Wangyi Fang1,2, Qingchen Li1,2, Jinyu Fan3, Ning Tang3, Jian Yu1,2, Huan Xu1,2, Yuan Zong1,2, Chunhui Jiang1,2, Guohua Shi3, Xinghuai Sun1,2.
Abstract
Purpose: To evaluate the potential value of microscope-integrated optical coherence tomography (MI-OCT) in anterior segment surgical maneuvers.Entities:
Keywords: intraoperative; microscope integrated; surgical maneuvers; surgical training; swept-source optical coherence tomography
Mesh:
Year: 2020 PMID: 32832225 PMCID: PMC7414702 DOI: 10.1167/tvst.9.7.18
Source DB: PubMed Journal: Transl Vis Sci Technol ISSN: 2164-2591 Impact factor: 3.283
Figure 1.The MI-OCT system. (A) Laboratory setup: the MI-OCT system and head-up display screen. (B) Image projected on the screen: the image from the microscope and the OCT (lower left corner, only displayed to group B in trial one). The red line in the microscope image indicates the OCT scanning beam, and the red line on the OCT image indicates the 2-mm length (only displayed to group B in trial one).
Figure 2.Representative OCT images of corneal tunnel from group A (A) and group B (B) in trial one.
Figure 3.Representative OCT images of scleral tunnel from group A (A) and group B (B) in trial one.
Figure 4.Representative OCT images of the simple corneal suture from group A (A) and group B (B). The linear hyper-reflective signal in the cornea is the 10-0 polyamide suture (white arrows). Depth of the suture was presented in percentage.
Figure 5.Representative OCT images of CLR from group A (A) and group B (B). The linear hyper-reflective signal in the cornea is the 10-0 polyamide suture (white arrows). Depth of the suture was presented in percentage.
Comparison of the Deviations Between Groups A and B
| Corneal Tunnel | Scleral Tunnel | Simple Corneal Suture | Corneal Laceration Suture | |||
|---|---|---|---|---|---|---|
| Length | Depth | Length | Depth | Depth | ||
| Trial 1 | A vs. B | 0.003 | 0.024 | 0.008 | 0.405 | 0.348 |
| Trial 2 | A vs. B | 0.006 | 0.092 | 0.020 | 0.134 | 0.621 |
| Trial 1/2 | B vs. B | 0.505 | 0.666 | 0.168 | 0.695 | 0.347 |
For each resident, every maneuver was performed on a new porcine eye. The values in all columns are absolute values of the deviation between the actual measured value and the target value. Group A = without MI-OCT and group B = with MI-OCT.
Statistically significant difference.