Clinician-collected urethra swab: A good alternative sample type using cobas 4800 system for Chlamydia trachomatis and Neisseria gonorrhoeae in men.

Qing-Yong Wang1, Lu-Qing Zheng1, Rong-Hai Li1, Ying Zheng1, Xun Chen1, Qi Li1.   

Abstract

BACKGROUND: Nucleic acid amplification tests (NAATs) are being used increasing to detection of CT (Chlamydia trachomatis) and NG (Neisseria gonorrhoeae) infections for superior sensitivity and specificity than other tests. Male first-void urine (FVU) sample is the optimal sample type for detection of CT and NG by NAATs. Although not being the recommended by NAATs, clinician-collected urethra swab (CCUS) is perhaps a good alternative sample type compared with the FVU sample in men.
METHODS: Paired samples (FVU and CCUS) from one hundred male outpatients were simultaneously detected by urine pattern and swab pattern using cobas 4800 CT/NG assay on cobas 4800 system for the detection of CT and NG, respectively. And twenty-one positive controls were also detected on cobas 4800 system.
RESULTS: The CT/NG cycle thresholds (Ct) value of urine pattern is lower than that of swab pattern for the same positive samples (clinical samples and positive controls) on the cobas 4800 CT/NG assay. The final CT/NG results of two sample patterns from patients were highly consistent except for four discordant results.
CONCLUSION: CCUS is validated for a good alternative sample type for the CT/NG detection on the cobas 4800 system in this study.
© 2020 The Authors. Journal of Clinical Laboratory Analysis published by Wiley Periodicals Inc.

Entities:  

Keywords:  zzm321990Chlamydia trachomatiszzm321990; zzm321990Neisseria gonorrhoeaezzm321990; cobas 4800 system

Mesh:

Substances:

Year:  2020        PMID: 32841434      PMCID: PMC7521308          DOI: 10.1002/jcla.23331

Source DB:  PubMed          Journal:  J Clin Lab Anal        ISSN: 0887-8013            Impact factor:   2.352


INTRODUCTION

Chlamydia trachomatis and NG are the most prevalent bacterial sexually transmitted infections globally, about 131 million and 78 million new cases each year, respectively. In 2017, a total of 17 million CT and 555 608 cases were reported by centers for Disease Control and Prevention (CDC) in the United States, 22% and 67% increasing since 2013. If not treated early, CT/NG can cause serious health problems, painful complications, emotional injury, and so on. NAATs are being used increasing to detection of CT/NG infections for superior sensitivity and specificity than other tests. Male FVU sample is currently the optimal sample type for detection of CT/NG by NAATs. Adopted widely in numerous settings, cobas CT/NG assay for use on the cobas 4800 system provides automated solutions for the detection of CT/NG with excellent sensitivity and specificity. But in practice, the CCUS sample type should be validated by the laboratories for the male CT/NG screening using cobas 4800 system. Because clinician workers could choose the CCUS sample type for male patients in some situation (offering urine/re‐sampling difficulty, multiple items simultaneously, not likely to return for test reports, and so on). So we compared the consistent of CCUS and FVU samples using cobas CT/NG assay on cobas 4800 system (Roche Molecular Systems).

MATERIALS AND METHODS

We retrospectively the CT/NG results of one hundred paired samples (100 FVU and 100 CCUS) from male outpatients attending urology department in the hospital (mostly symptomatic) from January, 2017 to December, 2018 in this study. Samples were collected in the cobas PCR Urine Sample Kits (Roche Molecular Systems, Inc.) for the detection of CT/NG using cobas 4800 system with patients consent. Collection of FVU followed collection of CCUS in this study, all of them were transported to the laboratory at 2‐30°C for the CT/NG detection. FVU was defined as the first 10 to 50 mL of the urine stream, about 4.5 mL was added to cobas PCR Urine Sample Kit; The CCUS (provided by Jiangsu Kangjie Medical Devices Co., Ltd) transported to the laboratory was firstly placed into sterile tubes with 3 mL cobas PCR media more than 4 hours, then all of which was added to the remaining cobas PCR Urine Sample Kit before being detected. The discordant results between two sample types for male patients were judged by final clinical judgment analysis in this study. In addition, twenty‐one cobas® 4800 CT/NG positive controls and positive samples were detected using cobas 4800 system by two patterns (urine and swab sample type), respectively. GraphPad Prism 8.0 software was used to plot the results. Statistical significance was set to P < .05.

