Lukas Streese1, Lukas Y Brawand1, Konstantin Gugleta2, Peter M Maloca2,3,4, Walthard Vilser5, Henner Hanssen1.
Abstract
Purpose: To compare measurement of wall-to-lumen ratio (WLR) by means of high-resolution adaptive optics imaging (AO) with intuitive to use retinal vessel wall (VW) analysis (VWA). Moreover, to validate the techniques by comparing WLR of healthy young (HY) with healthy older patients.Entities:
Keywords: microcirculation; retina; retinal vessel analysis; wall-to-lumen ratio
Mesh:
Year: 2020 PMID: 32821504 PMCID: PMC7408937 DOI: 10.1167/tvst.9.6.7
Source DB: PubMed Journal: Transl Vis Sci Technol ISSN: 2164-2591 Impact factor: 3.283
Figure 1.Retinal vessel imaging using AO and corresponding standard color fundus imaging by intuitive to use retinal VWA. Five maker lines were applied to measure retinal vessel microstructure in the same eye imaged with AO (A) and VWA (B). The original brightness profile for one analysis of the retinal vessel microstructure is shown (C).
Figure 2.Detailed description of retinal vessel microstructure assessment by means of brightness profiles. BB, mean brightness baseline, BP, mean brightness peak.
Clinical Characteristics
| HY ( | HO ( |
| |
|---|---|---|---|
| Age (years) | 24 ± 2 | 60 ± 8 | <.001 |
| Sex (female/male) | 8/5 | 7/9 | .358 |
| Weight (kg) | 64 ± 11 | 65 ± 8 | .713 |
| BMI (kg/m2) | 22 ± 2 | 23 ± 1 | .245 |
| Sys BP (mm Hg) | 118 ± 10 | 116 ± 11 | .465 |
| Dia BP (mm Hg) | 79 ± 3 | 77 ± 9 | .510 |
| IOP (mm Hg) | 14 ± 2 | 15 ± 2 | .633 |
P value, level of significance for two sample t-test. dia BP, diastolic blood pressure; IOP, intraocular pressure; SD, standard deviation; Sys BP, systolic blood pressure.
Figure 3.Two-tailed Pearson correlation of VW (A), VL (B), and WLR (C) measured by AO and VWA. Differences in VW (D), VL (E), and WLR (F) between AO and VWA as well as between HY and HO. ***P < .001 for two-sided paired sample t-tests; †P < .05 for two-sided independent t-tests.
Retinal VW, VL, and WLR in HY and HO Individuals Measured by AO and Retinal VWA
| HY | HO | |||||||
|---|---|---|---|---|---|---|---|---|
| AO | VWA |
| AO | VWA |
|
|
| |
| VW (MU) | 34 ± 9 | 20 ± 5 | <.001 | 40 ± 14 | 21 ± 4 | <.001 | .201 | .381 |
| VL (MU) | 226 ± 53 | 115 ± 27 | <.001 | 206 ± 49 | 104 ± 20 | <.001 | .295 | .245 |
| WLR (%) | 31 ± 8 | 36 ± 8 | <.001 | 39 ± 9 | 42 ± 5 | <.001 | .012 | .013 |
P value for paired sample t tests AO vs. VWA in HY
P value for paired sample t-tests AO vs. VWA in HO.
P value for two-sample t-tests of AO parameter HY vs. HO.
P value for two-sample t-tests of VWA parameter HY vs. HO. MU, measuring units.
Correlation Coefficients of Retinal Vessel Microstructure with Age, BMI, and Systolic and Diastolic Blood Pressure
| AO | VWA | |||||
|---|---|---|---|---|---|---|
| VW | VL | WLR | VW | VL | WLR | |
| Age | 0.31 | −0.13 | 0.47 | 0.24 | −0.16 | 0.43 |
| BMI | 0.25 | 0.24 | 0.08 | 0.44 | 0.34 | 0.16 |
| Sys BP | −0.14 | 0.13 | −0.27 | −0.17 | 0.27 | −0.20 |
| Dia BP | 0.07 | 0.34 | −0.24 | 0.16 | 0.28 | −0.17 |
p > .01.
p < .05. Sys BP, systolic blood pressure; Dia BP, diastolic blood pressure; VWA, retinal vessel wall analysis.