. has been used in this research consisting approximately of
9.7% α-, 1.1% β-, 66.7% γ-, and 22.5% δ-tocopherol,
determined by integration of their separate peaks in the UV chromatogram
(SI S1). A good separation can be achieved
within 2.5 min using a gradient of methanol in CO2, but
the β- and γ-isomers slightly coelute. The SFC column
can be loaded with up to 0.2 mg of tocopherol mixture during each
injection. Selecting 100 μL of the peak containing the highest
abundant isomer, γ-tocopherol (66.7% of the mixture, 0.13 mg),
and shuttling this compound directly to the NMR probe, without in-line
concentration, would result in an isomer concentration of 0.2 μg
(3.2 mM) in the 150 nL NMR detection volume of the stripline. As mentioned
in the introduction, the limit of detection (LOD) for the stripline
chip used in these experiments is 0.1 nmole 1H spins/.[20,22] This detection limit
is defined as the number of nuclear spins of a specific molecular
site needed for a signal-to-noise ratio of 1 in the FID, for a single
scan experiment. When more scans are acquired and averaged, a better
signal-to-noise ratio is achieved, scaling with the square root of
the number of scans. For a single scan experiment, with a receiver
bandwidth of 10 kHz, the concentration LOD is 67 mM in the 150 nL
detection volume of this stripline for a single resonance in the spectrum.
However, if 500 scans are acquired, then the concentration LOD would
be 3 mM for a signal-to-noise ratio of 1 in the FID for a single resonance.
The concentration of γ-tocopherol that comes from the SFC with
maximum loading of the column (3.2 mM) is therefore the bare minimum
required for NMR detection within reasonable time (3 mM in 500 scans,
taking approximately an hour of experiment time). The other, less
abundant isomers will not be visible within the same experiment time
when SFC and NMR are coupled directly. However, if an in-line concentration
step is implemented between SFC and NMR, then these isomers can also
be detected. On the basis of the detection limits and sample amounts
stated above, a concentration step of approximately 57 times would
be needed for the isomer with the lowest abundance (the β-isomer),
and of approximately 6.5 times for the second lowest abundant isomer
(the α-isomer) to obtain the detectable concentration of 3 mM.The calculations above set the minimum goal for the in-line concentration
step. The experiments in the following sections will show if this
goal can be achieved.The in-line concentration step can be
achieved by expanding the
scCO2 in which the sample of interest is dissolved after
chromatography, as explained below. The solubility of compounds in
CO2 depends on its density. By reducing the pressure, thus
expanding the scCO2 into a gas, its solvating power is
expected to decrease. Therefore, if the pressure drops, then the sample
and, if present, the cosolvent, will no longer dissolve in the CO2, thereby achieving an in-line separation of the sample from
the mobile phase. To test this hypothesis, NMR experiments were performed
at different pressures to monitor the expansion, of which the results
at a final pressure of 100 bar respectively 50 bar are shown in Figure . A mixture of toluene
in scCO2 is used as a model compound. The experiment was
designed to test if separating 100 μL of 8% toluene dissolved
in scCO2 at 120 bar is possible by expanding the CO2 in a broad tube. In Figure , an array of 1H NMR spectra is shown, zoomed
in on the CH3 peak of toluene (2.1 ppm). The spectra were
recorded every 0.7 s in flow.
Figure 2
Array of 1H NMR spectra, zoomed in
on the CH3-peak of toluene, recorded during the expansion
of a plug of 8% toluene
in 100 μL scCO2, in a water flow of 0.1 mL/min. The
spectra in both subfigures are scaled so that direct comparison of
peak intensity between the figures is possible. The spectra containing
water are indicated in blue, the other peaks originate from toluene
(red rectangle). The plug in (a) was expanded from 120 to 100 bar,
and in (b) the plug was expanded to 50 bar. The latter gives the best
results for concentrating toluene in scCO2.