RESULTS

The overall prevalence of CT was higher than that of NG infection (6.1% vs. 2.5%) for male patients attending the hospital during the past two years, and the median age of the positive patients was 32.7 years old. Then, the prevalence of CT (18% vs. 15%) was also higher than that of NG (9% vs. 8%) infection for CCUS and FVU of the recruited male patients in the study, respectively. And the median age of positive male patients was 33.8 years old. A total of 100 patients provided paired samples in the study, including 18 positive CCUS (15 positive FVU) for CT and 9 positive CCUS (8 positive FVU) for NG, respectively. Of them, we found four discordant results: three positive CT results from CCUS (the Ct value were 41.8, 40.2, and 39.8) but negative CT from FVU, and one positive NG result from CCUS (the Ct value was 41.2) but negative NG from FVU. The positive CT/NG CCUS from the four discordant results were right according to clinical further judgment. The sensitivity of cobas 4800 system (83.3% vs. 88.9%) for CT and NG in FVU samples showed in Table 1. The specificity and positive predictive value of CT and NG from FVU were all 100%. Then, the negative predictive value of those was 96.5% and 98.9%, respectively. Then, as shown in Figure 1, the mean Ct value of CT from positive controls/positive samples was 38.2 (range 37.0‐39.4 cycles)/31.3 (range 23.7‐40.2 cycles) from urine pattern was lower than 38.9 (range 37.1‐39.6 cycles) /33.8 (range 28.7‐41.8 cycles) from swab pattern, respectively (P < .05); and the mean Ct value of NG was 37.7 (range 35.8‐38.7 cycles) /30.3 (range 26.1‐37.3 cycles) from urine pattern was also lower than 38.2 (range 36.8‐39.5 cycles) /31.1 (range 26.6‐41.2 cycles) from swab pattern, respectively (P < .05).
Table 1

Performance characteristics of cobas 4800 system for clinician‐collected urethral swabs and first‐void urine samples

First‐void urine (n = 100)Clinician‐collected urethral swab (n = 100)Statistical data
Positive (n)Negative (n)Sensitivity (%)Specificity (%)PPV (%)NPV (%)
Chlamydia trachomatis Positive (n)15083.310010096.5
Negative (n)382
Neisseria gonorrhoeae Positive (n)8088.910010098.9
Negative (n)191

Abbreviations: NPV, negative predictive value; PPV, positive predictive value.

Figure 1

Comparison of the CT/NG Ct values from positive controls and positive samples by swab and urine patterns using cobas 4800 system, respectively

Performance characteristics of cobas 4800 system for clinician‐collected urethral swabs and first‐void urine samples Abbreviations: NPV, negative predictive value; PPV, positive predictive value. Comparison of the CT/NG Ct values from positive controls and positive samples by swab and urine patterns using cobas 4800 system, respectively

DISCUSSION

In the United States, the rates in men infected by CT increased 39.3% than 11.1% of women during 2013‐2017, while rates of NG among gay, bisexual, and other (men who have sex with men) MSM had higher positivity rates compare to women and men who have sex with women during 2010‐2017. The date of the US CDC reports indicated that a considerable increase in male CT/NG infections in past years. FVU is currently the male optimum sample type for the routine detection of CT/NG by NATTs, accompanied by some advantages (noninvasive, self‐taken, and high organism load). But it is clear that some CT/NG infections may be missed if urine is the only sample type for the male patients, so the semen sample protocol on cobas 4800 CT/NG assay was also a good successful method for the CT detection ; and pharyngeal and rectal swabs had been validated by some laboratories for the detection of CT/NG in MSM. , And it is also equally importance for us to know the strengths and weakness of other alternative male sample types in addition to FVU. We believe that the CDC recommendations on the optimal male specimen type (FVU) would discourage the use of other sample types. Self‐collected specimens for the detection of CT/NG were perhaps not an effective strategy without detailed practices information. Although painful taking due to deep insertion, CCUS might provide additional information being not offered only from FVU. Interestingly, CCUS are routinely tested for bacterial or mycoplasma in most laboratories, it should be developed for detection of CT/NG on the sample type with satisfactory performance specification for NAATs. In addition, the CCUS pretreatment protocol has good practicability and conveniences. In this study, the mean Ct value of CT and NG from positive controls/samples detected on urine pattern was 0.7/2.5 and 0.5/0.8 lower than those on swab pattern, respectively. Ct values versus DNA load are inverse, indicating that urine pattern is better than swab pattern for the same sample on cobas 4800 CT/NG assay. Obtaining one hundred paired samples is perhaps difficult for doctor workers due to expensive process and patient agreement in this study. For many years, CCUS collected from men may have posed a barrier for screening men. Cobas 4800 system offers high sensitivity, specificity, negative, and positive predictive values for the detection of CT/NG in urine specimens. And the detail preparation for CCUS in the study was not provided in the manufacturer's instructions of the cobas 4800 system. We described good concordance of CCUS with FVU on cobas 4800 CT/NG assay in this study. Four discordances of CT or NG results (CCUS samples showed Ct > 39 cycles but the corresponding FVU samples were all negative) showed in this study. Unfortunately, we could not use more methods to confirm the real CT/NG infection of the discordance samples due to the retrospective study. But we believed that the inhibition of amplification in urine specimens for PCR and a lack of monitoring self‐collection of FVU could provide false‐negative report in some cases, and a follow‐up study for that would be performed in future. Then, the supplementary testing of NG weak‐positive samples with Ct > 38 has limited utility ; and the detection of NG from rectal and urogenital samples using cobas 4800 system require no routine confirmatory testing. So being a good alternative sample type, the results from CCUS samples perhaps strengthen the identify and treatment of CT/NG infections using cobas 4800 system when applicable in the future. One limitation of this study was the specimen size. Another limitation was not rating the collection of CCUS as easy, difficult, or neither easy nor difficult. But we believe that date presented here may be also useful for clinician workers to choose the CCUS sample type for male patients in some situation, for example, offering urine/re‐sampling difficulty, multiple items simultaneously, not likely to return for test reports, and so on.

