Purpose: To investigate the repeatability and reproducibility of total retinal blood flow measurements using a custom-built dual-beam bidirectional Doppler optical coherence tomography (OCT) system in healthy subjects.Entities:
Keywords: healthy subjects; repeatability; retinal blood flow
Mesh:
Year: 2020 PMID: 32832239 PMCID: PMC7414639 DOI: 10.1167/tvst.9.7.34
Source DB: PubMed Journal: Transl Vis Sci Technol ISSN: 2164-2591 Impact factor: 3.283
Figure 1.Geometrical setting at the posterior pole of the eye. , probe beam wave vectors; , velocity vector; α1/2, Doppler angles; Δα, separation angle between probe beams; γ, angle between and the plane perpendicular to the detection plane.
Total Retinal Blood Flow as Obtained From Retinal Arteries and Veins Using Our Previous Algorithm (QA,Abs and QV,Abs) and Our New Algorithm (QA,New and QV,New) as Obtained in the Repeatability Cohort (n = 10)
| Values (µL/min) | Intraclass Correlation Coefficient | Coefficient of Variation (%) | |
|---|---|---|---|
| QA,abs | 41.3 ± 6.8 | 0.81 | 6.3 ± 3.7 |
| QV,abs | 40.6 ± 6.4 | 0.84 | 5.8 ± 3.4 |
| QA,new | 39.5 ± 6.2 | 0.91 | 3.8 ± 2.3 |
| Qv,new | 39.0 ± 5.9 | 0.93 | 3.3 ± 1.9 |
All data are the mean of the two measurements. Data are presented as means ± SD.
Figure 2.Bland-Altman plots for repeatability data. The x-axis shows the mean between the two measurements (Qmean), the y-axis shows the difference between the two measurements (Qdiff). The data for arteries are presented in red and the data for veins are presented in blue. The mean value of the difference (solid line) and the 95% confidence intervals are also shown (dashed line).
Total Retinal Blood Flow as Obtained From Retinal Arteries and Veins Using Our Previous Algorithm (QA,Abs and QV,Abs) and Our New Algorithm (QA,New and QV,New) as Obtained in the Reproducibility Cohort (n = 34)
| Values (µL/min) | Intraclass correlation coefficient | Coefficient of variation (%) | |
|---|---|---|---|
| QA,abs | 40.1 ± 6.5 | 0.78 | 6.7 ± 5.1 |
| QV,abs | 39.4 ± 6.3 | 0.79 | 7.0 ± 4.3 |
| QA,new | 38.4 ± 6.0 | 0.87 | 4.5 ± 2.9 |
| Qv,new | 38.1 ± 6.2 | 0.91 | 3.4 ± 2.6 |
All data are the mean of the two measurements. Data are presented as means ± SD.
Figure 3.Bland-Altman plots for reproducibility data. The x-axis shows the mean between the two measurements (Qmean), the y-axis shows the difference between the two measurements (Qdiff). The data for arteries are presented in red and the data for veins are presented in blue. The mean value of the difference (solid line) and the 95% confidence intervals are also shown (dashed line).