Array of 1H NMR spectra, zoomed in
on the CH3-peak of toluene, recorded during the expansion
of a plug of 8% toluene
in 100 μL scCO2, in a water flow of 0.1 mL/min. The
spectra in both subfigures are scaled so that direct comparison of
peak intensity between the figures is possible. The spectra containing
water are indicated in blue, the other peaks originate from toluene
(red rectangle). The plug in (a) was expanded from 120 to 100 bar,
and in (b) the plug was expanded to 50 bar. The latter gives the best
results for concentrating toluene in scCO2.When the toluene/CO2 plug is expanded
from 120 to 100
bar (Figure a), toluene
is spread over the whole CO2 plug, indicating that toluene
is still dissolved in CO2. If the pressure is decreased
to 50 bar (Figure b), then toluene is observed only at the trailing edge of the plug
in a much smaller volume and therefore in a higher concentration (based
on the peak intensities, a factor 10 higher). This indicates that
during the expansion to 50 bar, toluene no longer dissolves in CO2. In this way, a separation of toluene and CO2 is
achieved. Once the water flow passes, at the trailing edge of the
plug, the toluene is pushed forward by the water flow, since toluene
is insoluble in water. This results in a concentrated toluene front
at the trailing edge of the CO2 before the water, as shown
in Figure b. A small
amount of CO2 might still be dissolved in this toluene
front, but this cannot be observed with 1H NMR. In the
next section, the expansion mechanism will be investigated in more
detail. Due to the expansion into a gas, which has a lower density,
the CO2 plug at 50 bar has a larger volume than the plug
at 100 bar, even though before expansion the starting volume of scCO2 was the same for both experiments. This is why more NMR spectra
were recorded in the second experiment than in the first, to follow
the entire, expanded plug.Expansion to even lower pressures
should result in a higher concentration.
However, when the plug was expanded to 40 bar, the toluene front was
no longer intact and toluene droplets were observed in the water flow
after the CO2. We concluded that expanding CO2 must be controlled and cannot be too sudden.An increase in
temperature leads to an even lower CO2 density, so more
expansion of the gas and therefore better separation
of the CO2 from the toluene and thus to an even higher
concentration of toluene. Going from room temperature to 90 °C
results in approximately 2.5 times higher peaks in the spectra, as
shown in SI S2. In combination with the
concentration of ∼10 times that was already achieved by lowering
the pressure to 50 bar (Figure ), a total concentration factor of 20 to 25 can be achieved
by lowering the pressure and working at elevated temperatures. However,
since these results at room temperature were satisfactory and the
experiment is easier to perform, we decided to perform the following
experiments at room temperature. Moreover, in general terms, high
temperatures might not always be desirable as some compounds might
be thermally unstable at 90 °C.
Expansion Mechanism
In the previous section, it was
described how the separation of toluene and CO2 can be
achieved by expanding the plug. To research the expansion mechanism
of the plug, the forward pressure on the water pump was measured over
time (Figure ). The
pressure of the manual, low-pressure backpressure regulator at the
end of the flow line was kept constant during these experiments.
Figure 3
Forward
pressure on water pump recorded during the expansion experiment.
The backpressure regulator was originally set to 40 bar (red) and
46 bar (black). The first peak (*) originates from selecting a sample
fraction at 120 bar and injecting it into the flow at lower pressure,
40 or 46 bar.
Forward
pressure on water pump recorded during the expansion experiment.
The backpressure regulator was originally set to 40 bar (red) and
46 bar (black). The first peak (*) originates from selecting a sample
fraction at 120 bar and injecting it into the flow at lower pressure,
40 or 46 bar.The pressure of the water
flow increases after switching in the
scCO2 plug (indicated by a star in Figure ). This was expected as the plug coming from
the SFC column has a pressure of 120 bar, which is much higher than
the 40 or 46 bar of the water flow. An unexpected result was, however,
that after switching the plug into the flow, the pressure always dropped
to 57 bar, stayed at this level for about 1 min and then slowly decreased
to the set value of the backpressure regulator (40 or 46 bar). A pressure
of 57 bar which is observed is actually equal to the vapor pressure
of CO2 at room temperature. This leads to the conclusion
that after switching the valve, two phases of CO2 are present
simultaneously, a liquid and a gas phase. The decrease in pressure
indicates that the liquid phase slowly evaporates into the gas phase
until it is fully expanded. Since the gas phase has a lower solubility
than the liquid phase, the assumption is made that the sample and
cosolvent present in the CO2 will stay in the liquid CO2 phase, leading to an increased sample concentration.The expansion starts directly when injecting the plug into the
PEEK tubing at the beginning of the water flow-line. This PEEK tubing
was made long enough for the plug to fully expand before reaching
the stripline detector. The wide stainless steel expansion tube, that
is placed before the stripline detector, plays an essential role.