CONCLUSION

Effective laboratory diagnosis of CT/NG infection is essential, the study showed that CCUS was a good alternative male specimen type for the detection of CT/NG using cobas 4800 system. So clinical workers could select both male sample types (FVU and CCUS) detected by NAATs if impossible.
  11 in total

1.  Supplementary testing is not required in the cobas 4800 CT/NG test for Neisseria gonorrhoeae weak-positive urogenital samples.

Authors:  Collette Bromhead; Nadika Liyanarachchy; Julia Mayes; Arlo Upton; Michelle Balm
Journal:  J Clin Microbiol       Date:  2014-11-12       Impact factor: 5.948

2.  Screening methods for Chlamydia trachomatis and Neisseria gonorrhoeae infections in sexually transmitted infection clinics: what do patients prefer?

Authors:  Elizabeth J Howard; Fujie Xu; Stephanie N Taylor; Bradley P Stoner; Leandro Mena; M Jacques Nsuami; Suzanne Powell; Rebecca Lillis; David H Martin
Journal:  Sex Transm Infect       Date:  2010-11-12       Impact factor: 3.519

Review 3.  Cobas® 4800: a fully automated system for the detection of Chlamydia trachomatis and Neisseria gonorrhoeae.

Authors:  Barbara Van Der Pol
Journal:  Expert Rev Mol Diagn       Date:  2013-03       Impact factor: 5.225

4.  Recommendations for the laboratory-based detection of Chlamydia trachomatis and Neisseria gonorrhoeae--2014.

Authors: 
Journal:  MMWR Recomm Rep       Date:  2014-03-14

5.  Detection of Neisseria gonorrhoeae and Chlamydia trachomatis from pooled rectal, pharyngeal and urine specimens in men who have sex with men.

Authors:  David John Speers; I-Ly Joanna Chua; Justin Manuel; Lewis Marshall
Journal:  Sex Transm Infect       Date:  2017-10-24       Impact factor: 3.519

6.  Is confirmatory testing of Roche cobas 4800 CT/NG test Neisseria gonorrhoeae positive samples required? Comparison of the Roche cobas 4800 CT/NG test with an opa/pap duplex assay for the detection of N gonorrhoeae.

Authors:  Michael D Perry; Rachel N Jones; Sally A Corden
Journal:  Sex Transm Infect       Date:  2014-03-20       Impact factor: 3.519

Review 7.  Laboratory aspects of screening men for Chlamydia trachomatis in the new millennium.

Authors:  Charlotte A Gaydos; Dennis V Ferrero; John Papp
Journal:  Sex Transm Dis       Date:  2008-11       Impact factor: 2.830

8.  Swabs (dry or collected in universal transport medium) and semen can be used for the detection of Chlamydia trachomatis using the cobas 4800 system.

Authors:  Chloé Le Roy; Aline Papaxanthos; Oliver Liesenfeld; Virginie Mehats; Maïthé Clerc; Cécile Bébéar; Bertille de Barbeyrac
Journal:  J Med Microbiol       Date:  2012-11-01       Impact factor: 2.472

Review 9.  Global Estimates of the Prevalence and Incidence of Four Curable Sexually Transmitted Infections in 2012 Based on Systematic Review and Global Reporting.

Authors:  Lori Newman; Jane Rowley; Stephen Vander Hoorn; Nalinka Saman Wijesooriya; Magnus Unemo; Nicola Low; Gretchen Stevens; Sami Gottlieb; James Kiarie; Marleen Temmerman
Journal:  PLoS One       Date:  2015-12-08       Impact factor: 3.240

10.  Clinician-collected urethra swab: A good alternative sample type using cobas 4800 system for Chlamydia trachomatis and Neisseria gonorrhoeae in men.

Authors:  Qing-Yong Wang; Lu-Qing Zheng; Rong-Hai Li; Ying Zheng; Xun Chen; Qi Li
Journal:  J Clin Lab Anal       Date:  2020-08-25       Impact factor: 2.352

View more
  1 in total

1.  Clinician-collected urethra swab: A good alternative sample type using cobas 4800 system for Chlamydia trachomatis and Neisseria gonorrhoeae in men.

Authors:  Qing-Yong Wang; Lu-Qing Zheng; Rong-Hai Li; Ying Zheng; Xun Chen; Qi Li
Journal:  J Clin Lab Anal       Date:  2020-08-25       Impact factor: 2.352

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.