| Literature DB >> 32838208 |
Aditya Vasan1,2, Reiley Weekes1, William Connacher1,2, Jeremy Sieker3, Mark Stambaugh4, Preetham Suresh5, Daniel E Lee5,6, William Mazzei5, Eric Schlaepfer7, Theodore Vallejos8, Johan Petersen5, Sidney Merritt5, Lonnie Petersen1,9, James Friend1,2.
Abstract
The COVID-19 pandemic has produced critical shortages of ventilators worldwide. There is an unmet need for rapidly deployable, emergency-use ventilators with sufficient functionality to manage COVID-19 patients with severe acute respiratory distress syndrome. Here, we show the development and validation of a simple, portable and low-cost ventilator that may be rapidly manufactured with minimal susceptibility to supply chain disruptions. This single-mode continuous, mandatory, closed-loop, pressure-controlled, time-terminated emergency ventilator offers robust safety and functionality absent in existing solutions to the ventilator shortage. Validated using certified test lungs over a wide range of compliances, pressures, volumes and resistances to meet U.S. Food and Drug Administration standards of safety and efficacy, an Emergency Use Authorization is in review for this system. This emergency ventilator could eliminate controversial ventilator rationing or splitting to serve multiple patients. All design and validation information is provided to facilitate ventilator production even in resource-limited settings.Entities:
Keywords: COVID‐19 pandemic; acute respiratory distress syndrome; critical care; mass casualty incidents; mechanical ventilation; medical device design; respiratory insufficiency
Year: 2020 PMID: 32838208 PMCID: PMC7300530 DOI: 10.1002/mds3.10106
Source DB: PubMed Journal: Med Devices Sens ISSN: 2573-802X
FIGURE 1The ventilator was tested on a lung simulator. All parameters were tested to their stated limits (over 200 individual experiments) and according to International Standards Organization standards for pressure‐controlled ventilation. Notice that the dead space is kept to a minimum by reducing the length of tube between the bag and the lung simulator; this configuration was reproducible with a full‐sized simulator manikin and a standard adjustable overbed hospital bedside table. The system shown here is an early prototype with exposed electronics, but is to be supplied with housings as depicted in Figure 2
FIGURE 2Render of the final version of MADVent, with an electronics enclosure. The enclosure has an interface for the healthcare provider to adjust various ventilation settings such as target pressure, inspiratory time, respiratory rate and alarm thresholds. An liquid crystal display (LCD) screen displays ventilation parameters in real time. LED's and a built in alarm alert the healthcare provider in the event of an emergency
FIGURE 3Tidal volume is related to the rotation of the motor via compression of the bag, as indicated (A) by the experimental results compared with a model V tidal = V tidal()constructed from the geometry (see [Link], [Link], [Link], [Link] for the full derivation). Furthermore, a post hoc quadratic curve fit (3.47 × 10–4 + 0.322 − 52.5 with R 2 = .953) is provided showing a slightly improved fit, indicating that a quadratic function can adequately represent the tidal volume as a function of the angle . In B, the volume corresponding to a given motor rotation is seen to increase with compliance—accounting for the spread in the data along with experimental error. In C, the difference between peak pressure and PEEP is seen to increase along the model, as expected due to the ideal gas law
FIGURE 4The MADVent Mark V has alarms for high and low volume that may be set between 200 and 1,000 ml. In this example, the system was run at a rate of 13 breaths per minute (ventilation rate), a PEEP value of 15 cm H2O and the compliance on the lung simulator was initially set to 0.03 /cm H2O. A, The high‐volume alarm threshold was set to 500 ml for the first case. PEEP was decreased from 15 to 5 cm H2O in order to increase the tidal volume delivered to the lung simulator. A high‐volume alarm was triggered when the calculated tidal volume exceed the limit set by the healthcare provider. A relevant clinical scenario for this alarm would be a leak in the inspiratory circuit leading to an increase in volume delivered without the target pressure being reached. B, The low‐volume alarm is triggered once the calculated volume drops below the lower limit set by the healthcare provider. This was simulated by increasing the PEEP up to 17 cm H2O. A relevant clinical scenario for this alarm would be the inspiratory line being kinked. C, The high‐pressure scenario was simulated by interrupting the expansion of the lung simulator during inspiration to simulate a patient coughing. The high‐pressure alarm was triggered when the pressure exceeded the set value of 30 cm H2O. Other scenarios are provided in the [Link], [Link], [Link], [Link], including a 24‐hr operation test and 12 adverse ventilation situations per ISO80601‐2‐80:2018 table 201.105 (International Standards Organization, 2018)
Suitable MADVent Mark V operating parameter ranges
| Operating parameter | Tested range |
|---|---|
| Target inspiratory pressure | 10–35 cm H2O |
| Tidal volume ( | 200–1,000 ml |
| Respiratory rate (RR) | 6–35 bpm |
| Inspiratory time | 1–3.0 s |
| Low‐pressure alarm threshold | 0–20 cm H2O |
| High‐pressure alarm threshold | 30–60 cm H2O |
| High‐volume alarm threshold | 200–1,000 ml |
| Low‐volume alarm threshold | 200–1,000 ml |
FIGURE 5The volume‐driven version of the MADVent comes with alarms for high and low pressure that can be set between 0 and 50 cm H2O defined by the caregiver. The system was initially set at a rate of 34 breaths per minute, a PEEP value of 5 cm H2O was chosen and compliance on the lung simulator set to 0.03 /(cm H2O). A, The low‐ and high‐pressure alarm thresholds were set to 2 and 42 cm H2O, respectively. PEEP values were increased from 5 to 20 cm H2O and lowered back down to 5.0 cm H2O to ensure that the in‐line pressure sensor could detect and display changes in pressure values. A high‐pressure condition was simulated by decreasing patient lung compliance. The system triggered an alarm once the pressure went above 42 cm H2O. B, The low‐pressure alarm is triggered once the in‐line pressure value drops below the lower limit. A low‐pressure situation was simulated by disconnecting the endotracheal tube to trigger an alarm which results in the system immediately stopping. C, In the event that the tubing is kinked or there is a blockage in the endotracheal tube, the pressure begins to rise until the upper threshold is reached. This triggers a high‐pressure alarm and causes the system to resume ventilation at a lower volume, but at an increased rate according to the set minute ventilation