This tubing needs to be put in a vertical position in order for the
liquid phase containing the sample to be collected at the trailing
edge (bottom) of the plug. When the wide stainless steel tube is put
in a horizontal position, several droplets are observed before the
liquid sample front at the trailing edge of the plug. It is assumed
that gravity helps in separating the gaseous and liquid phase, so
that all the liquid containing the sample is collected at the bottom
of the expansion tube and the gas can flow to the top. Therefore,
the liquid phase is always observed at the trailing edge of the plug.
SFC-NMR of Tocopherol Isomers in Toluene/CO2
Now that we have shown that separating toluene when mixed with scCO2 is possible, the next step is to show that a sample, which
was separated by SFC, can be concentrated in the same way. This sample
is present in a much lower concentration than the toluene tested in
the previous sections. The experimental conditions where similar to
the model described in the first results section, where toluene is separated from scCO2 by expanding.
Instead of only having a mixture of toluene and CO2, now
a small amount of sample is added, i.e., toluene is now our cosolvent.
For this experiment 0.2 mg of a mixture of four tocopherol isomers
was separated in toluene/CO2. γ-tocopherol was selected
and shuttled into a 100 μL sample loop, together with the mobile
phase consisting of 25% toluene in CO2. This plug, consisting
of 25 μL toluene, a small amount of γ-tocopherol (less
than 0.13 mg) and 75 μL scCO2, was put into a water
flow at 50 bar at room temperature to expand the plug. When the plug
reached the center of the active detection volume in the spectrometer,
the flow was stopped and 500 scans were recorded in 90 min (Figure ). In the case of
a perfect separation of CO2 and toluene, 0.8 μg γ-tocopherol
should be present in the active detection volume of the stripline
(0.13 mg in 25 μL toluene, of which 150 nL can be detected).
The real amount is smaller since there is still some CO2 mixed in with the toluene at the trailing edge of the plug, so the
sample is therefore diluted in a larger volume than the 25 μL
pure toluene.
Figure 4
SFC–1H NMR spectrum of γ-tocopherol
in
toluene, obtained after selecting the correct isomer in SFC with 25%
toluene/CO2 and expanding the plug to 50 bar at room temperature,
in order to increase the concentration. The flow was stopped in the
NMR spectrometer and 500 scans were averaged to obtain this spectrum,
taking 90 min. The structure of γ-tocopherol and the corresponding
peaks are indicated in the figure. The aryl-H peak (red), which is
needed to distinguish between the isomers of tocopherol, is visible.
The aryl-CH3 (blue) peaks fall below the broad toluene
peak.
SFC–1H NMR spectrum of γ-tocopherol
in
toluene, obtained after selecting the correct isomer in SFC with 25%
toluene/CO2 and expanding the plug to 50 bar at room temperature,
in order to increase the concentration. The flow was stopped in the
NMR spectrometer and 500 scans were averaged to obtain this spectrum,
taking 90 min. The structure of γ-tocopherol and the corresponding
peaks are indicated in the figure. The aryl-H peak (red), which is
needed to distinguish between the isomers of tocopherol, is visible.
The aryl-CH3 (blue) peaks fall below the broad toluene
peak.One of the peaks needed to distinguish
between the isomers, i.e.,
the aryl-H peak, indicated in red in the figure, is clearly visible
after acquiring 500 scans. The aryl-CH3 peaks overlap with
the broad toluene peak at 2.1 ppm. On the basis of this spectrum it
can be concluded that either the γ- or β-tocopherol isomer
was isolated. To further distinguish between the two isomers the aryl-CH3 peaks must be visible. Concentrating the sample in 25 μL
toluene was however successful, since the sample can be detected by
NMR. A broad signal from 1 to 3 ppm can be observed in the baseline
of the spectrum, which originates from the coating of the fused-silica
capillary. A small water contamination is present as well, but does
not interfere with the tocopherol region in the spectrum. It can therefore
be concluded that concentration by expansion after SFC is successful
and sufficient for in-line detection by NMR. In principle, this concentration
method is applicable to any nonpolar sample separated in SFC with
CO2 and a nonpolar cosolvent as the mobile phase.
SFC-NMR
of Tocopherol Isomers in Methanol CO2
Although
we successfully separated the tocopherol isomers
by SFC using toluene as a cosolvent, separation of tocopherols and
many other samples, is usually achieved with methanol as cosolvent.
