Influence of Artificial Tears on Keratometric Measurements in Cataract Patients.

Veronika Röggla1, Christina Leydolt1, Daniel Schartmüller1, Luca Schwarzenbacher1, Elias Meyer2, Claudette Abela-Formanek1, Rupert Menapace3.   

Abstract

PURPOSE: To assess the influence of artificial tears of different viscosity on K-readings prior to cataract surgery.
DESIGN: Prospective randomized crossover, before-and-after clinical study.
METHODS: Setting: Department of Ophthalmology, Medical University of Vienna. PATIENT POPULATION: A total of 123 eyes of 80 patients prior to cataract surgery were assigned to 2 groups based on normal and dry eyes. INTERVENTION: Two native baseline keratometries were followed by instillation of either high- or low-viscosity eye drops. Keratometry was repeated 30 seconds, 2 minutes, and 5 minutes after instillation. MAIN OUTCOME MEASURES: Influence of eye drops of different viscosity in normal and dry eyes on short time K-readings.
RESULTS: Repeatability between native baseline measurements was high (standard deviation = 0.02 mm in normal and in dry eyes). In normal and dry eyes, a statistically significant increase in measurement variability after instillation of both low-viscosity and high-viscosity eye drops was observed (P < .01). Measurement variability was most pronounced between baseline measurement and 30 seconds and diminished over time. Variability of K-readings appeared higher in dry eyes compared with normal eyes. Astigmatism changed more than 0.5 diopters in 13.2% of normal eyes and 34.4% in dry eyes using eye drops of high viscosity.
CONCLUSION: Tear film-stabilizing eye drops prior to keratometry measurements influenced K-readings significantly, especially in dry eyes. A time period of more than 5 minutes should be allowed to pass after instillation of eye drops. The higher the viscosity of the eye drops, the stronger the influence and the longer its persistence.
Copyright © 2020 The Author(s). Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2020        PMID: 32828877     DOI: 10.1016/j.ajo.2020.08.024

Source DB:  PubMed          Journal:  Am J Ophthalmol        ISSN: 0002-9394            Impact factor:   5.258


  4 in total

1.  Difficulties in choosing the right intraocular lens in a previously vitrectomized patient - the role of the tear film.

Authors:  Monica Mălăescu; Bogdana Tăbăcaru; Horia Tudor Stanca
Journal:  Rom J Ophthalmol       Date:  2022 Jan-Mar

2.  Effects of Recombinant Human Epidermal Growth Factor Eye Drops Combined with Phacoemulsification on Short- and Long-Term Visual Acuity Recovery and Related Dry Eye Complications in Patients with Senile Cataract.

Authors:  Lingzhi Fan; Tao He; Jiguang Shi; Xianlin Ke; Xiaolei Lu
Journal:  Comput Math Methods Med       Date:  2022-06-28       Impact factor: 2.809

3.  Corneal irregularity and visual function using anterior segment optical coherence tomography in TGFBI corneal dystrophy.

Authors:  Yuito Abe; Takashi Omoto; Kohdai Kitamoto; Tetsuya Toyono; Junko Yoshida; Ryo Asaoka; Satoru Yamagami; Takashi Miyai; Tomohiko Usui
Journal:  Sci Rep       Date:  2022-08-12       Impact factor: 4.996

4.  Ocular surface optimization before cataract surgery.

Authors:  Nandini Venkateswaran; Regina D Luna; Preeya K Gupta
Journal:  Saudi J Ophthalmol       Date:  2022-08-29
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.