The chromatography of tocopherol in methanol/CO2 was optimized
using a gradient of 2–5% methanol/CO2 in 3 min.
However, when performing the SFC-NMR experiment described in the previous
section with methanol instead of toluene as cosolvent, no sample or
methanol were observed in the plug. This is due to the fact that methanol
is polar and dissolves into the water instead of making a front on
top of the water as toluene does. The sample is therefore not concentrated,
but diluted further into the water flow. One possible solution to
this problem is to replace the water by a nonpolar “transportation
medium”. This will be researched in the future. The other option
is an in-line solvent switch from methanol to a nonpolar solvent after
the separation by SFC. The later is described in the following section.To achieve this solvent switch, an adapted setup is used, where
a small 5 μL injection loop is inserted into the 100 μL
sample loop, which is achieved by adding a third 6-port binary valve
to the setup, as indicated in Figure .
Figure 5
Valve switching system for selecting the desired sample
from SFC
(valve 1). A 5 μL sample loop (pink) is added for injecting
a small amount of toluene (valve 3). The flow can be stopped in the
NMR by switching valve 2.
Valve switching system for selecting the desired sample
from SFC
(valve 1). A 5 μL sample loop (pink) is added for injecting
a small amount of toluene (valve 3). The flow can be stopped in the
NMR by switching valve 2.This injection loop can be filled with a nonpolar solvent,
e.g.,
toluene. After selecting the sample after SFC and switching in the
injection loop with toluene, the plug consist of 3% methanol/CO2, 5 μL toluene and γ-tocopherol. This plug is
expanded to 48 bar at room temperature and the flow stopped once the
plug reached the center of the stripline. The spectrum was recorded
during 50 min, averaging over 500 scans (Figure , middle spectrum).
Figure 6
SFC–1H NMR spectrum of γ-tocopherol in
methanol and toluene (green), obtained after selecting the correct
isomer in SFC with 3% methanol/CO2, switching solvents
to toluene and expanding the plug to 50 bar at room temperature, in
order to increase the concentration. 500 scans were averaged and compared
to 500 scans of the same experiment performed with 25% toluene as
SFC cosolvent (black, same as Figure ). On the right the same spectra are shown as on the
left, but zoomed in on the area around 2 ppm. The aryl-H peak (red),
which is needed to distinguish between the isomers of tocopherol,
is visible. The aryl-CH3 (blue) peaks overlap with the
broad toluene peak. When using 5 μL deuterated toluene for the
solvent switch the bottom spectrum (red) is obtained. Here the aryl-H
as well as the two aryl-CH3 peaks of γ-tocopherol
are observed.
SFC–1H NMR spectrum of γ-tocopherol in
methanol and toluene (green), obtained after selecting the correct
isomer in SFC with 3% methanol/CO2, switching solvents
to toluene and expanding the plug to 50 bar at room temperature, in
order to increase the concentration. 500 scans were averaged and compared
to 500 scans of the same experiment performed with 25% toluene as
SFC cosolvent (black, same as Figure ). On the right the same spectra are shown as on the
left, but zoomed in on the area around 2 ppm. The aryl-H peak (red),
which is needed to distinguish between the isomers of tocopherol,
is visible. The aryl-CH3 (blue) peaks overlap with the
broad toluene peak. When using 5 μL deuterated toluene for the
solvent switch the bottom spectrum (red) is obtained. Here the aryl-H
as well as the two aryl-CH3 peaks of γ-tocopherol
are observed.By injecting toluene
and expanding the plug, the solvent is switched
from methanol to toluene. The sample is separated in a methanol/CO2 mixture, but during the expansion the sample dissolves in
toluene, while methanol partially mixes into the water flow at the
trailing edge of the plug. In the middle spectrum in Figure , the aryl-H peak of tocopherol
is visible, as was the case without solvent switch (upper spectrum,
black). Although less toluene is now used (5 μL vs 25 μL)
to concentrate the sample, the toluene peak still has approximately
the same intensity. This can be expected, since the toluene intensity
is determined by the amount of toluene in the detection volume, not
by the total toluene volume in the flow line. The aryl-CH3 peaks of the sample therefore still overlap with the toluene signal.An additional advantage of the solvent switch procedure is that
deuterated toluene (Deutero GmbH, 99% deuterated) can be used at limited
cost, as the required volume is small, namely 5 μL. This avoids
intense solvent peaks in the spectrum, as is shown in the bottom spectrum
of Figure . Here the
aryl-H as well as the two aryl-CH3 peaks are clearly resolved,
confirming the correct selection of γ-tocopherol from the SFC
chromatogram. Most of the water contamination that was previously
present is removed as well, however a small methanol contamination
is now observed, originating from the cosolvent that was used during
chromatography. This does not interfere with the aryl-H and aryl-CH3 peaks, so the isomers can still be distinguished.The
experiment, including in-line expansion and solvent switch
to 5 μL toluene-d8, was repeated for α- and
δ-tocopherol (Figure ). The concentrations of these isomers in the mixture were
substantially lower than the γ-isomer (66.7%, determined by
integrating the UV chromatogram), namely 9.7% and 22.5% for α-
and δ-tocopherol, respectively. The β-isomer (1.1%, 2.2
μg) was too low in concentration to be detected in 500 scans
(50 min). The lowest amount of one single isomer that has to be injected
on the column is therefore between 2.2 μg (5.1 nmole) and 19
μg (45.3 nmole) to be able to detect it in the SFC-NMR setup
within 50 min of experiment time.
Figure 7
SFC–1H NMR spectrum
of γ-tocopherol in
methanol and toluene-d8 (red, same as Figure ), of δ-tocopherol in
methanol and toluene-d8 (black) and of α-tocopherol
in methanol and toluene-d8 (green) obtained after selecting
the correct isomer in SFC with 3% methanol/CO2 and expanding
the plug to 50 bar at room temperature, in order to increase the concentration.
500 scans were averaged, taking 50 min per spectrum. The aryl-H (red)
and aryl-CH3 (blue) peaks, which are needed to distinguish
between the isomers of tocopherol, are clearly visible. One of the
three methyl peaks of α-tocopherol partially overlaps with the
toluene peak, but can still be distinguished.
SFC–1H NMR spectrum
of γ-tocopherol in
methanol and toluene-d8 (red, same as Figure ), of δ-tocopherol in
methanol and toluene-d8 (black) and of α-tocopherol
in methanol and toluene-d8 (green) obtained after selecting
the correct isomer in SFC with 3% methanol/CO2 and expanding
the plug to 50 bar at room temperature, in order to increase the concentration.
500 scans were averaged, taking 50 min per spectrum. The aryl-H (red)
and aryl-CH3 (blue) peaks, which are needed to distinguish
between the isomers of tocopherol, are clearly visible. One of the
three methyl peaks of α-tocopherol partially overlaps with the
toluene peak, but can still be distinguished.As shown, the concentration method was successfully extended
to
enable concentrating a nonpolar sample separated in SFC with CO2 and a polar cosolvent as the mobile phase, by switching to
a nonpolar solvent in-line. This is applicable to other samples as
well, as long as the sample dissolves better in the nonpolar NMR switching
solvent than in the SFC cosolvent. The ability of switching solvents
is an advantage since the best suitable solvent can be selected for
SFC and NMR separately to match each sample. Care should be taken
in the solvent selection to avoid sample precipitation in the SFC
column or in the tubing toward the NMR spectrometer.Quantifying
the concentration step is possible by comparing the
toluene peak integrals for the expanded (48 bar) and unexpanded (120
bar) toluene/CO2 plug (SI S3). From these integrals it was calculated that due to the in-line
expansion, approximately 10 times higher sample concentrations are
observed, saving a factor of 100 in NMR experiment time. As stated
before, the tocopherol mixture consisted of 9.7% α-, 1.1% β-,
66.7% γ-, and 22.5% δ-tocopherol. Selecting the highest
abundant isomer, γ-tocopherol, and shuttling this compound directly
to the NMR probe, without in-line concentration, would result in an
isomer concentration of 0.2 μg (3.2 mM) in the 150 nL NMR detection
volume of the stripline. On the basis of the LOD of the stripline
chip mentioned in the introduction, this would
be the lowest concentration that can be detected with the stripline
probe within 500 scans (3 mM in 50 min). However, with the in-line
concentration step described in this work, a 10 times higher sample
concentration can be achieved in the stripline. Therefore, to end
up with a concentration of 3 mM in the stripline, 10 times less sample
is needed. A concentration of 3 mM corresponds to 0.19 μg tocopherol
in 150 nL, for which normally 0.13 mg would need to be injected onto
the SFC column, but with this concentration step only 13 μg
is needed. This coincides with the result that the α-isomer
can be detected (9.7%, 19 μg injected) after in-line concentration,
but not the β-isomer (1.1%, 2.2 μg injected). As mentioned
before, detecting the β-isomer would require concentrating the
sample ∼57 times, to 3 mM in the stripline.In the off-line
experiment described by Tayler et al.,[22] the tocopherol isomers were collected 10 times
after separation and redissolved in 0.5 μL methanol-d4. The concentration in the stripline of β-tocopherol, the lowest
abundant isomer, was therefore 3.2 mM, which is close to the detection
limit for this stripline probe in 500 scans. In this case, without
in-line concentration, the β-isomer concentration in the stripline
is 0.053 mM. Therefore, to match the results of this off-line SFC-NMR
experiment, an in-line concentration step, concentrating the sample
at least 61 times up to 3.2 mM, is needed. A higher concentration
can be achieved by working at a higher temperature (SI S2), concentrating the sample 20–25 times instead
of 10 times at room temperature, but this would still not be sufficient
to detect the β-isomer.As currently, the experimental
times of the NMR analysis exceed
the retention times in the SFC by far, storage loops are needed to
perform a full analysis for a single injection of a (mass-limited)
sample. To improve on this situation and to further bring down the
LOD, the NMR sensitivity needs to increase. An easily perceived option
is changing the detection volume of the stripline chip to better match
the volume of the sample plugs. The sample is now concentrated in
5 μL of toluene, of which only 150 nL can be detected with the
current stripline probe. By making the detection volume larger, which
is possible due to the scalability of the stripline,[20] more sample is detected in one scan leading to a better
signal-to-noise in the NMR spectrum. For example, increasing the volume
with a factor 10 to 1.5 μL leads to a 10 times higher signal,
with the sensitivity of the stripline decreasing with its width by
a factor of 10(1/3) (= 2.15),[23] this would overall save a factor ∼50 in experiment time.
So the spectra which now took 50 min to acquire can then be acquired
within 1 min, making SFC-NMR viable for higher throughput analysis.
For direct in flow detection of all fractions coming of the SFC columns
further sensitivity enhancements are needed. This might be achieved
by Overhauser dynamic nuclear polarization (DNP) for which scCO2 is a very efficient solvent.[24] Developments along these lines are in progress.
Conclusions
A technique for concentrating samples in-line, which are diluted
in scCO2 during chromatography was developed. This is achieved
by controlled in-line expansion of the scCO2 to ∼50
bar at room temperature. Even higher concentrations are achieved when
expanding at a higher temperature of ∼90 ◦C.In-line SFC-NMR detection of α-, γ-, and δ-tocopherol
is possible with this concentration step. Proof-of-principle experiments
have been performed on a tocopherol mixture separated with SFC in
toluene/scCO2. Separation and concentration in methanol/scCO2 is possible as well, but requires the addition of a small
amount of toluene (5 μL) after SFC separation to perform a solvent
switch. The β-isomer concentration was too low to be detected,
even after in-line concentration. Further research will focus on increasing
the detection volume of the stripline to better match the sample volumes
of ∼5 μL which are obtained after in-line concentration.
In this way more of the sample that is already present can be detected,
possibly also β-tocopherol.This in-line concentration
technique is essential for coupling
SFC and NMR in-line and could also be used in hyphenation of SFC with
other detectors than NMR, such as MS. The hyphenation of SFC and NMR
is interesting for studying supercritical behavior and mixing of solvents
in supercritical fluids as well. Currently, the presented method is
applicable for concentrating nonpolar samples in a polar or nonpolar
cosolvent in scCO2. Further research is currently done
to expand this method for concentrating and separating polar samples.
Authors: Manfred Krucker; Annette Lienau; Karsten Putzbach; Marc David Grynbaum; Paul Schuler; Klaus Albert Journal: Anal Chem Date: 2004-05-01 Impact factor: 6.986
Authors: Maximilian Kühnle; Diana Kreidler; Karsten Holtin; Harri Czesla; Paul Schuler; Walter Schaal; Volker Schurig; Klaus Albert Journal: Anal Chem Date: 2008-06-06 Impact factor: 6.986